← Product Code [KTT](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/KTT) · K024248

# MODIFICATION TO EBI XFIX DFS SYSTEM METAPHYSEAL CORRECTION SYSTEM (K024248)

_Ebi, L.P. · KTT · Jan 22, 2003 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/KTT/K024248

## Device Facts

- **Applicant:** Ebi, L.P.
- **Product Code:** [KTT](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/KTT.md)
- **Decision Date:** Jan 22, 2003
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 888.3030
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

The EBI XFIX® DFS® Metaphyseal Correction System is intended for use in the treatment of bone conditions including leg lengthening, osteotomies, arthrodesis, fracture fixation, and other bone conditions amenable to treatment by use of the external fixation modality.

## Device Story

Unilateral external fixation system; consists of external fixation components and implantable bone screws. Bone screws inserted through skin/soft tissue into bone; fixator frame attached to screw shanks. Addition of locking set screw to system. Used by clinicians for orthopedic bone correction/stabilization. Provides mechanical stability to bone segments to facilitate healing or correction.

## Clinical Evidence

No clinical data provided; bench testing only.

## Technological Characteristics

Unilateral external fixation device. Materials: titanium, stainless steel, aluminum. Components include fixator frame and implantable bone screws. Includes locking set screw mechanism.

## Regulatory Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- EBI XFIX® DFS® System

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

JAN 2 2 2003

Koa YQYP
page lofa

## 510(k) Summary of Safety & Effectiveness

This 510(k) Summary of Safety and Effectiveness for the EBI XFIX® DFS® Metaphyseal Correction System is provided as required per Section 513(3) of the Food, Drug and Cosmetic Act.

| 1. Submitter:         | EBI, L.P.<br>100 Interpace Parkway<br>Parsippany, NJ 07054                                      | Contact Person: | Jon Caparotta<br>Telephone: (973) 299-9022 |
|-----------------------|-------------------------------------------------------------------------------------------------|-----------------|--------------------------------------------|
| Date prepared:        | December 20, 2002                                                                               |                 |                                            |
| 2. Proprietary Name:  | EBI XFIX® DFS® Metaphyseal Correction System                                                    |                 |                                            |
| Common Name:          | External Fixation Device                                                                        |                 |                                            |
| Classification Names: | Single/Multiple Component Metallic Bone Fixation<br>Appliances and Accessories, 21 CFR 888.3030 |                 |                                            |

3. Predicate or legally marketed devices that are substantially equivalent:

- EBI XFIX® DFS® System -- EBI, L.P.
- 4. Description of the device: The EBI XFIX® DFS® Metaphyseal Correction System has not changed the Indications for Use or fundamental scientific technology of the previously cleared system. The system consists of external fixation components and implantable bone screws. The EBI XFIX® DFS® Metaphyseal Correction System is utilized in the following manner: bone screws are inserted through the patient's skin and soft tissue and into the bone. The fixator frame of the EBI XFIX® DFS® Metaphyseal Correction System is

{1}------------------------------------------------

K024448
page 2 of 2

attached to the shanks of the bone screws. This submission is for the addition of a locking set screw to the system.

- 5. Intended Use: The EBI XFIX® DFS® Metaphyseal Correction System is intended for use in the treatment of bone conditions including leg lengthening, osteotomies, arthrodesis, fracture fixation, and other bone conditions amenable to treatment by use of the external fixation modality.
- 6. Materials: The components of the System may be manufactured from materials such as titanium, stainless steel, and aluminum.
- 7. Comparison of the technological characteristics of the device to predicate devices: There are no significant differences between the EBI XFIX® DFS® Metaphyseal Correction System and other currently marketed external fixation systems. It is substantially equivalent* to the predicate devices in regards to intended use, materials, and function.

<sup>*</sup> Any statement made in conjunction with this submission regarding a determination of substantial equivalence to any other product is intended only to relate to whether the product can be lawfully marketed without pre-market approval or reclassification and is not intended to be interpreted as an admission or any other type of evidence in patent infringement litigation. [Establishment Registration and Premarket Notification Procedures, Final Regulation, Preamble, August 23, 1977, FR 42520 (Docket No. 76N-0355.)]

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with three wavy lines representing its wings.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

EBI. L.P. Frederic Testa Regulatory Affairs Specialist 100 Interpace Parkway Parsippany, New Jersey 07054

JAN 2 2 2003

Re: K024248

Trade/Device Name: EBI XFIX® DFS® Metaphyseal Correction System Regulation Number: 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: KTT Dated: December 20, 2002 Received: December 23, 2002

Dear Mr. Testa:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set

{3}------------------------------------------------

Page 2 - Mr. Frederic Testa

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address

http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Miriam C. Provost

for Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

## STATEMENT OF INDICATIONS FOR USE

1

Page 1 of

510(k) Number (if known): KOx 4040

Device Name: EBI® XFIX® DFS® Metaphyseal Correction System

Indications For Use:

The EBI® XFIX® DFS® Metaphyseal Correction System is a unilateral external fixation device intended for use in the treatment of bone conditions including leg lengthening, osteotomies, arthrodesis, fracture fixation, and other bone conditions amenable to treatment by use of the external fixation modality.

## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use (Per 21 CFR 801.109) OR

Over-The-Counter Use

Miriam C. Provost
(Optional Format 1-2-96)

ivision Sign-Off) Division of General, Restorative and Neurological Devices

Number K024248

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/KTT/K024248](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/KTT/K024248)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
