← Product Code [KRO](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/KRO) · K023087

# GLOBAL MODULAR REPLACEMENT SYSTEM (GMRS) (K023087)

_Howmedica Osteonics Corp. · KRO · Dec 16, 2002 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/KRO/K023087

## Device Facts

- **Applicant:** Howmedica Osteonics Corp.
- **Product Code:** [KRO](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/KRO.md)
- **Decision Date:** Dec 16, 2002
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3510
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

The Global Modular Replacement System (herein referred to as the GMRS) is intended to be used with the components of the Howmedica Osteonics Modular Replacement System (MRS), the Howmedica Osteonics Modular Rotating Hinge Knee System (MRH), and/or the Howmedica Osteonics Kinematic Rotating Hinge Knee System (KRH) in situations where replacement of extensive bone loss in the femur and/or proximal tibia is required. The Proximal Femoral Module and Distal Femoral Module of the GMRS may be used together, or with the components of the MRS, in replacement of the total femur. Indications and contraindications for this type of replacement are listed below:

## Device Story

GMRS is a modular orthopedic prosthesis system for limb salvage and oncology-related bone replacement. Components include proximal femoral modules, distal femoral/proximal tibial modules, and total femur assemblies. Used in surgical settings by orthopedic surgeons to replace extensive bone loss in femur/tibia. Proximal femoral modules allow cemented or press-fit fixation; distal/tibial modules require bone cement. System integrates with existing Howmedica Osteonics MRS, MRH, and KRH systems. Provides structural support for patients with severe bone loss, trauma, or failed prior implants; restores limb function through radical resection and prosthetic reconstruction.

## Clinical Evidence

Bench testing only. No clinical data provided.

## Technological Characteristics

Modular orthopedic prosthesis system. Materials include metal and polymer components. Proximal femoral stems feature PureFix HA coating for press-fit applications. Fixation methods include cemented and press-fit. System is modular, allowing assembly of proximal femoral, distal femoral, and proximal tibial modules. No software or electronic components.

## Regulatory Identification

A knee joint femorotibial metal/polymer constrained cemented prosthesis is a device intended to be implanted to replace part of a knee joint. The device limits translation or rotation in one or more planes and has components that are linked together or affined. This generic type of device includes prostheses composed of a ball-and-socket joint located between a stemmed femoral and a stemmed tibial component and a runner and track joint between each pair of femoral and tibial condyles. The ball-and-socket joint is composed of a ball at the head of a column rising from the stemmed tibial component. The ball, the column, the tibial plateau, and the stem for fixation of the tibial component are made of an alloy, such as cobalt-chromium-molybdenum. The ball of the tibial component is held within the socket of the femoral component by the femoral component's flat outer surface. The flat outer surface of the tibial component abuts both a reciprocal flat surface within the cavity of the femoral component and flanges on the femoral component designed to prevent distal displacement. The stem of the femoral component is made of an alloy, such as cobalt-chromium-molybdenum, but the socket of the component is made of ultra-high molecular weight polyethylene. The femoral component has metallic runners which align with the ultra-high molecular weight polyethylene tracks that press-fit into the metallic tibial component. The generic class also includes devices whose upper and lower components are linked with a solid bolt passing through a journal bearing of greater radius, permitting some rotation in the transverse plane, a minimal arc of abduction/adduction. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).

## Predicate Devices

- Modular Replacement System (Howmedica Osteonics Corp.)
- Distal Stems of the Restoration Hip System (Howmedica Osteonics Corp.)
- Modular Rotating Hinge Knee System (Howmedica Osteonics Corp.)
- Kinematic Rotating Hinge Knee System (Howmedica Osteonics Corp.)
- MOST System (Sulzer Orthopaedics)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Page 1 of 4

# 510(k) Summary K023087 Global Modular Replacement System

| Submission Information             | DEC 16 2002                                                                                                                                                                                     |
|------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Name and Address of Sponsor:       | Howmedica Osteonics Corp.<br>59 Route 17<br>Allendale, NJ 07401                                                                                                                                 |
| For Information contact:           | Margaret F. Crowe<br>Regulatory Affairs Consultant<br>Howmedica Osteonics Corp.<br>59 Route 17<br>Allendale, NJ 07401                                                                           |
| Device Identification              |                                                                                                                                                                                                 |
| Proprietary Name:                  | Global Modular Replacement System (GMRS)                                                                                                                                                        |
| Common Name:                       | Proximal Femoral Replacement<br>and<br>Modular Rotating Hinge Knee                                                                                                                              |
| Classification Name and Reference: | Prosthesis, Hip, Semi-constrained Metal/Polymer<br>Porous Uncemented<br>21 CFR §888.3358<br>and<br>Knee joint femorotibial metal/polymer<br>constrained cemented prosthesis<br>21 CFR §888.3510 |
| Proposed Regulatory Class:         | Class II                                                                                                                                                                                        |
| Device Product Code:               | OR(87) LPH and KRO                                                                                                                                                                              |

## Intended Use

The Global Modular Replacement System (herein referred to as the GMRS) is intended to be used with the components of the Howmedica Osteonics Modular Replacement System (MRS), the Howmedica Osteonics Modular Rotating Hinge Knee System (MRH), and/or the Howmedica Osteonics Kinematic Rotating Hinge Knee System

{1}------------------------------------------------

K023087

(KRH) in situations where replacement of extensive bone loss in the femur and/or proximal tibia is required. The Proximal Femoral Module and Distal Femoral Module of the GMRS may be used together, or with the components of the MRS, in replacement of the total femur. Indications and contraindications for this type of replacement are listed below:

# Indications

Femoral and/or proximal tibial replacement and total femoral replacement in Oncology cases where radical resection and replacement of bone is required, and in limb salvage procedures where radical resection and replacement of the bone is required. Limb salvage procedures would include surgical intervention for severe trauma, failed previous prosthesis, and/or Oncology indications.

#### Contraindications

#### As related to Bone Tumors A.

Not all bone tumors may be treated successfully by segmental resection. Any condition that may have already resulted in either local or distant spread of the tumor may be a contraindication. Examples of such conditions include:

- . pathological fracture;
- . overt infection;
- . inopportune placement of biopsy incision; and,
- rapid disease progression beyond a respectable margin. .
- As related to Failed Previous Prosthesis and Trauma B.
- Any active or suspected latent infection in or about the hip joint. ●
- Any mental or neuromuscular disorder which would create an unacceptable risk of . prosthesis instability, prosthesis fixation failure, or complication in postoperative care.
- Bone stock compromised by disease, infection, or prior implantation that cannot . provide adequate support and fixation of the prosthesis.

{2}------------------------------------------------

Ka.3087

page 3 of 4

For the use of GMRS Press Fit stems with PureFix™ HA Coating, the following additional contraindication should be noted:

- Inadequate bone stock to allow the use of a press fit stem .
The Proximal Femoral Module of the GMRS is intended to be used in a cemented or press fit mode. The components of the Distal Femoral/Proximal Tibial Module are intended to be used with bone cement.

## Device Description

The GMRS is comprised of a number of components that are intended to be used in conjunction with each other, or in conjunction with components of Howmedica Osteonics' Modular Rotating Hinge Knee, Kinematic Rotating Hinge Knee, and/or the Modular Replacement System. There are three modules of the GMRS:

- Proximal Femoral Module .
- . Distal Femoral/Proximal Tibial Module
- . Total Femur

Each of these modules is comprised of different components. The modules, and their individual components, are described below:

#### Proximal Femoral Module

The Proximal Femoral Module of the GMRS contains the following components:

- . Proximal Femoral Components
- Extension Pieces .
- . GMRS Press Fit Stems with PureFix™ HA

## Distal Femoral/Proximal Tibial Module

The Distal Femoral/Proximal Tibial Module of the GMRS contains the following new components:

- Distal Femoral Components .
- Extension Pieces (as described in the section above) .
- Small bushings

{3}------------------------------------------------

K023087

page 4 of 4

- Small Axle ●
- Proximal Tibial Component .
- Proximal Tibial Inserts .
- . Proximal Tibia Rotating Components

#### Total Femur

Replacement of the total femur is accomplished by combining the following components of the GMRS:

- Proximal Femoral Component .
- . Extension Pieces
- Connection Pieces ●
- Distal Femoral Component (with appropriate bushings, axle, etc.) .
- . Proximal Tibial Component (with appropriate tibial rotating component, tibial insert component)

Equivalent products include:

- 1. Modular Replacement System -Howmedica Osteonics Corp.
- Distal Stems of the Restoration™ Hip System Howmedica Osteonics Corp. 2.
- Modular Rotating Hinge Knee System Howmedica Osteonics Corp. 3.
- 4. Kinematic® Rotating Hinge Knee System - Howmedica Osteonics Corp.
- 5. MOST System - Sulzer Orthopaedics

Testing was presented to support a claim of substantial equivalence to the predicate devices.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. Inside the circle is an abstract symbol that resembles an eagle or a bird in flight, composed of three stylized human profiles facing to the right.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Mr. William J. Cymbaluk Vice President Stryker Howmedica Osteonics 59 Route 17 Allendale, New Jersey 07401

DEC 1 6 2002

Re: K023087

Trade/Device Name: Global Modular Replacement System Regulation Number: 21 CFR 888.3350 and 888.3510 Regulation Name: Hip joint metal/polymer semi-constrained cemented prosthesis; and Knee joint femorotibial metal/polymer constrained cemented prosthesis Regulatory Class: II Product Code: JDI and KRO Dated: September 16, 2002 Received: September 17, 2002

Dear Mr. Cymbaluk :

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set

{5}------------------------------------------------

# Page 2 - Mr. William J. Cymbaluk

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Muriam C. Provost

for Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

Distal Femoral/Proximal Tibial Module 510(k) Premarket Notific infidential

510(k) Number (if known): K023087

Device Name: Global Modular Replacement System (GMRS)

#### Intended Use

The Global Modular Replacement System (herein referred to as the GMRS)) is intended to be used with the components of Howmedica Osteonics' Modular Replacement System, Modular Rotating Hinge Knee System, and/or Kinematic® Rotating Hinge Knee System in situations where replacement of extensive bone loss in the femur and/or proximal tibia is required. The Proximal Femoral Module and Distal Femoral Module of the GMRS may be used together, or with the components of the MRS, in replacement of the total femur. Indications and contraindications for this type of replacement are listed below:

#### Indications

Femoral and/or proximal tibial replacement and total femoral replacement in Oncology cases where radical resection and replacement of bone is required, and in limb salvage procedures where radical resection and replacement of bone is required. Limb salvage procedures would include surgical intervention for severe trauma, failed previous prosthesis, and/or Oncology indications.

#### Contraindications

As Related to Bone Tumors A.

Not all bone turnors may be treated successfully by segmental resection. Any condition that may have already resulted in either local or distant spread of the tumor may be a contraindication. Examples of such conditions include:

- pathological fracture; .
- . overt infection;
- inopportune placement of biopsy incision; and, .
- rapid disease progression beyond a respectable margin. ●

1

{7}------------------------------------------------

K023087

- As related to Failed Previous Prosthesis and Trauma B.
- Any active or suspected latent infection in or about the hip joint. ●
- Any mental or neuromuscular disorder which would create an unacceptable risk of . prosthesis instability, prosthesis fixation failure, or complication in postoperative care.
- Bone stock compromised by disease, infection, or prior implantation that cannot . provide adequate support and fixation of the prosthesis.

For the use of GMRS Press Fit stems with PureFix™ HA Coating, the following additional contraindication should be noted:

- Inadequate bone stock to allow the use of a press fit stem ●
The Proximal Femoral Module of the GMRS is intended to be used in a cemented or press fit mode. The components of the Distal Femoral/Proximal Tibial Module are intended to be used with bone cement.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use _ र OR 801.109)

Over-the Counter-Use No (per 21 CFR

Miriam C. Provost

ion of General, Restorative and Neurological Devices

510(k) Number K023087

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/KRO/K023087](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/KRO/K023087)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
