← Product Code [KTT](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/KTT) · K143220

# SIDEKICK EZ FRAME External Fixation System (K143220)

_Wrightmedicaltechnologyinc · KTT · Jan 8, 2015 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/KTT/K143220

## Device Facts

- **Applicant:** Wrightmedicaltechnologyinc
- **Product Code:** [KTT](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/KTT.md)
- **Decision Date:** Jan 8, 2015
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3030
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

Triple arthrodesis Isolated rearfoot arthrodesis Midfoot arthrodesis Comminuted trauma Diabetic Charcot reconstruction Most foot pathology not requiring fixation above the ankle

## Device Story

External fixation system for foot stabilization; utilizes pins and wires for bone compression or distraction; includes boot component to secure tibia. Adjustable design accommodates patient anatomy. Used by orthopedic surgeons in clinical settings. Provides mechanical stabilization for arthrodesis, trauma, and reconstruction. Compatible with SIDEKICK FREEDOM system components to expand fixation options.

## Clinical Evidence

No clinical data provided; substantial equivalence based on non-clinical technological comparison.

## Technological Characteristics

Metallic bone fixation system; components include pins, wires, and a boot for tibial securement. Adjustable mechanical frame design. Compatible with SIDEKICK FREEDOM system components.

## Regulatory Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- SIDEKICK® EZ FRAME™ External Fixation System ([K130044](/device/K130044.md))

## Reference Devices

- SIDEKICK® FREEDOM External Fixation System ([K043289](/device/K043289.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an emblem featuring a stylized caduceus, which is a symbol often associated with healthcare and medicine. The caduceus is depicted with a staff entwined by two snakes and topped with wings.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

January 8, 2015

Wright Medical Technology, Incorporated Ms. Leslie Fitch Senior Regulatory Affairs Specialist 1023 Cherry Road Memphis, Tennessee 38117

Re: K143220

Trade/Device Name: SIDEKICK® EZ FRAME™ External Fixation System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: KTT Dated: November 6, 2014 Received: November 10, 2014

Dear Ms. Fitch:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing

{1}------------------------------------------------

# Page 2 - Ms. Leslie Fitch

(21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# Lori A. Wiggins -S

for Mark N. Melkerson Director. Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

# Indications for Use

510(k) Number (if known)

K143220

Device Name

SIDEKICK® EZ FRAME™ External Fixation System

Indications for Use (Describe) Wright's SIDEKICK® EZ FRAME™ External Fixation System is intended for:

- · Triple Arthrodesis
- · Isolated Rearfoot Arthrodesis
- · Midfoot Arthrodesis
- Comminuted Trauma
- · Diabetic Charcot Reconstruction
- · Most foot pathology not requiring fixation above the ankle

Type of Use (Select one or both, as applicable)

2 Prescription Use (Part 21 CFR 801 Subpart D)

_ Over-The-Counter Use (21 CFR 801 Subpart C)

### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

### FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Headquarters Wright Medical Technology, Inc.

1023 Cherry Road Memphis, TN 38117

01 867 9971 mt.com

### 510(K) SUMMARY

In accordance with the Food and Drug Administration Rule to implement provisions of the Safe Medical Devices Act of 1990 and in conformance with 21 CFR 807.92, this information serves as a Summary of Safety and Effectiveness for the use of the SIDEKICK® EZ FRAME™ External Fixation System.

| (a)(1). Submitted By:              | Wright Medical Technology, Inc. 1023 Cherry Road Memphis, TN 38117                                |
|------------------------------------|---------------------------------------------------------------------------------------------------|
| Date:                              | November 6, 2014                                                                                  |
| Contact Person:                    | Leslie Fitch, PhD Senior Regulatory Affairs Specialist Office: (901) 867-4120 Fax: (901) 867-4190 |
| (a)(2). Proprietary Name:          | SIDEKICK® EZ FRAME™ External Fixation System                                                      |
| Common Name:                       | External Fixation System                                                                          |
| Classification Name and Reference: | 21 CFR 888.3030 – Class II                                                                        |
| Device Product Code, Device Panel: | KTT, Orthopedic                                                                                   |
| (a)(3). Predicate Devices:         | K130044: SIDEKICK® EZ FRAME™ External Fixation System                                             |

# (a)(4). Device Description

The SIDEKICK® EZ FRAME™ External Fixation System uses a series of pins and wires for compression or distraction in the foot and uses a boot to secure the tibia. The system is adjustable to accommodate variations in patient anatomy. The SIDEKICK® EZ FRAMETM External Fixation System is compatible with the SIDEKICK® FREEDOM External Fixation System (K043289) components, and these components provide additional fixation options in the system.

{4}------------------------------------------------

## (a)(5). Intended Use

- · Triple arthrodesis
- Isolated rearfoot arthrodesis
- Midfoot arthrodesis
- · Comminuted trauma
- Diabetic Charcot reconstruction
- · Most foot pathology not requiring fixation above the ankle

## (a)(6). Technological Characteristics Comparison

The SIDEKICK® EZ FRAMETM External Fixation System is technologically substantially equivalent to predicate devices in material and design,

### (b)(1). Substantial Equivalence - Non-Clinical Evidence

Through technological comparison, the subject system does not represent a new worst-case.

- (b)(2). Substantial Equivalence Clinical Evidence N/A
### (b)(3). Substantial Equivalence - Conclusions

The design characteristics of the subject system do not raise any new types of questions of safety or effectiveness. From the evidence submitted in this 510(k), the subject devices can be expected to perform at least as well as the predicate device.

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/KTT/K143220](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/KTT/K143220)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
