BLACKSTONE ASCENT POSTERIOR OCCIPITAL CERVICO-THORACIC (POCT) SYSTEM

K080394 · Blackstone Medical, Inc. · KWP · Mar 13, 2008 · Orthopedic

Device Facts

Record IDK080394
Device NameBLACKSTONE ASCENT POSTERIOR OCCIPITAL CERVICO-THORACIC (POCT) SYSTEM
ApplicantBlackstone Medical, Inc.
Product CodeKWP · Orthopedic
Decision DateMar 13, 2008
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 888.3050
Device ClassClass 2
AttributesTherapeutic

Intended Use

When intended to promote fusion of the cervical spine and occipito-cervico-thoracic junction (occiput-T3), the Blackstone Ascent POCT System is indicated for: - degenerative disc disease (defined as neck pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies: - b) spondylolisthesis: - c) fracture/dislocation: - d) spinal stenosis: - e) atlanto-axial fracture with instability; - f) occipito-cervical dislocation: - g) tumors; - h) revision of previous cervical spine surgery The occipital bone screws are limited to occipital fixation only. The use of the multi-axial screws is limited to placement in the upper thoracic spine (T1-T3) for the treatment of thoracic conditions only. They are not intended to be placed in the cervical spine. The lateral offset adapter is indicated for use in the upper thoracic spine (T1-T3). The hooks are intended to be placed from C1 to T3. The Songer Cable (titanium) System to be used with the Blackstone Ascent POCT System allows for wire/cable attachment to the posterior cervical spine. The Blackstone Ascent POCT System can also be linked to the Blackstone Spinal Fixation System using the Blackstone Ascent Axial or Parallel Connector.

Device Story

Temporary, non-sterile, single-use titanium alloy spinal instrumentation system. Components include rods, setscrews, cross connectors, axial/parallel connectors, lateral offset adapters, multi-axial screws, hooks, plates, bone screws, and Songer cables. Used by surgeons in clinical settings to build spinal implant constructs for stabilization and fusion from occiput to T3. System allows modular assembly to address specific spinal pathologies. Provides mechanical support to the spine; facilitates fusion; stabilizes fractures/instability. Linked to Blackstone Spinal Fixation System via connectors.

Clinical Evidence

No clinical data provided. Substantial equivalence based on design and technological characteristics.

Technological Characteristics

Titanium alloy spinal instrumentation system. Components: rods, screws, hooks, plates, connectors. Non-sterile, single-use. Mechanical fixation principle. No software or energy source.

Indications for Use

Indicated for patients requiring spinal fusion in the occipito-cervico-thoracic junction (occiput-T3) due to degenerative disc disease, spondylolisthesis, fracture/dislocation, spinal stenosis, atlanto-axial fracture with instability, occipito-cervical dislocation, tumors, or revision of previous cervical spine surgery. Contraindications: Multi-axial screws not for cervical spine; occipital screws limited to occipital fixation.

Regulatory Classification

Identification

A spinal interlaminal fixation orthosis is a device intended to be implanted made of an alloy, such as stainless steel, that consists of various hooks and a posteriorly placed compression or distraction rod. The device is implanted, usually across three adjacent vertebrae, to straighten and immobilize the spine to allow bone grafts to unite and fuse the vertebrae together. The device is used primarily in the treatment of scoliosis (a lateral curvature of the spine), but it also may be used in the treatment of fracture or dislocation of the spine, grades 3 and 4 of spondylolisthesis (a dislocation of the spinal column), and lower back syndrome.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K080394 Premarket Notification Special 510(k) Blackstone Medical, Inc. Blackstone™ Ascent* POCT System Plate and Rod Modifications (System Modification) Confidential #### 510(K) SUMMARY | Name of Firm: | Blackstone Medical, Inc.<br>1211 Hamburg Turnpike<br>Wayne, NJ 07470 | MAR 13 2008 | |------------------------------|-------------------------------------------------------------------------------|-------------| | 510(k) Contact: | Whitney Törning, Senior Director of Regulatory Affairs &<br>Quality Assurance | | | Submitter: | Martin Sprunck, Regulatory Affairs Manager | | | Trade Name: | Blackstone™ Ascent® Posterior Occipital Cervico-<br>Thoracic (POCT) System | | | Common Name: | Rod and screw spinal instrumentation | | | Device Classification: | Class II | | | Classification Product Code: | KWP – 888.3050 – Spinal Interlaminal Fixation Orthosis | | ## Substantially Equivalent Devices: Blackstone™ Posterior Cervical System (K030197 SE 6-12-03) Blackstone™ Ascent POCT System Lateral Offset Adaptors (K040034 SE 2-11-04) Blackstone™ Ascent POCT System Hooks (K033980 SE 3-3-04) Blackstone™ Ascent POCT System 5.5mm/3.0mm Single Axial Connector (K042100 SE 9-20-04) Blackstone™ Ascent POCT System Parallel Rod Connectors (K073654 SE 1-24-08) ## Device Description: Description: The Blackstone Ascent POCT System is a temporary, titanium alloy, multiple component system comprised of a variety of non-sterile, single use components that allow the surgeon to build a spinal implant construct. The Ascent POCT System consists of an assortment of rods, setscrews, cross connectors, axial connectors, lateral offset adapters, multi-axial screws, hooks, plates, bone screws, and Songer Cables. Levels of Use: When used in the occipito-cervico-thoracic spine, the Blackstone Posterior Cervical System may be used from the occiput to T3. {1}------------------------------------------------ Indications: When intended to promote fusion of the cervical spine and occipito-cervicothoracic junction (occiput-T3), the Blackstone Ascent POCT System is indicated for: - a) degenerative disc disease (defined as neck pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies: - b) spondylolisthesis: - c) fracture/dislocation: - d) spinal stenosis: - e) atlanto-axial fracture with instability; - f) occipito-cervical dislocation: - g) tumors; - h) revision of previous cervical spine surgery The occipital bone screws are limited to occipital fixation only. The use of the multi-axial screws is limited to placement in the upper thoracic spine (T1-T3) for the treatment of thoracic conditions only. They are not intended to be placed in the cervical spine. The lateral offset adapter is indicated for use in the upper thoracic spine (T1-T3). The hooks are intended to be placed from C1 to T3. The Songer Cable (titanium) System to be used with the Blackstone Ascent POCT System allows for wire/cable attachment to the posterior cervical spine. The Blackstone Ascent POCT System can also be linked to the Blackstone Spinal Fixation System using the Blackstone Ascent Axial or Parallel Connector. ## Basis of Substantial Equivalence: The modified Blackstone™ Ascent Occipital Plate and Occipital-Transition Rod are substantially equivalent to the identified predicate systems, which have been cleared by FDA for the purpose of building a spinal implant construct in the occipito-cervico-thoracic spine. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized human figure embracing a globe, symbolizing the agency's mission to protect the health of all Americans. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the emblem. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 # MAR 1 3 2008 Blackstone Medical, Inc. % Ms. Whitney Törning Senior Director of Regulatory Affairs and Quality Assurance 1211 Hamburg Turnpike, Suite 300 Wayne, NJ 07470 K080394 Re: Trade/Device Name: Ascent® Posterior Occipital Cervico-Thoracic (POCT) System Regulation Number: 21 CFR 888.3050 Regulation Name: Spinal interlaminal fixation orthosis Regulatory Class: II Product Code: KWP Dated: February 11, 2008 Received: February 13, 2008 Dear Ms. Törning: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {3}------------------------------------------------ Page 2 - Ms. Whitney Törning This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at toll-free number (800) 638-2041 or (240) 276-3150 or the Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Mark N Millen Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # Indications for Use 510(k) Number (if known): Device Name: Blackstone™ Ascent Posterior Occipital Cervical Thoracic System (POCT) System #### Indications for Use: When intended to promote fusion of the cervical spine and occipito-cervico-thoracic junction (occiput-T3), the Blackstone Ascent POCT System is indicated for: - degenerative disc disease (defined as neck pain of discogenic origin with degeneration of a) the disc confirmed by patient history and radiographic studies; - spondylolisthesis; b) - fracture/dislocation; c) - spinal stenosis; d) - atlanto-axial fracture with instability; e) - occipito-cervical dislocation; f) - tumors; g) - h) revision of previous cervical spine surgery The occipital bone screws are limited to occipital fixation only. The use of the multi-axial screws is limited to placement in the upper thoracic spine (T1-T3) for the treatment of thoracic conditions only. They are not intended to be placed in the cervical spine. The lateral offset adapter is indicated for use in the upper thoracic spine (T1-T3). The hooks are intended to be placed from C1 to T3. The Songer Cable (titanium) System to be used with the Blackstone Ascent Posterior Occipital Cervical Thoracic System allows for wire/cable attachment to the posterior cervical spine. The Blackstone Ascent Posterior Occipital Cervical Thoracic System can also be linked to the Blackstone Spinal Fixation System using the Blackstone Ascent Axial or Parallel Rod Connector. Prescription Use X AND/OR Over-The-Counter Use (Part 21 CFR 801 Subpart D) (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Page 1 of Concurrence of CDRH, Office of Device Evaluation (ODE) Neil R. Dykman Division of General, Restorative, and Neurological Devices **510(k) Number** k080394
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...