When used as a pedicle screw fixation system in skeletally mature patients, the system is intended to provide immobilization and stabilization of spinal segments as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, spinal tumor, and failed previous fusion (pseudarthrosis). In addition, when used as a pedicle screw fixation system, the Silhouette™ Spinal Fixation System is indicated for use in patients: - having severe spondylolisthesis (Grade 3 and 4) at the L5-S1 joint - who are receiving fusions with autogenous graft only - who are having the device fixed or attached to the lumbar or sacral spine, and - who are having the device removed after the development of a solid fusion mass. When used as a hook and sacral screw system, the Silhouette™ Spinal Fixation System is intended for use in the treatment of degenerative disc disease (as defined by chronic back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), idiopathic scoliosis, spondylolisthesis, kyphotic or lordotic deformity of the spine, loss of stability due to tumors, spinal stenosis, vertebral fracture or dislocation, pseudoarthrosis, and previous failed spinal fusion. When used for this indication, screws of the Silhouette™ Spinal Fixation System are intended for sacral iliac attachment only. Hook and transverse connectors of the system are intended for posterior thoracic and/or lumbar use only. As a whole, the levels of use for hook and sacral screw fixation of this system are T1 to the sacrum.
Device Story
Temporary posterior spinal fixation system; consists of polyaxial/fixed screws, rods, hooks, and transverse connectors. Used in thoracic, lumbar, and sacral spine (T1-sacrum). Implants provide immobilization and stabilization to facilitate biological spinal fusion. Components are top-loading and top-tightening. Intended for removal after solid fusion. Used by surgeons in clinical settings. Benefits include correction of spinal deformity and stabilization of unstable segments.
Clinical Evidence
Bench testing only. Fatigue testing of the adjustable transverse connector construct was performed per ASTM Provisional Standard PS 5-94 (Static and Dynamic Spinal Implant Assembly in a Corpectomy Model). Results showed no failures at 156 N run-out load for 5,000,000 cycles.
Technological Characteristics
Titanium alloy (Ti-6Al-4V) per ASTM F-136. System includes polyaxial/fixed screws, rods, hooks, and transverse connectors. Top-loading/top-tightening mechanism. Mechanical fixation device; no energy source or software.
Indications for Use
Indicated for skeletally mature patients requiring spinal immobilization/stabilization as an adjunct to fusion for thoracic, lumbar, and sacral spine instabilities/deformities including degenerative spondylolisthesis, fracture, dislocation, scoliosis, tumor, and pseudarthrosis. Also indicated for severe spondylolisthesis (Grade 3-4) at L5-S1, degenerative disc disease, and spinal stenosis. Contraindicated for patients not receiving autogenous graft or those not requiring device removal post-fusion.
Regulatory Classification
Identification
A spinal interlaminal fixation orthosis is a device intended to be implanted made of an alloy, such as stainless steel, that consists of various hooks and a posteriorly placed compression or distraction rod. The device is implanted, usually across three adjacent vertebrae, to straighten and immobilize the spine to allow bone grafts to unite and fuse the vertebrae together. The device is used primarily in the treatment of scoliosis (a lateral curvature of the spine), but it also may be used in the treatment of fracture or dislocation of the spine, grades 3 and 4 of spondylolisthesis (a dislocation of the spinal column), and lower back syndrome.
{0}------------------------------------------------
Image /page/0/Picture/3 description: The image shows the words "SULZER MEDICA" in a bold, sans-serif font. The text is black and appears to be a logo or brand name. The words are arranged horizontally, with "SULZER" above "MEDICA".
NOV 1 6 1999
1.493067
Sulzer Spine-Tech
7375 Bush Lake Road Minneapolis, MN 55439-2027
Phone 612-832-5600 612-832-5620 Fax
## 510(k) Summary
| A. | Submitters' Information | |
|----|----------------------------------|-------------------------------------------------------------------------------------|
| | Date: | November 15, 1999 |
| | Name/Address: | Sulzer Spine-Tech<br>7375 Bush Lake Road<br>Minneapolis, Minnesota 55439 |
| | Telephone Number:<br>Fax Number: | (612) 830-6205<br>(612) 832-5620 |
| | Contact: | Janell A. Colley<br>Regulatory Affairs Specialist |
| B. | Device Information | |
| | Trade Name: | Silhouette™ Spinal Fixation System |
| | Common Name: | Rod, hook, and screw spinal<br>instrumentation |
| | Classification: | Class II, KWP, MNI, and MNH |
| | Predicate Device: | Silhouette™ Spinal Fixation System<br>K992276<br>(concurrence date October 1, 1999) |
## Device Description:
The Silhouette™ Spinal Fixation System is a temporary implant system used to correct spinal deformity and to facilitate the biological process of spinal fusion. This system is intended for posterior use in the thoracic, lumbar, and sacral areas of the spine. Implants in this system consist of hooks and/or screws connected to rods that are intended to be removed after solid fusion has occurred. The system includes polyaxial screws of varving diameters and lengths, fixed screws of varying diameters and lengths, rods in varying lengths, hooks in varying designs, and transverse connectors in fixed and adjustable widths. All implant components are top loading and top tightening. The implants in this system are manufactured from titanium alloy (Ti-6A1-4V) that conforms to ASTM F-136.
{1}------------------------------------------------
# Intended Use:
When used as a pedicle screw fixation system in skeletally mature patients, the system is intended to provide immobilization and stabilization of spinal segments as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, spinal tumor, and failed previous fusion (pseudarthrosis).
In addition, when used as a pedicle screw fixation system, the Silhouette™ Spinal Fixation System is indicated for use in patients:
- having severe spondylolisthesis (Grade 3 and 4) at the L5-S1 ೩) joint
- who are receiving fusions with autogenous graft only b)
- who are having the device fixed or attached to the lumbar or c) sacral spine, and
- d) who are having the device removed after the development of a solid fusion mass
When used as a hook and sacral screw system, the Silhouette™ Spinal Fixation System is intended for use in the treatment of degenerative disc disease (as defined by chronic back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), idiopathic scoliosis, spondylolisthesis, kyphotic or lordotic deformity of the spine, loss of stability due to tumors, spinal stenosis, vertebral fracture or dislocation, pseudoarthrosis, and previous failed spinal fusion. When used for this indication, screws of the Silhouette™ Spinal Fixation System are intended for sacral iliac attachment only. Hook and transverse connectors of the system are intended for posterior thoracic and/or lumbar use only. As a whole, the levels of use for hook and sacral screw fixation of this system are T1 to the sacrum.
{2}------------------------------------------------
#### C. Comparison of Required Technological Characteristics
The modified Silhouette™ Spinal Fixation System is substantially equivalent to the predicate Silhouette™ Spinal Fixation System. The table below provides a comparison of equivalency characteristics.
| Characteristics | Equivalency |
|-----------------------------------|--------------------------|
| Intended Use | Identical |
| Anatomical Sites | Identical |
| Target Population | Identical |
| Sterilization | Identical |
| Packaging | Identical |
| Operating Principle | Identical |
| Materials | Identical |
| Labeling | Substantially Equivalent |
| Physical Characteristics (Design) | Substantially Equivalent |
| Performance Testing | Substantially Equivalent |
| Safety Characteristics | Substantially Equivalent |
#### D. Summary of Non-Clinical Tests
The following is a summary of the testing performed on the modified Silhouette™ Spinal Fixation System:
| Modification | Test Performed | Acceptance<br>Criteria | Test Results |
|------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------|--------------|
| Addition of<br>Adjustable<br>Transverse<br>Connector | Fatigue Testing of Silhouette™<br>Spinal Fixation<br>System/adjustable transverse<br>connector construct per ASTM<br>Provisional Standard PS 5-94,<br>Test Method for Static and<br>Dynamic Spinal Implant<br>Assembly in a Corpectomy<br>Model | No Failures<br>156 N run-out<br>load<br>5,000,000 cycles | Passed |
#### Conclusions Drawn From Testing E.
Testing of the modified Silhouette™ Spinal Fixation System demonstrates that the device is substantially equivalent to the predicate Silhouette™ Spinal Fixation System and that the design modifications do not affect device safety and effectiveness
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird symbol, with three curved lines representing its body and wings. The bird is enclosed within a circular border, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is written around the border of the circle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
NOV 1 6 1999
Ms. Janell A. Colley Regulatory Affairs Specialist Sulzer Spine-Tech 7375 Bush Lake Road Minneapolis, Minnesota 55439-2027
Re: K993067
> Trade Name: Silhouette™ Spinal Fixation System Regulatory Class: II Product Codes: KWP, MNH and MNI Dated: October 21, 1999 Received: October 22, 1999
Dear Ms. Colley:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Ouality System Regulation (OS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{4}------------------------------------------------
Page 2 - Ms. Janell A. Colley
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or at (301) 443-6597, or at its Internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
James E. Dillard III Acting Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
### INDICATIONS FOR USE STATEMENT
510(K) Number: K993067
Silhouette™ Spinal Fixation System Device Name:
Indications for Use:
When used as a pedicle screw fixation system in skeletally mature patients, the Silhouette™ Spinal Fixation System is intended to provide immobilization of spinal segments as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, spinal tumor, and failed previous fusion (pseudarthrosis).
In addition, when used as a pedicle screw fixation system, the Silhouette™ Spinal Fixation System is indicated for use in patients:
- having severe spondylolisthesis (Grade 3 and 4) at the LS-S1 joint a)
- who are receiving fusions with autogenous graft only b)
- who are having the device fixed or attached to the lumbar or sacral spinc, and c)
- d) who are having the device removed after the development of a solid fusion mass
When used as a hook and sacral screw system, the Silhouette™ Spinal Fixation System is intended for use in the treatment of degenerative disc disease (as defined by chronic back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), idiopathic scoliosis, spondylolisthesis, kyphotic or lordotic deformity of the spine, loss of stability due to tumors, spinal stenosis, vertebral fracture or dislocation, pseudoarthrosis, and previous failed spinal fusion. When used for this indication, screws of the Silhouette™ Spinal Fixation System are intended for sacral iliac attachment only. Hook and transverse connectors of the system are intended for posterior thoracic and/or lumbar use only. As a whole, the levels of use for hook and sacral screw fixation of this system are T1 to the sacrum.
## PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED
### Concurrence of CDRH, Office of Device Evaluation (ODE)
OR
AR colep B
(Division Sign-Off)
Division of General Restorative Devices
510(k) Number K993067
**Prescription Use**
**(Per 21 CFR 801.109)**
Over-the-Counter-Use
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.