K243024 · United Orthopedic Corporation · JWH · Dec 20, 2024 · Orthopedic
Device Facts
Record ID
K243024
Device Name
Cellbrick Knee Spacer
Applicant
United Orthopedic Corporation
Product Code
JWH · Orthopedic
Decision Date
Dec 20, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3560
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
This product is indicated for temporary use (maximum of 180 days) as a total knee replacement (TKR) in skeletally mature patients undergoing a two-stage procedure due to a septic process. This product is not intended for use for more than 180 days, at which time it must be explanted and a permanent device implanted or another appropriate treatment performed (e.g., resection arthroplasty, fusion, etc.). During the implantation period, patients have to use traditional mobility assist devices (e.g. crutches, walkers, canes) for daily activities. With pre-clinical validations of antibiotic elution, tibial spacer wear test, and knee spacer wear test, only Palacos MV+G bone cement may be used for preparing the spacers.
Device Story
Cellbrick Knee Spacer is a temporary, multi-component implant system for two-stage revision knee surgery in septic patients. System includes femoral spacer, tibial spacer, and optional canal rod. Femoral and tibial spacers feature multi-aperture, scaffold-like designs to act as carriers for Palacos MV+G antibiotic-loaded bone cement, facilitating local antibiotic release at the infection site. Canal rod (Ti-6Al-4V) provides additional surface area for cement attachment in deeper infection cases. Device is implanted by orthopedic surgeons in a clinical/hospital setting. Patients must use mobility aids (crutches, walkers) during the 180-day maximum implantation period. After 180 days, the spacer must be explanted and replaced with a permanent prosthesis or alternative treatment. The device provides temporary joint function and infection management, bridging the gap between initial debridement and definitive reconstruction.
Clinical Evidence
Bench testing only. No clinical data provided. Testing included fatigue testing for tibial spacer, range of motion, constraint performance, wear testing, antibiotic release/in vitro antibiotic testing, usability evaluation, and endotoxin testing.
Technological Characteristics
Components: Femoral and Tibial spacers (UHMWPE, ASTM F648/ISO 5832-2), Canal Rod (Ti-6Al-4V, ASTM F136/ISO 5832-3). Design: Multi-aperture scaffold for antibiotic-loaded bone cement (Palacos MV+G). Fixation: Cemented. Sterilization: Not specified. Connectivity: None.
Indications for Use
Indicated for skeletally mature patients undergoing a two-stage total knee replacement procedure due to a septic process. Temporary use only (max 180 days). Requires use of mobility assist devices during implantation.
Regulatory Classification
Identification
A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the date December 20, 2024. The text is written in a clear, sans-serif font. The date is written in a standard format, with the month, day, and year clearly indicated. The image is simple and straightforward, with the focus on the date itself.
United Orthopedic Corporation Lois Ho Regulatory Affairs Manager No 57, Park Ave 2, Science Park Hsinchu, 30075 Taiwan
Re: K243024
Trade/Device Name: Cellbrick Knee Spacer Regulation Number: 21 CFR 888.3560 Regulation Name: Knee Joint Patellofemorotibial Polymer Semi-Constrained Cemented Prosthesis Regulatory Class: Class II Product Code: JWH Dated: September 13, 2024 Received: September 27, 2024
Dear Lois Ho:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{2}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{3}------------------------------------------------
Sincerely, JESSE MUIR - Digitally signed by JESSE
MUIR - MUIR - MUIR -s mo
Date: 2024.12.20 16:14:33
-05'00' S Jesse Muir, PhD
Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indications for Use
510(k) Number (if known) K243024
Device Name Cellbrick Knee Spacer
#### Indications for Use (Describe)
This product is indicated for temporary use (maximum of 180 days) as a total knee replacement (TKR) in skeletally mature patients undergoing a two-stage procedure due to a septic process.
This product is not intended for use for more than 180 days, at which time it must be explanted and a permanent device implanted or another appropriate treatment performed (e.g., resection arthroplasty, fusion, etc.). During the implantation period, patients have to use traditional mobility assist devices (e.g. crutches, walkers, canes) for daily activities. With pre-clinical validations of antibiotic elution, tibial spacer wear test, and knee spacer wear test, only Palacos MV+G bone cement may be used for preparing the spacers.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for United Orthopedic. The logo consists of a green letter U with a white circle inside of it. To the right of the U is the text "UNITED ORTHOPEDIC" in black, with a registered trademark symbol after the word "ORTHOPEDIC".
## 510(K) SUMMARY
# Traditional 510(k)
[as required by 21 CFR 807.92(c)]
## Contact Details
| Applicant Name | United Orthopedic Corporation |
|--------------------------------|-------------------------------------------------------------|
| Applicant Address | No 57, Park Ave 2, Science Park, Hsinchu City 30075, Taiwan |
| Applicant Contact<br>Telephone | +88635773351 |
| Applicant Contact | Mrs. Lois Ho |
| Applicant Contact Email | lois.ho@unitedorthopedic.com |
### Device Name
| Device Trade Name | Cellbrick Knee Spacer |
|---------------------|-----------------------------------------------------------------------------------------------|
| Common Name | Prosthesis, knee, patellofemorotibial, semi-constrained,<br>cemented, polymer/metal/polymer |
| Classification Name | Knee joint patellofemorotibial polymer/metal/polymer semi-<br>constrained cemented prosthesis |
| Requlation Number | 888.3560 |
| Product Codes | JWH |
### Predicate Device Information
| 510(k) Number | Predicate Trade Name | Product Code |
|--------------------------------|-----------------------------|--------------|
| Primary Predicate | | |
| K222570 | COPAL® knee moulds | JWH |
| Predicate or Reference Devices | | |
| K101356 | Spacer-K | JWH |
| K181732 | Spacer-K | JWH |
| K183017 | REMEDY® Stemmed Knee Spacer | JWH |
| K140073 | U2 Total Knee System | JWH |
| K132752 | U2 Total Knee System | JWH |
| K150829 | U2 Total Knee System | JWH |
| K222700 | U2 Total Knee System | JWH |
| Device Description | Cellbrick Knee Spacer is a temporary knee spacer product, it's<br>suitable for the patients with mature skeleton and required to<br>perform two-stage knee joint prosthesis procedure for infection<br>control. It consists of three components, which are (1) Femoral<br>Spacer, (2) Tibial Spacer, and (3) Canal Rod. The Femoral<br>Spacers and the Tibial Spacers are designed with multiaperture<br>features, which can act as carrier of antibiotic-loaded bone |
|----------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Summary | |
| | cement to release antibiotic at infection site. The Canal Rod is<br>optional.<br>- Femoral Spacer<br>The Femoral Spacer is made of UHMWPE (ASTM<br>F648/ISO5834-2). The geometrical appearance of Femoral<br>Spacer is designed and modified based on the 510(k) cleared<br>cruciate-retaining femoral component (K140073). Their<br>articular surface geometry and curvature is identical. In<br>addition, the Femoral Spacer is designed with multiaperture<br>feature. Such scaffold-like features can use as a carrier of<br>antibiotic-loaded cement to release antibiotic at the infection<br>site for infection control.<br>The Femoral Spacer can be used for both of left and right knee<br>because of the symmetric condyle design. Five femoral spacer<br>sizes have been prepared for different anatomical demands<br>with anteroposterior (AP) and mediolateral (ML) dimensions.<br>The Femoral Spacer is designed “same/ one up" sizing options<br>with the tibial spacers.<br>- Tibial Spacer<br>The Tibial Spacer is made of UHMWPE (ASTM F648/ISO5834-<br>2). The bearing surface's curvature and geometrical design of<br>the tibial spacer is the same as the 510(k) cleared ultra-<br>congruent tibial insert (K132752, K150829, and K222700). The<br>Tibial Spacer is designed with multiaperture feature as well.<br>This scaffold-like feature is used as a carrier of antibiotic-<br>loaded cement to release antibiotic at the infection site for<br>infection control.<br>Five Tibial Spacer sizes have been prepared for different<br>anatomical demands with anteroposterior (AP) and<br>mediolateral (ML) dimensions.<br>- Canal Rod<br>The canal rod is made of Ti-6Al-4V alloy (ASTM F136/ISO<br>5832-3). As a carrier, the rod provides the area for cement to<br>attach so that to increase the surface area of cement. The<br>canal rod is an optional device, it could be applied to patient<br>with deeper infection problem, applying Canal Rod can make<br>surgeon placing antibiotic-loaded cement into the deeper<br>infection site. The Canal Rod is provided with one dimension:<br>Ø 4(diameter) ×80mm (length). |
| Intended Use/<br>Indications for Use | This product is indicated for temporary use (maximum of 180<br>days) as a total knee replacement (TKR) in skeletally mature<br>patients undergoing a two-stage procedure due to a septic<br>process. |
| | This product is not intended for use for more than 180 days, at<br>which time it must be explanted and a permanent device<br>implanted or another appropriate treatment performed (e.g.,<br>resection arthroplasty, fusion, etc.). During the implantation<br>period, patients have to use traditional mobility assist devices<br>(e.g. crutches, walkers, canes) for daily activities.<br>With pre-clinical validations of antibiotic elution, tibial spacer<br>fatigue test, and knee spacer wear test, only Palacos MV+G<br>bone cement may be used for preparing the spacers. |
| Indications for Use<br>Comparison | The indication for use of the subject device is substantially<br>equivalent to the primary predicate.<br>Although the content of indication for use is not completely<br>identical, however, both of the subject device and primary<br>predicate are temporary implants (maximum of 180 days),<br>which is intended to apply for the patient with mature skeleton<br>and required a two-stage procedure due to a septic process. In<br>addition, both should use mobility assist devices during the<br>period of implantation.<br>Base on above mentioned, the indication for use of subject<br>device is considered as substantially equivalent to the primary<br>predicate. |
| Technological<br>Comparison | The technological characteristics of the subject device are<br>substantially equivalent to those of the primary predicate<br>(K222570), as the comparison given below shows.<br>- Their intended use is the same.<br>- The intended users, intended surgical procedure, body<br>contact and duration, and fixation method are identical to the<br>primary predicate.<br>- For the difference in the material, appearance design,<br>specification, device X-ray visibility, and sterilization method,<br>evaluation and verification activities were conducted, and the<br>equivalent safety and effectiveness compared to the predicate<br>devices were validated.<br>It established that no new risks arise compared to those of the<br>predicate devices. |
| Non-Clinical and/or<br>Clinical Tests Summary<br>& Conclusions | Based on the technological characteristics of the subject device,<br>the following evaluation and tests were conducted to validate<br>the safety and effectiveness of the subject device.<br>- Fatigue test for Tibial spacer<br>- Range of Motion<br>- Constraint performance on the tibiofemoral interface<br>- Wear test<br>- Antibiotic release test & In vitro antibiotic test<br>- Usability evaluation<br>- Endotoxin testing |
| | Not Applicable. |
| | Based upon equivalences in intended use, patient population,<br>site of application, conditions of use, and non-clinical<br>performance data, the Subject device has been shown to be<br>safe and effective and to perform equivalently as compared to<br>the legally marketed predicate devices.<br>Therefore, the Subject device is substantially equivalent to the<br>legally marketed predicate devices. |
{6}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for United Orthopedic. The logo consists of a green letter U with a white circle inside, next to the words "UNITED ORTHOPEDIC" in black. The word "ORTHOPEDIC" has a registered trademark symbol next to it.
{7}------------------------------------------------
Image /page/6/Picture/1 description: The image contains the logo for United Orthopedic. The logo consists of a green letter U with a white circle inside, followed by the words "UNITED ORTHOPEDIC" in black. The word "ORTHOPEDIC" has a registered trademark symbol next to it.
{8}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the logo for United Orthopedic. The logo consists of a green circle with a white circle inside, resembling the letter 'U'. To the right of the symbol, the words 'UNITED' and 'ORTHOPEDIC' are stacked on top of each other in a simple, sans-serif font. A small registered trademark symbol appears next to the word 'ORTHOPEDIC'.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.