← Product Code [KTT](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/KTT) · K012623

# REPROCESSED EXTERNAL FIXATION DEVICE (K012623)

_Alliance Medical Corp. · KTT · Jun 24, 2002 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/KTT/K012623

## Device Facts

- **Applicant:** Alliance Medical Corp.
- **Product Code:** [KTT](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/KTT.md)
- **Decision Date:** Jun 24, 2002
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3030
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

External Fixation Devices are intended to be used for the fixation of supracondylar, or condylar fractures of the femur: for fusion of a joint; for surgical procedures that involve cutting the bone, for fixation of bone fractures; bone reconstruction; as a guide pin for insertion of other implants: or may be implanted through the skin so that a pulling force or traction may be applied to the skeletal system.

## Device Story

Reprocessed external fixation devices consist of frames, clamps, rods, couplings, pins, posts, fasteners, bolts, washers, nuts, hinges, sockets, and connecting bars. Used for management of bone fractures, reconstructive, and corrective orthopedic surgery. Devices provide skeletal rigidity and stability; may be implanted through skin for traction. Operated by surgeons in clinical/OR settings. Reprocessing restores used devices to original performance specifications. Benefits include stabilization of fractures and correction of bony/soft tissue defects.

## Clinical Evidence

Bench testing only. Performance demonstrated through biocompatibility testing and validation of the reprocessing process to ensure devices perform as originally intended.

## Technological Characteristics

Materials include metal alloys, plastic, and composites. Mechanical design, shapes, and sizes are identical to original predicate devices. No energy source or software components.

## Regulatory Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- Synthes Large Fragment DCL System ([K000682](/device/K000682.md), [K010344](/device/K010344.md))
- Synthes (USA) Adjustable Large Fixator System
- Synthes TraumaFix™ External Fixator ([K914558](/device/K914558.md))
- Synthes (U.S.A.) Mini External Fixator ([K961350](/device/K961350.md), [K962484](/device/K962484.md))
- Synthes Titanium External Fixator Clamps
- Synthes ([K973017](/device/K973017.md))
- Combination Clamp and Dynamization Clip ([K963618](/device/K963618.md))
- Synthes Simple Small External Fixator

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K012623
10F3

## 510(k) Summary of Safety and Effectiveness

Submitter:

Alliance Medical Corporation 10232 South 51st Street Phoenix, Arizona 85044

Contact:

Don Selvey Vice President, Regulatory Affairs and Quality Assurance (480) 763-5300

| Date of preparation: | 8 August 2001                                                                                               |
|----------------------|-------------------------------------------------------------------------------------------------------------|
| Name of device:      | Reprocessed External Fixation Devices                                                                       |
| Common Name:         | External Fixation Devices, Fixation Appliance, Single/Multiple<br>Component and Invasive Traction Component |
| Classification Name: | Single/Multiple Component Metallic Bone Fixation Appliances and<br>Accessories                              |

Reprocessed device(s):

| MANUFACTURER  | MODEL NUMBER | DESCRIPTION                    |
|---------------|--------------|--------------------------------|
| SYNTHES (USA) | 392.88       | NUT FOR 8.0MM THREADED RODS    |
| SYNTHES (USA) | 393.36       | TUBE-TO-TUBE CLAMP             |
| SYNTHES (USA) | 393.361      | TITANIUM ROD-TO-ROD CLAMP      |
| SYNTHES (USA) | 393.436      | RING-TO-ROD CLAMP              |
| SYNTHES (USA) | 393.44       | EXTERNAL FIXATION CLAMP        |
| SYNTHES (USA) | 393.45       | EXTERNAL FIXATION CLAMP        |
| SYNTHES (USA) | 393.464      | ADJUSTABLE PIN CLAMP           |
| SYNTHES (USA) | 393.48       | ANGLED EXTERNAL FIXATION CLAMP |
| SYNTHES (USA) | 393.64       | ROD-TO-PIN ADJUSTABLE CLAMP    |
| SYNTHES (USA) | 393.647      | COMBINATION CLAMP              |
| SYNTHES (USA) | 393.71       | UNIVERSAL JOINT FOR TWO TUBES  |
| SYNTHES (USA) | 393.722      | 3/4 FIXATION RING              |
| SYNTHES (USA) | 393.732      | 3/4 STEEL RING                 |
| SYNTHES (USA) | 393.734      | 3/4 EXTERNAL FIXATION RING     |
| SYNTHES (USA) | 393.736      | 3/4 EXTERNAL FIXATION RING     |
| SYNTHES (USA) | 393.751      | ROD-TO-PIN UNIVERSAL CLAMP     |
| SYNTHES (USA) | 393.753      | ROD CLAMP                      |
| SYNTHES (USA) | 393.756      | ROD TO ROD CLAMP               |
| SYNTHES (USA) | 394.8        | CARBON FIBER ROD               |
| SYNTHES (USA) | 394.80       | CARBON FIBER ROD               |

{1}------------------------------------------------

| MANUFACTURER  | MODEL NUMBER | DESCRIPTION              |
|---------------|--------------|--------------------------|
| SYNTHES (USA) | 394.82       | CARBON FIBER ROD         |
| SYNTHES (USA) | 394.83       | CARBON FIBER ROD         |
| SYNTHES (USA) | 394.84       | CARBON FIBER ROD         |
| SYNTHES (USA) | 394.85       | CARBON FIBER ROD         |
| SYNTHES (USA) | 394.86       | CARBON FIBER ROD         |
| SYNTHES (USA) | 394.87       | CARBON FIBER ROD         |
| SYNTHES (USA) | 395.55       | OPEN CLAMP               |
| SYNTHES (USA) | 395.56       | PIN TO ROD COUPLER       |
| SYNTHES (USA) | 395.57       | ROD TO ROD COUPLER       |
| SYNTHES (USA) | 395.597      | ROD TO PIN COUPLER       |
| SYNTHES (USA) | 395.6        | CARBON FIBER ROD         |
| SYNTHES (USA) | 395.62       | CARBON FIBER ROD         |
| SYNTHES (USA) | 395.63       | CARBON FIBER ROD         |
| SYNTHES (USA) | 395.64       | CARBON FIBER ROD         |
| SYNTHES (USA) | 395.65       | CARBON FIBER ROD         |
| SYNTHES (USA) | 395.66       | CARBON FIBER ROD         |
| SYNTHES (USA) | 395.72       | CONNECTING BARS (S.S.)   |
| SYNTHES (USA) | 395.73       | CONNECTING BARS (S.S.)   |
| SYNTHES (USA) | 395.73C      | CONTRA ROD               |
| SYNTHES (USA) | 395.74       | CONNECTING BARS (S.S.)   |
| SYNTHES (USA) | 395.75       | CONNECTING BARS (S.S.)   |
| SYNTHES (USA) | 395.76       | CONNECTING BARS (S.S.)   |
| SYNTHES (USA) | 395.77       | CONNECTING BARS (S.S.)   |
| SYNTHES (USA) | 395.782      | CARBON FIBER ROD         |
| SYNTHES (USA) | 395.784      | CARBON FIBER ROD         |
| SYNTHES (USA) | 459.014      | EXTERNAL FIXATION DEVICE |
| SYNTHES (USA) | 633-200      | STEEL ROD                |
| SYNTHES (USA) | B18-200      | STEEL ROD                |
| SYNTHES (USA) | B18-300      | STEEL ROD                |
| SYNTHES (USA) | B18-350      | STEEL ROD                |
| SYNTHES (USA) | B18-400      | STEEL ROD                |

20F3

Predicate device(s): Synthes Large Fragment DCL System K000682 K010344 Synthes (USA) Adjustable Large Fixator System Synthes TraumaFix™ External Fixator K914558 Synthes (U.S.A.) Mini External Fixator K961350 K962484 Synthes Titanium External Fixator Clamps Synthes K973017 Combination Clamp and Dynamization Clip K963618 Synthes Simple Small External Fixator Device description: External fixation devices are specially designed frames, clamps, rods,

rod-to-rod couplings, pins, posts, fasteners, wire fixations, fixation bolts, washers, nuts, hinges, sockets, connecting bars and screws used

{2}------------------------------------------------

# 2623 30F 3

for the management of bone fractures and reconstructive, as well as corrective, orthopedic surgery. Materials used include metal alloys, plastic and composites. These materials are chosen to address a wide range of fractures and applications as well as to allow for the appropriate amount of rigidity and stability.

External Fixation Devices are intended to be used for the fixation of Intended use: supracondylar, or condylar fractures of the femur: for fusion of a joint; for surgical procedures that involve cutting the bone, for fixation of bone fractures; bone reconstruction; as a guide pin for insertion of other implants: or may be implanted through the skin so that a pulling force or traction may be applied to the skeletal system.

Reprocessed external fixation devices are indicated for use in patients Indications statement: requiring external skeletal fixation oand treatment of fractures, osteotomy, arthrodesis, correction of deformities, fracture revision, bone reconstruction procedures, limb lengthening, correction of bony or soft tissue deformities and segmental bony or soft tissue defects.

The design, materials, and intended use of the Reprocessed External Technological characteristics: Fixation Device s are identical to the predicate devices. The mechanism of action of the Reprocessed External Fixation Device is identical to the predicate devices in that the same standard mechanical design, materials, shapes and sizes are utilized. There are no changes to the claims, intended use, clinical applications, patient population, performance specifications, or method of operation.

#### Bench and laboratory testing was conducted to demonstrate Performance data: performance (safety and effectiveness) of the Reprocessed External Fixation Device s.

- Biocompatibility ●
- Validation of reprocessing

Performance testing demonstrates that Reprocessed External Fixation Devices perform as originally intended.

In accordance with the Federal Food, Drug and Cosmetic Act 21 CFR Conclusion: Part 807 and based on the information provided in this premarket notification, Alliance Medical Corporation concludes that the modified device (the Reprocessed External Fixation Device) is safe, effective and substantially equivalent to the predicate devices as described herein.

{3}------------------------------------------------

#### DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular emblem with the department's name encircling a stylized image. The image within the circle appears to be an abstract representation of a human figure or a bird-like shape. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA".

#### Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

### JUN 2 4 2002

Mr. Don Selvey Vice President Regulatory Affairs and Quality Assurance Alliance Medical Corporation 10232 South 51st Street Phoenix, Arizona 85044

#### Re: (K012623) K012634, K012645, K012648

Trade Name: Reprocessed External Fixation Devices Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: II Product Code: KTT, KTW, JEC Dated: April 3, 2002 Received: April 4, 2002

Dear Mr. Selvey:

We have reviewed your Section 510(k) premarket notifications of intent to market the devices referenced above and have determined the devices are substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the devices, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your devices are classified (see above) into either class II (Special Controls) or class III (PMA), they may be subject to such additional controls. Existing major regulations affecting your devices can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your devices in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your devices comply with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{4}------------------------------------------------

Page 2 - Mr. Don Selvey

This letter will allow you to begin marketing your devices as described in your Section 510(k) premarket notifications. The FDA finding of substantial equivalence of your devices to a legally marketed predicate device results in a classification for your devices and thus, permits your devices : to proceed to the market.

If you desire specific advice for your devices on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your devices, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours.

sincerely yours,

Mark McMillan

Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

## Indications for Use Statement

510(k) Number (if known):

(012623

Device Name: Alliance Medical Corporation Reprocessed External Fixation Devices

Indications for Use: Reprocessed external fixation devices are indicated for use in patients requiring external skeletal fixation and treatment of fractures, osteotomy, arthrodesis, correction of deformities, fracture revision, bone reconstruction procedures, limb lengthening, correction of bony or soft tissue deformities and segmental bony or soft tissue defects.

| MANUFACTURER  | MODEL<br>NUMBER | DESCRIPTION                    |
|---------------|-----------------|--------------------------------|
| SYNTHES (USA) | 392.88          | NUT FOR 8.0MM THREADED RODS    |
| SYNTHES (USA) | 393.36          | TUBE-TO-TUBE CLAMP             |
| SYNTHES (USA) | 393.361         | TITANIUM ROD-TO-ROD CLAMP      |
| SYNTHES (USA) | 393.436         | RING-TO-ROD CLAMP              |
| SYNTHES (USA) | 393.44          | EXTERNAL FIXATION CLAMP        |
| SYNTHES (USA) | 393.45          | EXTERNIAL FIXATION CLAMP       |
| SYNTHES (USA) | 393.464         | ADJUSTABLE PIN CLAMP           |
| SYNTHES (USA) | 393.48          | ANGLED EXTERNAL FIXATION CLAMP |
| SYNTHES (USA) | 393.64          | ROD-TO-PIN ADJUSTABLE CLAMP    |
| SYNTHES (USA) | 393.647         | COMBINATION CLAMP              |
| SYNTHES (USA) | 393.71          | UNIVERSAL JOINT FOR TWO TUBES  |
| SYNTHES (USA) | 393.722         | 3/4 FIXATION RING              |
| SYNTHES (USA) | 393.732         | 3/4 STEEL RING                 |
| SYNTHES (USA) | 393.734         | 3/4 EXTERNAL FIXATION RING     |
| SYNTHES (USA) | 393.736         | 3/4 EXTERNAL FIXATION RING     |
| SYNTHES (USA) | 393.751         | ROD-TO-PIN UNIVERSAL CLAMP     |
| SYNTHES (USA) | 393.753         | ROD CLAMP                      |
| SYNTHES (USA) | 393.756         | ROD TO ROD CLAMP               |
| SYNTHES (USA) | 394.8           | CARBON FIBER ROD               |
| SYNTHES (USA) | 394.80          | CARBON FIBER ROD               |
| SYNTHES (USA) | 394.82          | CARBON FIBER ROD               |
| SYNTHES (USA) | 394.83          | CARBON FIBER ROD               |
| SYNTHES (USA) | 394.84          | CARBON FIBER ROD               |
| SYNTHES (USA) | 394.85          | CARBON FIBER ROD               |
| SYNTHES (USA) | 394.86          | CARBON FIBER ROD               |
| SYNTHES (USA) | 394.87          | CARBON FIBER ROD               |
| SYNTHES (USA) | 395.55          | OPEN CLAMP                     |

CONFIDENTIAL

Alliance Medical Corporation Reprocessed External Fixation Devices Traditional 510(k)

{6}------------------------------------------------

| SYNTHES (USA) | 395.56  | PIN TO ROD COUPLER       |
|---------------|---------|--------------------------|
| SYNTHES (USA) | 395.57  | ROD TO ROD COUPLER       |
| SYNTHES (USA) | 395.597 | ROD TO PIN COUPLER       |
| SYNTHES (USA) | 395.6   | CARBON FIBER ROD         |
| SYNTHES (USA) | 395.62  | CARBON FIBER ROD         |
| SYNTHES (USA) | 395.63  | CARBON FIBER ROD         |
| SYNTHES (USA) | 395.64  | CARBON FIBER ROD         |
| SYNTHES (USA) | 395.65  | CARBON FIBER ROD         |
| SYNTHES (USA) | 395.66  | CARBON FIBER ROD         |
| SYNTHES (USA) | 395.72  | CONNECTING BARS (S.S.)   |
| SYNTHES (USA) | 395.73  | CONNECTING BARS (S.S.)   |
| SYNTHES (USA) | 395.73C | CONTRA ROD               |
| SYNTHES (USA) | 395.74  | CONNECTING BARS (S.S.)   |
| SYNTHES (USA) | 395.75  | CONNECTING BARS (S.S.)   |
| SYNTHES (USA) | 395.76  | CONNECTING BARS (S.S.)   |
| SYNTHES (USA) | 395.77  | CONNECTING BARS (S.S.)   |
| SYNTHES (USA) | 395.782 | CARBON FIBER ROD         |
| SYNTHES (USA) | 395.784 | CARBON FIBER ROD         |
| SYNTHES (USA) | 459.014 | EXTERNAL FIXATION DEVICE |
| SYNTHES (USA) | 633-200 | STEEL ROD                |
| SYNTHES (USA) | B18-200 | STEEL ROD                |
| SYNTHES (USA) | B18-300 | STEEL ROD                |
| SYNTHES (USA) | B18-350 | STEEL ROD                |
| SYNTHES (USA) | B18-400 | STEEL ROD                |

| Concurrence of CDRH, Office of Device Evaluation (ODE) |                                                                                                                                |                      |
|--------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------|----------------------|
|                                                        | <center>Mark n Milk</center> <center>(Division Sign-Off)</center> Division of General, Restorative<br>and Neurological Devices |                      |
| 510(k) Number                                          | K012623                                                                                                                        |                      |
| Prescription Use<br>(per 21 CFR 801.109)               | or                                                                                                                             | Over-the-Counter Use |

(per 21 CFR 801.109)

CONFIDENTIAL

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/KTT/K012623](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/KTT/K012623)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
