GMK HINGE

K130299 · Medacta International · KRO · Aug 15, 2013 · Orthopedic

Device Facts

Record IDK130299
Device NameGMK HINGE
ApplicantMedacta International
Product CodeKRO · Orthopedic
Decision DateAug 15, 2013
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3510
Device ClassClass 2
AttributesTherapeutic

Intended Use

The GMK HINGE® knee prosthesis is designed for cemented use in total knee arthroplasty when the preoperative diagnosis of the joint determines that the bone and stability situation require the implantation of a constrained prosthesis. The GMK HINGE® knee system is indicated in the following cases: - Severely painful and/or disabled joint as a result of arthritis, traumatic arthritis, . rheumatoid arthritis or polyarthritis associated with bone loss and/or severe joint instability - Considerable loss of function of the knee joint . - High-grade joint destruction requiring additional stabilization with stems and . reconstruction of bone defects with metal augmentation - Failure of a primary prosthesis (e.g. infection, loosening) . - Former revision arthroplasty . - Post traumatic loss of joint configuration - Avascular necrosis of femoral condyle Tibial augmentations are to be screwed to the tibial baseplate with both of the two provided fixing screws. When a GMK HINGE® implant is used it is mandatory to implant both the femoral and tibial components with an extension stem.

Device Story

GMK Hinge is a constrained, rotating total knee prosthesis for cemented implantation. System includes femoral component, tibial component, hinge post, hinge post extension, and UHMWPE bushing to prevent metal-on-metal contact. Tibial component features asymmetric design for bone coverage; tibial insert provides rotational control and varus/valgus constraint. Femoral and tibial augments allow restoration of joint line and filling of bone deficiencies; extension stems required for both components to ensure stable construct. Used by orthopedic surgeons in clinical settings for patients with poor bone quality or marked bone loss. Device restores joint function and stability through mechanical constraint and intramedullary fit and fill.

Clinical Evidence

Bench testing only. Mechanical performance evaluated against worst-case predicate devices per ISO 21534, ISO 7207-1, ASTM F1800, ASTM F2083-11, and ASTM F1223-08.

Technological Characteristics

Constrained, rotating knee prosthesis. Materials: CoCrMo (ISO 5832-4, ISO 5832-12), UHMWPE (ISO 5834-2 Type 1), Ti6Al4V (ISO 5832-3), M30NW (ISO 5832-9). Cemented fixation. Includes femoral/tibial components, hinge post, UHMWPE bushing, and optional augments. Requires extension stems.

Indications for Use

Indicated for patients requiring total knee arthroplasty with constrained prosthesis due to severe pain, joint disability, or destruction from arthritis (rheumatoid, traumatic, polyarthritis), avascular necrosis, post-traumatic loss of joint configuration, or failure of primary/revision arthroplasty associated with bone loss or severe instability.

Regulatory Classification

Identification

A knee joint femorotibial metal/polymer constrained cemented prosthesis is a device intended to be implanted to replace part of a knee joint. The device limits translation or rotation in one or more planes and has components that are linked together or affined. This generic type of device includes prostheses composed of a ball-and-socket joint located between a stemmed femoral and a stemmed tibial component and a runner and track joint between each pair of femoral and tibial condyles. The ball-and-socket joint is composed of a ball at the head of a column rising from the stemmed tibial component. The ball, the column, the tibial plateau, and the stem for fixation of the tibial component are made of an alloy, such as cobalt-chromium-molybdenum. The ball of the tibial component is held within the socket of the femoral component by the femoral component's flat outer surface. The flat outer surface of the tibial component abuts both a reciprocal flat surface within the cavity of the femoral component and flanges on the femoral component designed to prevent distal displacement. The stem of the femoral component is made of an alloy, such as cobalt-chromium-molybdenum, but the socket of the component is made of ultra-high molecular weight polyethylene. The femoral component has metallic runners which align with the ultra-high molecular weight polyethylene tracks that press-fit into the metallic tibial component. The generic class also includes devices whose upper and lower components are linked with a solid bolt passing through a journal bearing of greater radius, permitting some rotation in the transverse plane, a minimal arc of abduction/adduction. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for Medacta International. The logo consists of two black triangles on the left, followed by the word "Medacta" in black lowercase letters. Below the word "Medacta" is a horizontal line, and below that is the word "International" in a smaller font, followed by a small cross symbol. ### 510(k) Summary Applicant/Sponsor: Medacta International SA Strada Regina 6874 Castel San Pietro (CH) Switzerland Phone (+41) 91 696 60 60 Fax (+41) 91 696 60 66 AUG 1 5 2013 - Contact Person: Mr. Adam Gross Director of Regulatory and Quality Medacta USA 4725 Calle Quetzal, Unit B Camarillo, CA, 93012 Phone: (805)437-7085 Fax: (805)437-7553 Email: AGross@medacta.us.com Date Prepared: February 5, 2013 DEVICE INFORMATION Trade/Proprietary Name: GMK Hinge Common Name: Total Knee Prosthesis Classification Name: Prosthesis Knee, Femorotibial, Constrained, Cemented, Metal/Polymer > 21 CFR 888.3510 Class II Device Product Code: KRO #### Section 5 - Page 2 of 4 {1}------------------------------------------------ | 510(k) | Product | 510(k) Holder | Clearance<br>Date | |---------|-----------------------------|-------------------|-------------------| | K003504 | RT-Plus | Plus Orthopedics | 5/11/2001 | | K013385 | NexGen Rotating Hinge Knee | Zimmer | 1/9/2002 | | K002552 | Modular Rotating Hinge Knee | Stryker/Howmedica | 11/13/2000 | | K101815 | Enduro RHK | Aesculap | 12/20/2010 | | K896048 | S-ROM/NOILES | Depuy | 03/14/1990 | #### "Predicate Devices: ### Product Description The GMK Hinge System is a hinged-knee system for total knee replacement. The GMK Hinge femoral component is designed to preserve bone stock. The hinge-post system is comprised of the hinge post and the hinge post extension and has been developed in order to reduce the impingement with the patellar tendon and provide mechanical resistance. There is a UHMWPE bushing between the femoral component and the hinge post in order to avoid contact between metal parts. The tibial component has an asymmetric design to provide coverage of the bone after tibia resection. The surface is mirror polished and the holes for the tibial augmentations are on the bottom. The UHMWPE tibial insert is fixed to the tibial component through a screw and is designed to provide high rotational control and a high varus/valgus constraint. The femoral and tibial augments allow the surgeon to selectively fill bone deficiencies and to aid in restoring the ioint line. The fixation between the augmentations and the femoral component or tibial trav is assured by a screw fixation. The GMK Hinge provides two solutions for the patella prosthesis: an inset patella and a resurfacing patella. The GMK Hinge System would be used most often in patients with poor bone quality or marked bone loss, and in these circumstances, extension stems are helpful to obtain a stable construct. An extension stem in combination with the right offset connector provides intramedullary femoral and tibial fit and fill. #### Indications for Use The GMK HINGE® knee prosthesis is designed for cemented use in total knee arthroplasty when the preoperative diagnosis of the joint determines that the bone and stability situation require the implantation of a constrained prosthesis. The GMK HINGE® knee system is indicated in the following cases: - Severely painful and/or disabled joint as a result of arthritis, traumatic arthritis, . rheumatoid arthritis or polyarthritis associated with bone loss and/or severe joint instability - Considerable loss of function of the knee joint . - High-grade joint destruction requiring additional stabilization with stems and . reconstruction of bone defects with metal augmentation - Failure of a primary prosthesis (e.g. infection, loosening) . - Former revision arthroplasty . {2}------------------------------------------------ - · Post traumatic loss of joint configuration - Avascular necrosis of femoral condyle Tibial augmentations are to be screwed to the tibial baseplate with both of the two provided fixing screws. When a GMK HINGE® implant is used it is mandatory to implant both the femoral and tibial components with an extension stem. ### Comparison to Predicate Devices The indications for use, design features, and materials of the subject device are substantially equivalent to those of the predicate devices. The design features of both the GMK Hinge and the predicates are a constrained, rotating knee prosthesis that have similar sizes and are intended to be used for cemented use only. The GMK Hinge and the predicates use the same materials: CoCrMo (ISO 5832-4) UHMWPE (ISO 5834-2 Type 1) Ti6Al4V (ISO 5832-3) CoCrMo (ISO 5832-12) M30NW (ISO 5832-9) The fundamental scientific technology of the modified devices has not changed relative to the predicate devices. The safety and effectiveness of the GMK Hinge are adequately supported by the substantial equivalence information, materials information, and analysis data provided within this Premarket Notification. ## Performance Testing The GMK Hinge was compared to the worst case of the predicate devices. GMK Hinge was tested according to the following standards: - . ISO 21534 - ISO 7207-1 . - ASTM F1800 . - ASTM F2083-11 . - ASTM F1223-08 . The comparison to predicate devices and the mechanical testing performed demonstrate that the GMK Hinge does not introduce any issues in regards to safety and effectiveness. #### Conclusion: Based on the above information, the GMK Hinge can be considered as substantially equivalent to its predicate devices. {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized representation of a human figure embracing a globe, symbolizing the department's mission to protect and promote the health and well-being of the nation. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 August 15, 2013 Medacta International SA % Medacta United States of America Mr. Adam Gross Director of Regulatory and Quality 4725 Calle Quetzal, Unit B Camarillo, California 93012 Re: K130299 Trade/Device Name: GMK Hinge Regulation Number: 21 CFR 888.3510 Regulation Name: Knee joint femorotibial metal/polymer constrained cemented prosthesis Regulatory Class: Class II Product Code: KRO Dated: June 20, 2013 Received: June 21, 2013 Dear Mr. Gross: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical {4}------------------------------------------------ Page 2 - Mr. Adam Gross device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, # Erin Keith For Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ # Indications for Use K130299 510(k) Number (if known): Device Name: GMK Hinge Indications for Use: The GMK HINGE® knee prosthesis is designed for cemented use in total knee arthroplasty when the preoperative diagnosis of the joint determines that the bone and stability situation require the implantation of a constrained prosthesis. The GMK HINGE® knee system is indicated in the following cases: - Severely painful and/or disabled joint as a result of arthritis, traumatic arthritis, . rheumatoid arthritis or polyarthritis associated with bone loss and/or severe joint instability - Considerable loss of function of the knee joint . - High-arade joint destruction requiring additional stabilization with stems and . reconstruction of bone defects with metal augmentation - Failure of a primary prosthesis (e.g. infection, loosening) . - Former revision arthroplasty . - Post traumatic loss of joint configuration . - Avascular necrosis of femoral condyle . Tibial augmentations are to be screwed to the tibial baseplate with both of the two provided fixing screws. When a GMK HINGE® implant is used it is mandatory to implant both the femoral and tibial components with an extension stem. Prescription Use X (21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Casey L. Hanley, Ph.D. Division of Orthopedic Devices GMK Hinge 510(k) February 5, 2013 Section 4 - Page 2 of 2
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...