← Product Code [KWQ](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/KWQ) · K020244

# SYNTHES THORACOLUMBAR SPINE LOCKING PLATE (TSLP) (K020244)

_Synthes Spine · KWQ · Apr 22, 2002 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/KWQ/K020244

## Device Facts

- **Applicant:** Synthes Spine
- **Product Code:** [KWQ](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/KWQ.md)
- **Decision Date:** Apr 22, 2002
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3060
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

The Synthes TSLP System is indicated for use via the lateral or anterolateral surgical approach in the treatment of thoracic and thoracolumbar (T1 - L5) spine instability as a result of fracture (including dislocation and subluxation), tumor, degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies), scoliosis, kyphosis, lordosis, spinal stenosis, or a failed previous spine surgery.

## Device Story

Synthes TSLP System is a spinal fixation orthosis consisting of titanium alloy plates and 5.5 mm cancellous screws with locking heads. Device is implanted by surgeons via lateral or anterolateral approach to the thoracolumbar spine (T1-L5). Plates attach to the anterolateral aspect of the vertebral body to provide stabilization, facilitating biological spinal fusion. System is intended to treat instability caused by various pathologies including fractures, tumors, and degenerative conditions. Device provides mechanical support to the spine during the healing process.

## Clinical Evidence

No clinical data provided; bench testing only.

## Technological Characteristics

Titanium alloy plates and 5.5 mm cancellous screws with locking heads. Designed for anterolateral vertebral body fixation. Class II device, product code KWQ.

## Regulatory Identification

A spinal intervertebral body fixation orthosis is a device intended to be implanted made of titanium. It consists of various vertebral plates that are punched into each of a series of vertebral bodies. An eye-type screw is inserted in a hole in the center of each of the plates. A braided cable is threaded through each eye-type screw. The cable is tightened with a tension device and it is fastened or crimped at each eye-type screw. The device is used to apply force to a series of vertebrae to correct “sway back,” scoliosis (lateral curvature of the spine), or other conditions.

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# APR 2 2 2002

Synthes Spine 510(k) Premarket Notification Thoracolumbar Spine Locking Plate (TSLP) System

#### 11.0 510(K) SUMMARY

#### DESCRIPTION

The Synthes Thoracolumbar Spine Locking Plate (TSLP) System consists of a range of plate sizes and 5.5 mm cancellous screws with a locking head. The plates attach to the anterolateral aspect of the vertebral body of the thoracolumbar spine (levels T1-L5) and provide stabilization to permit the biological process of spinal fusion to occur. All components are manufactured from Titanium alloy.

#### INDICATIONS

The Synthes TSLP System is indicated for use via the lateral or anterolateral surgical approach in the treatment of thoracic and thoracolumbar (T1 - L5) spine instability as a result of fracture (including dislocation and subluxation), turnor, degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies), scoliosis, kyphosis, lordosis, spinal stenosis, or a failed previous spine surgery.

KO20244"P /4

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image is a black and white seal for the Department of Health & Human Services - USA. The seal is circular with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of what appears to be an eagle.

Public Health Service

APR 2 2 2002

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Ms. Vikki Hoffman Senior Regulatory Affairs Associate Synthes Spine 1380 Enterprise Drive West Chester, Pennsylvania 19380

Re: K020244

Trade Name: Synthes Thoracolumbar Spine Locking Plate (TSLP) System Regulatory Number: 21 CFR 888.3060 Regulation Name: Spinal Intervertebral Body Fixation Orthosis Regulatory Class: II Product Code: KWQ Dated: January 23, 2002 Received: January 24, 2002

Dear Ms. Hoffman:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{2}------------------------------------------------

### Page 2 - Ms. Vikki Hoffman

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Mark M. Milkerson

Celia Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Synthes Spine 510(k) Premarket Notification Thoracolumbar Spine Locking Plate (TSLP) System

## 3.0 FDA INDICATIONS FOR USE FORM

510(k) Number (if known): KOQ U244

Device Name:

Synthes Thoracolumbar Spine Locking Plate (TSLP) System

### INDICATIONS FOR USE:

The Synthes TSLP System is indicated for use via the lateral or anterolateral surgical approach in the treatment of thoracic and thoracolumbar (T1 - L5) spine instability as a result of fracture (including dislocation and subluxation), tumor, degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies), scoliosis, kyphosis, lordosis, spinal stenosis, or a failed previous spine surgery.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use (Per 21 CFR § 801.109) OR Over-The-Counter Use

Mark A. Milliken

(Division Sign-Off) Division Sign-Ont)
 Reston of General, Restorative (vision of General Devices rision of other on one of Neurological Neurological Neurological Neurological Neurological Neurological Neurological Neurological Neurological Neurological Neurological Neuro

(k) Number

の

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/KWQ/K020244](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/KWQ/K020244)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
