DEPUY MOTECH MOSS MIAMI SPINAL SYSTEM

K982011 · DePuy Orthopaedics, Inc. · KWP · Aug 6, 1998 · Orthopedic

Device Facts

Record IDK982011
Device NameDEPUY MOTECH MOSS MIAMI SPINAL SYSTEM
ApplicantDePuy Orthopaedics, Inc.
Product CodeKWP · Orthopedic
Decision DateAug 6, 1998
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3050
Device ClassClass 2
AttributesTherapeutic

Intended Use

The MOSS Miami Spinal System is intended for non-cervical use in the spine. When used with anterior screw fixation or posterior hook, non-pedicle screw fixation the MOSS Miami Spinal System is intended to treat scoliosis, kyphosis and lordosis, fracture, loss of stability due to tumor, spinal stenosis, spondylolisthesis, a previously failed fusion surgery or degenerative disc disease (i.e. discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies). When used with pedicle screw fixation, the MOSS Miami Spinal System is intended for use in patients with severe spondylolisthesis (Grades 3 and 4) at the L5-S1 vertebral joint, having fusions with autogenous bone graft, with the device fixed or attached to the lumbar and sacral spine (levels of pedicle screw fixation are L3 and below), and for whom the device system is intended to be removed after the development of a solid fusion mass.

Device Story

MOSS Miami Spinal System is a spinal instrumentation set comprising longitudinal rods, monoaxial/polyaxial screws, reduction screws, hooks, transverse/axial connectors, washers, and sacral extenders. Available in stainless steel or titanium. Used by surgeons in clinical settings to provide spinal stabilization. Components are implanted to treat various spinal pathologies; pedicle screw use is strictly limited to severe L5-S1 spondylolisthesis. System provides mechanical support to facilitate fusion; pedicle screw components are intended for removal post-fusion. Benefits include stabilization of the spinal column and support for bone graft healing. Risks include component fracture, loss of fixation, non-union, vertebral fracture, and neurological or vascular injury.

Clinical Evidence

No clinical data provided. Substantial equivalence is based on design, material, and mechanical similarity to previously cleared predicate devices.

Technological Characteristics

System components constructed from stainless steel or titanium. Includes longitudinal rods, monoaxial/polyaxial screws, reduction screws, hooks, and connectors. Mechanical fixation system; non-powered. No software or electronic components.

Indications for Use

Indicated for non-cervical spinal stabilization in patients with scoliosis, kyphosis, lordosis, fracture, tumor-related instability, spinal stenosis, spondylolisthesis, failed fusion, or degenerative disc disease. Pedicle screw fixation is restricted to patients with Grade 3 or 4 spondylolisthesis at L5-S1, requiring autogenous bone graft and subsequent device removal after fusion.

Regulatory Classification

Identification

A spinal interlaminal fixation orthosis is a device intended to be implanted made of an alloy, such as stainless steel, that consists of various hooks and a posteriorly placed compression or distraction rod. The device is implanted, usually across three adjacent vertebrae, to straighten and immobilize the spine to allow bone grafts to unite and fuse the vertebrae together. The device is used primarily in the treatment of scoliosis (a lateral curvature of the spine), but it also may be used in the treatment of fracture or dislocation of the spine, grades 3 and 4 of spondylolisthesis (a dislocation of the spinal column), and lower back syndrome.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ AUG 6 1998 #### MOTECH ™ 510(k) SUMMARY OF SAFETY AND EFFECTIVENE THE NATURAL BALANCE 1982011 ### NAME OF FIRM: DePuy Inc. 700 Orthopaedic Drive Warsaw, Indiana 46581-0988 510(K) CONTACT: Manager, Clinical and Regulatory Affairs TRADE NAME: COMMON NAME: DEVICE PRODUCT CODE AND CLASSIFICATION: SUBSTANTIALLY EQUIVALENT DEVICES: Lynnette Whitaker DePuy Motech MOSS Miami Spinal System Rod and screw spinal instrumentation 87 KWP. Classification 888.3050 87 KWQ, Classification 888.3060 87 MNH, Unclassified DePuy Motech MOSS Miami Spinal System, DePuy Motech MOSS Miami Spinal System, Pedicle Fixation DePuy Motech MOSS Miami Spinal System, Anterior Use. DePuy Motech Titanium MOSS Miami Spinal System # DEVICE DESCRIPTION AND INTENDED USE: The MOSS MIAMI System is available in either Stainless Steel or Titanium. The following components are available in Stainless Steel: longitudinal rods, monoaxial screws, polyaxial screws, reduction screws, hooks, reduction hooks, transverse connectors, axial connectors, staple washers and sacral extenders. The following components are available in Titanium: longitudinal rods, monoaxial screws, polyaxial screws, hooks, transverse connectors, axial connectors, washers, staple washers, and sacral extenders. The design, intended use, and material of these components have been cleared in previous 510(k) submissions. The MOSS Miami Spinal System is intended for non-cervical use in the spine. When used with anterior screw fixation or posterior hook, non-pedicle screw fixation the MOSS Miami Spinal System is intended to treat scoliosis, kyphosis and lordosis, fracture, loss of stability due to tumor, spinal stenosis, spondylolisthesis, a previously failed fusion surgery or degenerative disc disease (i.e. discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies). > 700 ORTHOPAEDIC DRIVE . P.O. BOX 988 WARSAW, IN 46581-0988 TELEPHONE 1-219-267-8143 {1}------------------------------------------------ 67/23/98 # DEPUY When used with pedicle screw fixation, the MOSS Miami Spinal System is intence which used with severe spondylolisthesis (Grades 3 and 4) at the L 5-ST vertebrad Journ having fusions with autogenous bone graft, with the device fixed or attached to the lumbar and sacral spine (levels of pedicle screw fixation are L3 and below), and for whom the device system is intended to be removed after the development of a solid fusion mass. ## BASIS OF SUBSTANTIAL EQUIVALENCE: The components of the MOSS Miami Spinal System are identical to the components of the other MOSS Miami Spinal Systems which have been cleared by FDA for certain anterior and posterior uses and for pedicle fixation use indications. An insertion tool has been included with the components for ease of implantation. > 700 ORTHOPAEDIC DRIVE • P.O. BDX 988 WARSAW, IN 46581-0988 TELEPHONE 1-219-267-8143 {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines representing hair or clothing. Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 AUG 6 1998 Ms. Lynnette Whitaker Manager, Clinical and Regulatory Affairs DePuy Motech™ 700 Orthopaedic Drive P.O. Box 988 Warsaw, Indiana 46581-0988 Re: K982011 > Reduction Screws and Hooks - to be used as part of the MOSS-Miami Spinal System Regulatory Class: II Product Codes: MNH, KWP, and KWQ Dated: June 4, 1998 Received: June 8, 1998 Dear Ms. Whitaker: We have reviewed your Section 510(k) notification of intent to market the device system referenced above and we have determined the device system is substantially equivalent (for the indications for use stated in the enclosure) to device systems marketed in interstate commerce prior to May 28, 1976 or to devices that have been reclassified in accordance with the provisions of the Federal, Food, Drug, and Cosmetic Act (Act). This decision is based on your device system being found equivalent only to similar device systems labeled and intended for patients: (a) having severe spondylolisthesis (Grades 3 and 4) of the fifth lumbar - first sacral (L5-S1) vertebral joint; (b) who are receiving fusions using autogenous bone graft only; (c) who are having the device fixed or attached to the lumbar and sacral spine; and (d) who are having the device removed after the development of a solid fusion mass. You may, therefore, market your device system subject to the general controls provisions of the Act and the limitations identified below. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Note that labeling or otherwise promoting this device system for pedicular screw fixation/attachment for indications other than severe spondylolisthesis, as described above, would cause the device system to be adulterated under 501(f)(1) of the Act. This device system, when intended for pedicular screw fixation/attachment to the spine for indications other than severe spondylolisthesis, as described above, is a class III device under Section 513(f) of the Act. Class III devices are required to have an approved premarket approval (PMA) application prior to marketing. Accordingly: {3}------------------------------------------------ ### Page 2 - Ms. Lynnette Whitaker - All labeling for this device, including the package label, must state that there are 1. labeling limitations. The package insert must prominently state that the device system using pedicle screws is intended only for patients: (a) having severe spondylolisthesis (Grades 3 and 4) of the fifth lumbar - first sacral (L5-S1) vertebral ioint: (b) who are receiving fusions using autogenous bone graft only; (c) who are having the device fixed or attached to the lumbar and sacral spine; and (d) who are having the device removed after the development of a solid fusion mass. - 2. You may not label or in anyway promote this device system for pedicular, screw fixation/attachment to the cervical, thoracic or lumbar vertebral column for intended uses other than severe spondylolisthesis, as described above. The package insert must include the following statements: ### WARNINGS: - . When used as a pedicle screw system, this device system is intended only for grade 3 or 4 spondylolisthesis at the fifth lumbar - first sacral (L5-S1) vertebral joint. - . The screws of this device system are not intended for insertion into the pedicles to facilitate spinal fusions above the L5-S1 vertebral joint. - . Benefit of spinal fusions utilizing any pedicle screw fixation system has not been adequately established in patients with stable spines. - . Potential risks identified with the use of this device system, which may require additional surgery, include: - device component fracture, loss of fixation, non-union, fracture of the vertebra. neurological injury, and vascular or visceral injury. See Warnings, Precautions, and Potential Adverse Events sections of the package insert for a complete list of potential risks. - 3. Any pedicular screw fixation/attachment for intended uses other than severe spondylolisthesis, as described by item 1, for this device is considered investigational and may only be investigated as a significant risk device in accordance with the investigational device exemption (IDE) regulations under 21 CFR. Part 812. All users of the device for pedicular screw fixation/attachment for intended uses other {4}------------------------------------------------ than severe spondylolisthesis, as described above, must receive approval from their respective institutional review boards (IRBs) and the Food and Drug Administration (FDA) prior to conducting an investigation. - Any previous warning statements identified as part of previous 510(k) clearances or 4. required by OC/Labeling and Promotion which stated that a component/system was not approved for screw fixation into the pedicles of the spine must be replaced by the warnings of items 1 and 2 above. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (OS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal Laws or Regulations. FDA advises that the use of your device system with any other device components but those identified in this 510(k) would require submission of a new 510(k) providing documentation of design, material, and labeling compatibility between the device components. Mechanical testing of a spinal system consisting of the subject device components and other device components, whether yours or those of other manufacturers, may also be required. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification immediately. An FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the {5}------------------------------------------------ Page 4 - Ms. Lynnette Whitaker Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or at (301) 443-6597, or at its Internet address "http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours, ન Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ 510(k) Number 982011 Device Name DePuy Motech MOSS Miami Spinal System, Additional Screw Components Indications for Use: The MOSS Miami Spinal System is intended for non-cervical use in the spine. When used with anterior screw fixation or posterior hook, non-pedicle screw fixation the MOSS Miami Spinal System is intended to treat scoliosis, kyphosis and lordosis, fracture, loss of stability due to tumor, spinal stenosis, spondylolisthesis, a previously failed fusion surgery or degenerative disc disease (i.e. discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies). When used with pedicle screw fixation, the MOSS Miami Spinal System is intended for use in patients with severe spondylolisthesis (Grades 3 and 4) at the L5-S1 vertebral joint, having fusions with autogenous bone graft, with the device fixed or attached to the lumbar and sacral spine (levels of pedicle screw fixation are L3 and below), and for whom the device system is intended to be removed after the development of a solid fusion mass. Concurrence of CDRH, Office of Device Evaluation boell Prescription Use OR (Per 21 CFR 801.109) Over-The Counter Use
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...