VERTEX RECONSTRUCTION SYSTEM, VERTEXT SELECT RECONSTRUCTION SYSTEM

K123906 · Medtronic Sofamor Danek USA, Inc. · KWP · Apr 1, 2013 · Orthopedic

Device Facts

Record IDK123906
Device NameVERTEX RECONSTRUCTION SYSTEM, VERTEXT SELECT RECONSTRUCTION SYSTEM
ApplicantMedtronic Sofamor Danek USA, Inc.
Product CodeKWP · Orthopedic
Decision DateApr 1, 2013
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3050
Device ClassClass 2
AttributesTherapeutic

Intended Use

When intended as an adjunct to fusion of the occipitocervical spine, cervical spine, and the thoracic spine (Occiput-T3), the VERTEX® Reconstruction System is indicated for skeletally mature patients using allograft and/or autograft for the following: DDD (neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, spinal stenosis, fracture, dislocation, failed previous fusion, and/or tumors. The occipitocervical plate rods, plates, rods, occipital screws, and hooks are intended to provide stabilization to promote fusion following reduction of fracture/dislocation or trauma in the occipitocervical junction and the cervical spine. When used to treat these occipitocervical and cervical conditions, these screws are limited to occipital fixation only. The screws are not intended to be placed in the cervical spine. Occipitocervical constructs require bilateral fixation to C2 and below. Note: Segmental fixation is recommended for these constructs. The hooks and rods are also intended to provide stabilization to promote fusion following reduction of fracture/dislocation or trauma in the cervical/upper thoracic (C1-T3) spine. The use of multi-axial screws is limited to placement in T1-T3. The screws are not intended to be placed in the cervical spine. Titanium ATLAS® Cable System to be used with the VERTEX® Reconstruction System allows for cable attachment to the posterior cervical or thoracic spine. In order to achieve additional levels of fixation, the VERTEX® Reconstruction System may be connected to the CD HORIZON® Spinal System rods with the VERTEX® rod connectors. Transition rods with differing diameters may also be used to connect the VERTEX® Reconstruction System to the CD HORIZON® Spinal System screws, hooks, and connectors. Refer to the CD HORIZON® Spinal System package insert for a list of the CD HORIZON® Spinal System indications of use.

Device Story

Posterior spinal fixation system; consists of plates, rods, hooks, screws, multi-axial screws, and connectors; used for stabilization to promote fusion in occipitocervical, cervical, and upper thoracic spine (Occiput-T3). Surgeon-implanted in clinical/OR setting; constructs are tailor-made for individual cases. Provides rigid fixation; may be used with ATLAS® cable system or connected to CD HORIZON® Spinal System. Benefits patient by providing mechanical stabilization during bone fusion process. Components fabricated from titanium, titanium alloy, cobalt chromium, and Nitinol (retaining rings); some components include silicone elastomeric stakes.

Clinical Evidence

Bench testing only. No clinical data provided. Non-clinical testing included compression fatigue, static compression, and static torsion performed in accordance with ASTM 1717-12 to demonstrate equivalence to predicate devices.

Technological Characteristics

Materials: medical grade titanium, titanium alloy, cobalt chromium, Nitinol (NiTi) for retaining rings, silicone adhesive for elastomeric stakes. Mechanical fixation system. Testing performed per ASTM 1717-12. No software or electronic components.

Indications for Use

Indicated for skeletally mature patients requiring fusion of occipitocervical, cervical, and thoracic (Occiput-T3) spine. Conditions include DDD, spondylolisthesis, spinal stenosis, fracture, dislocation, failed fusion, and tumors. Contraindicated for use with stainless steel components.

Regulatory Classification

Identification

A spinal interlaminal fixation orthosis is a device intended to be implanted made of an alloy, such as stainless steel, that consists of various hooks and a posteriorly placed compression or distraction rod. The device is implanted, usually across three adjacent vertebrae, to straighten and immobilize the spine to allow bone grafts to unite and fuse the vertebrae together. The device is used primarily in the treatment of scoliosis (a lateral curvature of the spine), but it also may be used in the treatment of fracture or dislocation of the spine, grades 3 and 4 of spondylolisthesis (a dislocation of the spinal column), and lower back syndrome.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ APR 0 1 2013 ## VERTEX® Reconstruction System 510(k) Summary February 25, 2013 Medtronic Sofamor Danek l. Company: 1800 Pyramid Place Memphis, TN 38132 Telephone: (901) 396-3133 Fax: (901) 346-9738 > Claire Evans Contact: Senior Regulatory Affairs Specialist - Proprietary Trade Name: VERTEX® Reconstruction System II. Classification Name: Spinal Interlaminal Fixation Orthosis, 21 CFR III. 888.3050 - IV. Classification: Class II - Product Codes: KWP V. - Product Description: The VERTEX SELECT® Reconstruction System is a VI. posterior system, which consists of a variety of shapes and sizes of plates, rods, hooks, screws, multi-axial screws, and connecting components, which can be rigidly locked to the rod in a variety of configurations, with each construct being tailor-made for the individual case. Titanium ATLAS® cable may be used with this system at the surgeon's discretion. See the package inserts of both of those systems for labeling limitations. The VERTEX® Reconstruction System is fabricated from medical grade titanium, medical grade titanium alloy, and medical grade cobalt chromium. Medical grade titanium, medical grade titanium alloy, and/or medical grade cobalt chromium may be used together. Never use titanium, titanium alloy, and/or cobalt chromium with stainless steel in the same construct. The VERTEX® Reconstruction System includes a retaining ring for the multi-axial screw made of Shape Memory Alloy (Nitinol -- NiTi). Shape Memory Alloy is compatible with titanium alloy, and cobalt chromium implants only. Some components contain elastomeric stakes made of silicone adhesive commonly used in implantable medical devices. Do not use with stainless steel. To achieve best results, do not use any of the VERTEX® Reconstruction System implant components with components from any other system or manufacturer unless specifically labeled to do so in this or another MEDTRONIC document. As with all orthopaedic and neurosurgical implants, {1}------------------------------------------------ none of the VERTEX® Reconstruction System components should ever be reused under any circumstances. - Indications: When intended as an adjunct to fusion of the occipitocervical VII. spine, cervical spine, and the thoracic spine (Occiput-T3), the VERTEX® Reconstruction System is indicated for skeletally mature patients using allograft and/or autograft for the following: DDD (neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, spinal stenosis, fracture, dislocation, failed previous fusion, and/or tumors. ## Occipitocervical Components: Plate Rod/Plates/Rods/Occipital Screws/Hooks The occipitocervical plate rods, plates, rods, occipital screws, and hooks are intended to provide stabilization to promote fusion following reduction of fracture/dislocation or trauma in the occipitocervical junction and the cervical spine. When used to treat these occipitocervical and cervical conditions, these screws are limited to occipital fixation only. The screws are not intended to be placed in the cervical spine. Occipitocervical constructs require bilateral fixation to C2 and below. Note: Segmental fixation is recommended for these constructs. ## Hooks and Rods The hooks and rods are also intended to provide stabilization to promote fusion following reduction of fracture/dislocation or trauma in the cervical/upper thoracic (C1-T3) spine. ### Multi-axial Screws The use of multi-axial screws is limited to placement in T1-T3. The screws are not intended to be placed in the cervical spine. Titanium ATLAS® Cable System to be used with the VERTEX® Reconstruction System allows for cable attachment to the posterior cervical or thoracic spine. ### Connectors In order to achieve additional levels of fixation, the VERTEX® Reconstruction System may be connected to the CD HORIZON® Spinal System rods with the VERTEX® rod connectors. Transition rods with differing diameters may also be used to connect the VERTEX® Reconstruction System to the CD HORIZON® Spinal System screws, hooks, and connectors. Refer to the CD HORIZON® Spinal System package insert for a list of the CD HORIZON® Spinal System indications of use. {2}------------------------------------------------ - Summary of Technological Characteristics: The purpose of this 510(k) is VIII. to add additional components to the VERTEX® Reconstruction System. The subject and predicate parts are identical in terms of material, indications for use, intended use, and performance specifications. The key differences between the subject and predicate devices are length and the addition of an angled open connection option. - Identification of Legally Marketed Devices: The design features, materials IX. and indications for use of the subject devices are substantially equivalent to predicate VERTEX® Reconstruction System devices cleared in K083071 (S.E. 11/14/2008) and K121191 (S.E. 06/29/12). The labeling is identical to that cleared in K121191 (S.E. 06/29/12). - Discussion of the Non-Clinical Testing: Medtronic performed compression X. fatigue, static compression and static torsion on the subject connectors in accordance with ASTM 1717-12. Non-clinical testing in the form of design verification and validation activities was performed on the subject devices to show equivalence to the previously listed predicate devices. - Conclusions: A risk analysis and non-clinical testing was completed for the XI. changes incorporated into the subject devices. The test results and additional supporting documentation provided in this submission demonstrated substantial equivalence of the subject devices to the predicate devices cleared in K083071 (S.E. 11/14/2008). {3}------------------------------------------------ ## DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged in a circular fashion around the symbol. The text is in all caps and is evenly spaced around the circle. Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 Medtronic Sofamor Danek USA, Incorporated % Ms. Claire Evans Senior Regulatory Affairs Specialist 1800 Pyramid Place Memphis, Tennessee 38132 Letter dated: April 1, 2013 Re: K123906 Trade/Device Name: VERTEX® Reconstruction System Regulation Number: 21 CFR 888.3050 Regulation Name: Spinal interlaminal fixation orthosis Regulatory Class: Class II Product Code: K WP Dated: February 25, 2013 Received: February 26, 2013 Dear Ms. Evans: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {4}------------------------------------------------ ## Page 2 - Ms. Claire Evans If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, # Mark N. Melkerson -S Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ K123906 510{k} Number (if known): Device Name: VERTEX® Reconstruction System When intended as an adjunct to fusion of the occipitocervical spine, cervical spine, and the thoracic spine (Occiput-T3), the VERTEX® Reconstruction System is indicated for skeletally mature patients using allograft and/or autograft for the following: DDD (neck pain of discogenic ongin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, spinal stenosis, fracture, dislocation, failed previous fusion, and/or tumors. ### Occipitocervical Components: Plate Rod/Plates/Rods/Occipital Screws/Hooks The occipitocervical plate rods, plates, rods, occipital screws, and hooks are intended to provide stabilization to promote fusion following reduction of fracture/dislocation or trauma in the occipitocervical junction and the cervical spine. When used to treat these occipitocervical and cervical conditions, these screws are limited to occipital fixation only. The screws are not intended to be placed in the cervical spine. Occipitocervical constructs require bilateral fixation to C2 and below. Note: Segmental fixation is recommended for these constructs. ### Hooks and Rods . The hooks and rods are also intended to provide stabilization to promote fusion following reduction of fracture/dislocation or trauma in the cervical/upper thoracic (C1-T3) spine. ### Multi-axial Screws The use of multi-axial screws is limited to placement in T1-T3. The screws are not intended to be placed in the cervical spine. Titanium ATLAS® Cable System to be used with the VERTEX® Reconstruction System allows for cable attachment to the posterior cervical or thoracic spine. #### Connectors In order to achieve additional levels of fixation, the VERTEX® Reconstruction System may be connected to the CD HORIZON® Spinal System rods with the VERTEX® rod connectors. Transition rods with differing diameters may also be used to connect the VERTEX® Reconstruction System to the CD HORIZON® Spinal System screws, hooks, and connectors. Refer to the CD HORIZON® Spinal System package insert for a list of the CD HORIZON® Spinal System indications of use. Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) # Ronald P. Jean -S (Division Sign-Off) Division of Orthopedic Devices 510(k) Number: K123906
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...