← Product Code [JWH](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/JWH) · K242046

# EXULT Knee Replacement System (K242046)

_Corentec Co., Ltd. · JWH · Aug 1, 2024 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/JWH/K242046

## Device Facts

- **Applicant:** Corentec Co., Ltd.
- **Product Code:** [JWH](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/JWH.md)
- **Decision Date:** Aug 1, 2024
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 888.3560
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

EXULT Knee Replacement System is indicated for the treatment of diseases as follows: · Painful, disabling joint disease of the knee resulting from non-inflammatory degenerative joint disease (including osteoarthritis traumatic arthritis or avascular necrosis) or rheumatoid arthritis. · Post-traumatic loss of knee joint configuration and function. • Moderate varus, valgus, or flexion deformity in which the ligamentous structures can be returned to adequate function and stability. · Correction or revision of unsuccessful osteotomy, arthrodesis, or failure of previous arthroplasty procedure. EXULT Knee Replacement System is intended for cemented application only.

## Device Story

The EXULT Knee Replacement System is a total knee arthroplasty system. This submission introduces an ultra-congruent (UC-type) tibial insert as a line extension to the existing system. The UC-type insert features high conformity and increased anterior/posterior edge height to limit anterior/posterior motion, functioning without the post structure found in PS-type inserts. It is designed for use with existing femoral components and tibial baseplates from the predicate system. The device is used by orthopedic surgeons in a clinical/surgical setting to restore knee joint function and stability in patients with degenerative or inflammatory joint disease. The surgeon selects the appropriate component size and configuration to address the patient's specific deformity. The device is implanted via cemented fixation, providing a mechanical bearing surface to replace damaged natural joint surfaces, thereby reducing pain and improving mobility.

## Clinical Evidence

Bench testing only. No clinical data provided. Testing included wear testing (ISO 14242-1&2, ASTM F2003-02), constraint testing (ASTM F1223), contact pressure testing (ASTM F2083, ASTM F1672), and range of motion testing (ISO 21536).

## Technological Characteristics

Total knee joint replacement system. UC-type tibial insert provides ultra-congruent articulation with increased A/P edge height for stability. Materials and design are consistent with predicate system. Standards: ISO 14242-1&2 (wear), ASTM F2003-02 (wear), ASTM F1223 (constraint), ASTM F2083/F1672 (contact pressure), ISO 21536 (ROM). Cemented application only. Symmetrical design for left/right knee use.

## Regulatory Identification

A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).

## Predicate Devices

- EXULT Knee Replacement System ([K192507](/device/K192507.md))

## Reference Devices

- LOSPA Total Knee System ([K110404](/device/K110404.md), [K121037](/device/K121037.md), [K130673](/device/K130673.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

August 1, 2024

Corentec Co., Ltd. Seo Sanghee Senior Manager 33-2, Banpo-daero 20-gil, Seocho Gu, Seoul, 06649 Republic of Korea

Re: K242046

Trade/Device Name: EXULT Knee Replacement System Regulation Number: 21 CFR 888.3560 Regulation Name: Knee Joint Patellofemorotibial Polymer Semi-Constrained Cemented Prosthesis Regulatory Class: Class II Product Code: JWH Dated: July 12, 2024 Received: July 12, 2024

Dear Seo Sanghee:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

{2}------------------------------------------------

Sincerely,

Image /page/2/Picture/3 description: The image shows the name "Lixin Liu-S" in a simple, sans-serif font. The text is black against a light background. The letters are evenly spaced and clearly legible. The overall impression is clean and professional.

Lixin Liu, Ph.D. Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

## Indications for Use

Submission Number (if known)

K242046

Device Name

EXULT Knee Replacement System

Indications for Use (Describe)

EXULT Knee Replacement System is indicated for the treatment of diseases as follows:

· Painful, disabling joint disease of the knee resulting from non-inflammatory degenerative joint disease (including osteoarthritis traumatic arthritis or avascular necrosis) or rheumatoid arthritis.

· Post-traumatic loss of knee joint configuration and function.

• Moderate varus, valgus, or flexion deformity in which the ligamentous structures can be returned to adequate function and stability.

· Correction or revision of unsuccessful osteotomy, arthrodesis, or failure of previous arthroplasty procedure.

EXULT Knee Replacement System is intended for cemented application only.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

# 510(k) Summary

## ADMINISTRATIVE INFORMATION

Manufacturer: Corentec Co., Ltd. 12, Yeongsanhong 1-gil, Ipjang-Myeon, Seobuk-Gu, Cheonansi, Chungchongnam-do, Republic of Korea (31056) Telephone: +82-41-585-7114 Fax: +82-41-585-7113 Official Contact: Sanghee Seo RA Senior Manager 33-2, Banpo-daero 20-gil, Seocho-gu, Seoul, Republic of Korea (06649) Tel: +82-2-6937-4902 Fax: +82-2-3445-5497 E-mail: sh.seo@corentec.com Date Prepared: July 12, 2024

## DEVICE NAME AND GENERAL INFORMATION

| Trade/Proprietary Name: | EXULT Knee Replacement System |
|-------------------------|-------------------------------|
| Common Name:            | Total Knee Joint Replacement  |
| Classification          | 21 CFR 888.3560               |
| Regulations: Regulatory | Class II                      |
| Class: Product Codes:   | JWH                           |
| Classification Panel:   | Orthopedic Products Panel     |

#### Predicate device

Primary predicate device: EXULT Knee Replacement System, Corentec Co., Ltd. (K192507).

Reference device: LOSPA Total Knee System, Corentec Co., Ltd. (K110404, K121037, K130673)

## Device Description

The purpose of the submission is a line extension of the EXULT Tibial Insert (UC-type) to the EXULT Knee Replacement System (K192507). The UC-type Tibial Inserts are designed based on the design of the other Tibial Inserts of the EXULT Knee Replacement System (PS-type and CR-type). The UC-type, which is ultra-congruent, has high conformity and increased anterior/posterior edge height, which limits anterior/posterior motion instead of a post structure, like the PS-type. The UC-type Tibial Insert matches Femoral Components of one-size up and down. The UC-type Tibial Insert is symmetrical and used in both the right and left knee. It is only used in combination with the Femoral component/Tibial baseplate of the predicate device (K192507).

{5}------------------------------------------------

#### Indications for Use

EXULT Knee Replacement System is indicated for the treatment of diseases as follows:

- Painful, disabling joint disease of the knee resulting from non-inflammatory degenerative . joint disease (including osteoarthritis, traumatic arthritis or avascular necrosis) or rheumatoid arthritis.
- . Post-traumatic loss of knee joint configuration and function.
- . Moderate varus, valgus, or flexion deformity in which the ligamentous structures can be returned to adequate function and stability.
- . Correction or revision of unsuccessful osteotomy, arthrodesis, or failure of previous arthroplasty procedure.

EXULT Knee Replacement System is intended for cemented application only.

#### Summary of Technological Characteristics

Device Comparisons and performance testing indicate that EXULT Tibial Insert (UC-type) of the EXULT Knee Replacement System is substantially equivalent to the predicates in terms of intended use, indications, design, materials, performance characteristics and operational principles. It offers an additional option of constraint to the existing Tibial Inserts available with the EXULT Knee Replacement System.

#### Non-Clinical Testing

The Corentec Co., Ltd. has evaluated the subject EXULT Tibial Insert (UC-type) to demonstrate substantial equivalence to the predicate devices.

- . Wear Testing as per ISO 14242-1&2, ASTM F2003-02
- Constraint test as per ASTM F1223 ·
- . Contact pressure as per ASTM F2083, ASTM F1672
- . Range of Motion as per ISO 21536

#### Substantial Equivalence Conclusion

Based on above, the subject device is determined to be Substantially Equivalent (SE) to the predicate device in respect of safety and effectiveness.

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/JWH/K242046](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/JWH/K242046)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
