← Product Code [KWQ](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/KWQ) · K062110

# VERTEBRON SCP CERVICAL PLATE SYSTEM (K062110)

_Vertebron, Inc. · KWQ · Aug 16, 2006 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/KWQ/K062110

## Device Facts

- **Applicant:** Vertebron, Inc.
- **Product Code:** [KWQ](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/KWQ.md)
- **Decision Date:** Aug 16, 2006
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 888.3060
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

The modified VERTEBRON SSP™ Cervical Plate System is intended for non pedicle fixation for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed bv history and radiographic studies): spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion in skeletally mature patients. The VERTEBRON SSP™ Cervical Plate System is intended for non cervical pedicle fixation for for the following indications: spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion in skeletally mature patients.

## Device Story

System comprises non-sterile, single-use titanium alloy plates and self-tapping/self-drilling screws; used for anterior cervical spine stabilization during bone graft fusion. Operative site stabilization achieved via mechanical fixation to vertebral bodies. Intended for use by surgeons in clinical settings. Output is physical stabilization of spinal segments. Benefits include support for fusion process in patients with degenerative, traumatic, or neoplastic spinal conditions. Submission introduces reduced stature plate design.

## Clinical Evidence

Bench testing only. Mechanical testing performed in accordance with ASTM F1717 to demonstrate substantial equivalence.

## Technological Characteristics

Titanium alloy components conforming to ASTM F136. System includes plates and self-tapping/self-drilling screws. Mechanical fixation device. Non-sterile, single-use.

## Regulatory Identification

A spinal intervertebral body fixation orthosis is a device intended to be implanted made of titanium. It consists of various vertebral plates that are punched into each of a series of vertebral bodies. An eye-type screw is inserted in a hole in the center of each of the plates. A braided cable is threaded through each eye-type screw. The cable is tightened with a tension device and it is fastened or crimped at each eye-type screw. The device is used to apply force to a series of vertebrae to correct “sway back,” scoliosis (lateral curvature of the spine), or other conditions.

## Predicate Devices

- VERTEBRON SSP™ Cervical Plate System ([K051815](/device/K051815.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

VIII. Summary of Safety and Effectiveness

| SUBMITTER: | VERTEBRON Inc.<br>400 Long Beach Blvd.<br>Stratford, CT 06615<br>(203) 380-9340 | AUG 16 2006 |
|------------|---------------------------------------------------------------------------------|-------------|
|------------|---------------------------------------------------------------------------------|-------------|

CONTACT PERSON: Luis Nesprido Regulatory Affairs, Manager

July 21, 2006 DATE PREPARED:

CLASSIFICATION NAME: Spinal Intervertebral Body Fixation Orthosis

COMMON NAME: Cervical Plating System

PROPRIETARY NAME: VERTEBRON SSP™ Cervical Plate System

PREDICATE DEVICES: VERTEBRON SSP™ Cervical Plate System - K051815

DEVICE DESCRIPTION:

The modified VERTEBRON SSP™ Cervical Plate System is comprised of non-sterile, single-use, titanium alloy components. The modified VERTEBRON SSP™ Cervical Plate System attaches to the vertebral body by means of self tapping and self drilling cervical screws. This system's design is intended to stabilize the spinal operative site during the fusion process of a bone graft in the disc space. This submission adds a reduced stature cervical plate.

INTENDED USE: The modified VERTEBRON SSP™ Cervical Plate System is intended for non pedicle fixation for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed bv history and radiographic studies): spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion in skeletally mature patients. The VERTEBRON SSP™ Cervical Plate System is intended for non cervical pedicle fixation for for the following indications: spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion in skeletally mature patients.

MATERIALS: The material used is titanium Alloy material that conforms to ASTM F136.

{1}------------------------------------------------

Substantial Equivalence: Testing in accordance with ASTM F1717 was performed and demonstrated that the modified VERTEBRON SSP™ Cervical Plate System is substantially equivalent to the VERTEBRON SSP™ Cervical Plate System (K051815) which have received market clearance by the FDA.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features the department's name in a circular arrangement around a stylized emblem. The emblem consists of three abstract human figures, possibly representing health, hope, and well-being.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

## AUG 1 6 2006

Vertebron, Inc. % Mr. Luis Nesprido Manager, Regulatory Affairs 400 Long Beach Boulevard Stratford, Connecticut 06615

Re: K062110

Trade/Device Name: VERTEBRON SSP™ Cervical Plate System Regulation Number: 21 CFR 888.3060 Regulation Name: Spinal intervertebral body fixation orthosis Regulatory Class: II Product Code: KWQ Dated: July 21, 2006 Received: July 24, 2006

Dear Mr. Nesprido:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA

{3}------------------------------------------------

Page 2 - Mr. Luis Nesprido

finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtains, other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index html .

Sincerely vours.

for P.J. Pma

Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

## V. Indications for Use

510(k) Number (if known):

Device Name:

VERTEBRON SSP™ Cervical Plate System

Indications For Use:

The Modified VERTEBRON SSP™ Cervical Plate System is intended for anterior interbody fixation of the cervical spine. The system is indicated for use in the temporary stabilization of the anterior spine during the development of cervical soins fue tempor patients with degenerative disc disease (as defined by neck pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic shudies), trauma (including fractures), tumors, deformity (defined as kyphosis, lordnois, ar scoliosis), pseudoarthriosis, and/or failed previous fusions. The SSP™ Cervical Plate System can be implanted in the sub-axial cervical spine from C3 through C7 levels.

Prescription Use: X (Per 21 CFR 801 Subpart D)

AND / OR

Over-The-Counter Use: (21 CFR 801 Subpart C)

(Please do not write below this line - continue on another page if needed)

Concurrence of CDRH, Office of Device Evaluation (ODE)

P. Mthme

(Division Sign-O Division of General, Restorative, and Neurological Devices

Kob2110 510(k) Number

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/KWQ/K062110](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/KWQ/K062110)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
