Smith & Nephew VISIONAIRE Patient Matched Cutting Blocks

K183010 · Smith & Nephew, Inc. · JWH · Nov 28, 2018 · Orthopedic

Device Facts

Record IDK183010
Device NameSmith & Nephew VISIONAIRE Patient Matched Cutting Blocks
ApplicantSmith & Nephew, Inc.
Product CodeJWH · Orthopedic
Decision DateNov 28, 2018
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 888.3560
Device ClassClass 2
AttributesTherapeutic

Intended Use

Smith & Nephew's VISIONAIRE Patient Matched Cutting Blocks are intended to be used as patient-specific surgical instrumentation to assist in the positioning of total knee replacement components intra-operatively and in guiding the marking of bone before cutting provided that anatomic landmarks necessary for alignment and positioning of the implant are identifiable on patient imaging scans. The Smith & Nephew VISIONAIRE Patient Matched Cutting Blocks are intended for use with the following existing Smith & Nephew, Inc. Knee Systems and their cleared indications for use: - Genesis II Knee System - Legion Knee System - Journey BCS Knee System - Journey II Knee System The Smith & Nephew VISIONAIRE Patient Matched Cutting Blocks are intended for single use only.

Device Story

VISIONAIRE Patient Matched Cutting Blocks are patient-specific surgical instruments used during total knee replacement. Input data consists of patient imaging (MRI, X-ray) processed via VISIONAIRE software to design and manufacture custom blocks. These blocks assist surgeons in intra-operative positioning of knee implants and guiding bone marking prior to resection. Used in clinical settings by orthopedic surgeons; the output is a physical, single-use guide that ensures alignment based on patient-specific anatomy. This process aims to improve surgical accuracy and implant positioning, potentially enhancing patient outcomes. This submission specifically covers software component updates, including a new X-ray measurement tool and an upgraded alignment tool.

Clinical Evidence

No clinical data was needed to support safety and effectiveness. Bench testing was limited to software verification and validation in accordance with FDA guidance.

Technological Characteristics

Patient-specific surgical cutting blocks. No changes to material composition, packaging, or manufacturing processes. Software-driven design based on MRI/X-ray imaging. Software components classified as 'major' level of concern.

Indications for Use

Indicated for patients undergoing total knee replacement requiring patient-specific surgical instrumentation for positioning implants and marking bone, provided anatomic landmarks are identifiable on imaging scans. Compatible with Genesis II, Legion, Journey BCS, and Journey II Knee Systems. Single use only.

Regulatory Classification

Identification

A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and the word "ADMINISTRATION" in a smaller font size below. November 28, 2018 Smith & Nephew, Inc. Brad Sheals Regulatory Affairs Manager 7135 Goodlett Farms Parkway Cordova, Tennessee 38016 Re: K183010 Trade/Device Name: Smith & Nephew Patient Matched Cutting Blocks Regulation Number: 21 CFR 888.3560 Regulation Name: Knee Joint Patellofemorotibial Polymer Semi-Constrained Cemented Prosthesis Regulatory Class: Class II Product Code: JWH. MBH. OOG Dated: October 30, 2018 Received: October 31, 2018 Dear Brad Sheals: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal {1}------------------------------------------------ statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803). please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Image /page/1/Figure/5 description: The image shows a digital signature. The signature is for Peter G. Allen -S. The date of the signature is 2018.11.28, and the time is 11:43:36 -05'00'. FOR Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration # Indications for Use 510(k) Number (if known) # K183010 Device Name Smith & Nephew VISIONAIRE Patient Matched Cutting Blocks Indications for Use (Describe) Smith & Nephew's VISIONAIRE Patient Matched Cutting Blocks are intended to be used as patient-specific surgical instrumentation to assist in the positioning of total knee replacement components intra-operatively and in guiding the marking of bone before cutting provided that anatomic landmarks necessary for alignment and positioning of the implant are identifiable on patient imaging scans. The Smith & Nephew VISIONAIRE Patient Matched Cutting Blocks are intended for use with the following existing Smith & Nephew, Inc. Knee Systems and their cleared indications for use: - Genesis II Knee System - Legion Knee System - Journey BCS Knee System - Journey II Knee System The Smith & Nephew VISIONAIRE Patient Matched Cutting Blocks are intended for single use only. | Type of Use (Select one or both, as applicable) | |-------------------------------------------------| |-------------------------------------------------| > Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below. {3}------------------------------------------------ Image /page/3/Picture/1 description: The image contains the Smith & Nephew logo. The logo consists of an orange flower-like symbol on the left, followed by the text "We are" in gray and "smith&nephew" in orange. The text is aligned horizontally, with the flower symbol positioned to the left of the text. | Submitted by: | Smith & Nephew, Inc.<br>Orthopaedic Division<br>1450 East Brooks Road<br>Memphis, Tennessee 38116 | |----------------------------------------------|-----------------------------------------------------------------------------------------------------------------| | Date of Summary: | 11/27/2018 | | Contact Person and Address: | Brad Sheals<br>Requlatory Affairs Manager<br>T 901-399-6897<br>F 901-566-7911 | | Name of Device: | Smith & Nephew VISIONAIRE Patient Matched Cutting<br>Blocks | | Common Name: | Knee Prosthesis | | Device Classification Name and<br>Reference: | 21 CFR 888.3560- Knee joint patellofemorotibial<br>polymer/metal/polymer non-constrained cemented<br>prosthesis | | Device Class: | Class II | | Panel Code: | Orthopaedics/87 | | Product Code: | JWH, MBH, OOG | ## Device Description The subject of this premarket notification is to seek FDA clearance of software components to be used in the design and manufacture of the VISIONAIRE Patient Matched Cutting Blocks. Patient Matched Cutting Blocks were previously cleared for market via premarket notifications- K172336. The blocks are designed utilizing the VISIONAIRE Patient Matched Technology software components and patient imaging data (MRI, X-Ray). The blocks are intended to be used as patient-specific surgical instrument to assist in the positioning of total knee replacement implant components intra-operatively and in guiding the marking of bone before cutting. ### Intended Use Smith & Nephew's VISIONAIRE Patient Matched Cutting Blocks are intended to be used as patient-specific surgical instrumentation to assist in the positioning of total knee replacement components intra-operatively and in guiding the marking of bone before cutting provided that anatomic landmarks necessary for alignment and positioning of the implant are identifiable on patient imaging scans. The Smith & Nephew VISIONAIRE Patient Matched Cutting Blocks are intended for use with the following existing Smith & Nephew, Inc. Knee Systems and their cleared indications for use: - Genesis II Knee System - Legion Knee System - Journey BCS Knee System - Journey II Knee System {4}------------------------------------------------ The Smith & Nephew VISIONAIRE Patient Matched Cutting Blocks are intended for single use only. #### Technological Characteristics No new mechanical testing was performed. No implants or new blocks are being introduced in this premarket notification. There are no changes to the block design, packaging, material composition or manufacturing of Smith & Nephew VISIONAIRE Patient Matched Cutting Blocks as a result of these changes described in the premarket notification. Clinical data was not needed to support the safety and effectiveness of the subject device(s). The following technological differences exist between the subject device and predicate device, which are related to the manufacturing and design process: -The use of a different X-ray measurement tool. -The upgrade of an alignment tool. Software verification and validation testing was completed in line with FDA's guidance document entitled, "General Principles of Software Validation; Final Guidance for Industry and FDA Staff," dated January 11, 2002. A review of this testing has demonstrated that there are no new issues related to the safety and effectiveness of the subject device and the software will perform as intended. Based on FDA's recommendations, the VISIONAIRE Patient Matched Technology software components for this device were considered as a "major" level of concern, since a failure in the software could directly result in serious injury or death to the patient or operator. Based on the documentation within this premarket notification, there are no new issues related to the safety and effectiveness of the subject device. Clinical data was not needed to support the safety and effectiveness of the subject device. #### Substantial Equivalence Information The Smith & Nephew VISIONAIRE Patient Matched Cutting Blocks are identical in function, equivalent design features, intended use, indications for use, operational principles, manufacturing processes, and materials as the predicate device- Patient Matched Cutting Blocks (K172336, S.E. 01/10/2018). #### Conclusion This premarket notification is being submitted to request clearance of the subject Smith & Nephew VISIONAIRE Patient Matched Cutting Blocks. Based on the similarities to the predicate cutting blocks and a review of the testing performed, the subject device is substantially equivalent to the predicate device- Patient Matched Cutting Blocks (K172336, S.E. 01/10/2018).
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...