DIAMOND RATTLESNAKE SYSTEM

K133194 · Eminent Spine · KWQ · Jun 26, 2014 · Orthopedic

Device Facts

Record IDK133194
Device NameDIAMOND RATTLESNAKE SYSTEM
ApplicantEminent Spine
Product CodeKWQ · Orthopedic
Decision DateJun 26, 2014
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3060
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Eminent Spine Black Diamond Rattlesnake Lumbar Plating System is indicated for use via a lateral / anterolateral surgical approach above the great vessels in the treatment of the thoracolumbar spine (T1-L5). The Eminent Spine Red Diamond Rattlesnake Lumbar Plating System is indicated for use via an anterior surgical approach below the bifurcation of the great vessels in the treatment of the lumbosacral (L1-S1) spine. The Diamond Rattlesnake System is intended to provide temporary stabilization as an adjunct to fusion using a lateral approach in skeletally mature patients for the treatment of the following acute and chronic instabilities and deformities: degenerative disc disease (DDD, defined as back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies), pseudarthrosis, spondylolysis, spondylolisthesis, deformity (scoliosis, kyphosis, and or lordosis), spinal stenosis, fracture (including dislocation and subluxation), tumor, or failed previous spine surgery.

Device Story

System consists of titanium alloy plates and non-locking screws; provides temporary spinal stabilization as adjunct to fusion. Surgeon selects implant size based on patient anatomy; utilizes standard orthopedic instruments for placement. Used in surgical settings (OR) by orthopedic or neurosurgeons. Provides mechanical fixation to stabilize spinal segments; facilitates fusion process; benefits patients by addressing acute/chronic instabilities, deformities, or degenerative conditions.

Clinical Evidence

Bench testing only. Static and dynamic axial compression and static torsion testing performed per ASTM F1717. Results demonstrate performance meets or exceeds predicate devices.

Technological Characteristics

Materials: Titanium alloy 6AL-4V-ELI (ASTM F-136). Components: Plates (single level 21-33mm, two level 64-76mm) and non-locking screws (5.5mm diameter, 25-50mm length). Mechanical fixation via standard orthopedic instrumentation.

Indications for Use

Indicated for skeletally mature patients requiring temporary stabilization as an adjunct to fusion in the thoracolumbar (T1-L5) or lumbosacral (L1-S1) spine. Treats DDD, pseudarthrosis, spondylolysis, spondylolisthesis, deformity (scoliosis, kyphosis, lordosis), spinal stenosis, fracture, tumor, or failed previous surgery.

Regulatory Classification

Identification

A spinal intervertebral body fixation orthosis is a device intended to be implanted made of titanium. It consists of various vertebral plates that are punched into each of a series of vertebral bodies. An eye-type screw is inserted in a hole in the center of each of the plates. A braided cable is threaded through each eye-type screw. The cable is tightened with a tension device and it is fastened or crimped at each eye-type screw. The device is used to apply force to a series of vertebrae to correct “sway back,” scoliosis (lateral curvature of the spine), or other conditions.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ 2014 2 6 2014 In accordance with 21 CFR 807.87 (h) and 21 CRF 807.92, the 510(k) summary for the Eminent Spine Black / Red Diamond Rattlesnake Lumbar Plating System is provided below. | Date Summary Prepared | June 11, 2014 | |-------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Manufacturer/Distributor/Spons or | Eminent Spine, LLC<br>7200 N 1-35 Building #1<br>Georgetown, TX 46037<br>Phone 512-868-5980<br>Fax 512-864-1462 | | 510(k) Contact | Secure BioMed Evaluations<br>Linda Braddon, Ph.D.<br>7828 Hickory Flat Highway<br>Suite 120<br>Woodstock, GA 30188<br>770-837-2681 (direct)<br>855-MED-DEV1 (office)<br>LGB@SecureBME.com | | Trade Name | Black / Red Diamond Rattlesnake Lumbar Plating System | | Common Name | Spinal intervertebral body fixation orthosis | | Code -Classification | KWQ 21 CFR 888.3060: Class II | | Predicate Devices | Medtronic Z-Plate (K922543)<br>Acromed M-2 Plate (K972718)<br>Nuvasive Lateral Plate System (K061789, K082070, K091071) | | Predicate Device - 2 Hole<br>Lateral Plate | Nuvasive Lateral Plate System K091071 | | Predicate - Anterior Plate | Medtronic Z-Plate (K922543)<br>Acromed M-2 Plate (K972718) | | Device Description | The Eminent Spine Black / Red Diamond Rattlesnake Lumbar Plating System<br>consists of a variety of plates and screws. All components are fabricated<br>and manufactured from titanium alloy 6AL-4V-ELI per ASTM F-136.<br><br>The plates are manufactured in a variety of configurations with lengths of<br>21 to 33mm in a single level plate and 64 to 76mm in a two level plate.<br>Additionally, the non-locking screws are provided in diameters of 5.5mm<br>with lengths of 25 - 50mm. The responsible surgeon will determine the<br>correct size of the implant in accordance with the size of the individual<br>patient.<br><br>The Eminent Spine Black / Red Diamond Rattlesnake Lumbar Plating System<br>utilizes a variety of standard orthopedic instruments to assist in the<br>placement of the devices. | | Intended Use | The Eminent Spine Black Diamond Rattlesnake Lumbar Plating System is indicated<br>for use via a lateral / anterolateral surgical approach above the great vessels in the<br>treatment of the thoracolumbar spine (T1-L5). The Eminent Spine Red Diamond<br>Rattlesnake Lumbar Plating System is indicated for use via an anterior surgical<br>approach below the bifurcation of the great vessels in the treatment of the<br>lumbosacral (L1-S1) spine. The Diamond Rattlesnake System is intended to provide<br>temporary stabilization as an adjunct to fusion using a lateral approach in skeletally<br>mature patients for the treatment of the following acute and chronic instabilities<br>and deformities: degenerative disc disease (DDD, defined as back pain of discogenic<br>origin with degeneration of the disc confirmed by patient history and radiographic<br>studies), pseudarthrosis, spondylolysis, spondylolisthesis, deformity (scoliosis,<br>kyphosis, and or lordosis), spinal stenosis, fracture (including dislocation and<br>subluxation), tumor, or failed previous spine surgery. | | Technological Characteristics | As was established in this submission, the subject Eminent Spine Black / Red<br>Diamond Rattlesnake Lumbar Plating System is substantially equivalent to<br>the predicate systems in terms of design, intended use, material<br>composition, function, and range of sizes. | | Non-Clinical Performance<br>Testing Conclusion | Nonclinical testing was performed to demonstrate that the Eminent Spine<br>Black / Red Diamond Rattlesnake Lumbar Plating System is substantially<br>equivalent to other predicate devices. The following testing was performed:<br>Static and dynamic axial compression testing per ASTM F1717 Static torsion testing per ASTM F 1717 The results of these studies show that the subject Eminent Spine Black / Red<br>Diamond Rattlesnake Lumbar Plating System meets or exceeds the<br>performance of the predicate device, and the device was therefore found to be substantially equivalent. | | Substantial Equivalence<br>Summary (Conclusion) | The information presented demonstrates the substantial equivalency of the<br>Eminent Spine Black / Red Diamond Rattlesnake Lumbar Plating System. | {1}------------------------------------------------ . : . : . ) {2}------------------------------------------------ Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three tail feathers, representing the three levels of government: federal, state, and local. The eagle is enclosed in a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. #### DEPARTMENT OF HEALTH & HUMAN SERVICES Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 June 26, 2014 Eminent Spine, LLC % Linda Braddon, Ph.D. Secure BioMed Evaluations 7828 Hickory Flat Highway, Suite 120 Woodstock, Georgia 30188 Re: K133194 Trade/Device Name: Black / Red Diamond Rattlesnake Lumbar Plating System Regulation Number: 21 CFR 888.3060 Regulation Name: Spinal intervertebral body fixation orthosis Regulatory Class: Class II Product Code: KWQ Dated: May 22, 2014 Received: May 28, 2014 Dear Dr. Braddon: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set {3}------------------------------------------------ ## Page 2 - Linda Braddon, Ph.D. forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please . contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please-note the requlation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safetv/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours. # Ronald P. Jean -S for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ #### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration # Indications for Use 510(k) Number (if known) K133194 #### Device Name Black / Red Diamond Rattlesnake Lumbar Plating System #### Indications for Use (Describe) The Eminent Spine Black Diamond Rattlesnake Lumbar Plating System is indicated for use via a lateral / anterolateral surgical approach above the great vessels in the thoracolumbar spine (TI-LS). The Eminent Spine Red Diamond Rattlesnake Lumbar Plating System is indicated for use via an anterior surgical approach below the great vessels in the treatment of the lumbosacral (LI-SI) spine. The Diamond Rattlesnake System is intended to provide temporary stabilization as an adjunct to fusjon using a lateral approach in skeletally mature patients for the following acute and chronic instabilities and deformities: degenerative disc disease (DDD, defined as back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies), poeudarthrosis, spondylolisthesis, deformity (scollosis, kyphosis, and or lordosis), spinal stenosis, fracture (including dislocation and subluxation), tumor, or failed previous spine surgery. Type of Use (Select one or both, as applicable) > Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) # PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON A SEPARATE PAGE IF NEEDED. FOR FDA USE ONLY Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) .. . . . **A. W. FABER** **Division of Orthopedic Devices** : Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last page. {5}------------------------------------------------ This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." ้า
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...