MODIFICATION TO CASPIAN SPINAL SYSTEM

K093534 · K2m, Inc. · KWP · Dec 16, 2009 · Orthopedic

Device Facts

Record IDK093534
Device NameMODIFICATION TO CASPIAN SPINAL SYSTEM
ApplicantK2m, Inc.
Product CodeKWP · Orthopedic
Decision DateDec 16, 2009
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 888.3050
Device ClassClass 2
AttributesTherapeutic

Intended Use

The Caspian Spinal System is indicated for the following: DDD (neck pain of discogenic origin with degeneration of the disc as confirmed by patient history and radiographic studies) Spondylolisthesis Spinal stenosis Fracture/dislocation Revision of previous cervical spine surgery Tumors The implants are intended to provide stabilization as an adjunct to fusion when used with autogenous bone graft or allograft following the reduction of fractures/dislocations or trauma in the spine. The rod and hook components are intended for use in the cervical/upper thoracic (C1-T3) spine. The pedicle screws are limited to placement in T1-T3 in treating thoracic conditions only. The pedicle screws are not intended to be placed in or treat conditions involving the cervical spine. The Caspian Spinal System can also be linked to the Range Spinal System using the 3.5mm/5.5mm rod connectors or transitional rods.

Device Story

Caspian Spinal System is a top-loading, posterior cervical-thoracic spinal fixation system. Components include pedicle screws, rods, locking set screws, hooks, and rod connectors. System provides spinal segment immobilization as an adjunct to fusion. Used by surgeons in clinical settings to treat degenerative disc disease, spondylolisthesis, stenosis, fractures, dislocations, tumors, or revision cases. Implants are used with autogenous bone graft or allograft. System can be linked to Range Spinal System via transitional rods or connectors. Benefits include spinal stabilization following trauma or surgical intervention.

Clinical Evidence

No clinical data provided; bench testing only.

Technological Characteristics

Top-loading posterior spinal fixation system. Materials: Ti6Al4V Eli and Cobalt Chrome per ASTM and ISO standards. Components: pedicle screws, rods, locking set screws, hooks, rod connectors. Connectivity: mechanical linkage to Range Spinal System. Non-software device.

Indications for Use

Indicated for patients with DDD, spondylolisthesis, spinal stenosis, fracture/dislocation, tumors, or requiring revision of previous cervical spine surgery. Used for stabilization as an adjunct to fusion in the cervical/upper thoracic (C1-T3) spine. Pedicle screws restricted to T1-T3 thoracic use only; contraindicated for cervical spine placement.

Regulatory Classification

Identification

A spinal interlaminal fixation orthosis is a device intended to be implanted made of an alloy, such as stainless steel, that consists of various hooks and a posteriorly placed compression or distraction rod. The device is implanted, usually across three adjacent vertebrae, to straighten and immobilize the spine to allow bone grafts to unite and fuse the vertebrae together. The device is used primarily in the treatment of scoliosis (a lateral curvature of the spine), but it also may be used in the treatment of fracture or dislocation of the spine, grades 3 and 4 of spondylolisthesis (a dislocation of the spinal column), and lower back syndrome.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ #### 510(k) Summary for the Caspian Spinal System DEC 1 6 2009 This 510(k) summary for the Caspian Spinal System is provided as required per Section 513(i(i)(i) of the Food, Drug and Cosmetic Act. - 1. Submitter : K2M, Inc. 751 Miller Drive SE, Suite F1 Leesburg, VA 20175 Contact Person : Nancy Giezen K2M, Inc. 751 Miller Drive SE, Suite F1 Leesburg, VA 20175 Telephone: 703-777-315 ذ Date Prepared: 11/13/09 | 2. Tradename: | Caspian Spinal System | |----------------------|--------------------------------------------------------| | Common Name: | Spinal Fixation System | | Classification Name: | Spinal Interlaminal Fixation Orthosis (21CFR 888.3050) | | Device Product Code: | KWP | | Regulatory Class: | Class II | - 3. Predicate or legally marketed devices which are substantially equivalent : - K2M Caspian Spinal System � - . DePuy Mountaineer - Biomet Altius . ### 4. Description of the device: The Caspian Spinal System is a top-loading, multiple component, posterior (cervical-thoracic) spinal fixation system which consists of pedicle screws, rods, locking set screws, hooks, and rod connectors. Materials: The devices are manufactured from Ti6Al4V Eli and Cobalt Chrome per ASTM and ISO standards. Function: The system functions as an adjunct to fusion to provide immobilization of spinal segments of the cervical and thoracic (T1-T3) spine. #### 5. Intended Use: The Caspian Spinal System is indicated for the following: - DDD (neck pain of discogenic origin with degeneration of the disc as confirmed by patient history and . radiographic studies) - Spondylolisthesis . - Spinal stenosis . - . Fracture/dislocation - . Revision of previous cervical spine surgery - . Tumors The implants are intended to provide stabilization as an adjunct to fusion when used with autogenous bone graft or allograft following the reduction of fractures/dislocations or trauma in the spine. The rod and hook components are intended for use in the cervical/upper thoracic (C1-T3) spine. The pedicle screws are limited to placement in T1-T3 in treating thoracic conditions only. The pedicle screws are not intended to be placed in or treat conditions involving the cervical spine. The Caspian Spinal System can also be linked to the Range Spinal System using the 3.5mm/5.5mm rod connectors or transitional rods. {1}------------------------------------------------ - 6. Comparison of the technological characteristics of the device to predicate and legally marketed devices : The Caspian Spinal System is considered substantially equivalent to other legally marketed devices. They are similar in design, material, and indications for use and are expected to be equivalent in safety and effectiveness. . {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States of America. The seal features a stylized eagle with outstretched wings, symbolizing protection and care. Encircling the eagle is the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA", indicating the department's name and national affiliation. The seal is presented in black and white, giving it a formal and official appearance. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-066-0609 Silver Spring, MD 20993-0002 K2M, Inc. % Ms. Nancy Giezen Manager, Regulatory Affairs 751 Miller Drive, SE, Suite F-1 Leesburg, Virginia 20175 DEC 1 6 2009 Re: K093534 Trade/Device Name: Caspian Spinal System Regulation Number: 21 CFR 888.3050 Regulation Name: Spinal interlaminal fixation orthosis Regulatory Class: Class II Product Code: KWP, MNI Dated: November 13, 2009 Received: November 16, 2009 Dear Ms. Giezen: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set {3}------------------------------------------------ Page 2 - Ms. Nancy Giezen forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Jonita B Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # Indications for Use ## 510(k) Number (if known): K093534 Device Name: Caspian Spinal System Indications For Use: The Caspian Spinal System is indicated for the following: - DDD (neck pain of discogenic origin with degeneration of the disc as confirmed . by patient history and radiographic studies) - . Spondylolisthesis - . Spinal stenosis - . Fracture/dislocation - . Revision of previous cervical spine surgery - . Tumors The implants are intended to provide stabilization as an adjunct to fusion when used with autogenous bone graft or allograft following the reduction of fractures/dislocations or trauma in the spine. The rod and hook components are intended for use in the cervical/upper thoracic (C1-T3) spine. The pedicle screws are limited to placement in T1-T3 in treating thoracic conditions only. The pedicle screws are not intended to be placed in or treat conditions involving the cervical spine. The Caspian Spinal System can also be linked to the Range Spinal System using the 3.5mm/5.5mm rod connectors or transitional rods. Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Concurrence of --- (Division Sign-Off) Division of Surgical, Orthopedic, and Restorative ينان viets Page 1 of 510(k) Number K093534
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...