ALPHA POSTERIOR SPINAL SYSTEM

K964275 · Sofamor Danek USA,Inc. · KWP · Jan 14, 1997 · Orthopedic

Device Facts

Record IDK964275
Device NameALPHA POSTERIOR SPINAL SYSTEM
ApplicantSofamor Danek USA,Inc.
Product CodeKWP · Orthopedic
Decision DateJan 14, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3050
Device ClassClass 2
AttributesTherapeutic

Intended Use

The ALPHA™ Posterior Spinal System is intended to provide temporary stabilization and to help augment the development of a solid spinal fusion. The system is intended only patients (a) having severe spondylolisthesis (Grades 3 and 4) of the fifth lumbar – first sacral (L5-S1) vertebral joint; (b) who are receiving fusions using autogenous bone graft only; (c) who are having the device fixed or attached to the lumbar and sacral spine (L3 and below); and (d) who are having the device removed after the development of a solid fusion mass. The ALPHA™ Posterior Spinal System is limited to non-cervical use. All implant components of the ALPHA™ Posterior Spinal System are intended for posterior spinal fixation. The ALPHA™ screws are indicated only for insertion no higher than L3 and not lower than the sacrum. CD HORIZON™ Low Profile MULTI-SPAN™ CROSSLINK® plates, ALPHA™ Low Profile MULTI-SPAN™ CROSSLINK® plates, TSRH® Low Profile CROSSLINK® plates, and TSRH® Low Profile CROSSLINK® Offset plates are intended for posterior thoracic, lumbar, and/or sacral use only.

Device Story

ALPHA™ Posterior Spinal System is a rod-based spinal fixation system; provides temporary stabilization to facilitate solid spinal fusion. Components include rods, pedicle screws, connectors, and cross-connectors; components are rigidly locked into configurations tailor-made for individual cases. Used in posterior spinal surgery; implanted by surgeons. System components are rod-size specific (5.5mm or 6.35mm). Device intended for removal after fusion mass development. Benefits include stabilization of severe spondylolisthesis (Grades 3 and 4) at L5-S1.

Clinical Evidence

Bench testing only.

Technological Characteristics

Spinal rod-based fixation system. Components include rods, screws, connectors, and cross-connectors. Rod diameters: 5.5mm and 6.35mm. Screw diameters: 5.5mm, 6.5mm, 7.5mm. Rigid locking mechanism via set screws.

Indications for Use

Indicated for patients with severe spondylolisthesis (Grades 3 and 4) of the L5-S1 vertebral joint requiring temporary stabilization to augment solid spinal fusion using autogenous bone graft. Device fixed to lumbar and sacral spine (L3 and below). Non-cervical use only.

Regulatory Classification

Identification

A spinal interlaminal fixation orthosis is a device intended to be implanted made of an alloy, such as stainless steel, that consists of various hooks and a posteriorly placed compression or distraction rod. The device is implanted, usually across three adjacent vertebrae, to straighten and immobilize the spine to allow bone grafts to unite and fuse the vertebrae together. The device is used primarily in the treatment of scoliosis (a lateral curvature of the spine), but it also may be used in the treatment of fracture or dislocation of the spine, grades 3 and 4 of spondylolisthesis (a dislocation of the spinal column), and lower back syndrome.

Reference Devices

Related Devices

Submission Summary (Full Text)

{0} JAN 09 '97 01:50PM SOFAMOR DANEK P.16/17 # ALPHA™ Posterior Spinal System 510(k) Summary K964275 K964275 January, 1997 *JAN.14, 1997* I. Company: Sofamor Danek USA 1800 Pyramid Place Memphis, TN 38132 901-396-3133 II. Proprietary Trade Name: ALPHA™ Posterior Spinal System Classification Name: Pedicle Screw Fixation - Spondylolisthesis spinal fixation device system. III. The ALPHA™ Posterior Spinal System is a spinal rod based system. The ALPHA™ implant components, as well as implant components from the other systems, can be rigidly locked into a variety of configurations, with each construct being tailor-made for the individual case. Implant components comprising the ALPHA system and those components from other Sofamor Danek spinal systems which can be used with the ALPHA™ Posterior Spinal System are listed in the following table. Please note that while the set screws can be used on both size rods, all of the other components are rod-size specific. | Component | 5.5mm Diameter Rod | 6.35mm Diameter Rod | | --- | --- | --- | | Rods: TSRH® Rods | | ☑ | | GDLH™ Rods | ☑ | | | ALPHA™ Rod | | ☑ | | Screws: ALPHA™ Screws 5.5mm, 6.5mm, 7.5mm diameter | ☑ | ☑ | | Connectors and Cross Connectors: CD HORIZON™ M-10 Break-Off Set Screw | ☑ | ☑ | | ALPHA™ M-10 Break-Off Set Screw | ☑ | ☑ | | CD HORIZON™ Low Profile MULTI-SPAN™ CROSSLINK® Plates | ☑ | ☑ | | ALPHA™ Low Profile MULTI-SPAN™ CROSSLINK® Plates | ☑ | ☑ | | TSRH® Low Profile CROSSLINK® Offset Plates | ☑ | ☑ | | TSRH® Low Profile CROSSLINK® Plates | ☑ | ☑ | | TSRH® CROSSLINK® Plate Set Screws | ☑ | ☑ | IV. The ALPHA™ Posterior Spinal Spinal System is intended to provide temporary stabilization and to help augment the development of a solid spinal fusion. The system is intended only patients (a) having severe spondylolisthesis (Grades 3 and 4) of the fifth lumbar – first sacral (L5-S1) vertebral joint; (b) who are receiving fusions using autogenous bone graft only; (c) who are having the device fixed or attached to the lumbar and sacral spine (L3 and below); and (d) who are having the device removed after the development of a solid fusion mass. {1} JAN 09 '97 01:51PM SOFAMOR DANEK P.17/17 NOTA BENE: The ALPHA™ Posterior Spinal System is limited to non-cervical use. All implant components of the ALPHA™ Posterior Spinal System are intended for posterior spinal fixation. The ALPHA™ Posterior Spinal System is intended only for patients: (a) having severe spondylolisthesis (Grades 3 and 4) of the fifth lumbar – first sacral (L5-S1) vertebral joint; (b) who are receiving fusions using autogenous bone graft only; (c) who are having the device fixed or attached to the lumbar and sacral spine; and (d) who are having the device removed after the development of a solid fusion mass. The ALPHA™ screws are indicated only for insertion no higher than L3 and not lower than the sacrum. CD HORIZON™ Low Profile MULTI-SPAN™ CROSSLINK® plates, ALPHA™ Low Profile MULTI-SPAN™ CROSSLINK® plates, TSRH® Low Profile CROSSLINK® plates, and TSRH® Low Profile CROSSLINK® Offset plates are intended for posterior thoracic, lumbar, and/or sacral use only. V. Mechanical test data were supplied in support of the ALPHA™ Posterior Spinal System 510(k) notification. The ALPHA™ Posterior Spinal System was declared to be substantially equivalent to several commercially available devices. ©1997 Sofamor Danek
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...