ANTERIOR APPROACH HIP SURGERY INSTRUMENTS

K102565 · Wrightmedicaltechnologyinc · KWA · Dec 22, 2010 · Orthopedic

Device Facts

Record IDK102565
Device NameANTERIOR APPROACH HIP SURGERY INSTRUMENTS
ApplicantWrightmedicaltechnologyinc
Product CodeKWA · Orthopedic
Decision DateDec 22, 2010
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 888.3330
Device ClassClass 3
AttributesTherapeutic

Intended Use

The Wright Total Hip Implant Stems for use with these instruments are indicated for use in total hip arthroplasty for reduction or relief of pain and/or improved hip function in skeletally mature patients with the following conditions: 1. non-inflammatory degenerative joint disease such as osteoarthritis, avascular necrosis, ankylosis, protrusio acetabuli, and painful hip dysplasia; 2. inflammatory degenerative joint disease such as rheumatoid arthritis; 3. correction of functional deformity; and, 4. revision procedures where other treatments or devices have failed

Device Story

Anterior Approach Hip Surgery Instruments consist of stainless steel surgical tools including broaches, inserters/impactors, chisels, and rasps. Used by orthopedic surgeons in clinical settings to prepare the proximal femur and impact hip implants during total hip arthroplasty. Instruments feature specific geometries (e.g., lateral surface curvature) to facilitate access to the proximal femur via the anterior approach without interfering with the greater trochanter. Modular components allow for familiar orientation and handling. These manual instruments assist the surgeon in bone removal and implant positioning, directly impacting the accuracy of the arthroplasty procedure and potentially improving patient outcomes by facilitating less invasive surgical access.

Clinical Evidence

No clinical data provided. Safety and effectiveness supported by substantial equivalence analysis and bench evaluation by design surgeons in a surgical setting.

Technological Characteristics

Manufactured from stainless steel. Manual surgical instruments including broaches, chisels, inserters, and rasps. Designed for anterior approach hip surgery. No energy source, connectivity, or software components.

Indications for Use

Indicated for skeletally mature patients undergoing total hip arthroplasty for non-inflammatory degenerative joint disease (osteoarthritis, avascular necrosis, ankylosis, protrusio acetabuli, painful hip dysplasia), inflammatory degenerative joint disease (rheumatoid arthritis), functional deformity correction, or revision procedures.

Regulatory Classification

Identification

A hip joint metal/metal semi-constrained, with an uncemented acetabular component, prosthesis is a two-part device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that consist of a femoral and an acetabular component, both made of alloys, such as cobalt-chromium-molybdenum. The femoral component is intended to be fixed with bone cement. The acetabular component is intended for use without bone cement (§ 888.3027).

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K102565 (pg 1/2) ## 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS DEC 2 2 2010 In accordance with the Food and Drug Administration Rule to implement provisions of the Safe Medical Devices Act of 1990 and in conformance with 21 CFR 807, this information serves as a Summary of Safety and Effectiveness for the Anterior Approach Hip Surgery Instruments. | (a)(1) Submitted By<br>Submitter's Name: | Wright Medical Technology, Inc.<br>5677 Airline Rd. Arlington, TN 38002<br>800-238-7188 (phone), 901-867-4190 (fax) | |---------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------| | Date:<br>Contact Person: | December 20, 2010<br>Gregory Neal<br>Regulatory Affairs Specialist II | | (a)(2) Device Name<br>Proprietary Name:<br>Common Name: | Anterior Approach Hip Surgery Instruments<br>Broach, Inserter/Impactor, Chisel, Rasp | | Classification Name and Reference: | 21 CFR 888.3330 Hip joint metal/ metal<br>semi-constrained, with an uncemented<br>acetabular component prosthesis - Class III | | Subject Device Product Code and Panel Code: | Orthopedics/87/KWA, JDI, JDL, LWJ, LZO | | (a)(3) Predicate Device<br>Predicate Proprietary Name: | PROFEMUR® LX 5/8 Hip Stem (K081090)<br>PROFEMUR® TL Hip Stem (K060358) | | Predicate Classification and Number: | Orthopedics/87/ KWA, 888.3330 | #### (a)(4) Device Description The design features of the Anterior Approach Hip Surgery Instruments are summarized below: - Manufactured from Stainless Steel . - Modular PROFEMUR® Starter Broach provides means to begin preparation of . proximal femur; overall size envelope is smaller than any existing PROFEMUR® stem, allowing usage with all designs; curvature of lateral surface allows insertion without interfering with greater trochanter; engagement with broach handle allows usage in familiar orientation - Modular PROFEMUR® Box Chisel provides means to remove bone in calcar region, . creating opening for other instruments; curvature of lateral surface allows insertion without interfering with greater trochanter; engagement with broach handle allows usage in familiar orientation {1}------------------------------------------------ - Modular PROFEMUR® Dimple Stem Inserter provides means by which implant may . be impacted via engagement of dimple on lateral shoulder; engagement with broach handle allows usage in familiar orientation - Modular PROFEMUR® Pocket Stem Inserter provides means by which implant may . be impacted via engagement of modular neck pocket; engagement with broach handle allows usage in familiar orientation - Lateralizing Rasp provides rigid, aggressively-toothed surface at end of handle that . can be used to contour the proximal region of the femoral canal. - Dual-Offset Broach Handles Right and Left broach handles provide improved access . to the proximal femur ## (a)(5) Intended Use The Wright Total Hip Implant Stems for use with these instruments are indicated for use in total hip arthroplasty for reduction or relief of pain and/or improved hip function in skeletally mature patients with the following conditions: - 1. non-inflammatory degenerative joint disease such as osteoarthritis, avascular necrosis, ankylosis, protrusio acetabuli, and painful hip dysplasia; - 2. inflammatory degenerative joint disease such as rheumatoid arthritis; - 3. correction of functional deformity; and, - 4. revision procedures where other treatments or devices have failed #### (a)(6) Technological Characteristics of the Device The indications for implants used with the Anterior Approach Hip Surgery Instruments are identical to instruments used with implants cleared by PROFEMUR® LX 5/8 Hip Stem and PROFEMUR® TL Hip Stem. Materials for the Anterior Approach Hip Surgery Instruments are identical to instruments used with implants cleared by PROFEMUR® TL Hip Stem. The design features of the Anterior Approach Hip Surgery Instruments are substantially equivalent to the predicate devices. The fundamental scientific technology of the modified device has not changed relative to the predicate devices. #### (b)(1) Nonclinical Testing Prototypes were used and further evaluated by the design surgeons in a surgical setting. #### (b)(2) Clinical Testing Clinical data was not provided for the Class I instruments. ## (b)(3) Conclusions The safety and effectiveness of the Anterior Approach Hip Surgery Instruments are adequately supported by the substantial equivalence information, materials information, and analysis data provided within this Premarket Notification. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized graphic of an eagle or bird-like figure with three curved lines representing its wings or body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the graphic. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 Wright Medical Technology, Inc. % Mr. Gregory M. Neal Regulatory Affairs Specialist II 5677 Airline Road Arlington, TN 38002 DEC 2 2 2910 K102565 Trade/Device Name: Anterior Approach Hip Surgery Instruments Regulation Number: 21 CFR 888.3330 Regulation Name: Hip joint metal/metal semi-constrained, with an uncemented acetabular component, prosthesis Regulatory Class: Class III Product Code: KWA, JDI, JDL, LWJ, LZO Dated: November 24, 2010 Received: November 26, 2010 Dear Mr. Neal: Re: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set {3}------------------------------------------------ Page 2 - Mr. Gregory M. Neal CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours. A.S. B. R. fu Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ DEC 2 2 2010 # Indications for Use 510(k) Number (if known): Device Name: Anterior Approach Hip Surgery Instruments Indications For Use: The Wright Total Hip Implant Stems for use with these instruments are indicated for use in total hip arthroplasty for reduction or relief of pain and/or improved hip function in skeletally mature patients with the following conditions: - 1. non-inflammatory degenerative joint disease such as osteoarthritis, avascular necrosis, ankylosis, protrusio acetabuli, and painful hip dysplasia; - 2. inflammatory degenerative joint disease such as rheumatoid arthritis; - 3. correction of functional deformity; and, - 4. revision procedures where other treatments or devices have failed Modular necks can be used during either cemented or uncemented femoral and acetabular arthroplasty. . Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) for M. Melkersm (Diviston/Sign-Oft) (Division of Surgical, Orthopedic, and Restorative Devices 510(k) Number K102565 Page 1 of 1 l
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...