← Product Code [KWP](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/KWP) · K111183

# ASCENT POSTERIOR OCCIPITAL THORACIC (POCT) SYSTEM (TITANIUM & COBALT CHROME RODS) (K111183)

_Orthofix, Inc. · KWP · May 24, 2011 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/KWP/K111183

## Device Facts

- **Applicant:** Orthofix, Inc.
- **Product Code:** [KWP](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/KWP.md)
- **Decision Date:** May 24, 2011
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 888.3050
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

When intended to promote fusion of the cervical spine and occipito-cervico-thoracic junction (Occiput-T3), the Ascent POCT System is indicated for: a) degenerative disc disease (defined as neck pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies); b) spondylolisthesis; c) fracture/dislocation; d) spinal stenosis; e) atlanto-axial fracture with instability: f) occipito-cervical dislocation; g) tumors; h) revision of previous cervical spine surgery. The occipital bone screws are limited to occipital fixation only. The use of the multi-axial screws is limited to placement in the upper thoracic spine (T1-T3) for the treatment of thoracic conditions only. They are not intended to be placed in the cervical spine. The lateral offset adapter is indicated for use in the upper thoracic spine (T1-T3). The hooks are intended to be placed from C1 to T3. The Songer Cable (titanium) System to be used with the Ascent Posterior Occipital Cervical Thoracic System allows for wire/cable attachment to the posterior cervical spine. The Ascent Posterior Occipital Cervical Thoracic System can also be linked to the Orthofix Spinal Fixation System using the Ascent Axial or Parallel Rod Connector.

## Device Story

The Ascent POCT System is a temporary, non-sterile, single-use spinal fixation system used by surgeons to build implant constructs for immobilization and stabilization of the cervical and upper thoracic spine (Occiput-T3). Components include rods, setscrews, cross connectors, axial/parallel rod connectors, lateral offset adapters, multi-axial screws, hooks, bone screws, and Songer cables. The system can be linked to the Orthofix Spinal Fixation System. Surgeons assemble the construct intraoperatively to provide structural support during the fusion process. The device benefits patients by stabilizing spinal segments, facilitating fusion, and addressing instability or deformity. It is intended for use in clinical settings by trained orthopedic or neurosurgeons.

## Clinical Evidence

Bench testing only. Mechanical performance was validated through static and dynamic compression bending and torsion testing in accordance with ASTM F2706-08 standards. No clinical data was required or provided.

## Technological Characteristics

System components are constructed from Titanium Alloy (ASTM F136) and Cobalt Chrome Alloy (ASTM F1537). The system utilizes a modular rod-and-screw fixation design. It is non-sterile and intended for single use. No software or electronic components are involved.

## Regulatory Identification

A spinal interlaminal fixation orthosis is a device intended to be implanted made of an alloy, such as stainless steel, that consists of various hooks and a posteriorly placed compression or distraction rod. The device is implanted, usually across three adjacent vertebrae, to straighten and immobilize the spine to allow bone grafts to unite and fuse the vertebrae together. The device is used primarily in the treatment of scoliosis (a lateral curvature of the spine), but it also may be used in the treatment of fracture or dislocation of the spine, grades 3 and 4 of spondylolisthesis (a dislocation of the spinal column), and lower back syndrome.

## Predicate Devices

- Blackstone Posterior Cervical System ([K030197](/device/K030197.md))
- Ascent POCT System ([K080394](/device/K080394.md))
- Firebird Spinal Fixation System, Cobalt Chrome Rods ([K092624](/device/K092624.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# 183

## MAY 2 4 2011

Premarket Notification 510(k) Orthofix Inc. Orthofix Ascent POCT System

#### 510(k) SUMMARY

### Orthofix Ascent POCT System

| Submitter Information      |                                                                                                                                                                                                                  |
|----------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Name:                      | Orthofix Inc.                                                                                                                                                                                                    |
| Address:                   | 3451 Plano Parkway<br>Lewisville, TX 75056                                                                                                                                                                       |
| Telephone Number:          | 214.937.2000                                                                                                                                                                                                     |
| Fax Number:                | 214.937.3322                                                                                                                                                                                                     |
| Email:                     | darlachew@orthofix.com                                                                                                                                                                                           |
| Registration Number:       | 3008524126                                                                                                                                                                                                       |
| Contact Person:            | Darla Chew<br>Director of Regulatory Affairs                                                                                                                                                                     |
| Date Prepared:             | April 26, 2011                                                                                                                                                                                                   |
| Name of Device             |                                                                                                                                                                                                                  |
| Trade Name / Proprietary   | Ascent® Posterior Occipital Cervical Thoracic (POCT) System                                                                                                                                                      |
| Common Name:               | Spinal Fixation System                                                                                                                                                                                           |
| Product Code:              | KWP – Spinal Interlaminal Fixation Orthosis                                                                                                                                                                      |
| Regulatory Classification: | Class II – 888.3050 – Spinal Interlaminal Fixation Orthosis                                                                                                                                                      |
| Review Panel:              | Orthopedic Device Panel                                                                                                                                                                                          |
| Predicate Devices:         | Blackstone Posterior Cervical System, K030197, SE 6-12-03<br>Ascent POCT System, K080394, SE 3-13-08 (Plate and Rod Modifications)<br>Firebird Spinal Fixation System, Cobalt Chrome Rods, K092624, SE 9-25-2009 |

Reason for 510(k) Submission: New product offering

Orthofix inc., 3451 Plano Parkway, Lewisville, Texas 75056 | 214-937-2061 www.orthofix.com

Pg 1 of 4

{1}------------------------------------------------

KIIII183

Premarket Notification 510(k) Orthofix Inc. Orthofix Ascent POCT System

#### Device Description

The Ascent® POCT System is a temporary, multiple component system comprised of a variety of non-sterile, single use components made of Titanium and Cobalt Chrome allov that allow the surgeon to build a spinal implant construct. The system design is intended to promote immobilization and stabilization of spinal segments as an adjunct to fusion of the cervical and for upper thoracic spine. Ascent POCT System consists of an assortment of rods, setscrews, cross connectors, axial connectors, lateral offset adapters, multi-axial screws, hooks, plates, bone screws, and Songer Cables. Ascent POCT system can also be linked to Orthofix Spinal Fixation System (SFS) using the Ascent Axial or Parallel Rod Connector or Transition Rods.

#### Intended Use / Indications for Use

When intended to promote fusion of the cervical spine and occipito-cervico-thoracic junction (Occiput-T3), the Ascent POCT System is indicated for:

- a) degenerative disc disease (defined as neck pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies);
- b) spondylolisthesis;
- c) fracture/dislocation;
- d) spinal stenosis;
- e) atlanto-axial fracture with instability:
- f) occipito-cervical dislocation;
- g) tumors;
- h) revision of previous cervical spine surgery

The occipital bone screws are limited to occipital fixation only. The use of the multi-axial screws is limited to placement in the upper thoracic spine (T1-T3) for the treatment of thoracic conditions only. They are not intended to be placed in the cervical spine. The lateral offset adapter is indicated for use in the upper thoracic spine (T1-T3). The hooks are intended to be placed from C1 to T3. The Songer Cable (titanium) System to be used with the Ascent Posterior Occipital Cervical Thoracic System allows for wire/cable attachment to the posterior cervical spine.

The Ascent Posterior Occipital Cervical Thoracic System can also be linked to the Orthofix Spinal Fixation System using the Ascent Axial or Parallel Rod Connector.

> Orthofix inc., 3451 Plano Parkway, Lewisville, Texas 75056 | 214-937-2061 www.orthofix.com

Page 2 of 4

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the text "Premarket Notification 510(k)" at the top, followed by "Orthofix Inc." and "Orthofix Ascent POCT System" below. The text is arranged in a centered format. The image appears to be a snippet from a document or presentation.

#### Summary of the Technological Characteristics of the Device Compared to the Predicate Device

| Characteristic    | Orthofix Ascent Rods                                              | Ascent POCT System<br>(K030197 & K080394)                         | Firebird Spinal<br>Fixation System,<br>Cobalt Chrome Rods<br>(K092624) |
|-------------------|-------------------------------------------------------------------|-------------------------------------------------------------------|------------------------------------------------------------------------|
| Function / Design | This system allows a surgeon to build a spinal implant construct. | This system allows a surgeon to build a spinal implant construct. | This system allows a surgeon to build a spinal implant construct.      |
| Rod Configuration | Straight & Occipital                                              | Straight & Occipital                                              | Straight                                                               |
| Rod Sizes         | Straight: 3.0 mm x 70, 120, 200 mm long; Occipital Rod            | Straight: 3.0 mm x 70, 120, 200 mm long; Occipital Rod            | Not relevant                                                           |
| Material          | 1) Cobalt Chrome Alloy, ASTM F1537<br>2) Titanium Alloy ASTM F136 | Titanium Alloy ASTM F136                                          | Cobalt Chrome Alloy, ASTM F1537                                        |

#### PERFORMANCE DATA - Summary of Non-Clinical Test Conducted for Determination of Substantial Equivalence

| Characteristic                   | Standard / Test/ FDA Guidance |
|----------------------------------|-------------------------------|
| Static Compression Bending Test  | ASTM F2706-08                 |
| Static Torsion Test              | ASTM F2706-08                 |
| Dynamic Compression Bending Test | ASTM F2706-08                 |
| Dynamic Torsion Test             | ASTM F2706-08                 |

#### Performance Data Summary

Mechanical testing of the Orthofix Ascent POCT System was conducted in accordance to ASTM F2706-08 standard for Static & Dvnamic Compression Bending test and ASTM F2706-08 standards for Static & Dynamic Torsion test. Test results demonstrated that the new, proposed Ascent rods are substantially equivalent to predicate devices that have the same intended use, similar indications, technological characteristics and principles of operation.

#### Substantial Equivalence

The Orthofix proposed Ascent rods are substantially equivalent in design, function, and intended use to the Orthofix Ascent POCT system rods. The original Ascent POCT System received 510(k) clearance under K030197 on June 12, 2003 with subsequent changes to the Ascent Occipital rod under K080394 SE March 13, 2008.

The Orthofix proposed Ascent rods are substantially equivalent in material composition to Firebird Spinal Fixation System, Cobalt Chrome Rods 510(k) K092624 (SE 9-25-2009).

> Orthofix inc., 3451 Plano Parkway, Lewisville, Texas 75056 | 214-937-2061 www.orthofix.com

Pg 3 of 4

Page 3 of 4

{3}------------------------------------------------

KIIII183

Premarket Notification 510(k) Orthofix Inc. Orthofix Ascent POCT System

The difference between the modified rods and the predicate devices consist of minor dimensional changes for the Ascent Occipital rod, and the addition of Cobalt Chrome rods. Based on test results, these changes do not present any new / additional issues of safety or effectiveness therefore; the Orthofix proposed Ascent rods (part of the Ascent POCT System) are substantially equivalent to the predicate devices.

> Orthofix inc., 3451 Plano Parkway, Lewisville, Texas 75056 | 214-937-2061 www.orthofix.com

> > Page 4 of 4

Rs 4 of 4

{4}------------------------------------------------

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Orthofix, Inc. % Ms. Darla Chew - - -Director of Regulatory Affairs 3451 Plano Parkway Lewisville, TX 75056

MAY 2 4 2011

Re: K111183

Trade/Device Name: Ascent Posterior Occipital Cervical Thoracic System (Titanium &

Cobalt Chrome Rods) Regulation Number: 21 CFR 888.3050 Regulation Name: Spinal interlaminal fixation orthosis Regulatory Class: Class II Product Code: K WP Dated: April 26, 2011 Received: April 27, 2011

Dear Ms. Chew:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21

{5}------------------------------------------------

#### Page 2 - Ms. Darla Chew

CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

E.S. Keith

2 Mark N. Melkerson Director Division of Surgical, Orthopedic And Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

#### INDICATION FOR USE STATEMENT

#### 510(k) Number (if known): K111183

Device Name: Ascent Posterior Occipital Cervical Thoracic (POCT) System (Titanium & Cobalt Chrome Rods)

#### Indications for Use:

When intended to promote fusion of the cervical spine and occipito-cervico-thoracic junction (Occiput-T3), the Ascent POCT System is indicated for:

- a. degenerative disc disease (defined as neck pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies);
- b. spondvlolisthesis;
- c. fracture/dislocation;
- d. spinal stenosis;
- e. atlanto-axial fracture with instability;
- f. occipito-cervical dislocation;
- g. tumors;
- h. revision of previous cervical spine surgery

The occipital bone screws are limited to occipital fixation only. The use of the multi-axial screws is limited to placement in the upper thoracic spine (T1-T3) for the treatment of thoracic conditions only. They are not intended to be placed in the cervical spine. The lateral offset adapter is indicated for use in the upper thoracic spine (T1-T3). The hooks are intended to be placed from C1 to T3. The Songer Cable (titanium) System to be used with the Ascent Posterior Occipital Cervical Thoracic System allows for wire/cable attachment to the posterior cervical spine.

The Ascent Posterior Occipital Cervical Thoracic System can also be linked to the Orthofix Spinal Fixation System using the Ascent Axial or Parallel Rod Connector.

Prescription Use: X (Part 21 CFR 801 Subpart D) And / Or

Over-The-Counter (Part 21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Div/sion Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices

510(k) Number_________________________________________________________________________________________________________________________________________________________________ K ! ! ! ! ! ! ! ! ! ! ! ! ! ! ! ! ! ! ! ! ! ! ! ! ! ! ! ! ! ! ! ! ! ! ! ! ! ! ! ! ! ! ! ! ! ! ! ! ! ! ! ! ! ! ! ! ! ! ! ! ! ! ! ! ! ! ! ! ! ! ! ! ! ! ! ! ! ! ! ! ! ! ! ! ! ! www.orthofix.com

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/KWP/K111183](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/KWP/K111183)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
