← Product Code [KWQ](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/KWQ) · K022344

# TRINICA SELECT ANTERIOR CERVICAL PLATE SYSTEM (K022344)

_Centerpulse Spine-Tech, Inc. · KWQ · Sep 24, 2002 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/KWQ/K022344

## Device Facts

- **Applicant:** Centerpulse Spine-Tech, Inc.
- **Product Code:** [KWQ](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/KWQ.md)
- **Decision Date:** Sep 24, 2002
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 888.3060
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

The Trinica Select Anterior Cervical Plate System is intended for anterior interbody screw fixation of the cervical spine at levels C2-T1. The System is indicated for use in the temporary stabilization of the anterior spine during the development of cervical spinal fusions in patients with degenerative disc disease (as defined by neck pain of discogenic origin of the disc confirmed by patient history and radiographic studies), trauma (including fractures), tumors, deformity (defined as kyphosis, lordosis, or scoliosis), pseudoarthrosis, and/or failed previous fusions. WARNING: This device is not approved for screw attachment to the posterior elements (pedicles) of the cervical, thoracic, or lumbar spine.

## Device Story

Trinica Select Anterior Cervical Plate System provides temporary anterior spinal stabilization during fusion. System components include titanium alloy cervical plates, preassembled locking caps, and fixed/variable angle bone screws. Locking caps feature tabs to prevent screw back-out. Plates available in one-level to four-level configurations; screws available in various lengths and diameters. Device used by surgeons in clinical/OR settings to fixate cervical spine segments. Surgeon implants hardware to maintain alignment and stability while bone fusion occurs. Benefits include mechanical support for spinal segments during healing process.

## Clinical Evidence

Bench testing only. No clinical data provided. Testing evaluated design modifications against established criteria to ensure safety and effectiveness.

## Technological Characteristics

Materials: Titanium alloy (ASTM F136). Surface treatment: Titanium anodization (AMS 2488 Type II). Components: Cervical plates, locking caps, fixed/variable angle bone screws. Form factor: 22mm-92mm plate lengths; 10mm-18mm screw lengths. Mechanical fixation principle.

## Regulatory Identification

A spinal intervertebral body fixation orthosis is a device intended to be implanted made of titanium. It consists of various vertebral plates that are punched into each of a series of vertebral bodies. An eye-type screw is inserted in a hole in the center of each of the plates. A braided cable is threaded through each eye-type screw. The cable is tightened with a tension device and it is fastened or crimped at each eye-type screw. The device is used to apply force to a series of vertebrae to correct “sway back,” scoliosis (lateral curvature of the spine), or other conditions.

## Predicate Devices

- Trinica Anterior Cervical Plating System ([K012305](/device/K012305.md))
- Michelson Anterior Cervical Plate ([K974435](/device/K974435.md))
- Synthes Anterior Cervical Spine Locking Plate ([K945700](/device/K945700.md))
- Synthes Spine Small Stature Anterior Cervical Vertebrae Plate System ([K971883](/device/K971883.md))
- Medtronic Sofamor-Danek Zephir Anterior Cervical Plate System ([K994239](/device/K994239.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

U 22344 p1/3

#### 5. 510(k) Summary

SEP 2 4 2002

#### Submitter's Information

Date:

Name/Address:

July 24, 2002

Centerpulse Spine-Tech 7375 Bush Lake Road Minneapolis, Minnesota 55439

Telephone Number:

Fax Number:

Contact:

# Device Information

Trade Name:

Common Name:

Classification:

Predicate Devices:

Minneapolis, Minnesota

(952) 830-6284

(952) 832-5620

Tim Miller Director of Clinical and Regulatory Affairs

Trinica Select™ Anterior Cervical Plate System

Spinal Intervertebral Body Fixation Orthosis

Class II, KWQ

Trinica" Anterior Cervical Plating System K012305, concurrence date August 22, 2001

Michelson Anterior Cervical Plate K974435, concurrence date February 19, 1998

Synthes Anterior Cervical Spine Locking Plate K945700, concurrence date July 20, 1995

Synthes Spine Small Stature Anterior Cervical Vertebrae Plate System K971883, concurrence date October 16, 1997

Medtronic Sofamor-Danek Zephir" Anterior Cervical Plate System K994239, concurrence date June 19, 2000

{1}------------------------------------------------

#### Device Description:

The Trinica Select Anterior Cervical Plate System is a fixation device consisting of cervical plates, locking caps, fixed bone screws and variable angle bone screws made from titanium alloy in conformance with ASTM F136. The locking cap is preassembled onto the plate and is designed with tabs that prevent bone screws from backing out. The plates and locking caps are treated with titanium anodization per AMS (Aerospace Material Specification) 2488 Type II. Bone screws are subjected to a color anodizing process to differentiate the screw type and diameter

Plates are offered in one-level, two-level, and four-level fusion configurations (22 mm through 92 mm). Bone screws are available in lengths from 10mm through 18mm in 1mm increments. The screws have either a 4.2 inch 4.6 inch diameter. Fixed angle and variable angle screws are available.

#### Intended Use:

The Trinica Select Anterior Cervical Plate System is intended for anterior interbody screw fixation of the cervical spine at levels C2-T1. The System is indicated for use in the temporary stabilization of the anterior spine during the development of cervical spinal fusions in patients with degenerative disc disease (as defined by neck pain of discogenic origin of the disc confirmed by patient history and radiographic studies), trauma (including fractures), tumors, deformity (defined as kyphosis, lordosis, or scoliosis), pseudoarthrosis, and/or failed previous fusions.

WARNING: This device is not approved for screw attachment to the posterior elements (pedicles) of the cervical, thoracic, or lumbar spine.

#### Comparison of Required Technological Characteristics

The Trinica Select™ Anterior Cervical Plate System is substantially equivalent to the predicate Synthes anterior cervical spine system (K945700, concurrence date July 20, 1995). The table below provides a comparison of equivalency characteristics.

| Characteristics                   | Equivalency              |
|-----------------------------------|--------------------------|
| Intended Use                      | Identical                |
| Anatomical Sites                  | Identical                |
| Target Population                 | Identical                |
| Sterilization                     | Identical                |
| Packaging                         | Identical                |
| Operating Principle               | Identical                |
| Materials                         | Identical                |
| Labeling                          | Substantially Equivalent |
| Physical Characteristics (Design) | Substantially Equivalent |
| Performance Testing               | Substantially Equivalent |
| Safety Characteristics            | Substantially Equivalent |

{2}------------------------------------------------

#### Summary of Non-Clinical Tests

Based on risk analysis, appropriate testing was conducted to evaluate the impact of the changes to ensure that the modified device meets established criteria and that identified potential risks were mitigated. Results of the testing demonstrated that the modified device meets established criteria.

#### Conclusions Drawn From Testing

Testing of the Trinica Select Anterior Cervical Plate System demonstrates that the device is substantially equivalent to the devices and that the design modifications do not affect device safety and effectiveness.

2022344 ply

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services. The logo consists of a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or a stylized human figure. The logo is black and white.

### SEP 2 4 2002

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Timothy Miller Director of Clinical and Regulatory Affairs Centerpulse Spine-Tech, Inc. 7375 Bush Lake Road Minneapolis. Minnesota 55439-2027

Re: K022344

> Trade/Device Name: Trinica Select Anterior Cervical Plate System Regulatory Number: 21 CFR 888.3060 Regulation Name: Spinal Intervertebral Body Fixation Orthosis Regulatory Class: II Product Code: KWQ Dated: August 28, 2002 Received: August 29, 2002

Dear Mr. Miller:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adultcration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labcling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{4}------------------------------------------------

Page 2 -- Mr. Timothy Miller

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of vour device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

O. Mark N. Millen

Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

- Indications for Use Statement 2.
## 510(K) Number: K (222) 44

Device Name: Trinica/Trinica Select™ Anterior Cervical Plate System

### Indications for Use:

The Trinica/Trinica Select™ Anterior Cervical Plate System is intended for anterior interbody screw fixation of the cervical spine at levels C2-T1. The system is indicated for use in the temporary stabilization of the anterior spine during the development of cervical spinal fusions in patients with degenerative disc disease (as defined by neck pain of discogenic origin of the disc confirmed by patient history and radiographic studies), trauma (including fractures), tumors, deformity (defined as kyphosis, lordosis, or scoliosis), pseudoarthrosis, and/or failed previous fusions.

WARNING: This device is not approved for screw attachment to the posterior elements (pedicles) of the cervical, thoracic, or lumbar spine.

PLEASE DO NOT WRITE BELOW THIS LINE-------------------------------------------------------------------------------------------------------------------------------------------

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use

MarkRR M. Mckers

al. Restorative

510(k) Number K022344

Over-the-Counter-Use (Per 21 CFR 801.109)

(Pending)

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/KWQ/K022344](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/KWQ/K022344)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
