The Reflex™ ACP System is intended for anterior intervertebral screw fixation of the cervical spine at levels C2-T1. The system is indicated for temporary stabilization of the anterior spine during the development of cervical spine fusions in patients with the following indications: Degenerative Disc Disease (as defined by neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), trauma (including fractures), tumors, deformities or curvatures (including kyphosis, lordosis or scoliosis), pseudoarthrosis, failed previous fusion, decompression of the spinal cord following total or partial cervical vertebrectomy, spondylolisthesis and spinal stenosis. WARNING: This device is not approved for screw attachment to the posterior elements (pedicles) of the cervical, thoracic, or lumbar spine.
Device Story
Reflex™ Anterior Cervical Plate System consists of titanium alloy bone plates and screws; used for temporary anterior intervertebral fixation of cervical spine (C2-T1) during fusion procedures. Components available in various lengths to accommodate patient physiology; provided non-sterile. System implanted by surgeons in clinical/OR settings to stabilize spine following trauma, degenerative disease, or deformity correction. Mechanical fixation provides structural support to facilitate bone fusion; output is physical stabilization of spinal segments. Benefit is restoration of spinal alignment and support during healing process.
Clinical Evidence
Bench testing only. Static and fatigue test results show the Reflex ACP System demonstrated comparable mechanical properties to the predicate device.
Technological Characteristics
Titanium alloy spinal intervertebral body fixation orthosis. System comprises bone plates and screws in various lengths. Non-sterile. Mechanical fixation principle.
Indications for Use
Indicated for patients requiring temporary anterior cervical spine stabilization (C2-T1) during fusion. Conditions include degenerative disc disease, trauma/fractures, tumors, deformities (kyphosis, lordosis, scoliosis), pseudoarthrosis, failed previous fusion, post-vertebrectomy decompression, spondylolisthesis, and spinal stenosis. Contraindicated for posterior element (pedicle) attachment.
Regulatory Classification
Identification
A spinal intervertebral body fixation orthosis is a device intended to be implanted made of titanium. It consists of various vertebral plates that are punched into each of a series of vertebral bodies. An eye-type screw is inserted in a hole in the center of each of the plates. A braided cable is threaded through each eye-type screw. The cable is tightened with a tension device and it is fastened or crimped at each eye-type screw. The device is used to apply force to a series of vertebrae to correct “sway back,” scoliosis (lateral curvature of the spine), or other conditions.
Predicate Devices
Synthes' Cervical Spine Locking Plate (CSLP) System
K020650 — MODIFICATION TO REFLEX ANTERIOR CERVICAL PLATE SYSTEM · Howmedica Osteonics Corp. · Mar 29, 2002
K093776 — INDUS ANTERIOR CERVICAL PLATE SYSTEM · Spinefrontier, Inc. · Mar 26, 2010
K251965 — Proximity Anterior Cervical Plate System; Segmental Plating System (SPS) · Alphatec Spine, Inc. · Aug 28, 2025
Submission Summary (Full Text)
{0}------------------------------------------------
Reflex™ Anterior Cervical Plate System -- Additional Indications
510(k) Premarket Notification
## 510(k) Summary for Reflex" Anterior Cervical Plate System - Additional Indications
| Proprietary Name: | Reflex™ Anterior Cervical Plate System |
|------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name: | Anterior Cervical Plate System |
| Proposed Regulatory Class: | Class II |
| Classification Name and Reference: | Spinal Invertebral Body Fixation Orthosis<br>21 CFR §888.3060 |
| Device Product Code: | 87 KWQ: Appliance, Fixation, Spinal Intervertebral Body |
| For Information contact: | Karen Ariemma<br>Regulatory Affairs Specialist<br>Howmedica Osteonics Corp.<br>59 Route 17<br>Allendale, NJ 07401-1677<br>Telephone: (201) 831-5718<br>Fax: (201) 831-6038<br>Email: kariemma@howost.com |
| Date Summary Prepared: | May 30, 2003 |
The purpose of this premarket notification is to add indications to the Reflex " Anterior Cervical Plate (ACP) System.
### Predicate Device Information:
The Reflex " ACP System consists of bone plates and screws. Both components are available in a variety of lengths in order to accommodate patient physiology. The components are fabricated from titanium alloy. The implants will be provided non-sterile. The following system was already determined substantially equivalent for the desired indications: Synthes' Cervical Spine Locking Plate (CSLP) System.
#### Intended Use:
The Reflex™ ACP System is intended for anterior intervertebral screw fixation of the cervical spine at levels C2-T1. The system is indicated for temporary stabilization of the anterior spine during the development of cervical spine fusions in patients with the following indications: Degenerative Disc Disease (as defined by neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), trauma (including fractures), tumors, deformities or curvatures (including kyphosis, lordosis or scoliosis), pseudoarthrosis, failed previous fusion, decompression of the spinal cord following total or partial cervical vertebrectomy, spondylolisthesis and spinal stenosis.
WARNING: This device is not approved for screw attachment to the posterior elements (pedicles) of the cervical, thoracic, or lumbar spine.
#### Performance Data:
Static and fatigue test results show the Reflex ACP System demonstrated comparable mechanical properties to the predicate device.
{1}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/1/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird in flight, composed of three curved lines.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
AUG - 8 2003
Ms. Karen Ariemma Regulatory Affairs Specialist Howmedica Osteonics Corporation 59 Route 17 Allendale, NJ 07401-1677
Re: K031702
Trade/Device Name: Reflex™ Anterior Cervical Plate System Regulation Number: 21 CFR 888.3060 Regulation Name: Spinal intervertebral body fixation orthosis Regulatory Class: II Product Code: KWO Dated: May 30, 2003 Received: June 2, 2003
Dear Ms. Ariemma:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{2}------------------------------------------------
Page 2 - Ms. Karen Ariemma
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Miriam C. Provost
O Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# 510(k) Number (if known): K ∞ 3 170 x
#### Device Name: Reflex Anterior Cervical Plate System
The Reflex ACP System is intended for anterior intervertebral screw fixation of the cervical spine at levels C2-T1. The system is indicated for temporary stabilization of the anterior spine during the development of cervical spine fusions in patients with the following indications:
- Degenerative Disc Disease (as defined by neck pain of discogenic origin with degeneration of . the disc confirmed by history and radiographic studies)
- Trauma (including fractures) .
- Tumors ●
- Deformities or curvatures (including kyphosis, lordosis or scoliosis) .
- Pseudoarthrosis .
- Failed previous fusion .
- 6 Decompression of the spinal cord following total or partial cervical vertebrectomy.
- Spondylolisthesis ●
- Spinal Stenosis ●
WARNING: This device is not approved for screw attachment to the posterior elements (pedicles) of the cervical, thoracic, or lumbar spine.
## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use OR Over-The-Counter Use
(Per 21 CFR 801.109)
(Optional Format 1-2-96)
Muriam C. Provost
Division of General, Restorative and Neurological Devices
510(k) Number K031702
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.