The LTD Polyaxial Fixation System is intended to help provide immobilization and stabilization of spinal segments as an adjunct to fusion of the lumbar and/or sacral spine, specifically as follows: When used as a pedicle screw fixation system of the non-cervical posterior spine in skeletally mature patients, the LTD Polyaxial Fixation System is indicated for one or more of the following: (1) degenerative spondylolisthesis with objective evidence of neurologic impairment, (2) fracture, (3) dislocation, (4) spinal tumor, and/or (5) failed previous fusion (pseudarthrosis). In addition, when used as a pedicle screw fixation system, the LTD Polyaxial Fixation System is indicated for skeletally mature patients: (a) having severe spondylolisthesis (Grades 3 and 4) of the fifth lumbar-first sacral (L5-S1) vertebral joint; (b) who are receiving fusions using autogenous bone graft only; (c) who are having the device fixed or attached to the lumbar and sacral spine (L3 and below); and (d) who are having the device removed after the development of a solid fusion mass. When used as a posterior, non-cervical, non-pedicle screw fixation system, the LTD Polyaxial Fixation System is intended for the following indications: (1) degenerative disc disease (as defined by back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies), (2) spinal stenosis, (3) spondylolisthesis, (4) fracture, (5) pseudarthrosis, (6) tumor resection, and/or (7) failed previous fusion.
Device Story
LTD Polyaxial Fixation System comprises titanium rods, screws, and connecting components; used for spinal immobilization and stabilization as an adjunct to fusion. Components are surgically implanted by physicians in the lumbar and sacral spine. System features top-loading polyaxial screws with >40° variability; screws use threaded set screws to lock rods. Construct configurations are tailor-made for individual patient cases. Device provides mechanical support to stabilize spinal segments, facilitating fusion and potentially reducing pain or neurologic impairment associated with spinal conditions.
Clinical Evidence
Bench testing only. Testing performed per ASTM F1717-01, including static axial load-to-failure, static torsion, and cyclical axial compression bending (5.0 x 10^6 cycles). Results confirm compliance with voluntary performance standards and demonstrate equivalent static and fatigue strength to predicate devices.
Technological Characteristics
Materials: Medical grade titanium or titanium alloy (ASTM F67, ASTM F136, ISO 5832-3, ISO 5832-2). Design: Top-loading polyaxial screw system with >40° variability; two-piece assembly (screw shank and head body). Locking mechanism: Threaded set screw. Connectivity: N/A. Sterilization: Not specified.
Indications for Use
Indicated for skeletally mature patients requiring spinal immobilization/stabilization as an adjunct to fusion in the lumbar and/or sacral spine. Conditions include degenerative spondylolisthesis with neurologic impairment, fracture, dislocation, spinal tumor, pseudarthrosis, degenerative disc disease, and spinal stenosis. Specific criteria for pedicle screw use include severe L5-S1 spondylolisthesis (Grades 3-4) and fixation at L3 and below using autogenous bone graft.
Regulatory Classification
Identification
A spinal intervertebral body fixation orthosis is a device intended to be implanted made of titanium. It consists of various vertebral plates that are punched into each of a series of vertebral bodies. An eye-type screw is inserted in a hole in the center of each of the plates. A braided cable is threaded through each eye-type screw. The cable is tightened with a tension device and it is fastened or crimped at each eye-type screw. The device is used to apply force to a series of vertebrae to correct “sway back,” scoliosis (lateral curvature of the spine), or other conditions.
Predicate Devices
Moss Miami system
Xia Spine System
Optima Spine System
Submission Summary (Full Text)
{0}------------------------------------------------
# SEP - 8 2004 510(k) Summary
Submitted on behalf of:
U.S. Spinal Technologies, Inc. 3701 FAU Blvd. Ste 210 Boca Raton, FL 33431, USA Telephone: 561-367-7463 Fax: 561-367-9627
by:
Elaine Duncan, M.S.M.E., RAC President, Paladin Medical, Inc. PO Box 560 Stillwater. MN 55082 715-549-6035 715-549-5380
Kot 1225
CONTACT PERSON:
Telephone:
Fax:
DATE PREPARED: June 14, 2004
Elaine Duncan
| TRADE NAME: | LTD Polyaxial Fixation System |
|--------------|-------------------------------|
| COMMON NAME: | spinal fixation system |
#### SUBSTANTIALLY EQUIVALENT TO:
The LTD, the Moss Miami system, the Xia Spine System and the Optima Spine System are all top-loading polyaxial screws with greater than 40° variability and all four systems are made from titanium. All four systems have screws that accept a 6.0mm titanium rod. All four pedicle screws use a threaded Set Screw to lock the rod to the pedicle screw. All four systems are indicated from L5-S1 for 1) Degenerative Disc Disease, 2) Spinal Stenosis, and 3) Spondylolisthesis. [The Moss Miami and the Xia Spine System are also indicated for deformities (kyphosis/scoliosis)]. However, the Moss Miami is assembled in three pieces w screw shank, bushing and head body, where as the LTD and Xia are assembled using two pieces: screw shank and head body.
#### DESCRIPTION of the DEVICE:
The LTD Polyaxial Fixation System consists of a variety of shapes and sizes of rods, screws and connecting components, sold with or without the surgical instrument tray. These components can be rigidly locked into a variety of configurations, with each construct being tailor-made for the individual case. The LTD Polyaxial Fixation System implant components are fabricated from medical grade titanium or titanium alloy described by such standards as ASTM F67 or ASTM F136 or ISO 5832-3 or 5832-2
#### INDICATIONS FOR USE:
The LTD Polyaxial Fixation System is intended to help provide immobilization and stabilization of spinal segments as an adjunct to fusion of the lumbar and/or sacral spine, specifically as follows:
When used as a pedicle screw fixation system of the non-cervical posterior spine in skeletally mature patients, the LTD Polyaxial Fixation System is indicated for one or more of the following: (1) degenerative spondylolisthesis with objective evidence of neurologic impairment, (2) fracture, (3) dislocation, (4) spiral tumor, and/or (5) failed previous fusion (pseudarthrosis).
510(k) Submission
l of i
13
{1}------------------------------------------------
#### 510(k) Summary-Continued
In addition, when used as a pedicle screw fixation system, the LTD Polyaxial Fixation System is indicated for skeletally mature patients: (a) having severe spondylolisthesis (Grades 3 and 4) of the fifth lumbar-first sacral (L5-S1) vertebral joint; (b) who are receiving fusions using autogenous bone graft only; (c) who are having the device fixed or attached to the lumbar and sacral spine (L3 and below); and (d) who are having the device removed after the development of a solid fusion mass.
When used as a posterior, non-cervical, non-pedicle screw fixation system, the LTD Polyaxial Fixation System is intended for the following indications: (1) degenerative disc disease (as defined by bark nin of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies), (2) spinal stenosis, (3) spondylolisthesis, (4) fracture, (5) pseudarthrosis, (6) turnor resection, and/or (7) failed previous fusion.
#### SUMMARY of TESTING:
The LTD Polyaxial Fixation System was tested for: 1. Static axial testing in a load to failure mode in compression bending, 2. Static torsion testing in a load to failure mode in torsion, and 3. Cyclical axial compression bending testimate the maximum run out load value at 5.0 x 106 cycles. The the the protocol followed ASTM Standard F1717-01, "Standard Test Methods for Static and Fatisyer or Spinal Implant Constructs in a Vertebrectomy Model." Results demonstrate that the LTD Polyaxial Fixation System complies with the recognized, voluntary performance standards for spinal implant systems. In addition, test results indicate that the LTD Polyaxial Fixation System is substantially equivalert in staic and fatigue strength to predicate devices previously cleared by FDA.
f 7
{2}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird in flight, composed of three curved lines.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
SEP = 8 2004
U.S. Spinal Technologies, LLC. C/o Ms. Elaine Duncan, M.S.M.E., RAC President Paladin Medical, Inc. PO Box 560 Stillwater, Minnesota 55082
Re: K041625
Trade/Device Name: LTD Polyaxial Fixation System Regulation Number: 21 CFR 888.3050, 21 CFR 888.3060, 21 CFR888.3070 Regulation Name: Spinal interlaminal fixation orthosis, Spinal intervertebral body fixation orthosis, Pedicle screw spinal system Regulatory Class: II Product Code: KWQ, KWP, MNH, MNI Dated: June 14, 2004
Dear Ms. Duncan:
Received: June 15, 2004
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must
{3}------------------------------------------------
## Page 2 -- Ms. Elaine Duncan, M.S.M.E., RAC
comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally premated predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Sincerely, yours,
Mark A. Melkerson
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indications for Use
# 510(k) Number (if known): KO41625
Device Name: LTD Polyaxial Fixation System
### INDICATIONS:
The LTD Polyaxial Fixation System is intended to help provide immobilization of spinal segments as an adjunct to fusion of the lumbar and/or sacral spine, specifically as follows:
When used as a pedicle screw fixation system of the non-cervical posterior spine in skeletally mature patients, the LTD Polyaxial Fixation System is indicated for one or more of the following: (1) degenerative spondylolisthesis with objective evidence of neurologic impairment, (2) fracture, (3) dislocation, (4) spinal tumor, and/or (5) failed previous fusion (pseudarthrosis).
In addition, when used as a pedicle screw fixation system, the LTD Polyaxial Fixation System is indicated for skeletally mature patients: (a) having severe spondylolisthesis (Grades 3 and 4) of the fifth lumbarfirst sacral (L5-S1) vertebral joint; (b) who are receiving fusions using autogenous bone graft only; (c) who are having the device fixed or attached to the lumbar and sacral spine (L3 and below); and (d) who are having the device removed after the development of a solid fusion mass.
When used as a posterior, non-cervical, non-pedicle screw fixation system, the LTD Polyaxial Fixation System is intended for the following indications: (1) degenerative disc disease (as defined by back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiosraphic studies), (2) spinal stenosis, (3) spondylolisthesis, (4) fracture, (5) pseudarthrosis, (6) tumor resection, and/or (7) failed previous fusion.
Prescription Use J J J _______________________________________________________________________________________________________________________________________________________ AND/OR Over-The-Counter Use .........................................................................................................................................................
(Part 21 CFR 801 Subpart D)
(21 CFR 807 Subpart C)
ﻠﺴﻴﺴﺎﺕ
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of General, Restorative,
Page 1 of 1 and Neurological Devices
510(k) Number K04 1625
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.