GREAT TOE IMPLANT

K021063 · Signal Medical Corp. · KWD · Jul 1, 2002 · Orthopedic

Device Facts

Record IDK021063
Device NameGREAT TOE IMPLANT
ApplicantSignal Medical Corp.
Product CodeKWD · Orthopedic
Decision DateJul 1, 2002
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3730
Device ClassClass 2
AttributesTherapeutic

Intended Use

Osteoarthritis . 2. Where the use of a more conservative procedure has failed or is unacceptable. The device is intended for non-cemented use.

Device Story

Great Toe Implant is a hemi-toe prosthesis designed for non-cemented implantation. The device stem is inserted into the proximal phalangeal bone, while the convex surface articulates with the first metatarsal head. Used by orthopedic surgeons in clinical settings to replace the joint surface in patients with osteoarthritis. The implant provides a mechanical bearing surface to restore joint function and alleviate pain when conservative treatments are ineffective.

Clinical Evidence

No clinical data provided; substantial equivalence is based on design, material, and intended use comparisons to legally marketed predicate devices.

Technological Characteristics

Material: ASTM F75 Cobalt Chrome. Surface: ASTM F75 sintered beads (-45+60 mesh) for porous coating. Design: Hemi-toe prosthesis with convex articulating surface and stem for non-cemented fixation. Dimensions: Multiple sizes available.

Indications for Use

Indicated for patients with osteoarthritis where conservative procedures have failed or are unacceptable.

Regulatory Classification

Identification

A toe joint phalangeal (hemi-toe) polymer prosthesis is a device made of silicone elastomer intended to be implanted to replace the base of the proximal phalanx of the toe.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ JUL 1 2002 K021063 page loft ## 510K SUMMARY OF SAFETY AND EFFECTIVENESS The Signal Medical Corporation Great Toe Implants are manufactured from ASTM F75 Cobalt Chrome. The porus coating is ASTM F75 sintered beads, -45+60 mesh. The device is intended for non-cemented use. The stem of the device is intended to be inserted into the proximal end or base of the first phalangel bone. The convex surface of the implant is intended to articulate with the distal end or head of the first metatarsal bone. The design is made available in multiple sizes equivalent to the toe implants in the Townley Great Toe Joint 510K (K911378) and the Zimmer Great Toe 510K (K823405). Indications for use - Osteoarthritis . 2. Where the use of a more conservative procedure has failed or is unacceptable. ## Contact Information If further information is required please contact Dr. Louis Serafin, Signal Medical Corporation, 3777 Lapper Road Suite 3C, Port Huron, Michigan 48060, phone (810) 966-3917. {1}------------------------------------------------ Image /page/1/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo consists of a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with three curved lines above them. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 ## 1 2002 JUL Louis A. Serafin, Jr., M.D. Signal Medical Corporation 3777 Lapeer Road. Suite 3C Port Huron, Michigan 48060 Re: K021063 Trade/Device Name: Great Toe Implant Regulation Number: 21 CFR 888.3730 Regulation Name: Toe joint phalangeal (hemi-toe) polymer prosthesis Regulatory Class: Class II Product Code: KWD Dated: March 19, 2002 Received: April 2, 2002 Dear Dr. Serafin, Jr .: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {2}------------------------------------------------ Page 2 - Dr. Louis A. Serafin This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours, Stipt Rhodes Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ Page i of 1 510(k) Number (if known):__________________________________________________________________ Device Name:__________________________________________________________________________________________________________________________________________________________________ Indications For Use: Osteoarthritis Where the use of a more conservative procedure has failed or is unacceptable failed or is unacceptable. The device is intended for non-cemented use. (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Stupt Rurly Division of General, Restorative and Neurological Devices Over-The-Counter Use___ 510(k) Number (Optional Formal 1-2-96) Prescription Use_ (Per 21 CFR 801.109)
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...