← Product Code [JWH](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/JWH) · K172936

# Destiknee Total Knee System (K172936)

_Meril Healthcare Pvt. , Ltd. · JWH · Nov 30, 2017 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/JWH/K172936

## Device Facts

- **Applicant:** Meril Healthcare Pvt. , Ltd.
- **Product Code:** [JWH](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/JWH.md)
- **Decision Date:** Nov 30, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3560
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

The Destiknee™ Total Knee System is indicated for patients with severe knee pain and disability due to: · Severe knee joint pain, loss of mobility, and disability due to: rheumatoid arthritis, traumatic arthritis, polyarthritis. · Correction of functional deformities. · Post-traumatic loss of knee joint contour, particularly when there is patellofemoral erosion, dysfunction, or prior patellectomy. · Moderate valgus, varus, or flexion trauma. · Knee fractures untreatable by other methods. The Destiknee™ Total Knee System is intended for cemented use only. This device is for single use only.

## Device Story

Destiknee™ Total Knee System replaces femoral, tibial, and patellar articular surfaces of knee joint. System comprises metallic and non-metallic components in various styles/sizes. Intended for cemented fixation in surgical setting. Device restores joint function, corrects deformities, and alleviates pain in patients with severe knee pathology. Surgeon implants components during total knee arthroplasty. Benefits include improved mobility and pain relief.

## Clinical Evidence

Bench testing only. Fatigue testing of the Tibial Base Plate was performed per ASTM F1800-12 and ISO 14879-1:2000. Sterility and endotoxin testing (USP) were conducted to ensure manufacturing safety.

## Technological Characteristics

Semi-constrained cemented knee prosthesis. Components manufactured from metallic and non-metallic materials. Fatigue testing performed per ASTM F1800-12 and ISO 14879-1:2000. Sterilization via validated methods. System is a modular implant set.

## Regulatory Identification

A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).

## Predicate Devices

- Destiknee™ Total Knee System ([K160771](/device/K160771.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

November 30, 2017

Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

Meril Healthcare Pvt. Ltd. % Linda Braddon President & Chief Executive Officer Secure BioMed Evaluations 7828 Hickory Flat Highway, Suite 120 Woodstock, Georgia 30188

Re: K172936

Trade/Device Name: Destiknee™ Total Knee System Regulation Number: 21 CFR 888.3560 Regulation Name: Knee Joint Patellofemorotibial Polymer Semi-Constrained Cemented Prosthesis Regulatory Class: Class II Product Code: JWH Dated: September 27, 2017 Received: October 04, 2017

Dear Linda Braddon:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good

{1}------------------------------------------------

manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Mark N. Melkerson -S

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

### Indications for Use

510(k) Number (if known)

### K172936

Device Name DestikneeTM Total Knee System

Indications for Use (Describe)

The DestikneeTM Total Knee System is indicated for patients with severe knee pain and disability due to:

· Severe knee joint pain, loss of mobility, and disability due to: rheumatoid arthritis, traumatic arthritis, polyarthritis.

· Correction of functional deformities.

· Post-traumatic loss of knee joint contour, particularly when there is patellofemoral erosion, dysfunction, or prior patellectomy.

· Moderate valgus, varus, or flexion trauma.

· Knee fractures untreatable by other methods.

The Destiknee™ Total Knee System is intended for cemented use only. This device is for single use only.

Type of Use (Select one or both, as applicable)

> Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## 5. 510(k) SUMMARY

## 510(k) Summary

This 510(k) Summary is submitted in accordance with the requirements of 21 CFR Part 807. Section 92.

#### 5.1 Applicant:

Meril Healthcare Private Limited Survey No. 135/2/B & 174/2, First Floor, H1-H3, Meril Park Muktanand Marg, Chala, Vapi - 396 191, Gujarat, INDIA

#### 5.2 Regulatory Correspondent:

Linda Braddon, Ph.D President and CEO Secure BioMed Evaluations 7828 Hickory Flat Highway, Suite 120 Woodstock, GA, 30188 Email: lgb@securebme.com Phone: +-770-837-2681

### Meril Healthcare Contact Information:

Bhavik Gondaliya Manager- regulatory Affairs Meril Healthcare Private Limited Survey No. 135/2/B & 174/2, First Floor, H1-H3, Meril Park, Muktanand Marg, Chala, Vapi - 396 191, Gujarat, INDIA Mobile: +91 – 9687604713 Fax: +91 260-3052125 E mail: Bhavik.gondaliya@merillife.com

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the word "Meril" in a sans-serif font. The word is in a light blue color, and there is a small yellow circle above the "i" in "Meril". The background is also light blue.

| 5.3 | Date prepared:       | 27, September 2017                                                                                                  |
|-----|----------------------|---------------------------------------------------------------------------------------------------------------------|
| 5.4 | Device information:  |                                                                                                                     |
|     | Proprietary Name:    | DestikneeTM Total Knee System                                                                                       |
|     | Common / Usual Name: | Total Knee Replacement                                                                                              |
|     | Regulation name:     | Knee joint patellofemorotibial<br>polymer/metal/polymer semi-constrained cemented<br>prosthesis per 21 CFR 888.3560 |
|     | Product Code:        | JWH                                                                                                                 |
|     | Device Class:        | Class II                                                                                                            |

#### 5.5 Predicate Devices

| Component                                                                                                                                     | Predicate<br>Device                 | Manufacturer                  | Submission<br>Number         |
|-----------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------|-------------------------------|------------------------------|
| Femoral Component,<br>Tibial Component (Tibial Base<br>Plate and Tibial Articular<br>Surface), All Poly Tibial<br>Component, All Poly Patella | DestikneeTM<br>Total Knee<br>System | Meril Healthcare<br>Pvt. Ltd. | 510(k)<br>Number:<br>K160771 |

### 5.6 Device Description:

The Destiknee™ Total Knee System is a system of components intended to replace the Femoral, Tibial and Patella articular surfaces of the knee joint. The components are

{5}------------------------------------------------

available in many styles and sizes and are manufactured from metallic and non-metallic materials.

The Destiknee TM Total Knee System was originally cleared under 510(k): K160771. The purpose of current submission is to add Tibial Base Plate (size 7 and size 8) and All Poly Patella (size 37mm and 40mm) in Destiknee™ Total Knee System.

#### 5.7 Indications for Use:

The Destiknee™ Total Knee System is indicated for patients with severe knee pain and disability due to:

- Severe knee joint pain, loss of mobility, and disability due to: rheumatoid . arthritis, osteoarthritis, traumatic arthritis, polyarthritis.
- Correction of functional deformities. .
- Post-traumatic loss of knee joint contour, particularly when there is . patellofemoral erosion, dysfunction, or prior patellectomy.
- Moderate valgus, varus, or flexion trauma. .
- . Knee fractures untreatable by other methods.

The Destiknee™ Total Knee System is intended for cemented use only. This device is for single use only.

#### 5.8 Comparison of Technological characteristics:

The Destiknee™ Total Knee system is substantially equivalent to the previously cleared predicate device based on similarities in intended use, device design/technological characteristics, materials and sterilization method.

#### Non clinical Performance data: 5.9

The components of Destiknee™ Total Knee System were subjected to following nonclinical performance testing to evaluate device function/ mechanical performance.

- Fatigue test of Tibial Base Plate (ASTM F1800-12; ISO 14879-1:2000) ●

{6}------------------------------------------------

Endotoxin testing has demonstrated that the manufacturing process does not introduce Endotoxin as a bi-product of the manufacturing and cleaning process.

Destiknee™ Total Knee System is subjected to 14 day USP sterility testing and USP Endotoxin testing prior to release.

Meril Healthcare Pvt. Ltd. Confirms that the Bacterial Endotoxin Test [(BET)/LAL test] will be conducted on final sterilized device of every batch prior to releasing to US market.

### 5.10 Conclusion

Based on performance testing results and similarities in intended use, device design/technological characteristics, materials and sterilization method, the Destiknee™ Total Knee System is considered substantially equivalent to the previously cleared predicate device.

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/JWH/K172936](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/JWH/K172936)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
