RELIANCE ANTERIOR CERVICAL PLATE SYSTEM

K122216 · Reliance Medical Systems, LLC · KWQ · Sep 5, 2012 · Orthopedic

Device Facts

Record IDK122216
Device NameRELIANCE ANTERIOR CERVICAL PLATE SYSTEM
ApplicantReliance Medical Systems, LLC
Product CodeKWQ · Orthopedic
Decision DateSep 5, 2012
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3060
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Reliance Anterior Cervical Plate System is indicated for stabilization of the anterior cervical spine from C2 to C7 employing unicortical screw fixation at the anterior face of the vertebral bodies. Specific clinical indications for anterior plating include: - instability caused by trauma or fracture; - instability associated with correction of cervical lordosis and kyphosis deformity; - instability associated with pseudoarthosis as a result of previously failed cervical spine surgery; - instability associated with major reconstructive surgery for primary tumors or metastatic malignant tumors of the cervical spine; - instability associated with single or multiple level corpectomy in advanced degenerative disk disease (defined as discogenic pain with degeneration of the disc confirmed by history and radiographic studies), spinal canal stenosis and cervical myelopathy. This device is not approved for screw attachment or fixation to the posterior elements (pedicles) of the cervical, thoracic, or lumbar spine.

Device Story

Reliance Anterior Cervical Plate System is a spinal intervertebral body fixation orthosis. System components include plates, cover plates, set screws, and bone screws. Used by surgeons in clinical settings to stabilize the anterior cervical spine (C2-C7) via unicortical screw fixation. Device provides mechanical support to vertebral bodies to address instability from trauma, tumors, or degenerative conditions. Surgeon selects and implants components to achieve fixation; device remains in situ to facilitate stabilization. No software or electronic components.

Clinical Evidence

Bench testing only. Mechanical testing performed per ASTM F-1717, including static compression bending, static torsion, and dynamic compression bending. No clinical data provided.

Technological Characteristics

Titanium alloy (ASTM F-136). System includes plate, cover plate, set screw, and bone screw components. Mechanical fixation via unicortical screws. No software, electronics, or energy sources.

Indications for Use

Indicated for stabilization of anterior cervical spine (C2-C7) in patients requiring unicortical screw fixation for instability due to trauma, fracture, deformity (lordosis/kyphosis), pseudoarthrosis from failed surgery, tumors, or advanced degenerative disc disease (with stenosis/myelopathy). Contraindicated for posterior element (pedicle) fixation.

Regulatory Classification

Identification

A spinal intervertebral body fixation orthosis is a device intended to be implanted made of titanium. It consists of various vertebral plates that are punched into each of a series of vertebral bodies. An eye-type screw is inserted in a hole in the center of each of the plates. A braided cable is threaded through each eye-type screw. The cable is tightened with a tension device and it is fastened or crimped at each eye-type screw. The device is used to apply force to a series of vertebrae to correct “sway back,” scoliosis (lateral curvature of the spine), or other conditions.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K122216 Page 115 SEP 5 2012 ### 510(k) Summary 21 August 2012 Reliance Medical Systems, LLC 545 West 500 South. Suite 100 Bountiful, UT 84010 Telephone: 801-295-3280 Fax: 801-294-0079 Contact: Bret M. Berry Member-Manager 510(k) Number: Common or Usual Name: Proposed Proprietary or Trade Name: Classification Name: Regulation Number: Product Code: Anterior Cervical Plate Reliance Anterior Cervical Plate System Spinal Intervertebral Body Fixation Orthosis 21 CFR 888.3060 KWQ . ### Substantial Equivalence The Reliance Anterior Cervical Plate is substantially equivalent to the legally marketed Synthes Anterior Cervical Vertebral Plate (K030866, K945700), the Interpore Cross C-Tek (K042798), the Altiva cerView ACP (K070891), the Blackstone Hallmark Anterior Cervical Plate System (K100614), Pioneer PAC Anterior Cervical Plate (K083663), X-Spine Spider Plate (K052292), and the Aesculap Caspar Plate (K953720). The Reliance Anterior Cervical Plate is equivalent to these commercially available devices in terms of material, intended use, levels of attachment, size range, and strength. ### Device Description The Reliance Anterior Cervical Plate System device is intended to be used as an anterior cervical plate device. The Reliance Anterior Cervical Plate System is comprised of implant and instrument components. The Reliance Anterior Cervical Plate implant device is manufactured from Titanium alloy as specified by ASTM F-136. The Reliance Anterior Cervical Plate is a combination of the plate, cover plate, set screw, and bone screw components. The cover plate is attached to the plate by means of the set screw. ### Intended Use/Indications for Use The Reliance Anterior Cervical Plate System is indicated for stabilization of the anterior cervical spine from C2 to C7 employing unicortical screw fixation at the anterior face of the vertebral bodies. Specific clinical indications for anterior plating include: {1}------------------------------------------------ K122216 Page 2/7 - instability caused by trauma or fracture; . - instability associated with correction of cervical lordosis and kyphosis deformity; ● - . instability associated with pseudoarthosis as a result of previously failed cervical spine surgery; - instability associated with major reconstructive surgery for primary tumors or metastatic . malignant tumors of the cervical spine; - instability associated with single or multiple level corpectomy in advanced degenerative . · disk disease (defined as discogenic pain with degeneration of the disc confirmed by history and radiographic studies), spinal canal stenosis and cervical myelopathy. This device is not approved for screw attachment or fixation to the posterior elements (pedicles) of the cervical, thoracic, or lumbar spine. ## Performance Data and Substantial Equivalence Mechanical testing was performed on the Reliance Anterior Cervical Plate System following ASTM F-1717 (static compression bending, static torsion, and dynamic compression bending). The Reliance Anterior Cervical Plate System was found to be substantially equivalent to the predicate devices. Additionally, the Reliance Cervical Plate System is substantially equivalent to the predicate devices in terms of sterilization and biocompatibility. {2}------------------------------------------------ # DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized symbol resembling an eagle or bird in flight, composed of three curved lines. To the left of the symbol, there is a circular arrangement of text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES USA". Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 Reliance Medical Systems, LLC % Mr. Bret M. Berry Member-Manager 545 West 500 South, Suite 100 Bountiful, Utah 84010 SEP 5 2012 Re: K122216 Trade/Device Name: Reliance Anterior Cervical Plate System Regulation Number: 21 CFR 888.3060 Regulation Name: Spinal intervertebral body fixation orthosis Regulatory Class: Class II Product Code: KWQ Dated: July 24, 2012 Received: July 25, 2012 Dear Mr. Berry: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set {3}------------------------------------------------ # Page 2 - Mr. Bret M. Berry forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, Mark N. Molkerson Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ### Indications for Use ### 510(k) Number (if known): K122216 Device Name: Reliance Anterior Cervical Plate System Indications for Use: The Reliance Anterior Cervical Plate System is indicated for stabilization of the anterior cervical spine from C2 to C7 employing unicortical screw fixation at the anterior face of the vertebral bodies. Specific clinical indications for anterior plating include: - . instability caused by trauma or fracture; - instability associated with correction of cervical lordosis and kyphosis deformity; . - instability associated with pseudoarthosis as a result of previously failed cervical spine . surgery; - instability associated with major reconstructive surgery for primary turnors or metastatic . malignant tumors of the cervical spine; - instability associated with single or multiple level corpectomy in advanced degenerative . disk disease (defined as discogenic pain with degeneration of the disc confirmed by history and radiographic studies), spinal canal stenosis and cervical myelopathy. This device is not approved for screw attachment or fixation to the posterior elements (pedicles) of the cervical, thoracic, or lumbar spine. Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use _ (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) # Concurrence of CDRH, Office of Device Evaluation (ODE) Carol-Rti- (Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices 510(k) Number K122216 Page 1 of 1
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...