← Product Code [KWP](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/KWP) · K983353

# SCS SPINAL SYSTEM (K983353)

_Orthotec, Inc. · KWP · Dec 18, 1998 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/KWP/K983353

## Device Facts

- **Applicant:** Orthotec, Inc.
- **Product Code:** [KWP](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/KWP.md)
- **Decision Date:** Dec 18, 1998
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3050
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

When used as a nonpedicle posterior system, the SCS system is indicated for patients with: degenerative disk disease defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies spondylolisthesis fracture spinal stenosis deformities (i.e., scoliosis, kyphosis, lordosis) tumors pseudarthrosis failed previous fusion (pseudarthrosis) The system is indicated for thoracic, lumbar and sacral level. When used as an anterolateral/anterior system the SCS system is indicated for patients with: degenerative disk disease defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies spondylolisthesis fracture spinal stenosis deformities (i.e., scoliosis, kyphosis, lordosis) tumors pseudarthrosis failed previous fusion (pseudarthrosis) The system is indicated for thoracic and lumbar level. When used as a posterior pedicle system, the device is indicated for use in skeletally mature patients who are: having severe spondylolisthesis (Grades 3 and 4) at the L5-S1 joint receiving fusions using autogenous bone graft only having the device fixed or attached to the lumbar and sacral spine having the device removed after the development of a solid fusion mass. Screw fixation is limited to L3 and below. Posterior pedicle systems are intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: degenerative spondylolisthesis with objective evidence of neurologic impairment fracture dislocation scoliosis kyphosis spinal tumor failed previous fusion (pseudarthrosis)

## Device Story

SCS Spinal System consists of implantable components for spinal fusion surgery; includes hooks, rods, vertebral/pedicle screws, and connectors. Surgeon assembles construct to fit patient anatomy. Used in OR by spinal surgeons to immobilize/stabilize thoracic, lumbar, and sacral spine segments. Components made of titanium or stainless steel. System supplied non-sterile; requires high-temperature steam autoclave sterilization before use. Provides mechanical support to allow bone graft fusion. Clinical benefit includes stabilization of spinal instabilities or deformities.

## Clinical Evidence

Bench testing only. Fatigue testing conducted on stainless steel and titanium samples per ASTM F-1717 guidelines. At least two run-out points exceeded 5,000,000 cycles for each material.

## Technological Characteristics

Implantable spinal fixation system. Materials: Titanium alloy (ASTM F136-92, ISO 5832-3), Titanium (ASTM F67 GR2-95, ISO 5832-2), 316LVM stainless steel (ASTM F138 GR2, ISO 5832-1). Components: hooks, rods, vertebral/pedicle screws, connectors. Mechanical principle: rigid fixation/immobilization. Sterilization: high-temperature steam autoclave (121°C, 30 mins).

## Regulatory Identification

A spinal interlaminal fixation orthosis is a device intended to be implanted made of an alloy, such as stainless steel, that consists of various hooks and a posteriorly placed compression or distraction rod. The device is implanted, usually across three adjacent vertebrae, to straighten and immobilize the spine to allow bone grafts to unite and fuse the vertebrae together. The device is used primarily in the treatment of scoliosis (a lateral curvature of the spine), but it also may be used in the treatment of fracture or dislocation of the spine, grades 3 and 4 of spondylolisthesis (a dislocation of the spinal column), and lower back syndrome.

## Predicate Devices

- Synergy™ Spinal System ([K974749](/device/K974749.md), [K950709](/device/K950709.md) & [K950099](/device/K950099.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

DEC 18

Image /page/0/Picture/4 description: The image shows the number "k983353" at the top. Below that, the text "1 of 8" is written. The text and numbers are written in a dark, bold font.

Regulatory Authority: Safe Medical Devices Act of 1990, 21 CFR 807.92

#### 1. Device Name:

| Trade Name: | SCS Spinal System |
|-------------|-------------------|
|-------------|-------------------|

## Common

Anterior spine system, posterior spine system, pedicle screw spine system, hook Name(s): and rod spine system

### Classification

Spinal Interlaminal Fixation Orthosis Name(s): نهنجي جي

- 2. Spinal Intervertebral Body Fixation Orthosis
- ന Pedicle Screw Spinal System

#### 2. Establishment Name & Registration Number:

| Name:   | ORTHOTEC, INC. |
|---------|----------------|
| Number: | Pending        |

#### ని. Classification:

## § 888.3050 - Spinal Interlaminal Fixation Orthosis

& (a) Identification. A spinal interlaminal fixation orthosis is a device intended to be implanted made of an alloy, such as stainless steel, that consists of various hooks and a posteriorly placed compression or distraction rod. The device is implanted, usually across three adjacent vertebrae, to straighten and immobilize the spine to allow bone grafts to unite and fuse the vertebrae together. The device is used primarily in the treatment of scoliosis (a lateral curvature of the spine), but it also may be used in the treatment of fracture or dislocation of the spine, grades 3 and 4 of spondylolisthesis (a dislocation of the spinal column), and lower back syndrome.

## (b) Classification. Class II.

## & 888.3060 - Spinal Intervertebral Body Fixation Orthosis

(a) Identification. A spinal intervertebral body fixation orthosis is a device intended to be implanted made of titanium. It consists of various vertebral plates that are punched into each of a series of vertebral bodies. An eye-type screw is inserted in a hole in the center of each of the plates. A braided cable is threaded through each eye-type screw. The cable is tightened with a tension device and it is fastened or crimped at each eye-type screw. The device is used to apply force to a series of vertebrae to correct sway back, scoliosis (lateral curvature of the spine), or other conditions.

## (b) Classification. Class II.

{1}------------------------------------------------

Image /page/1/Picture/3 description: The image contains handwritten text. The top line reads "K983353". The bottom line reads "a of 5". The text appears to be written with a dark marker or pen, and the handwriting is somewhat messy.

# & 888.3070 - Pedicle Screw Spinal System

(a) Pedicle screw spinal systems-(1) Identification. Pedicle screw spinal systems are multiple component devices, made of a variety of materials, including alloys such as 316L stainless steel, 316LVM stainless steel. 22Cr-13Ni-5Mn stainless steel, Ti-6AI-4V, and unalloyed titanium, that allows the surgeon to build an implant system to fit the patients anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of anchors (e.g., bolts, hooks, and/or screws); interconnection mechanisms incorporating nuts, screws, sleeves, or bolts; longitudinal members (e.g., plates, rods, and/or plate/rod combinations); and/or transverse connectors. The devices are intended to provide immobilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute or chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: degenerative spondvlolisthesis with objective evidence of neurologic impairment, fracture, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudoathrosis). (2) Classification. Class II (special controls). Pedicle screw spinal systems must comply with the following special controls:

- Compliance with material standards, (i)
- Compliance with mechanical testing standard, (ii)
- (iii) Compliance with biocompatability standard, and
- Labeling which contains the following statements in addition to other appropriate labeling (iv) information.

"Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, spinal tumor, and failed previous fusion (pseudathrosis). The safety and effectiveness of these devices for any other conditions are unknown."

"Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient."

(b) Pedicle screw spinal systems for all other uses (1) Identification. Pedicle screw spinal systems for all other uses are multiple component devices, made from a variety of materials, including alloys such as 316L stainless steel, 316LVM stainless steel, 22Cr-13Ni-5Mn stainless steel. Ti-6Al-4V, and unalloyed titanium, that allows the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of anchors (e.g., bolts, hooks, and/or screws); interconnection mechanisms incorporating nuts, screws, sleeves, or bolts; longitudinal members (e.g., plates, rods, and/or plate/rod combinations); and/or transverse connectors.

(2) Classification. Class III (premarket approval)

(c) Date PMA or notice of completion of a PDP is required. An approved PMA or a declared completed PDP must be in effect before placing the device in commercial distribution. See Sec. 888.3.

Device Class: Class II

Classification Panel: Orthopaedic and Rehabilitation Devices Panel

Product Code(s): KWP KWQ & MNH respectively

{2}------------------------------------------------

#### 4. Equivalent Predicate Device:

Synergy™ Spinal System: K974749, K950709 & K950099 - Cross Medical Products, Inc.

The Synergy™ Spinal System is cleared for marketing as a system described in all of the following device classifications:

Spinal Interlaminal Fixation Orthosis Spinal Intervertebral Body Fixation Orthosis Pedicle Screw Spinal System

#### เก่ Device Description:

General system description. The SCS Spinal System includes implantable components which fit together to form a construct for use during spinal fusion surgery. The system contains components of various designs and sizes that allow the surgeon to build an implant system for each of three defined intended uses and to fit the patient's anatomical and physiological requirements.

Materials. Materials for components of the SCS Spinal System are implant grade materials of titanium or stainless steel. The materials comply with applicable standards shown below:

| Titanium Alloy | ASTM F136-92    | ISO 5832-3 |
|----------------|-----------------|------------|
| Titanium       | ASTM F67 GR2-95 | ISO 5832-2 |
| 316LVM         | ASTM F138 GR2   | ISO 5832-1 |

# The system components include:

Lumbar, thoracic, and pedicular hooks Vertebral screws, open vertebral screws, set screws, and pedicle screws (including self-locking) Rods in various lengths Connectors with set screws (sacral, anterior, transverse, lateral) Connecting elements (longitudinal connectors) Instruments Sterilizer trays

Testing Summary. Fatigue testing of a typical system configuration was conducted on samples of both stainless steel and titanium. Samples were tested using ASTM F-1717 as a guideline. Samples were tested for each load configuration. At least two run out points exceeded 5,000,000 cycles for each material.

#### 6. Applicant Name & Address:

ORTHOTEC, INC. 1284 Hillgreen Drive Los Angeles, CA 90035-1021 310.557.2000~310.843.9500 - fax

#### 7. Company Contact:

Regulatory Affairs ORTHOTEC, INC. 1284 Hillgreen Drive Los Angeles, CA 90035-1021 310.557.2000 ~ 310.843.9500 - fax

{3}------------------------------------------------

Image /page/3/Picture/3 description: The image shows the text "K983353" on the first line and "4 of 5" on the second line. The text is written in a bold, sans-serif font. The numbers are clearly visible and the text is easy to read. The image appears to be a close-up of a document or label.

#### Submission Correspondent: 8.

Mr. David W. Schlerf Buckman Company, Inc. 200 Gregory Lane, Suite C-100 Pleasant Hill, CA 94523-3389 925.356.2640 - 925.356.2654 - fax

#### 9. Performance Standards:

United States Food and Drug Administration mandated performance standards for this device do not exist. Various voluntary performance standards are utilized. Voluntary standards utilized include ASTM, Standard Operating Procedures, vendor & process certification and qualification procedures, Quality Systems Regulations, ISO materials standards and ISO 9000 series quality requlations. ORTHOTEC. INC. also meets appropriate general controls authorized under Sections 501, 502, 510, 516, 518, 519, and 520 of the Food, Drug, and Cosmetic Act.

#### 10. Special Controls:

Special controls were published in the Federal Register. Vol. 63, No. 143, July 27, 1998 (Orthopedic Devices: Classification and Reclassification of Pedicle Screw Spinal Systems). The following special controls apply to the marketing of this device when used as a posterior pedicle system:

- Compliance with material standards, (i)
- Compliance with mechanical testing standard, (II)
- Compliance with biocompatability standard, and (iii)
- Compliance with specified labeling requirements. (iv)

#### Storage, Packaging & Sterilization Information: 11.

The SCS Spinal System is supplied "NON-STERILE" and must be sterilized prior to use. The recommended sterilization process is high temperature steam autoclave sterilization. The recommended sterilization cycle will produce a Sterility Assurance Level (SAL) of at least 10"

# The validated cycle is:

Method: Steam Cycle: Gravity Temperature: 250°F (121°C) Exposure Time: 30 minutes

All packages containing implants or instruments should be intact upon receipt. Damaged packaging may indicate the presence of unsafe product. If the package or product is damaged, the product should not be used and should be returned.

Product must be handled, stored and opened in such a way that it is protected from inadvertent damage or contamination. When used, the product must be placed into use following cleaning, sterilization and accepted surgical sterile technique.

{4}------------------------------------------------

#### 12. Summary Comparison Table:

| FEATURE              | SCS Spinal System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Synergy™ Spinal System:    | SE? |
|----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------|-----|
| Indications for Use: | As a Non-pedicle posterior system: spondylolisthesis, fracture, spinal stenosis, deformities (i.e., scoliosis, kyphosis, lordosis), tumors, pseudarthrosis, failed previous fusion (pseudarthrosis).<br><br>As a Anterolateral/anterior system: spondylolisthesis, fracture, spinal stenosis, deformities (i.e., scoliosis, kyphosis, lordosis), tumors, pseudarthrosis, failed previous fusion (pseudarthrosis).<br><br>As a Posterior pedicle system: degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis spinal tumor, failed previous fusion (pseudarthrosis). severe spondylolisthesis (Grades 3 and 4) at the L5-S1 joint, Fusions using autogenous bone graft only, Device fixed or attached to the lumbar and sacral spine, device removed after the development of a solid fusion mass. | SAME                       | YES |
| Design:              | Hook, rod, vertebral screw, sacral screw, pedicle screw & cross link system                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | SAME                       | YES |
| Sterile:             | Non-sterile                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | SAME                       | YES |
| Sizes:               | Rods: 5.8-6.0mm dia. X 40-540mm length<br>Hooks: Small, Medium, Large, Offset, Pedicular, Transverse, Thoracic<br>Vertebral Screws: 5-10mm dia. X 35-50mm length<br>Sacral Screws: 6-7mm dia. X 30-55mm length                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | EQUIVALENT                 | YES |
| Material:            | Stainless steel, titanium alloy, pure titanium                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | SAME                       | YES |
| Origin:              | EUROPEAN                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | USA                        | YES |
| Manufacturer:        | OrthoTec, Inc.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | CROSS MEDICAL              | YES |
| Product Code:        | KWP KWQ & MNH                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | SAME                       | YES |
| K-Number.            | Pending                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | K974749, K950709 & K950099 | YES |

#### 13. Intended Use:

# When used as a nonpedicle posterior system, the SCS system is indicated for patients with:

degenerative disk disease defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies

spondylolisthesis

fracture

spinal stenosis

deformities (i.e., scoliosis, kyphosis, lordosis)

tumors

pseudarthrosis

failed previous fusion (pseudarthrosis)

The system is indicated for thoracic, lumbar and sacral level.

### When used as an anterolateral/anterior system the SCS system is indicated for patients with:

deqenerative disk disease defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies spondylolisthesis fracture spinal stenosis deformities (i.e., scoliosis, kyphosis, lordosis) tumors pseudarthrosis failed previous fusion (pseudarthrosis) The system is indicated for thoracic and lumbar level.

# When used as a posterior pedicle system, the device is indicated for use in skeletally mature patients who are:

having severe spondylolisthesis (Grades 3 and 4) at the L5-S1 joint

receiving fusions using autogenous bone graft only

having the device fixed or attached to the lumbar and sacral spine

having the device removed after the development of a solid fusion mass.

Screw fixation is limited to L3 and below.

Posterior pedicle systems are intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine:

degenerative spondylolisthesis with objective evidence of neurologic impairment

fracture dislocation scoliosis kyphosis spinal tumor failed previous fusion (pseudarthrosis

{5}------------------------------------------------

Image /page/5/Picture/2 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of an eagle or bird with three curved lines representing its body and wings. The logo is surrounded by text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular pattern. The text is in all caps and is evenly spaced around the logo.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

DEC 18 1998

Mr. David W. Schlerf Buckman Company, Inc. Representing OrthoTec, Inc. 200 Greqory Lane Suite C-100 Pleasant Hill, California 94523-3389

K983353 Re : SCS Spinal System Trade Name: Requlatory Class: II Product Codes: MNI, MNH, KWQ, and KWP Dated: September 21, 1998 Received: September 23, 1998

Dear Mr. Schlerf:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major requlations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. ਸ substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory In addition, FDA may publish further announcements action. concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations.

{6}------------------------------------------------

## Page 2 - Mr. David W. Schlerf

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,

a M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{7}------------------------------------------------

Page 1 of

# 610(k) Number : K983353

# Device Name(s): SCS Spinal System

# Intended Use(s) of the Device:

used as a nonpedicle posterior When · system, the SCS system is indicated for patients with:

degenerative disk disease defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies spondylolisthesis fracture spinal stenosis deformities (i.e., scoliosis, kyphosis, lordosis) tumors pseudarthrosis failed previous fusion (pseudarthrosis)

The system is indicated for thoracic, lumbar and sacral level.

When used as an anterolateral/anterior system the SCS system is indicated for patients with:

degenerative disk disease defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies spondylolisthesis fracture spinal stenosis deformities (i.e., scoliosis, kyphosis, lordosis) tumors pseudarthrosis failed previous fusion (pseudarthrosis)

The system is indicated for thoracic and lumbar level.

## When used as a posterior pedicle system, the device is indicated for use in skeletally mature patients who are:

having severe spondylolisthesis (Grades 3 and 4) at the L5-S1 joint

receiving fusions using autogenous bone graft only having the device fixed or attached to the lumbar and sacral spine

having the device removed after the development of a solid fusion mass.

Screw fixation is limited to L3 and below.

Posterior pedicle systems are intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine:

degenerative spondylolisthesis with objective evidence of neurologic impairment fracture dislocation scoliosis kyphosis spinal tumor failed previous fusion (pseudarthrosis)

# PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NECESSARY Concurrence of CDRH, Office of Device Evaluation (ODE)

| Prescription Use     | X | Over-The-Counter Use     |  |
|----------------------|---|--------------------------|--|
| (Per 21 CFR 801.109) |   | (Optional format 1-2-96) |  |

 (Division Sign-Off)  

 Division of General Restorative Devices

| 510(k) Number | K983353 |
|---------------|---------|
|---------------|---------|

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/KWP/K983353](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/KWP/K983353)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
