← Product Code [KWQ](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/KWQ) · K161524

# Tangis Anterior Cervical Plate (K161524)

_Nvision Biomedical Technologies, LLC · KWQ · Sep 2, 2016 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/KWQ/K161524

## Device Facts

- **Applicant:** Nvision Biomedical Technologies, LLC
- **Product Code:** [KWQ](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/KWQ.md)
- **Decision Date:** Sep 2, 2016
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3060
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

The Tangis Anterior Cervical Plate is intended for anterior screw fixation to the cervical spine. It is to be used in skeletally mature patients as an adjunct to fusion of the cervical spine (C2 to T1). The system is indicated for use in the temporary stabilization of the anterior spine during the development of cervical spinal fusion in patients with: · degenerative disc disease (as defined by neck pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies), - · spondylolisthesis. - · trauma (i.e. fractures or dislocations), - tumors, - · deformity (defined as kyphosis, lordosis, or scoliosis), - · pseudarthrosis, - · failed previous fusion, - · spinal stenosis

## Device Story

Tangis Anterior Cervical Plate system; provides temporary anterior stabilization of cervical spine (C2-T1) during fusion. System comprises titanium alloy plates (20-92mm) and self-drilling/self-tapping bone screws (3.75-4.75mm diameter; 10-20mm length). Surgeon performs anterior approach; attaches plate to vertebral bodies using screws. Plates feature central graft visualization holes for intraoperative monitoring. Used in clinical settings by orthopedic or neurosurgeons. Provides mechanical support to facilitate spinal fusion; benefits patients by stabilizing spine during healing process.

## Clinical Evidence

Bench testing only. Static and dynamic compression and static torsion testing performed per ASTM F1717 to demonstrate mechanical equivalence.

## Technological Characteristics

Titanium alloy (ASTM F136). Plate lengths 20-92mm; screw diameters 3.75-4.75mm. Features graft visualization holes. Mechanical fixation via self-drilling/self-tapping screws. No software or energy source.

## Regulatory Identification

A spinal intervertebral body fixation orthosis is a device intended to be implanted made of titanium. It consists of various vertebral plates that are punched into each of a series of vertebral bodies. An eye-type screw is inserted in a hole in the center of each of the plates. A braided cable is threaded through each eye-type screw. The cable is tightened with a tension device and it is fastened or crimped at each eye-type screw. The device is used to apply force to a series of vertebrae to correct “sway back,” scoliosis (lateral curvature of the spine), or other conditions.

## Predicate Devices

- Anterior Cervical Plate System / Osteomed Implantes, LTDA ([K151553](/device/K151553.md))
- ACLP System / Synthes Spine ([K031276](/device/K031276.md))
- Blackstone™ “Classic” Anterior Cervical Plate / Blackstone Medical, Inc. ([K022965](/device/K022965.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

September 2, 2016

Nvision Biomedical Technologies, LLC % Allison Komiyama, Ph.D., RAC Principal Consultant AcKnowledge Regulatory Strategies 2834 Hawthorn Street San Diego, California 92104

Re: K161524

Trade/Device Name: Tangis Anterior Cervical Plate Regulation Number: 21 CFR 888.3060 Regulation Name: Spinal intervertebral body fixation orthosis Regulatory Class: Class II Product Code: KWQ Dated: August 2, 2016 Received: August 3, 2016

Dear Dr. Komiyama:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{1}------------------------------------------------

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# Mark N. Melkerson -S

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) к161524

Device Name Tangis Anterior Cervical Plate

#### Indications for Use (Describe)

The Tangis Anterior Cervical Plate is intended for anterior screw fixation to the cervical spine. It is to be used in skeletally mature patients as an adjunct to fusion of the cervical spine (C2 to T1). The system is indicated for use in the temporary stabilization of the anterior spine during the development of cervical spinal fusion in patients with: · degenerative disc disease (as defined by neck pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies),

- · spondylolisthesis.
- · trauma (i.e. fractures or dislocations),
- tumors,
- · deformity (defined as kyphosis, lordosis, or scoliosis),
- · pseudarthrosis,
- · failed previous fusion,
- · spinal stenosis

Type of Use (Select one or both, as applicable)

2 Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

#### FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for nVision Biomedical Technologies. The "n" in "nVision" is gold and stylized, while the rest of the word is in black. Below "nVision" is the text "BIOMEDICAL TECHNOLOGIES" in a smaller, gold font.

# 510(k) Summary

DATE PREPARED

May 31, 2016

### MANUFACTURER AND 510(k) OWNER

Nvision Biomedical Technologies, LLC 1350 N Loop 1604 E, Suite 103, San Antonio, TX 78232, USA Telephone: (210) 545-3713 Fax: (866) 764-1139 Official Contact: Diana L. Langham, Director of Regulatory and Corporate Compliance

### REPRESENTATIVE/CONSULTANT

Allison C. Komiyama, Ph.D., R.A.C. AcKnowledge Regulatory Strategies +1 (619) 208-7888 Telephone: Email: akomiyama@acknowledge-rs.com

### PROPRIETARY NAME OF SUBJECT DEVICE

Tangis Anterior Cervical Plate

COMMON NAME

Anterior Cervical Plate

### DEVICE CLASSIFICATION

Spinal intervertebral body fixation orthosis (21 CFR 888.3060, Product Code KWQ, Class II)

### PREMARKET REVIEW

ODE/DOD/ASDB Orthopedic Panel

### INDICATIONS FOR USE

The Tangis Anterior Cervical Plate is intended for anterior screw fixation to the cervical spine. It is to be used in skeletally mature patients as an adjunct to fusion of the cervical spine (C2 to T1). The system is indicated for use in the temporary stabilization of the anterior spine during the development of cervical spinal fusion in patients with:

- o degenerative disc disease (as defined by neck pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies),
- . spondylolisthesis,
- trauma (i.e. fractures or dislocations),
- tumors,
- . deformity (defined as kyphosis, lordosis, or scoliosis),

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Nvision Biomedical Technologies. The logo features a stylized letter 'n' in gold, followed by the word 'vision' in black. Below the word 'vision' are the words 'BIOMEDICAL TECHNOLOGIES' in a smaller, gold font. The logo is clean and modern, reflecting the company's focus on biomedical technology.

- pseudarthrosis,
- failed previous fusion,
- spinal stenosis

## DEVICE DESCRIPTION

The Tangis Anterior Cervical Plate is intended for anterior screw fixation of the plate to the cervical spine. The fixation construct consists of a cervical plate that is attached to the vertebral body of the cervical spine with self-drilling and self-tapping bone screws using an anterior approach. Plates are available in a variety of lengths (20 mm – 92 mm), addressing multiple levels of fixation (one to four). The plate incorporates graft visualization holes on the longitudinal center line for intraoperative visualization. Bone screws are available in three diameters (3.75 mm, 4.25 mm, and 4.75 mm) and a variety of lengths (10 mm - 20 mm). All components are made from titanium alloy per ASTM F136.

## PREDICATE DEVICE IDENTIFICATION

The Tangis Anterior Cervical Plate is substantially equivalent to the following predicates:

| 510(k)<br>Number | Predicate Device Name / Manufacturer                                     | Primary<br>Predicate |
|------------------|--------------------------------------------------------------------------|----------------------|
| K151553          | Anterior Cervical Plate System / Osteomed Implantes, LTDA                | ✓                    |
| K031276          | ACLP System / Synthes Spine                                              |                      |
| K022965          | Blackstone™ “Classic” Anterior Cervical Plate / Blackstone Medical, Inc. |                      |

### SUMMARY OF NON-CLINICAL TESTING

No FDA performance standards have been established for the Tangis Anterior Cervical Plate. The following tests were performed to demonstrate safety based on current industry standards:

- . Static and dynamic compression (per ASTM F1717)
- . Static torsion (per ASTM F1717)

The results of these tests indication that the Tangis Anterior Cervical Plate is substantially equivalent to the predicate devices.

## EQUIVALENCE TO PREDICATE DEVICES

Nvision believes that the Tangis Anterior Cervical Plate is substantially equivalent to the predicate devices based on the information summarized here:

The subject device has a similar design and dimensions, and uses similar or identical materials as the devices cleared in K151553, K031276, and K022965. The subject device has the same intended use and similar technological characteristics to the devices cleared in K151553 and K022965. The device has similar instrumentation to the device cleared in K151553, and K031276.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Nvision Biomedical Technologies. The logo features a stylized letter 'n' in gold, followed by the word 'vision' in black, bold letters. Below the word 'vision' are the words 'BIOMEDICAL TECHNOLOGIES' in a smaller, gold font.

## CONCLUSION

The Tangis Anterior Cervical Plate is considered substantially equivalent to the predicate devices based on the testing performed, the identical indications for use, and similar technological characteristics. Based on the testing performed, including static and dynamic compression as well as static torsion (per ASTM F1717), it can be concluded that the subject device does not raise new issues of safety or efficacy compared to the predicate devices.

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/KWQ/K161524](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/KWQ/K161524)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
