EMPOWR VVC TIBIAL INSERT

K180930 · Encore Medical L.P. · JWH · Jul 9, 2018 · Orthopedic

Device Facts

Record IDK180930
Device NameEMPOWR VVC TIBIAL INSERT
ApplicantEncore Medical L.P.
Product CodeJWH · Orthopedic
Decision DateJul 9, 2018
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3560
Device ClassClass 2
AttributesTherapeutic

Indications for Use

Joint replacement is indicated for patients suffering from disability due to: - degenerative, post-traumatic or rheumatoid arthritis; - avascular necrosis of the femoral condyle; - post-traumatic loss of joint configuration, particularly when there is patellofemoral erosion, dysfunction or prior patellectomy; - moderate valgus, varus or flexion deformities; - treatment of fractures that are unmanageable using other techniques. This device may also be indicated in the salvage of previously failed surgical attempts. All devices are intended for cemented applications except for the 3DKNEE™ Porous Coated Femur, EMPOWR Porous™ Knee Femur and EMPOWR Porous™ Knee Tibia which are intended for cementless applications. While knee replacements are not intended to withstand activity levels and loads of normal healthy bone, they are a means of restoring mobility and reducing pain for many patients.

Device Story

EMPOWR VVC Tibial Insert is a line extension to the EMPOWR Knee platform; provides increased internal/external and varus/valgus rotational stability for total knee arthroplasty. Device consists of a tibial insert component; implanted by orthopedic surgeons in clinical settings. Functions as a semi-constrained prosthetic component to restore joint mobility and reduce pain in patients with severe knee disability. Output is physical joint stability; used by surgeons to address complex patient demographics requiring higher constraint than standard posterior stabilized inserts. Benefits include improved joint configuration and stability in salvage or complex deformity cases.

Clinical Evidence

No clinical data. Substantial equivalence supported by non-clinical mechanical testing, including femoral-tibial contact area, intrinsic stability, post fatigue, locking mechanism static/fatigue stability, reinforcement pin taper characterization, and varus/valgus constraint characterization.

Technological Characteristics

Material: Highly crosslinked UHMWPE with vitamin E (ASTM F648). Design: Semi-constrained tibial insert with modified proximal post geometry for increased rotational stability. Connectivity: None (mechanical implant). Sterilization: Not specified.

Indications for Use

Indicated for patients with disability due to degenerative, post-traumatic, or rheumatoid arthritis; avascular necrosis of femoral condyle; post-traumatic loss of joint configuration (patellofemoral erosion/dysfunction/prior patellectomy); moderate valgus, varus, or flexion deformities; or fractures unmanageable by other techniques. Also indicated for salvage of failed surgical attempts.

Regulatory Classification

Identification

A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. Encore Medical, L.P. Teffany Hutto Manager, Regulatory Affairs 9800 Metric Blvd. Austin, Texas 78758 Re: K180930 Trade/Device Name: EMPOWR VVC Tibial Insert Regulation Number: 21 CFR 888.3560 Regulation Name: Knee Joint Patellofemorotibial Polymer Semi-Constrained Cemented Prosthesis Regulatory Class: Class II Product Code: JWH, OIY Dated: April 9, 2018 Received: April 10, 2018 Dear Teffany Hutto: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good July 9, 2018 {1}------------------------------------------------ manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803). please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, # Mark N. Melkerson -S Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K180930 Device Name EMPOWR VVC Tibial Insert Indications for Use (Describe) Joint replacement is indicated for patients suffering from disability due to: - · degenerative, post-traumatic or rheumatoid arthritis; - · avascular necrosis of the femoral condyle; - · post-traumatic loss of joint configuration, particularly when there is patellofemoral erosion, dysfunction or prior patellectomy; - · moderate valgus, varus or flexion deformities; - · treatment of fractures that are unmanageable using other techniques. This device may also be indicated in the salvage of previously failed surgical attempts. All devices are intended for cemented applications except for the 3DKNEE™ Porous Coated Femur, EMPOWR Porous™ Knee Femur and EMPOWR Porous™ Knee Tibia which are intended for cementless applications. While knee replacements are not intended to withstand activity levels and loads of normal healthy bone, they are a means of restoring mobility and reducing pain for many patients. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------|---------------------------------------------| | X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary Date: July 9, 2018 Manufacturer: DJO Surgical (Legal Name: Encore Medical, L.P.) 9800 Metric Blvd Austin, TX 78758 Contact Person: Teffany Hutto Manager, Regulatory Affairs Phone: (512) 834-6255 Fax: (760) 597-3466 Email: teffany.hutto@djoglobal.com | Product | Common Name | Classification | Product Code | |--------------------------|--------------------|----------------|--------------| | EMPOWR VVC Tibial Insert | Total Knee Implant | Class II | JWH, OIY | | Product Code | Regulation and Classification Name | |--------------|------------------------------------------------------------------------------------------------------------------| | JWH | Knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis<br>per 21 CFR 888.3560 | | OIY | Knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis<br>per 21 CFR 888.3560 | #### Description: The EMPOWR Varus / Valgus Constrained (VVC) Tibial Insert and accessory is a line extension to the EMPOWR Knee platform (cleared via K143242), to include a more constrained tibial insert component that provides additional internal / external and varus / valgus rotational stability, which provide surgeons with comprehensive surgical solutions to address a variety of patient demographics and surgeon preferences. #### Indications for Use: Joint replacement is indicated for patients suffering from disability due to: - degenerative, post-traumatic or rheumatoid arthritis; ● - . avascular necrosis of the femoral condyle; - post-traumatic loss of joint configuration, particularly when there is patellofemoral erosion, ● dysfunction or prior patellectomy; - moderate valgus, varus or flexion deformities; ● - treatment of fractures that are unmanageable using other techniques. ● This device may also be indicated in the salvage of previously failed surgical attempts. All devices are intended for cemented applications except for the 3DKNEE™ Porous Coated Femur. EMPOWR Porous™ Knee Femur and EMPOWR Porous™ Knee Tibia which are intended for cementless applications. While knee replacements are not intended to withstand activity levels and loads of normal healthy bone, they are a means of restoring mobility and reducing pain for many patients. {4}------------------------------------------------ ### Predicate Devices: - EMPOWR PS Knee System (K160342) - Exprt Knee System (K140830) - Foundation Constrained Posterior Stabilized Tibial Insert (K933539) ● ## Reference Devices: - Encore Medical Acetabular System (K072154) - Highly Cross-Linked Vitamin E UHMWPE Tibial Insert (K091956) ● # Comparable Features to Predicate Device(s): The EMPOWR VVC Tibial Insert Implant is a line extension also includes a modified version of the EMPOWR PS Tibial Insert Implants (cleared via K160342). The sizing architecture, material (highly crosslinked UHMWPe with vitamin E per ASTM F648), minimum thickness offering (10mm), and distal locking mechanism geometry with the mating tibial baseplate remains identical to EMPOWR PS Tibial Insert Implants. # Key Differences in Subject Device to Predicate: A primary difference between the EMPOWR VVC Tibial Insert compared to the EMPOWR PS Tibial Insert Implants is the proximal side of the tibial insert and post geometry was modified to configure an additional amount of internal / external and varus / valgus constraint when compared to the EMPOWR PS Tibial Insert Implants. Non-Clinical Testing: Mechanical testing has demonstrated the device's ability to perform under expected conditions. This testing includes: - Femoral Tibial Contact Area ● - Femoral – Tibial Intrinsic Stability - VVC Post Fatigue - Locking mechanism Static Stability and Fatige ● - Reinforcement Pin Taper Characterization - Varus / Valgus Constraint Characterization ● All testing has determined that the device is substantially equivalent to the predicate devices. Endotoxin Assessment: DJO Surgical conducts device testing to assure that pyrogen limit specifications are met via the Kinetic Chromogenic method for bacterial endotoxin testing. Clinical Testing: Clinical testing was not required Conclusions: All testing and evaluations demonstrate that the device is substantially equivalent to the predicates identified.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...