The Signature Orthopaedics NOOSA, NAMBUCCA, and CAIRNS Anterior Lumbar Plate Systems are indicated for use via the lateral or anterolateral surgical approach above the bifurcation of the great vessels or via the anterior surgical approach, below the bifurcation of the great vessels. The devices are intended as temporary fixation is achieved. The subject systems are indicated in the treatment of lumbosacral (L I-S I ) fixation for the following indications: degenerative disc disease (DDD)(as define by back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e. fracture or dislocation), deformities or curvatures (i.e. scoliosis, kyphosis, and/or lordosis), tumor, pseudarthrosis, and failed previous fusion.
Device Story
Temporary spinal fixation system consisting of plates and screws; used for stabilization during spinal fusion procedures. Inputs: manual surgical placement by orthopedic/neurosurgeon. Operation: plates contoured to vertebral body; screws hand-tightened; NOOSA/NAMBUCCA systems utilize anti-angulation locking cams to restrict screw movement. Output: mechanical stabilization of spinal segments. Used in clinical settings; provides supplemental fixation to augment fusion. Benefits: correction/stabilization of lumbosacral spine; temporary support until solid fusion achieved.
Clinical Evidence
No clinical data. Bench testing only: static and dynamic compression bending, static torsion, screw insertion, screw pull-out, and screw torque-to-failure testing performed on worst-case plate and screw sizes.
Technological Characteristics
Materials: Ti6Al4V alloy (ISO 5832-3, ASTM-F136). Principle: Mechanical temporary fixation via plates and screws. Design: Low-profile, anatomically contoured plates; manual implantation. Locking: Anti-angulation locking cams (NOOSA/NAMBUCCA). Sterilization: Not specified.
Indications for Use
Indicated for patients requiring lumbar or lumbosacral (L1-S1) fixation for degenerative disc disease, spondylolisthesis, trauma (fracture/dislocation), deformities (scoliosis, kyphosis, lordosis), tumor, pseudarthrosis, or failed previous fusion.
Regulatory Classification
Identification
A spinal intervertebral body fixation orthosis is a device intended to be implanted made of titanium. It consists of various vertebral plates that are punched into each of a series of vertebral bodies. An eye-type screw is inserted in a hole in the center of each of the plates. A braided cable is threaded through each eye-type screw. The cable is tightened with a tension device and it is fastened or crimped at each eye-type screw. The device is used to apply force to a series of vertebrae to correct “sway back,” scoliosis (lateral curvature of the spine), or other conditions.
Synthes Spine Anterior Tension Band (ATB) (K022791)
Reference Devices
Signature Orthopaedics Logical Bone Screw (K121297)
Related Devices
K253201 — Lumbar Spine Truss System - Plating Solution (LSTS-PS) · 4Web, Inc. · Feb 12, 2026
K042702 — SILHOUETTE SPINAL FIXATION SYSTEM · Zimmer Spine, Inc. · Nov 23, 2004
K100061 — CAYMAN SCREWS MODEL K2-12-1000-XX, K2-12-1001-XX · K2m, Inc. · Feb 12, 2010
K101255 — ALPHATEC SPINE ANTERIOR LUMBAR PLATING SYSTEM · Alphatec Spine, Inc. · Jul 21, 2010
K230424 — Omnia Medical Coupler-A Anterior Lumbar Plate System; Omnia Medical Coupler-L Lateral Lumbar Plate System · Omnia Medical, LLC · Apr 12, 2023
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus or a representation of the human form.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 11, 2017
Signature Orthopaedics Pty Ltd. Dr. Declan Brazil Managing Director 7 Sirius Road Lane Cove, NSW 2066 Australia
Re: K163625
Trade/Device Name: NOOSA Anterior Lumbar Plate System NAMBUCCA Anterior Lumbar Plate System CAIRNS Anterior Lumbar Plate System Regulation Number: 21 CFR 888.3060 Regulation Name: Spinal Intervertebral Body Fixation Orthosis Regulatory Class: Class II Product Code: KWO Dated: July 3. 2017 Received: July 14, 2017
Dear Dr. Brazil:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K163625
#### Device Name
NOOSA Anterior Lumbar Plate System NAMBUCCA Anterior Lumbar Plate System CAIRNS Anterior Lumbar Plate System
#### Indications for Use (Describe)
The Signature Orthopaedics NOOSA, NAMBUCCA, and CAIRNS Anterior Lumbar Plate Systems are indicated for use via the lateral or anterolateral surgical approach above the bifurcation of the great vessels or via the anterior surgical approach, below the bifurcation of the great vessels. The devices are intended as temporary fixation is achieved. The subject systems are indicated in the treatment of lumbosacral (L I-S I ) fixation for the following indications: degenerative disc disease (DDD)(as define by back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e. fracture or dislocation), deformities or curvatures (i.e. scoliosis, kyphosis, and/or lordosis), tumor, pseudarthrosis, and failed previous fusion.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
{3}------------------------------------------------
# 510(K) SUMMARY
| Manufacturer: | Signature Orthopaedics Pty Ltd<br>7 Sirius Road<br>Lane Cove, NSW 2066<br>Australia |
|-----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Trade<br>Name: | NOOSA Anterior Lumbar Plate System<br>NAMBUCCA Anterior Lumbar Plate System<br>CAIRNS Anterior Lumbar Plate System |
| Common Name: | Spinal Intervertebral Body Fixation Orthosis |
| Contact: | Dr. Declan Brazil<br>Managing Director of Signature Orthopaedics |
| Prepared By: | Signature Orthopaedics Pty Ltd<br>7 Sirius Road<br>Lane Cove, NSW 2066<br>Australia<br>Phone: +61 (2) 9428 5181<br>Fax: +61 (2) 8456 6065 |
| Date Prepared: | August 11th, 2017 |
| Classification: | Class II per 21 CFR 888.3060: Spinal Intervertebral Body Fixation<br>Orthosis (KWQ) |
| Predicate Devices: | Primary Predicate<br>• DePuy Spine AEGIS (K052546)<br>Additional Predicate<br>• Medtronic Sofamor Danek PYRAMIDTM (K013665)<br>• Synthes Spine Anterior Tension Band (ATB) (K022791)<br>Reference Predicate<br>• Signature Orthopaedics Logical Bone Screw (K121297) |
# Device Description:
The Signature Orthopaedics' NOOSA, NAMBUCCA and CAIRNS Anterior Lumbar Plate Systems' components are manufactured from Ti6Al4V alloy per ISO 5832-3 and ASTM-F136.
The NOOSA, NAMBUCCA and CAIRNS Anterior Lumbar Plate Systems are temporary supplemental fixation devices consisting of variety of shapes and sizes of plates and screws. The subject systems are used as an implant for the correction and stabilisation of the lumbosacral spine. The systems provide temporary stablisation and augments the development of a solid fusion. Additionally, the subject systems can be used in conjunction with interbody fusion devices with anterior plate fixation. The subject plates are low profile and anatomically designed to provide optimal fit from anterior, lateral or
{4}------------------------------------------------
anterolateral approach. The NOOSA and NAMBUCCA systems feature anti-angulation locking cams to help restrict the angular movement of the screw within the screw holes of the plates.
## Indications for Use:
The Signature Orthopaedics NOOSA. NAMBUCCA and CAIRNS Anterior Lumbar Plate Systems are indicated for use via the lateral or anterolateral surgical approach above the bifurcation of the great vessels or via the anterior surgical approach, below the bifurcation of the great vessels. The devices are intended as a temporary fixation devices until fusion is achieved. The subject systems are indicated in the treatment of lumbar or lumbosacral (L1-S1) fixation for the following indications: degenerative disc disease (DDD)(as define by back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies) spondylolisthesis, trauma (i.e. fracture or dislocation), deformities or curvatures (i.e. scoliosis, and/or lordosis), tumor, pseudoarthrosis, and failed previous fusion.
# Performance Testing:
Non-clinical testing and engineering evaluations were conducted to verify that the performance of the NOOSA. NAMBUCCA and CAIRNS Anterior Lumbar Plates Systems are adequate for anticipated in-vivo use. The following non-clinical testings were carried out on the worst case sizes of the plates and screws:
- . Static and dynamic compression bending testing
- Static torsion testing
- . Screw insertion testing
- Screw pull-out testing
- . Screw torque to failure testing
## Substantial Equivalence:
That Signature Orthopaedics' NOOSA, NAMBUCCA and CAIRNS Anterior Lumbar Plate Systems have the same intended use, indication for use, materials and similar design features as the DePuy Spine AEGIS (K052546), Medtronic Sofamor Danek PYRAMID™ (K013665), Synthes Spine ATB (K022791) Anterior Lumbar Plate Systems and Signature Orthopaedics Logical Bone Screw (K121297). Non-clinical testing results support the substantial equivalence to the predicate devices.
## Comparison of technological characteristics
Temporary immobilisation and stabilisation of spinal segments as supplemental fixation in spinal fusion procedure is the technological principle for both the subject devices and the predicate devices. The subject and predicate devices are based on the following same technological elements:
- . The indication for use of the subject devices is the same as one or more of the predicate devices
- . The intended surgery sites of the subject devices matches the intended surgery sites of at least one of the predicates
- . The subject devices are manufactured from the same material as two of the
{5}------------------------------------------------
predicates
- . The locking mechanism on NOOSA and NAMBUCCA are identical to primary predicate
- The profile thickness of the subject devices are within the profile thickness of the . predicates
- . The size range of the subject devices fall within the size range of the predicates
- . The subject devices accommodate the same number of supplemental screws as two of the predicates
- . Both subject devices and predicates are contoured to match the vertebral body
- . Both subject devices and predicates are implanted manually and the screws are hand tightened
The following technological differences exist between the subject and predicate devices:
- Some of the design features are different between the subject and predicate ● devices
- The material used on one of the predicates are slightly different than the subject devices
- . The screw locking mechanism on two of the predicates are different the subject devices
- . The indicated surgery location of two of the predicates are limited compared to the subject devices
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.