iTotal® Identity Posterior Stabilized Knee Replacement System

K201023 · Conformis, Inc. · JWH · Jun 16, 2020 · Orthopedic

Device Facts

Record IDK201023
Device NameiTotal® Identity Posterior Stabilized Knee Replacement System
ApplicantConformis, Inc.
Product CodeJWH · Orthopedic
Decision DateJun 16, 2020
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3560
Device ClassClass 2
AttributesTherapeutic

Intended Use

The iTotal Identity Posterior Stabilized Knee Replacement System is intended for use as a total knee replacement in patients with knee joint pain and disability whose conditions cannot be solely addressed by the use of a prosthetic device that treats only one or two of the three compartments, such as a unicondylar, patellofemoral or bicompartmental prosthesis. The Indications for Use include: · Painful joint disease due to osteoarthritis, traumatoid arthritis, polyarthritis, polyarthritis or osteonecrosis of the knee. - · Post traumatic loss of joint function. - · Moderate varus, valgus or flexion deformity. - · Failed osteotomies, hemiarthroplasties, and unicondylar, patellofemoral or bicompartmental implants. · Revision procedures provided that anatomic landmarks necessary for alignment and positioning of the implant are identifiable on patient imaging scans. This implant is intended for cemented use only.

Device Story

Patient-specific semi-constrained knee replacement system; includes femoral, tibial, and patellar components. Input: patient CT or MR scans. Process: design of patient-specific implants matching individual anatomy/geometry. Output: custom-fit femoral (CoCrMo), tibial (titanium tray, polyethylene insert), and patellar (UHMWPE) components. Used in orthopedic surgery; implanted by surgeons. Includes patient-specific single-use manual instruments for bone cutting/implant positioning. Benefits: improved fit/alignment for patients with severe knee pain/disability.

Clinical Evidence

Bench testing only. Performance evaluation included insert interlock/modularity (ASTM F1814, F2083), tibial tray fatigue (ASTM F1800, F2083), simulated use testing, and MR compatibility (ASTM F2052-14, F2213-06, F2182-11a, F2119-07).

Technological Characteristics

Patient-specific semi-constrained knee implant. Materials: CoCrMo alloy (femoral), titanium alloy (tibial tray), UHMWPE (patellar/inserts). Standards: ASTM F1814, F2083, F1800, F2052-14, F2213-06, F2182-11a, F2119-07. Manual surgical instrumentation provided. Cemented fixation.

Indications for Use

Indicated for patients with knee joint pain/disability requiring total knee replacement due to osteoarthritis, traumatic/rheumatoid arthritis, polyarthritis, osteonecrosis, post-traumatic loss of function, moderate varus/valgus/flexion deformity, or failed prior partial/hemi-arthroplasty/osteotomy. Suitable for revision if anatomic landmarks are identifiable on imaging. Cemented use only.

Regulatory Classification

Identification

A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ June 16, 2020 Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a larger font size than the full name. The full name of the agency is "U.S. Food & Drug Administration". Conformis, Inc. Nancy Giezen Manager Regulatory Affairs 600 Technology Park Drive Billerica, Massachusetts 01821 Re: K201023 Trade/Device Name: iTotal Identity Posterior Stabilized Knee Replacement System Regulation Number: 21 CFR 888.3560 Regulation Name: Knee Joint Patellofemorotibial Polymer Semi-Constrained Cemented Prosthesis Regulatory Class: Class II Product Code: JWH, OIY, OOG Dated: April 15, 2020 Received: April 20, 2020 Dear Nancy Giezen: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. Ting Song, Ph.D., R.A.C. Acting Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K201023 Device Name iTotal Identity Posterior Stabilized Knee Replacement System Indications for Use (Describe) The iTotal Identity Posterior Stabilized Knee Replacement System is intended for use as a total knee replacement in patients with knee joint pain and disability whose conditions cannot be solely addressed by the use of a prosthetic device that treats only one or two of the three compartments, such as a unicondylar, patellofemoral or bicompartmental prosthesis. The Indications for Use include: · Painful joint disease due to osteoarthritis, traumatoid arthritis, polyarthritis, polyarthritis or osteonecrosis of the knee. - · Post traumatic loss of joint function. - · Moderate varus, valgus or flexion deformity. - · Failed osteotomies, hemiarthroplasties, and unicondylar, patellofemoral or bicompartmental implants. · Revision procedures provided that anatomic landmarks necessary for alignment and positioning of the implant are identifiable on patient imaging scans. This implant is intended for cemented use only. Type of Use (Select one or both, as applicable) | Prescription Use (Part 21 CFR 801 Subpart D) | | |----------------------------------------------|--| | Over-The-Counter Use (21 CFR 801 Subpart C) | | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) SUMMARY | Submitter's Name and Address: | Conformis, Inc.<br>600 Technology Park Drive<br>Billerica, MA 01821 | |---------------------------------------|-----------------------------------------------------------------------| | Establishment Registration Number(s): | 3009844603 and 3004153240 | | Date Summary was Prepared: | June 10th, 2020 | | Contact Person: | Nancy Giezen<br>Manager Regulatory Affairs<br>Telephone: 781-345-9058 | Trade/Device Name(s): iTotal Identity Posterior Stabilized Knee Replacement System Common Name: Knee Replacement System Device Class: Class II Regulation Numbers: 888.3560 ## Classification Names and Product Codes: Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/PolymerKnee Arthroplasty Implantation System; JWH, OIY, OOG Legally Marketed Predicate Device (Primary Predicate): iTotal Posterior Stabilized (PS) Knee Replacement System (K161668, K193105) ## (Secondary Predicates): iTotal Identity Cruciate Retaining (CR) Knee Replacement System (K190562) iTotal Cruciate Retaining (CR) Knee Replacement System (K193105) DePuy PFC Sigma Knee Prosthesis (K952830, K060515) {4}------------------------------------------------ # Device Description: Conformis knee replacement systems are patient-specific semi-constrained knee implants which consist of a femoral, tibial, and/or patellar components. The products are intended for treatment of severe pain and/or disability of the knee damaged by osteoarthritis or trauma. Using patient imaging (either CT or MR scans), a patient-specific implant is designed that best meets the geometric and anatomic requirements of the specific patient. The femoral components of the subject devices are manufactured from cobalt chromium molybdenum (CoCrMo) alloy. The tibial component includes a metal tray manufactured from titanium alloy and polyethylene inserts. The patellar components are manufactured from UHMWPE. For user convenience, single-use, patient-specific ancillary orthopedic manual surgical instruments designed for use with the proposed iTotal Identity Posterior Stabilized Knee Replacement System are provided to assist in the positioning of the total knee replacement components intraoperatively and in guiding the cutting of bone. In addition, reusable orthopedic manual surgical instruments are provided separately. # Indications for Use: The iTotal Identity Posterior Stabilized Knee Replacement System is intended for use as a total knee replacement in patients with knee joint pain and disability whose conditions cannot be solely addressed by the use of a prosthetic device that treats only one or two of the three knee compartments, such as a unicondylar, patellofemoral or bicompartmental prosthesis. The Indications for Use include: - · Painful joint disease due to osteoarthritis, traumatic arthritis, rheumatoid arthritis, polyarthritis or osteonecrosis of the knee. - · Post traumatic loss of joint function. - · Moderate varus, valgus or flexion deformity. - · Failed osteotomies, hemiarthroplasties, and unicondylar, patellofemoral or bicompartmental implants. - · Revision procedures provided that anatomic landmarks necessary for alignment and positioning of the implant are identifiable on patient imaging scans. This implant is intended for cemented use only. # Technological Characteristics: The subject devices incorporate updates to the material and design which are consistent with previously cleared devices. The operating principle, fundamental technology, manufacturing methods and sterilization options are the same as the predicates. {5}------------------------------------------------ ## Non-Clinical Performance Evaluation: The following confirmatory testing was performed: - Insert interlock/Modularity per ASTM F1814 and ASTM F2083 ● - Tibial Tray Fatigue per ASTM F1800 and ASTM F2083 ● - Simulated Use Testing ● - MR Compatibility per ASTM F2052-14, ASTM F2213-06, ASTM F2182-11a, and ASTM F2119-07 ## Conclusion: Based on a comparison of the intended use and technological characteristics to predicate devices and on the results of confirmatory testing it is concluded that the proposed iTotal Identity Posterior Stabilized Knee Replacement System is substantially equivalent.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...