ULTIMA TPS CEMENTED STEM

K965198 · Johnson & Johnson International · KWL · Jul 23, 1997 · Orthopedic

Device Facts

Record IDK965198
Device NameULTIMA TPS CEMENTED STEM
ApplicantJohnson & Johnson International
Product CodeKWL · Orthopedic
Decision DateJul 23, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3360
Device ClassClass 2
AttributesTherapeutic

Intended Use

The ULTIMA* TPS Cemented Stem is indicated for use in total or partial hip replacement procedures for patients suffering severe pain and disability due to: 1. structural damage in the hip joint from rheumatoid arthritis, 2. osteoarthritis, 3. post-traumatic arthritis, 4. collagen disorders, 5. avascular necrosis, 6. nonunion of femoral fractures, 7. congenital hip dysplasia, 8. protrusio acetabuli, 9. slipped capital femoral epiphysis, and 10. disability due to previous fusion where bone stock is inadequate for other reconstruction techniques. The ULTIMA* TPS Cemented Stem is indicated for use only with PMMA bone cement.

Device Story

ULTIMA* TPS Cemented Stem is a hip joint metal/polymer semi-constrained cemented prosthesis; used in total or partial hip replacement procedures. Device forged from Cobalt-Chromium alloy; features straight, collarless, double-taper geometry; polished finish; 10/12 Morse taper. Available in six sizes with varying offsets and neck options; used with PMMA bone cement and distal centralizer to ensure orientation in medullary canal. Implanted by orthopedic surgeons in clinical settings. Provides structural support for hip joint; alleviates pain and disability; restores function. Substantial equivalence established via mechanical testing (2,300N force, 5,000,000 cycles).

Clinical Evidence

Bench testing only. Mechanical testing demonstrated the device withstands a maximum load of 2,300N force for 5,000,000 cycles, meeting established criteria for substantial equivalence.

Technological Characteristics

Cobalt-Chromium alloy (ASTM F799); PMMA distal centralizer. 10/12 Morse taper; collarless; straight, double-tapered stem. Neck angle 125°. Available in six sizes with multiple offset and neck length options. Non-sterile; intended for use with PMMA bone cement.

Indications for Use

Indicated for patients suffering severe pain and disability requiring total or partial hip replacement due to rheumatoid arthritis, osteoarthritis, post-traumatic arthritis, collagen disorders, avascular necrosis, femoral fracture nonunion, congenital hip dysplasia, protrusio acetabuli, slipped capital femoral epiphysis, or previous fusion with inadequate bone stock. Contraindicated for use without PMMA bone cement.

Regulatory Classification

Identification

A hip joint femoral (hemi-hip) metallic cemented or uncemented prosthesis is a device intended to be implanted to replace a portion of the hip joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum. This generic type of device includes designs which are intended to be fixed to the bone with bone cement (§ 888.3027) as well as designs which have large window-like holes in the stem of the device and which are intended for use without bone cement. However, in these latter designs, fixation of the device is not achieved by means of bone ingrowth.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} JUL-22-37 15:45 FROM:J & J/REG AFFAIRS ID:5088283212 PAGE 5/7 # K965198 ## Summary of Safety & Effectiveness Data for the ULTIMA* TPS Cemented Stem JUL 23 1997 Johnson & Johnson Professional, Inc. 325 Paramount Drive Raynham, Massachusetts 02767 1. Contact Person Janet G. Johnson, Assoc. Regulatory Affairs Specialist, (508) 828-3466. 2. Device Name Proprietary Name: ULTIMA* TPS Cemented Stem Common Name: Hip joint metal/polymer semi-constrained cemented prosthesis Classification: Prosthesis, Hip Name: Regulatory Class: Class II by 21 CFR § 888.3350 Product Code: 87JDI Owner/Operator #: 9001269 3. Device Classification Classification for ULTIMA* TPS Cemented Stem has been placed in Class II by 21 CFR § 888.3350 4. Statement of Substantial Equivalence The safety and effectiveness of the ULTIMA* TPS Cemented Stem is substantially equivalent in terms of function to the Howmedica Exeter hip stem (#K891454) and to the Zimmer CPT hip stem. Furthermore, testing results demonstrate that the ULTIMA* TPS Cemented Stem is substantially equivalent by withstanding a maximum load of 2,300N force, 5,000,000 cycles, and, as such meets the set criteria for the establishment of "substantial equivalence". {1} ID:5088283212 PAGE 6/7 # 5. Indications for Use The ULTIMA* TPS Cemented Stem is indicated for use in total or partial hip replacement procedures for patients suffering severe pain and disability due to: 1. structural damage in the hip joint from rheumatoid arthritis, 2. osteoarthritis, 3. post-traumatic arthritis, 4. collagen disorders, 5. avascular necrosis, 6. nonunion of femoral fractures, 7. congenital hip dysplasia, 8. protrusio acetabuli, 9. slipped capital femoral epiphysis, and 10. disability due to previous fusion where bone stock is inadequate for other reconstruction techniques. The ULTIMA* TPS Cemented stem is indicated for use only with PMMA bone cement. # 6. Physical Description The ULTIMA* TPS Cemented Stem is forged from a Cobalt-Chromium alloy which conforms to ASTM F799 (EXHIBIT IV); this material has high strength and good hardness properties and is also highly immune to fatigue. The ULTIMA* TPS Cemented Stem is polished to a bright finish on all surfaces below the spigot and can be used interchangeably in either the left or the right hip. The stem is straight, tapering towards the distal tip in both the mediolateral and anteriorposterior sections. It does not have a proximal collar. The ULTIMA* TPS Cemented Stem has a 10/12 Morse taper; it will be available in six sizes (0, 1 - 5), each of the 1 - 5 sizes have a choice of either a 37.5 mm (standard) offset, or a 44.0 mm (medialized "M") offset. In addition to this central range, there shall be two long stem options. The stem can be used with various neck heads (22.225, 26, 28 and 32 mm) as well as neck options (-5 (32), -3 (28), 0, +5 and +10). All of the stems in the range have a neck angle of 125°. | Stem Size | Trunnion Designs | Offset | Neck Heads | Neck Options | Neck Angle | | --- | --- | --- | --- | --- | --- | | 0,1,2,3,4,5 | 10/12 taper | 37.5 / 44.0mm | 22.225, 26, 28, 32 mm | -5 (32), -3 (28), 0, +5, +10 | 125° | {2} JUL-22-97 15:46 FROM: J & J/REG AFFAIRS ID:5088283212 PAGE 7/7 ## Distal Centralizer The distal centralizer that is used with the ULTIMA® TPS stem comes in a range of sizes and is manufactured from PMMA. There are three wings on the centralizer which ensure the orientation of the stem distally. These increase in size through the centralizer range to deal with the variation in anatomy of the medullary canal. In revision surgery, a "wingless" centralizer is required, thus a 7mm diameter version is available which simply acts to provide the void, and does not centralize the stem. Table 1. Similarities and Differences Matrix | | ULTIMA® TPS Cemented Stem | Howmedica Exeter hip stem | Zimmer CPT hip stem | | --- | --- | --- | --- | | DESIGN | | | | | Polished | Yes | Yes | Yes | | Collarless | Yes | Yes | Yes | | Double-taper geometry | Yes | Yes | Yes | | INTENDED USE | | | | | Total / partial hip replacement | Yes | Yes | Yes | | Used with distal centralizer | Yes | Yes | Yes | | MATERIALS | | | | | Orthinox™ (stainless steel alloy) | No | Yes | No | | Zimtron™ (stainless steel alloy) | No | No | Yes | | Cobalt-Chromium-Molybdenum | Yes | No | No | {3} DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Ms. Janet G. Johnson *Associate Regulatory Affairs Specialist Johnson & Johnson Professional, Inc. 325 Paramount Drive Raynham, Massachusetts 02767-0350 JUL 23 1997 Re: K965198 Ultima* TPS Cemented Stem Regulatory Class: II Product Codes: JDI and KWL Dated: April 22, 1997 Received: April 25, 1997 Dear Ms. Johnson: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {4} Page 2 - Ms. Janet G. Johnson This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours, ![img-0.jpeg](img-0.jpeg) Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5} # Indications for Use for the ULTIMA* TPS Cemented Stem Johnson & Johnson Professional, Inc. 325 Paramount Drive Raynham, MA 02767-0350 *Indications for Use* K965198 The ULTIMA* TPS Cemented Stem is indicated for use in total or partial hip replacement procedures for patients suffering severe pain and disability due to: 1. structural damage in the hip joint from rheumatoid arthritis, 2. osteoarthritis, 3. post-traumatic arthritis, 4. collagen disorders, 5. avascular necrosis, 6. nonunion of femoral fractures, 7. congenital hip dysplasia, 8. protrusio acetabuli, 9. slipped capital femoral epiphysis, and 10. disability due to previous fusion where bone stock is inadequate for other reconstruction techniques. The ULTIMA* TPS Cemented Stem is indicated for use only with PMMA bone cement. Prescription Use ☑ (Per 21 CFR 801.109) ![img-1.jpeg](img-1.jpeg)
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...