Pyrenees Cervical Plate System

K150983 · K2m · KWQ · Jul 14, 2015 · Orthopedic

Device Facts

Record IDK150983
Device NamePyrenees Cervical Plate System
ApplicantK2m
Product CodeKWQ · Orthopedic
Decision DateJul 14, 2015
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3060
Device ClassClass 2
AttributesTherapeutic

Indications for Use

K2M Cervical Plate Systems are indicated for use in anterior screw fixation to the cervical spine (C2-C7) for the following indications: degenerative disc disease (DDD) (defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (including fractures), spinal stenosis and tumors (primary and metastatic), failed previous fusions (pseudarthrosis) and deformity (defined as scoliosis. kyphosis or lordosis).

Device Story

Pyrenees Cervical Plate System provides support and stabilization of cervical vertebrae (C2-C7) via anterior screw fixation. System comprises titanium alloy (ASTM F1472/F136) and CP titanium (ASTM F67) plates (1-5 level, 18-110mm) and screws (10-22mm). Available in constrained and translational designs. Used by surgeons in clinical settings to treat spinal pathologies including DDD, trauma, and deformity. Components provided non-sterile; require sterilization before use. Mechanical performance validated via static compression, static torsion, and dynamic compression testing per ASTM F1717.

Clinical Evidence

Bench testing only. Mechanical performance evaluated via static compression, static torsion, and dynamic compression testing in accordance with ASTM F1717 standards compared to predicate devices.

Technological Characteristics

Materials: Titanium alloy (ASTM F1472, ASTM F136) and CP titanium (ASTM F67). Components: Plates (1-5 level, 18-110mm) and self-tapping/self-starting screws (10-22mm). Designs: Constrained and translational. Energy source: None (mechanical). Connectivity: None. Sterilization: Non-sterile (requires user sterilization).

Indications for Use

Indicated for patients requiring anterior screw fixation to the cervical spine (C2-C7) for degenerative disc disease, spondylolisthesis, trauma (fractures), spinal stenosis, tumors, pseudarthrosis, or spinal deformity (scoliosis, kyphosis, lordosis).

Regulatory Classification

Identification

A spinal intervertebral body fixation orthosis is a device intended to be implanted made of titanium. It consists of various vertebral plates that are punched into each of a series of vertebral bodies. An eye-type screw is inserted in a hole in the center of each of the plates. A braided cable is threaded through each eye-type screw. The cable is tightened with a tension device and it is fastened or crimped at each eye-type screw. The device is used to apply force to a series of vertebrae to correct “sway back,” scoliosis (lateral curvature of the spine), or other conditions.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 July 14, 2015 K2M, Incorporated % Ms. Nancy Giezen Manager, Regulatory Affairs 751 Miller Drive South East Leesburg, Virginia 20175 Re: K150983 Trade/Device Name: Pyrenees Cervical Plate System Regulation Number: 21 CFR 888.3060 Regulation Name: Spinal intervertebral body fixation orthosis Regulatory Class: Class II Product Code: KWQ Dated: June 16, 2015 Received: June 17, 2015 Dear Ms. Giezen: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set {2}------------------------------------------------ Page 2 - Ms. Nancy Giezen forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely vours. # Mark N. Melkerson -S Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K150983 Device Name Pyrenees Cervical Plate System #### Indications for Use (Describe) K2M Cervical Plate Systems are indicated for use in anterior screw fixation to the cervical spine (C2-C7) for the following indications: degenerative disc disease (DDD) (defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (including fractures), spinal stenosis and tumors (primary and metastatic), failed previous fusions (pseudarthrosis) and deformity (defined as scoliosis. kyphosis or lordosis). | Type of Use (Select one or both, as applicable) | | |-----------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------| | <span> <span style="text-decoration: overline;">X</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> | <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ ## 510(k) SUMMARY Pyrenees Cervical Plate System #### Submitter K2M, Inc. 751 Miller Drive SE Leesburg, VA 20175 Contact Person: Nancy Giezen Telephone: 703-777-3155 Date Prepared: 06/16/2015 ## Classification Pyrenees Cervical Plate System Trade Name: Common Name: Spinal Fixation System Regulatory Class: Class II Classification Name(s): Spinal Intervertebral Body Fixation Orthosis (21 CFR 888.3060, Product Code KWQ) ## Predicate Device(s) Primary Predicate: K2M Pyrenees Cervical Plate System (K063544) Additional Predicates: DePuy Uniplate (K042544) Synthes CSLP (K971883, K000536) EBI Cyprus (K060379) DePuy Peak (K971730) ## Device Description The Pyrenees Cervical Plate System consists of plates (1-5 level) and screws (self-tapping and self-starting) made of titanium alloy per ASTM F1472 and F136 and CP titanium per ASTM F67. The plates range in length from 18-110mm and the screws from 10-22mm. The Pyrenees Cervical Plates are offered in both constrained and translational designs. The subject 510(k) adds plates and screws to the system. The Pyrenees system components are provided non-sterile. Function: The system functions as a spinal fixation device to provide support and stabilization of the cervical vertebrae. ## Intended Use K2M Cervical Plate Systems are indicated for use in anterior screw fixation to the cervical spine (C2-C7) for the following indications: degenerative disc disease (DDD) (defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (including fractures), spinal stenosis and tumors (primary and metastatic), failed previous fusions (pseudarthrosis) and deformity (defined as scoliosis, kyphosis or lordosis). {5}------------------------------------------------ #### Technological Comparison to Predicate(s) The Pyrenees Cervical Plate System was compared to predicate systems and the design features, materials and sizes were found to be substantially the same as these systems. #### Non-clinical Performance Evaluation The Pyrenees plates were mechanically tested and compared to predicate devices. The plates performed equally to or better than these systems in static compression, static torsion, and dynamic compression in accordance with ASTM F1717. #### Conclusion There are no significant differences between the Pyrenees Cervical Plate System and other systems currently being marketed which would adversely affect the use of the product. It is substantially equivalent to these other devices in design, function, material and intended use.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...