← Product Code [KWQ](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/KWQ) · K150700

# ACP 1 Anterior Cervical Plating System (K150700)

_Stryker Corporation · KWQ · Jul 15, 2015 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/KWQ/K150700

## Device Facts

- **Applicant:** Stryker Corporation
- **Product Code:** [KWQ](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/KWQ.md)
- **Decision Date:** Jul 15, 2015
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3060
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

The ACP 1™ Anterior Cervical Plating System is intended for use as an aid in cervical spinal fusion and is intended for unilateral fixation. The ACP 1™ system is intended for anterior intervertebral screw fixation of the cervical spine at levels C2-T1. The system is indicated for temporary stabilization of the anterior spine during the development of cervical spine fusions in patients with the following indications: Degenerative Disc Disease (as defined by neck pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies) Trauma (including fractures) Tumors Deformities or curvatures (including kyphosis, lordosis, or scoliosis) Pseudarthrosis Failed previous fusion Decompression of the spinal cord following total or partial cervical vertebrectomy Spondylolisthesis Spinal stenosis

## Device Story

ACP 1™ Anterior Cervical Plating System; spinal fixation hardware for anterior cervical spine stabilization. System comprises titanium alloy plates (1-4 levels) and bone screws (fixed/variable angle, self-tapping/self-drilling). Locking screw mechanism prevents bone screw backout; provides visual/tactile confirmation of secure locking. Used by surgeons in clinical settings to provide temporary stabilization during fusion development. Implants are single-use and sterile. Device facilitates spinal alignment and stability; aids in fusion process for various pathologies.

## Clinical Evidence

Bench testing only. Risk analysis and mechanical testing performed per ASTM F1717-14, including static and dynamic compression testing and static torsion testing. All predefined acceptance criteria were met.

## Technological Characteristics

Titanium alloy (Ti6Al4V) per ASTM F136 and ISO 5832-3. Plate configurations: 1-4 levels (10mm-96mm lengths). Bone screws: 4.0mm/4.5mm diameters, fixed/variable angles, self-tapping/self-drilling. Locking screw anti-backout mechanism. Sterile, single-use. Manual surgical instrumentation.

## Regulatory Identification

A spinal intervertebral body fixation orthosis is a device intended to be implanted made of titanium. It consists of various vertebral plates that are punched into each of a series of vertebral bodies. An eye-type screw is inserted in a hole in the center of each of the plates. A braided cable is threaded through each eye-type screw. The cable is tightened with a tension device and it is fastened or crimped at each eye-type screw. The device is used to apply force to a series of vertebrae to correct “sway back,” scoliosis (lateral curvature of the spine), or other conditions.

## Predicate Devices

- Stryker Spine, Aviator® Anterior Cervical Plating (ACP) System ([K142237](/device/K142237.md))
- Synthes Anterior CSLP System ([K000742](/device/K000742.md))
- BioMet MaxAn® Anterior Cervical Plate System ([K133518](/device/K133518.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an emblem featuring a stylized image of three human profiles facing to the right, stacked one behind the other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

July 15, 2015

Stryker Corporation Ms. Tina Mornak Regulatory Affairs Specialist 2 Pearl Court Allendale, New Jersey 07401

Re: K150700

Trade/Device Name: ACP 1™ Anterior Cervical Plating System Regulation Number: 21 CFR 888.3060 Regulation Name: Spinal Intervertebral Body Fixation Orthosis Regulatory Class: Class II Product Code: KWQ Dated: June 18, 2015 Received: June 19, 2015

Dear Ms. Mornak:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{1}------------------------------------------------

Page 2 - Ms. Tina Mornak

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Mark N. Melkerson -S

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

## 510(k) Number (if known)

K150700

#### Device Name

ACP 1TM Anterior Cervical Plating System

#### Indications for Use (Describe)

The ACP 1™ Anterior Cervical Plating System is intended for use as an aid in cervical spinal fusion and is intended for unilateral fixation.

The ACP 1™ system is intended for anterior intervertebral screw fixation of the cervical spine at levels C2-T1. The system is indicated for temporary stabilization of the anterior spine during the development of cervical spine fusions in patients with the following indications:

· Degenerative Disc Disease (as defined by neck pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies)

- · Trauma (including fractures)
- Tumors
- · Deformities or curvatures (including kyphosis, lordosis, or scoliosis)
- Pseudarthrosis
- · Failed previous fusion
- · Decompression of the spinal cord following total or partial cervical vertebrectomy
- · Spondylolisthesis
- · Spinal stenosis

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|-------------------------------------------------|--|

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

# CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

| 510(k) Summary                                     |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
|----------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter:                                         | Stryker Spine<br>2 Pearl Court<br>Allendale, New Jersey 07401                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Contact Person                                     | Tina Mornak<br>Regulatory Affairs Specialist<br>Phone: 201-760-8193<br>Fax: 201-962-4193<br>E-mail: tina.mornak@stryker.com                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Date Prepared                                      | June 30, 2015                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Trade Name                                         | ACP 1TM Anterior Cervical Plating System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Common Name                                        | Appliance, fixation, spinal intervertebral body                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Proposed Class                                     | Class II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Classification Name and<br>Number                  | Spinal intervertebral body fixation orthosis<br>21 CFR §888.3060                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Product Code                                       | KWQ                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Predicate Devices                                  | <b>Primary Predicate:</b><br>Stryker Spine, Aviator® Anterior Cervical Plating (ACP) System, K142237<br><b>Additional Predicates:</b><br>Synthes Anterior CSLP System, K000742<br>BioMet MaxAn® Anterior Cervical Plate System, K133518                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Device Description                                 | The ACP 1TM Anterior Cervical Plating System are one-, two-, three- and four-level<br>plate configurations ranging in lengths 10mm to 22mm for the one-level plates, 24mm<br>to 46mm for the two-level plates, 39mm to 69mm for the three-level plates and 56mm<br>to 96mm for the four-level plates. A locking screw is utilized as an anti-backout<br>mechanism to lock the bone screws in place. The bone screws are provided with either<br>fixed or variable angles available in self-tapping or self-drilling designs. The variable<br>angle bone screws allow the screw to be placed into bone at various degrees of<br>angulation, while the fixed bone screws are inserted at a defined angle. Any<br>combination of bone screws can be used to secure the cervical plate. The bone screws<br>are offered in 4.0mm and 4.5mm diameters in lengths 10mm - 20mm. The implants<br>(bone screws and cervical plates) are provided as single-use, sterile devices<br>manufactured from titanium alloy (Ti6Al4V) per ASTM F136 and ISO 5832-3, as well<br>as associated manual general surgical instrumentation. The implants are available in a<br>variety of sizes to accommodate various patient anatomies. |
| Indications for Use                                | The ACP 1TM Anterior Cervical Plating System is intended for use as an aid in cervical<br>spinal fusion and is intended for unilateral fixation.<br>The ACP 1TM system is intended for anterior intervertebral screw fixation of the<br>cervical spine at levels C2-T1. The system is indicated for temporary stabilization of<br>the anterior spine during the development of cervical spine fusions in patients with the<br>following indications:<br>Degenerative Disc Disease (as defined by neck pain of discogenic origin with<br>degeneration of the disc confirmed by patient history and radiographic studies)Trauma (including fractures)TumorsDeformities or curvatures (including kyphosis, lordosis, or scoliosis)PseudarthrosisThe associated instrumentation are Class I / 510(k) exempt devices under 21 CFR<br>§888.4540.                                                                                                                                                                                                                                                                                                                                                                        |
| 510(k) Summary                                     |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
|                                                    | Failed previous fusion Decompression of the spinal cord following total or partial cervical vertebrectomy Spondylolisthesis Spinal stenosis                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Summary of the<br>Technological<br>Characteristics | The purpose of this Traditional 510(k) is to summarize the design changes and sterile<br>labeling that have been implemented to the Aviator® Anterior Cervical Plating (ACP)<br>System. These modifications did not alter the fundamental scientific technology or<br>change/introduce an energy source. The modified devices retained by previously<br>FDA cleared indications/intended use and mode of operation as presented in 510(k)<br>#K142237.<br><br>The ACP 1TM System introduces a locking screw as an anti-backout mechanism. The<br>primary screw locking mechanism is a locking screw, which is designed to lock over<br>the bone screw heads. Insertion of the locking screw locks the bone screws in place<br>and allows for additional visual and tactile feedback that the bone screws are locked.<br>The ACP 1TM System will be provided sterile to the end-user. The implants continue<br>to be manufactured from surgical grade implantable titanium alloy.<br><br>The implemented modifications did not raise new questions of safety or efficacy.                                                                                                                                          |
| Summary of the<br>Performance Data                 | Risk analysis was performed to demonstrate that the ACP 1TM System is substantially<br>equivalent to its predicate devices. The risk analysis determined that the predefined<br>acceptance criteria associated with the following mechanical testing was met:  Static and dynamic compression testing per ASTM F1717-14 Static torsion testing per ASTM F1717-14                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Conclusions                                        | The ACP 1TM System has identical indications, technological characteristics, and<br>principles of operation as its predicates. The risk analysis performed demonstrates<br>that any minor differences do not impact device performance as compared to the<br>predicates. The ACP 1TM System was shown to be substantially equivalent to its<br>predicate devices.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |

{4}------------------------------------------------

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/KWQ/K150700](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/KWQ/K150700)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
