← Product Code [KWL](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/KWL) · K172906

# REMEDY PLUS Hip Spacer, UNITE PLUS Bone Cement (K172906)

_Osteoremedies, LLC · KWL · Aug 29, 2018 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/KWL/K172906

## Device Facts

- **Applicant:** Osteoremedies, LLC
- **Product Code:** [KWL](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/KWL.md)
- **Decision Date:** Aug 29, 2018
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3360
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

The REMEDY PLUS Hip Spacer, which consists of a modular head and stem, is indicated for temporary use (maximum 180 days) as an adjunct to total hip replacement (THR) in skeletally mature patients undergoing a two-stage procedure due to a septic process and where gentamicin and vancomycin are the most appropriate antibiotics based on the susceptibility pattern of the infecting micro-organism(s). The head and stem components are inserted into the acetabular cavity and femoral medullary canal, respectively, following removal of the existing acetabular and femoral components and radical debridement. The device is intended for use in conjunction with systemic antimicrobial antibiotic therapy (standard treatment approach to an infection). The REMEDY PLUS Hip Spacer is not intended for use for more than 180 days, at which time it must be explanted and a permanent device implanted or another appropriate treatment performed (e.g., resection arthroplasty, fusion, etc.). The UNITE PLUS Bone Cement is intended for the fixation of a REMEDY PLUS spacer device to the host bone.

## Device Story

REMEDY PLUS Hip Spacer is a modular, sterile, single-use temporary hip prosthesis; consists of PMMA head and stem with AISI 316 ESR stainless steel reinforcing structure; contains gentamicin and vancomycin. Used in two-stage revision surgery for periprosthetic joint infection (PJI). Surgeon removes existing components, performs radical debridement, and implants spacer using UNITE PLUS Bone Cement (antibiotic-loaded PMMA). Device acts as temporary joint replacement and local antibiotic delivery system for max 180 days. Facilitates patient mobility and infection management while awaiting definitive reimplantation. Benefits include reduced infection recurrence and improved clinical outcomes compared to single-antibiotic spacers.

## Clinical Evidence

Prospective/retrospective study of 130 subjects (49 GV Spacer, 81 G Spacer). Composite success (Stage 1/Stage 2) for GV Spacer (89.2%) significantly higher than G Spacer (71.6%, p=0.01054). Stage 2 success for GV Spacer (97.7%) superior to G Spacer (84.1%, p=0.0198). AKI rates (10.2%) within literature ranges (8.5-20%).

## Technological Characteristics

Materials: Polymethylmethacrylate (PMMA) and AISI 316 ESR stainless steel reinforcement. Antibiotics: Gentamicin and vancomycin. Form factor: Modular head and stem. Sterilization: Sterile. Connectivity: None. Standards: ISO 7206-4 (stem fatigue), ISO 7206-6 (neck fatigue), ISO 7206-13 (disassembly).

## Regulatory Identification

A hip joint femoral (hemi-hip) metallic cemented or uncemented prosthesis is a device intended to be implanted to replace a portion of the hip joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum. This generic type of device includes designs which are intended to be fixed to the bone with bone cement (§ 888.3027) as well as designs which have large window-like holes in the stem of the device and which are intended for use without bone cement. However, in these latter designs, fixation of the device is not achieved by means of bone ingrowth.

## Predicate Devices

- DePuy Prostalac Hip Temporary Prosthesis System
- NP Cements Genta HV Bone Cements ([K143100](/device/K143100.md))

## Reference Devices

- REMEDY Hip Spacer (2GC Hip Modular Spacer) ([K112470](/device/K112470.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

August 29, 2018

OsteoRemedies, LLC % Hollace Rhodes Senior Director, Orthopedic Regulatory Affairs Musculoskeletal Clinical Regulatory Advisers, LLC 1050 K Street. Suite 1000 Washington, District of Columbia 20001

## Re: K172906

Trade/Device Name: REMEDY PLUS Hip Spacer, UNITE PLUS Bone Cement Regulation Number: 21 CFR 888.3360 Regulation Name: Hip joint femoral (hemi-hip) metallic cemented or uncemented prosthesis Regulatory Class: Class II Product Code: KWL, KWY, MBB Dated: September 22, 2017 Received: September 22, 2017

## Dear Hollace Rhodes:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Mark N. Melkerson -S

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

| DEPARTMENT OF HEALTH AND HUMAN SERVICES<br>Food and Drug Administration | Form Approved: OMB No. 0910-0120<br>Expiration Date: 06/30/2020<br>See PRA Statement below. |
|-------------------------------------------------------------------------|---------------------------------------------------------------------------------------------|
|-------------------------------------------------------------------------|---------------------------------------------------------------------------------------------|

**Indications for Use**

| 510(k) Number (if known)       | K172906                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
|--------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name                    | REMEDY PLUS Hip Spacer                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Indications for Use (Describe) | The REMEDY PLUS Hip Spacer, which consists of a modular head and stem, is indicated for temporary use (maximum 180 days) as an adjunct to total hip replacement (THR) in skeletally mature patients undergoing a two-stage procedure due to a septic process and where gentamicin and vancomycin are the most appropriate antibiotics based on the susceptibility pattern of the infecting micro-organism(s).<br>The head and stem components are inserted into the acetabular cavity and femoral medullary canal, respectively, following removal of the existing acetabular and femoral components and radical debridement. The device is intended for use in conjunction with systemic antimicrobial antibiotic therapy (standard treatment approach to an infection).<br>The REMEDY PLUS Hip Spacer is not intended for use for more than 180 days, at which time it must be explanted and a permanent device implanted or another appropriate treatment performed (e.g., resection arthroplasty, fusion, etc.). |

| Type of Use (Select one or both, as applicable)                                                       |                                                                                    |
|-------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------|
| <label><input checked="checked" type="checkbox"/>Prescription Use (Part 21 CFR 801 Subpart D)</label> | <label><input type="checkbox"/>Over-The-Counter Use (21 CFR 801 Subpart C)</label> |

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***

The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services  
Food and Drug Administration  
Office of Chief Information Officer  
Paperwork Reduction Act (PRA) Staff  
[PRAStaff@fda.hhs.gov](mailto:PRAStaff@fda.hhs.gov)

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."

| FORM FDA 3881 (7/17) | Page 1 of 1 |
|----------------------|-------------|
|----------------------|-------------|

{3}------------------------------------------------

| ປະໄດ້ ປະຈຸບັນ (10) ປະລາ ບາວຣ໌ ສູງແນນ (ປະມານ ປະມານ ປະເທດ<br>국립                                             | (TIT)<br>f To I agay                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
|-----------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| গা<br>other september of the start of the start of the first of the first for the first for the first and | a mit<br>and with the nebrud sint pribategor stremmon business on the works in the was be<br>of this information collection, including suggestions for reducing this burden, to.<br>əri balıbuları şəhəd və qalında çəkilməsi ilə sinfinin əsasən və bir nə əmir nəbud əri<br>"<br>. WOJB 223900A JIAMS 47ATS ASS 3HT OT MAO3 237319MOJ AUOY GM32 TON OG*<br>See t to to to to to to to see a se to streme supper of vino se ilgas sin T<br>niformation unless it displays a currently valid OMB number<br>Food and Drug Administration<br>Department of Health and Human Services<br>Paperwork Reduction Act (PRA) Star<br>Office of Chief Information Officer<br>vog .zrld.leb1@fdb.ltdbg.gdg |
|                                                                                                           | CONTINUE ON A SEPARATE PAGE IF NEEDED.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| O Dueddus The Counter Use (2) CFR CFR 800 Support C)                                                      | aldesildde se (yod ro eno or other), as applically a<br>(0 heqdns 108 月刊2 12 ກຸຣິ 772 FFB) ອະດີ ການປະຊາວຂອງ (2                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
|                                                                                                           |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
|                                                                                                           | Indications for Use (Describe)<br>erod tron of essivel result SUI9 YCHNER & To nother it insme Ono8 SUA EUTH STIVIU and                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
|                                                                                                           | UNITE PLUS Bane Coment<br>Device Name                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
|                                                                                                           | 510(k) Number (if known<br>XXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXX                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Ported : 00 8MO :bevorqqA more<br>See PRA Statenent below<br>0202108100 Dete: 06300/2020                  | DEPATTAL<br>ndications for Use<br>For Forministration                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
|                                                                                                           |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |

{4}------------------------------------------------

# 510(k) Summary for REMEDY PLUS Hip Spacer and UNITE PLUS Bone Cement

| Submitter:                  | OsteoRemedies, LLC<br>6800 Poplar Avenue,<br>#120<br>Germantown, TN 38138<br>Phone: (901) 453-3141<br>Email: info@osteoremedies.com                                                                                                        |
|-----------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Official Correspondent:     | Ms. Hollace Saas Rhodes<br>Senior Director, Orthopedic Regulatory Affairs<br>Musculoskeletal Clinical Regulatory Advisers, LLC<br>1050 K Street NW, Suite 1000<br>Washington, DC 20001<br>Phone: (202) 552-5807<br>Email: hrhodes@mcra.com |
| Date Prepared:              | August 29, 2018                                                                                                                                                                                                                            |
| Trade/Proprietary Names:    | REMEDY PLUS Hip Spacer<br>UNITE PLUS Bone Cement                                                                                                                                                                                           |
| Common/Usual Names:         | Hip Spacer<br>PMMA bone cement with antibiotic for orthopedics                                                                                                                                                                             |
| Classification Information: | Class II<br>21 CFR 888.3360, Hip joint femoral (hemi-hip) metallic<br>cemented or uncemented prosthesis<br>21 CFR 888.3027, Polymethylmethacrylate (PMMA) bone cement                                                                      |
| Product Codes:              | KWL, KWY, and MBB for the REMEDY PLUS Hip Spacer<br>MBB for the UNITE PLUS Bone Cement                                                                                                                                                     |
| Predicate Devices:          | DePuy Prostalac Hip Temporary Prosthesis System<br>NP Cements Genta HV Bone Cements (K143100)                                                                                                                                              |
| Reference Device:           | REMEDY Hip Spacer cleared as the 2GC Hip Modular Spacer<br>(K112470)                                                                                                                                                                       |

{5}------------------------------------------------

#### Device Description:

The REMEDY PLUS Hip Spacer is comprised of modular components, which are available in different sizes to accommodate variations in patient anatomy. The REMEDY PLUS Hip Spacer includes stems in various sizes and lengths, and femoral heads in various diameters. The modular design of the REMEDY PLUS Hip Spacer allows the surgeon to choose a femoral head offset which best matches the patient's anatomy.

The REMEDY PLUS Hip Spacer is a sterile, single-use device intended for temporary use (maximum 180 days) as a joint replacement. The implants are made of fully formed polymethylmethacrylate (PMMA), which is radiopaque, and contains gentamicin and vancomycin. The hip device has an inner stainless steel (AISI 316 ESR stainless steel) reinforcing structure.

The UNITE PLUS Bone Cement is a PMMA, radiopaque bone cement, containing gentamicin and vancomycin, designed for the fixation of prosthesis to the living bone. The UNITE PLUS Bone Cement is provided sterile.

#### Indications for Use and Intended Use:

#### REMEDY PLUS Hip Spacer

The REMEDY PLUS Hip Spacer, which consists of a modular head and stem, is indicated for temporary use (maximum 180 days) as an adjunct to total hip replacement (THR) in skeletally mature patients undergoing a two-stage procedure due to a septic process and where gentamicin and vancomycin are the most appropriate antibiotics based on the susceptibility pattern of the infecting micro-organism(s).

The head and stem components are inserted into the acetabular cavity and femoral medullary canal, respectively, following removal of the existing acetabular and femoral components and radical debridement. The device is intended for use in conjunction with systemic antimicrobial antibiotic therapy (standard treatment approach to an infection).

The REMEDY PLUS Hip Spacer is not intended for use for more than 180 days, at which time it must be explanted, and a permanent device implanted or another appropriate treatment performed (e.g., resection arthroplasty, fusion, etc.).

#### UNITE PLUS Bone Cement

The UNITE PLUS Bone Cement is intended for the fixation of a REMEDY PLUS spacer device to the host bone.

{6}------------------------------------------------

#### Substantial Equivalence Comparison Overview

#### REMEDY PLUS Hip Spacer

The REMEDY PLUS Hip Spacer is similar to the predicate device with respect to the following:

- 1) the same intended use in the same surgical procedures in the same target patient population
- 2) similar overall design, having same geometry and sizes as the REMEDY Hip Spacer
- 3) similar manufacturing processes and steps at the same manufacturing facility as the REMEDY Hip Spacer
- 4) similar mechanical performance characteristics
- 5) similar clinical performance, and
- 6) similar instructions for use and draft labeling.

There are differences between the REMEDY PLUS Hip Spacer and the predicate device, though these differences, shown through pre-clinical and clinical information, do not adversely impact the substantial equivalence of the subject device. Specifically:

- 1) the REMEDY PLUS Hip Spacer has gentamicin and vancomycin; the REMEDY Hip Spacer contains gentamicin; and the 2-antibiotic predicate contains tobramycin and vancomycin added at time of implantation.
- 2) the REMEDY PLUS Hip Spacer and REMEDY Hip Spacer are pre-made spacers; the 2-antibiotic predicate is molded intra-operatively.

Please see the Performance Testing Overview and Clinical Study Overview for comparison data supporting the substantial equivalence of the REMEDY PLUS Hip Spacer implanted with UNITE PLUS Bone Cement compared to predicate spacers.

#### UNITE PLUS Bone Cement

The UNITE PLUS Bone Cement device is similar to the predicate device with respect to the following:

- 1) both are antibiotic-loaded bone cements
- 2) similar main cement components
- 3) similar manufacturing processes and steps at the same manufacturing facility
- 4) similar mechanical performance characteristics, and
- 5) similar in vitro and in vivo elution profiles.

There are differences between the UNITE PLUS Bone Cement, NP Cements Genta HV Bone Cement, and cement used to fix the 2-antibiotic predicate, though these differences do not impact the substantial equivalence of the subject device relative to the predicate. Specifically:

{7}------------------------------------------------

- 1) UNITE PLUS Bone Cement has gentamicin and vancomycin; the NP Cements Genta HV Bone Cement contains only gentamicin; and the cement used to fix the 2antibiotic spacer contains tobramycin and vancomycin
- 2) UNITE PLUS Bone Cement is intended for the fixation of REMEDY PLUS Spacer device to the host bone; NP Cements Genta HV Bone Cement is intended for the fixation of the prosthesis to the living bone in the second stage of a two stage revision for total joint arthroplasty after the initial infection has been cleared; cement mixed intra-operatively with tobramycin and vancomycin is used for the fixation of the 2antibiotic spacer, and
- 3) UNITE PLUS Bone Cement does not have the colorants used in NP Cements Genta HV Bone Cement.

Please see the Performance Testing Overview and Clinical Study Overview for comparison data supporting the substantial equivalence of the REMEDY PLUS Hip Spacer implanted with UNITE PLUS Bone Cement compared to the predicate spacer.

## Performance Testing Overview:

Performance data for the REMEDY PLUS Hip Spacer, the UNITE PLUS Bone Cement, and their predicate devices demonstrate their substantial equivalence. The following testing was performed:

- Stem fatigue per ISO 7206-4
- Neck fatigue per ISO 7206-6 ●
- Femoral head/stem disassembly based on ISO 7206-13 ●
- Chemical and physical properties of cement ●
- Antibiotic elution kinetics
- . Clinical data
- Biocompatibility data
- Sterilization, pyrogenicity, bacterial endotoxin, and shelf life testing ●

## Clinical Study Overview:

Prospective and retrospective data were collected outside of the United States on subjects with a diagnosis of periprosthetic joint infection (PII) who were implanted with an antibiotic spacer during the first stage of a two-stage revision due to a PJI. One antibiotic spacer contained gentamicin and vancomycin ("GV Spacer") having data collected prospectively. The other one contained gentamicin only ("G Space") having the majority of its data collected retrospectively, but some prospectively. The antibiotic formulation and PMMA in the GV Spacer are identical to the REMEDY PLUS Spacers. The antibiotic formulation and PMMA in the G Spacer are identical to the REMEDY Spacers. The study included 49 subjects (26 hips and 23 knees) implanted with the GV Spacer, and 81 subjects (51 hips and 31 knees) implanted with the G Spacer which is 510(k)-cleared though has only one antibiotic. The GV and G spacer data were also compared to the predicate hip spacer and literature data.

{8}------------------------------------------------

#### Composite Endpoint (Stage 1/Stage 2 Success)

Success for the composite endpoint required Stage 1 success (i.e., absence of Girdlestone (hip fusion), arthrodesis (knee fusion), amputation, or spacer-related death) and Stage 2 success (i.e., absence of two or more positive cultures of microorganisms at the time of reimplantation with the definitive implant).

| All Subjects                        | GV Spacer |     |       | G Spacer |    |       | Dif.  |         |
|-------------------------------------|-----------|-----|-------|----------|----|-------|-------|---------|
|                                     | N         | n   | %     | N        | n  | %     | %     | p-value |
| Composite – All Evaluable (N=129) 1 | 48        | 43  | 89.6% | 81       | 58 | 71.6% | 18%   | 0.01262 |
| Composite – MI (N=130)3             | 49        | --- | 89.2% | 81       | 58 | 71.6% | 17.6% | 0.01054 |
| Hip Subjects                        | GV Spacer |     |       | G Spacer |    |       | Dif.  |         |
|                                     | N         | n   | %     | N        | n  | %     | %     | p-value |
| Composite (N=76)                    | 25        | 21  | 84%   | 51       | 34 | 66.7% | 17.3% |         |

## Composite Success (Stage 1 / Stage 2 Success)

1 For this comparison, one patient with good clinical status but with a definitive implant was defined as a composite success and one patient with unrelated death prior to reimplantation with a definitive implant were censored.

2 One-sided Fisher's Exact Test.

3 The number of successes is not reported for GV Spacer under MI (multiple imputation). 89.2% is the average across 20 MIS. For this comparison, the outcome for one GV patient with unrelated death prior to reimplantation with a definitive implant was imputed.

4 One-sided T-test from MI taking into account between and within imputation variance.

ª N = All subjects implanted with each type of spacer. Definition applies to all tables.

b n = All subjects meeting the specified endpoint. Definition applies to all tables.

For the "All Subjects" cohort, the composite success rate for the GV Spacer is significantly higher than the composite success rate for the G Spacer for both the "all evaluable" and "multiple imputation" analyses. For the "Hip Subjects" cohort, the composite success rate for the GV Spacer is 17.3% higher than the composite success rate for the G Spacer.

A composite endpoint based on Stage 1 and Stage 2 outcomes was not defined for the retrospective study of the 2-antibiotic predicate. Nonetheless, it is reasonable to consider the absence of persistent/recurrent infection at both Stage 1 and Stage 2 as a composite endpoint. At Stage 1, the absence of infection rate is 91.1% (123/135); at Stage 2, the rate is 89.3% (100/112). Combined, the predicate success rate is 82.2% for Stage 1 and Stage 2 which is comparable to the composite success rate for the GV Spacer.

In the Rothman Institute publication (Gomez, et al., 2015), "Retrospective multi-center study on outcomes of PJI patients treated with various antibiotic cement spacers", outcomes were captured for 504 hip and knee PJI subjects. A Stage 1 success rate was calculated using the same definition used in the GV vs. G study (i.e., absence of Girdlestone, arthrodesis, amputation, or death). When the lostto-follow-up subjects (n=19) are excluded, the Stage 1 success rate was calculated to be 89.5% (434/485) which includes 417 subjects who were reimplanted and 17 patients with retained spacers who were considered to be treatment successes. When the lost-to-follow-up subjects (n=85) are excluded, the Stage 2 success rate was calculated to be 80.7% (268/332). Overall composite success requires success at both stages. When the lost-to-follow-up subjects are excluded at both stages, the overall

{9}------------------------------------------------

composite success rate for Stage 1 and Stage 2 was calculated to be 71.3% (285/400) which is lower than the composite success rate for the GV Spacer.

### Stage 1 Outcomes

Success at Stage 1 is defined as the absence of Girdlestone (hip fusion), arthrodesis (knee fusion), amputation, or spacer-related death.

| All Subjects          | GV Spacer |    |       | G Spacer |    |       | Diff. | Exact    |
|-----------------------|-----------|----|-------|----------|----|-------|-------|----------|
|                       | N         | n  | %     | N        | n  | %     | %     | p-value* |
| Total Enrolled        | 49        |    |       | 81       |    |       |       |          |
| Unrelated deaths      | 1         |    |       | 0        |    |       |       |          |
| Stage 1 Successes     | 48        | 44 | 91.7% | 81       | 69 | 85.2% | 6.5%  | 0.213    |
| THA                   | 25        | 22 | 88%   | 51       | 43 | 84.3% |       |          |
| TKA                   | 23        | 22 | 95.7% | 30       | 26 | 86.7% |       |          |
| Stage 1 Failures      | 48        | 4  | 8.3%  | 81       | 12 | 14.8% | -6.5% |          |
| Girdlestone           | 48        | 2  | 4.2%  | 81       | 8  | 9.9%  |       |          |
| Arthrodesis           | 48        | 0  | 0%    | 81       | 3  | 3.7%  |       |          |
| Amputation            | 48        | 1  | 2.1%  | 81       | 1  | 1.2%  |       |          |
| Death                 | 48        | 1  | 2.1%  | 81       | 0  | 0%    |       |          |
| Hip Subjects          | GV Spacer |    |       | G Spacer |    |       | Diff. |          |
|                       | N         | n  | %     | N        | n  | %     | %     |          |
| Total Enrolled        | 26        |    |       | 51       |    |       |       |          |
| Unrelated deaths      | 1         |    |       | 0        |    |       |       |          |
| Stage 1 THA Successes | 25        | 22 | 88%   | 51       | 43 | 84.3% | 3.7%  |          |
| Stage 1 Failures      | 25        | 3  | 12%   | 51       | 8  | 15.7% | -3.7% |          |
| Girdlestone           | 25        | 2  | 8%    | 51       | 8  | 15.7% |       |          |
| Death                 | 25        | 1  | 4%    | 51       | 0  | 0%    |       |          |

|  | Stage 1 Outcomes |
|--|------------------|

*Two-sided Fisher's Exact p-value.

For the "All Subjects" cohort, the rate of Stage 1 success for the GV Spacer cohort is 91.7% which is higher than the Stage 1 success rate of 85.2% for the G Spacer cohort, though this difference is not significant (two-sided Fisher's Exact p=0.213). For the "Hip Subjects" cohort, the Stage 1 success rate for the GV Spacer is 3.7% higher than the Stage 1 success rate for the G Spacer.

The outcomes for the GV Spacer were also compared to: (1) a retrospective study of 135 patients implanted with the 2-antibiotic predicate, and (2) literature.

At Stage 1, 6 (4.4%) of the 135 2-antibiotic predicate subjects had a Girdlestone procedure, 12 (8.9%) had persistent/recurrent deep infections, 8 (5.9%) had an intra-operative bone fracture, and 5 (3.7%) had a post-operative bone fracture. Seventeen (17) of the 135 subjects (12.6%) did not have a reimplantation with a definitive implant. These 17 subjects had a retained predicate, received a second spacer device, or died.

{10}------------------------------------------------

A publication from the Rothman Institute reports on 504 hip and knee PJI subjects. When the lost-tofollow-up subjects (n=19) are excluded, the Stage 1 success rate was calculated to be 89.5% which is comparable to the Stage 1 success rate for the GV Spacer. The 10.5% failure rate included 6 amputations, 5 Girdlestones, 4 arthrodeses, and 36 deaths (Gomez, et al., 2015).

### Safety - Inter-Stage Reoperations

Inter-stage reoperations included spacer exchanges and debridements.

| All Subjects       | GV Spacer |   |       | G Spacer |   |       |
|--------------------|-----------|---|-------|----------|---|-------|
|                    | N         | n | %     | N        | n | %     |
| Spacer exchange    | 48        | 5 | 10.4% | 81       | 8 | 9.9%  |
| Spacer debridement | 48        | 3 | 6.25% | 81       | 0 | 0%    |
| Hip Subjects       | GV Spacer |   |       | G Spacer |   |       |
|                    | N         | n | %     | N        | n | %     |
| Spacer exchange    | 25        | 3 | 12%   | 51       | 6 | 11.8% |

## Inter-stage Reoperations

For both the "All Subjects" and "Hip Subjects" cohorts, the GV and G cohorts exhibit similar rates of spacer exchange. The rate of spacer debridement for the GV Spacer is higher than what is reported for the G spacer.

The inter-stage reoperation rates were also compared to the 2-antibiotic predicate study and literature.

## Inter-stage Reoperations

|                          | 2-Antibiotic Predicate |  |      | Gomez (2015) |    |       | Cancienne (2017)** |     |       |
|--------------------------|------------------------|--|------|--------------|----|-------|--------------------|-----|-------|
| Inter-stage Reoperations |                        |  |      |              |    |       |                    |     | 0/0   |
| Spacer exchange          | 118                    |  | 6.8% | 504          | 60 | 11.9% |                    |     |       |
| Spacer debridement       | 135                    |  |      |              |    |       | 7146               | 775 | 10.8% |

* Although the number of debridenent procedures was not identified for the predicate, a publication (Wentworth, et d., 2002) reports that 12 subjects (8.9%) had persistent/recurrent deep infections at Stage 1.

** Cancienne, J. M. et al. Removal of an Infected Total Hip Arthroplasy; Risk Factors for Repeat Debridenent, Long-term Spacer Retention, and Mortality The Journal of Arthroplasty , Volume 32 , Issue 8 , 2519 - 2522.

The inter-stage reoperation rates for the GV Spacer, including spacer exchanges and debridements, are similar to those reported for the predicate and/or literature.

## Safety - Acute Kidney Injury (AKI)

In the "All Subjects" cohort, acute kidney injury (AKI) was reported in 5 of the 49 subjects (10.2%) implanted with the GV Spacer, and 3 of the 81 subjects (3.7%) implanted with the G Spacer. These rates are within the range of what is reported in the literature for patients being treated for PJI with antibiotic-loaded spacers (i.e., 8.5%) to 20%2) (Reed et al., 2014; Aeng et al., 2015). Published literature indicates that high-dose antibiotic spacers made in the operating room, such as the predicate, may be associated with a higher risk of nephrotoxicity when compared to preformed, low-dose antibiotic spacers such as the GV Spacer (Luu et al., 2013; Edelstein et al., 2018). In the "Hip Subjects"

{11}------------------------------------------------

cohort, AKI was reported in 4 of 25 subjects (16%) implanted with a GV Hip Spacer, and 3 of the 51 subjects (5.9%) of the subjects implanted with the G Spacer. Again, these are within the range of AKI rates reported in the literature for patients undergoing treatment for PJI. One patient with chronic kidney disease, hypertension, and cardiopathy experienced an AKI-related death.

Patients should be monitored for AKI while undergoing treatment for PJI, as the combination of systemic antibiotics, drugs prescribed to treat any comorbidities, and the antibiotics present in the spacer can all contribute to the risk of AKI. Please refer to the warnings and precautions in the package insert for the REMEDY PLUS Hip Spacer for additional information.

#### Effectiveness - Stage 2 Outcomes

Success at the second stage of a two-stage revision is defined as the absence of two or more positive cultures of microorganisms at the time of reimplantation.

| All Subjects       | GV Spacer |    |       | G Spacer |    |       | Diff. | Exact    |
|--------------------|-----------|----|-------|----------|----|-------|-------|----------|
|                    | N         | n  | %     | N        | n  | %     | %     | p-value* |
| Success at Stage 2 | 43        | 42 | 97.7% | 69       | 58 | 84.1% | 13.6% | 0.0198   |
| Hip Subjects       | GV Spacer |    |       | G Spacer |    |       | Diff. |          |
|                    | N         | n  | %     | N        | n  | %     | %     |          |

## Stage 2 Outcomes

*One-sided Fisher's Exact Test.

For the "All Subjects" cohort, the Stage 2 success rate for the GV Spacer of 97.7% is statistically superior to the success rate for the G Spacer of 84.1% (one-sided Fisher's Exact p = 0.0198). For the "Hip Subjects" cohort, the Stage 2 success rate for the GV Spacer was 95.5%, which is 16.4% higher than the Stage 2 success rate for the G Spacer subjects of 79.1%.

In the retrospective study of the 2-antibiotic predicate, a successful treatment outcome was defined as no growth of a microorganism on any culture obtained from the operative site during the second stage surgery. A successful clinical outcome was defined as implantation of a total hip prosthesis at the second stage surgery with no growth of a microorganism on any culture obtained from the operative site and no reoperations resulting from a recurrent or persistent infection of the affected hip at a minimum of one month postoperatively. Of the 135 subjects enrolled in the retrospective study, 116 were included in the evaluation of Stage 2 success. Of these, 96 (83%) were considered a treatment success. Note: Seventeen (17) of the 135 subjects (12.6%) did not have a reimplantation with a definitive implant. These 17 subjects had a retained spacer device, received a second spacer device, or died. Two additional subjects were excluded from the evaluation of Stage 2 outcomes because their intraoperative cultures could not be confirmed. Successful clinical outcomes were reported for 91.7 percent (89/97) of the cases with the minimum required one-month follow-up.

{12}------------------------------------------------

In the Rothman Institute publication (Gomez, et al., 2015), the Stage 2 success rate was calculated to be 80.7% (268/332). The Stage 2 success rate for the GV Spacer compares favorably to the rate reported by Gomez.

#### Secondary Outcome - Antibiotic Usage

Success at Stage 2 was associated with a shorter mean total duration of systemic antibiotic usage after the second stage surgery. 93% of the GV subjects were finished with their antibiotic course within a week of their second stage surgery. In contrast, only 25% of the G subjects were completed with antibiotics within the first post-operative week after the second stage surgery.

#### Conclusions:

The nonclinical and clinical testing demonstrates that the REMEDY PLUS Hip Spacer and UNITE PLUS Bone Cement are substantially equivalent to the predicate device.

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/KWL/K172906](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/KWL/K172906)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
