ACP4 SPINAL IMPLANT SYSTEM

K092360 · X-Spine Systems, Inc. · KWQ · Jan 22, 2010 · Orthopedic

Device Facts

Record IDK092360
Device NameACP4 SPINAL IMPLANT SYSTEM
ApplicantX-Spine Systems, Inc.
Product CodeKWQ · Orthopedic
Decision DateJan 22, 2010
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3060
Device ClassClass 2
AttributesTherapeutic

Intended Use

The X-Spine ACP4 System is intended for anterior screw fixation to the cervical spine (C2-C7 inclusive). The system is indicated for use in the temporary stabilization of the anterior spine during the development of cervical spinal fusion in patients with: - Degenerative disc disease (as defined by neck pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies), - Spondylolisthesis, - Trauma (i.e., fractures or dislocations), - Tumors, - Deformity (defined as kyphosis, lordosis, or scoliosis) . - Pseudoarthrosis, and/or - Failed previous fusions.

Device Story

ACP4 Spinal Implant System consists of titanium alloy anterior cervical plates and bone screws; designed for anterior screw fixation to cervical spine (C2-C7). Device provides temporary stabilization of anterior spine during cervical spinal fusion. Used by surgeons in clinical settings. Implant components provided clean and non-sterile. Biomechanical testing confirms performance for intended use.

Clinical Evidence

Bench testing only; no clinical data provided.

Technological Characteristics

Titanium alloy anterior cervical plates and bone screws. Designed for anterior fixation to cervical spine (C2-C7). Provided clean and non-sterile.

Indications for Use

Indicated for patients requiring temporary anterior cervical spine stabilization (C2-C7) during fusion. Applicable for degenerative disc disease, spondylolisthesis, trauma (fractures/dislocations), tumors, deformity (kyphosis, lordosis, scoliosis), pseudoarthrosis, and failed previous fusions.

Regulatory Classification

Identification

A spinal intervertebral body fixation orthosis is a device intended to be implanted made of titanium. It consists of various vertebral plates that are punched into each of a series of vertebral bodies. An eye-type screw is inserted in a hole in the center of each of the plates. A braided cable is threaded through each eye-type screw. The cable is tightened with a tension device and it is fastened or crimped at each eye-type screw. The device is used to apply force to a series of vertebrae to correct “sway back,” scoliosis (lateral curvature of the spine), or other conditions.

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K092360 Page 1/2 # JAN 22 2010 ## 510(k) Summary ## ADMINISTRATIVE INFORMATION Manufacturer Name: X-spine Systems, Inc. 452 Alexandersville Rd. Miamisburg, OH 45342 Telephone (937) 847-8400 FAX (937) 847-8410 Official Contact: David Kirschman, MD Chief Medical Officer # DEVICE NAME Trade/Proprietary Name: ACP4 ™ Spinal Implant System Spinal Intervertebral Body Fixation Orthosis Classification Name: Common Name: Cervical plating system # ESTABLISHMENT REGISTRATION NUMBER X-spine Systems, Inc. has submitted an Establishment Registration to FDA. The Establishment Registration number is 3005031160. The owner/operator number for X spine Systems; Inc. is 9063903. #### DEVICE CLASSIFICATION FDA has classified anterior cervical plating devices as a Class II (21 CFR 888.3060). The product code for Spinal Intervertebral Body Fixation Orthosis is KWQ. These device classifications are reviewed by the Orthopedic Devices Branch. {1}------------------------------------------------ K092360 Page 2/2 ### INTENDED USE ## Indications for Use: The X-Spine ACP4 System is intended for anterior screw fixation to the cervical spine (C2-C7 inclusive). The system is indicated for use in the temporary stabilization of the anterior spine during the development of cervical spinal fusion in patients with: - Degenerative disc disease (as defined by neck pain of discogenic origin with ● degeneration of the disc confirmed by patient history and radiographic studies), - Spondylolisthesis, � - Trauma (i.e., fractures or dislocations), ● - � Tumors, - Deformity (defined as kyphosis, lordosis, or scoliosis) . - Pseudoarthrosis, and/or - Failed previous fusions. ## DEVICE DESCRIPTION The X-Spine ACP4 System includes titanium alloy anterior cervical plates and bone screws. The plate attaches to the anterior portion of the vertebral body of the cervical spine, levels C2 to C7. The implant components are provided clean and non-sterile. #### EQUIVALENCE TO MARKETED PRODUCT X-spine Systems, Inc. has submitted information to demonstrate that, for the purposes of FDA's regulation of medical devices, the ACP4 Spinal implant System is substantially equivalent in indications and design principles to predicate devices. #### PERFORMANCE DATA Biomechanical testing indicates that the ACP4 system is capable of performing in accordance with its intended use. Pg. 2 of 2 {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the seal is an emblem of an eagle with its wings spread, symbolizing protection and service. The eagle is stylized and abstract, with flowing lines representing its feathers. #### Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-O66-0609 Silver Spring, MD 20993-0002 # JAN 22 2010 X-Spine Systems, Inc. % David Kirschman, M.D. Chief Medical Officer 452 Alexandersville Road Miamisburg, Ohio 45342 Re: K092360 Trade/Device Name: ACP4" Spinal Implant System Regulation Number: 21 CFR 888.3060 Regulation Name: Spinal intervertebral body fixation orthosis Regulatory Class: Class II Product Code: KWQ Dated: January 15, 2010 Received: January 19, 2010 Dear Dr. Kirschman: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical {3}------------------------------------------------ Page 2 - David Kirschman, M.D. device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Page 2 - Mr. Dave Kirschman If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours. Calvare Bneelup Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ Indications for Use 510(k) Number (if known): K092360 Device Name: ACP4™ Spinal Implant System #### Indications for Use: The X-Spine ACP4 System is intended for anterior screw fixation to the cervical spine (C2-C7 inclusive). The system is indicated for use in the temporary stabilization of the anterior spine during the development of cervical spinal fusion in patients with: - Degenerative disc disease (as defined by neck pain of discogenic origin with ♣ degeneration of the disc confirmed by patient history and radiographic studies), - Spondylolisthesis, � - Trauma (i.e., fractures or disfocations), . AND/OR - Tumors, . - Deformity (defined as kyphosis, lordosis, or scoliosis) - Pseudoarthrosis, and/or - Failed previous fusions. Prescription Use X (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) # (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CORH, Office of Device Evaluation (ODE) . (Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices 510(k) Number_________________________________________________________________________________________________________________________________________________________________ KO92360 12360
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...