← Product Code [KWQ](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/KWQ) · K112748

# SLIMPLICITY SOLO ANTERIOR CERVICAL PLATE SYSTEM (K112748)

_Spinal USA, LLC · KWQ · Jul 11, 2012 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/KWQ/K112748

## Device Facts

- **Applicant:** Spinal USA, LLC
- **Product Code:** [KWQ](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/KWQ.md)
- **Decision Date:** Jul 11, 2012
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3060
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

The Slimplicity Solo Anterior Cervical Plate System is intended for anterior cervical fixation to the cervical spine C2-C7 for the following indications: degenerative disc disease (as defined by neck pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies), trauma (including fractures), tumors, deformity (defined as kyphosis, lordosis, or scoliosis), pseudarthrosis, failed previous fusion, spondylolisthesis, and spinal stenosis.

## Device Story

System consists of titanium alloy (ASTM F136) plates and bone screws; used for anterior cervical fixation (C2-C7). Implants attached to anterior vertebral body to provide stabilization. Used by surgeons in clinical settings. Mechanical performance validated via static axial compression, static torsion, and cyclical axial compression testing per ASTM F1717. Provides structural support to facilitate spinal fusion; benefits patients by stabilizing cervical spine segments affected by degenerative disease, trauma, or deformity.

## Clinical Evidence

Bench testing only. No clinical data provided. Mechanical testing performed per ASTM F1717, including static axial compression, static torsion, and cyclical axial compression (run out load at 5,000,000 cycles).

## Technological Characteristics

Titanium alloy (ASTM F136) plates and screws. Mechanical fixation via bone screws. Provided non-sterile. Dimensions and form factor designed to accommodate cervical anatomy (C2-C7).

## Regulatory Identification

A spinal intervertebral body fixation orthosis is a device intended to be implanted made of titanium. It consists of various vertebral plates that are punched into each of a series of vertebral bodies. An eye-type screw is inserted in a hole in the center of each of the plates. A braided cable is threaded through each eye-type screw. The cable is tightened with a tension device and it is fastened or crimped at each eye-type screw. The device is used to apply force to a series of vertebrae to correct “sway back,” scoliosis (lateral curvature of the spine), or other conditions.

## Predicate Devices

- Uniplate Anterior Cervical Plate System ([K042544](/device/K042544.md))
- Synthes Anterior CSLP System ([K030866](/device/K030866.md))
- Synthes Vectra-One System ([K071667](/device/K071667.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the text "K112748." above the logo for Spinal USA. The logo is a black rectangle with the words "SPINAL" and "USA" in white. A graphic of a spine is between the two words. 

JUL 1 1 2012

2050 Executive Drive Pearl, MS 39208

Ph: 601-420-4244 Fax: 601-420-5501

## 510(k) Summary

| Date Summary Prepared: | July 18, 2011                                   |
|------------------------|-------------------------------------------------|
| Submitter Information: | Spinal USA                                      |
|                        | 2050 Executive Drive                            |
|                        | Pearl, MS 39208                                 |
| Contact Name:          | Frankie Cummins                                 |
| Phone:                 | 601-420-4244                                    |
| Fax:                   | 601-420-5501                                    |
| E-mail:                | fc@spinalusa.com                                |
| Device Trade Name:     | Slimplicity Solo Anterior Cervical Plate System |
| Common Name:           | Spinal Intervertebral Body Fixation Orthosis    |
| Regulatory Number:     | 888.3060                                        |
| Classification:        | Class II                                        |
| Product Code:          | KWQ                                             |
| INTENDED USE:          |                                                 |

The Slimplicity Solo Anterior Cervical Plate System is intended for anterior cervical fixation to the cervical spine C2-C7 for the following indications: degenerative disc disease (as defined by neck pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies), trauma (including fractures), tumors, deformity (defined as kyphosis, lordosis, or scoliosis), pseudarthrosis, failed previous fusion, spondylolisthesis, and spinal stenosis.

### DEVICE DESCRIPTION:

The Slimplicity Solo Anterior Cervical Plate System consists of an assortment of plates and screws. Each Slimplicity Solo plate attaches to the anterior portion of the vertebral body of the cervical spine (levels C2-C7). Slimplicity Solo implants are composed of titanium alloy, as specified in ASTM F136. The components will be provided non-sterile.

{1}------------------------------------------------

KII2748

#### MECHANICAL TESTING:

The Slimplicity Solo Anterior Cervical Plate System was mechanically tested per ASTM F1717. The objective of this test battery is to mechanically test the Slimplicity Solo Anterior Cervical Plate System using the following test methods:

Static Axial Compression Bending testing in a load to failure mode

Static Torsion testing in a load to failure mode

Cyclical axial compression testing to estimate the maximum run out load value at 5,000,000 cycles

#### Mechanical Test Conclusion:

The Slimplicity Solo Anterior Cervical Plate System is as safe as the predicates based on the following:

- 1) The design of the Slimplicity Solo Anterior Cervical Plate System is similar to the predicates in thickness, width, and length.
- 2) The Slimplicity Solo Anterior Cervical Plate System bone screw is similar to the predicates in type, size, and length,
- 3) The material used in the Slimplicity Solo Anterior Cervical Plate System is the same as used in the predicate systems.
- 4) The mechanical strength of the Slimplicity Solo Anterior Cervical Plate System is substantially equivalent to the predicate based on mechanical test results. The Slimplicity Solo Anterior Cervical Plate System test results are provided in section 18 of this 510(k) submittal. Mechanical strength substantial equivalence to predicates is shown in section 12.

The Slimplicity Solo Anterior Cervical Plate System is as effective as the predicate based on the following:

- 1) The intended use of the Slimplicity Solo Anterior Cervical Plate System is the same as the predicates.
- 2) The indication for use of the Slimplicity Solo Anterior Cervical Plate System is the same as the predicates.

The Slimplicity Solo Anterior Cervical Plate System is as safe and effective as its predicates.

#### EQUIVALENT DEVICE:

Documentation was provided which demonstrated the Slimplicity Solo Anterior Cervical Plate System to be substantially equivalent to previously cleared devices. The substantial equivalence is based upon equivalence in intended use, indications, anatomic sites, performance and material of manufacture.

The overall design of the Slimplicity Solo Anterior Cervical Plate System consists of plates and bone screws. The components are available in a variety of lengths in order to accommodate patient anatomy and are substantially equivalent to the predicate devices.

Components of the Slimplicity Solo Anterior Cervical Plate System are similar to the predicate systems. The size and shape of the Slimplicity Solo Anterior Cervical Plate System and predicate systems are similar. The Slimplicity Solo Anterior Cervical Plate System bone screw thread type, size and length are similar to the predicates. Material of both the plate and screws is the same as the predicates. Sterilization is the same as the predicates.

#### Predicate Devices:

Uniplate Anterior Cervical Plate System K042544 Synthes Anterior CSLP System K030866 Synthes Vectra-One System K071667

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread, depicted with three thick, curved lines forming the body and wings. The eagle is facing towards the right side of the image.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Spinal USA, LLC % Mr. Frankie Cummins Engineering Manager 2050 Executive Drive Pearl, Mississippi 39208

JUL 1 1 2012

Re: K112748

Trade/Device Name: Slimplicity Solo Anterior Cervical Plate System Regulation Number: 21 CFR 888.3060 Regulation Name: Spinal intervertebral body fixation orthosis Regulatory Class: Class II Product Code: KWQ Dated: July 05, 2012 Received: July 05, 2012

Dear Mr. Cummins:

We have reviewed your Section 510(k) premarket notification of intent to market the device we nave rollowed your your be device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The I ou may, therefore, mains of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it your device to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean 1 toase be devised that I hat your device complies with other requirements of the Act than - Diederal statutes and regulations administered by other Federal agencies. You must or uny I vith all the Act's requirements, including, but not limited to: registration and listing (21

{3}------------------------------------------------

## Page 2 – Mr. Frankie Cummins

CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical CI I all 007); doomig (21 OFF 803); good manufacturing practice requirements as set Gorth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please rr you doo!!e bpv.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part note the regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Tou may of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

for
Mark N. Melkerson

Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

K112748

# Indications for Use

510(k) Number (if known): __ Ku2748

Device Name: Slimplicity Solo Anterior Cervical Plate System

Indications for Use:

The Slimplicity Solo Anterior Cervical Plate System is intended for anterior cervical fixation to the cervical spine C2-C7 for the following indications: degenerative disc disease (as defined by neck pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies), trauma (including fractures), tumors, deformity (defined as kyphosis, lordosis, or scoliosis), pseudarthrosis, failed previous fusion, spondylolisthesis, and spinal stenosis.

Prescription Use_ X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off) . Division of Surgical. Orthopedic, and Restorative Devices

KII2748 510(k) Number

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/KWQ/K112748](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/KWQ/K112748)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
