← Product Code [JWH](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/JWH) · K171389

# EVOLUTION Revision CCK System (K171389)

_Microport Orthopedics, Inc. · JWH · Aug 8, 2017 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/JWH/K171389

## Device Facts

- **Applicant:** Microport Orthopedics, Inc.
- **Product Code:** [JWH](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/JWH.md)
- **Decision Date:** Aug 8, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3560
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

The EVOLUTION® Revision CCK system is indicated for use in knee arthroplasty in skeletally mature patients with the following conditions: 1. Non-inflammatory degenerative joint disease: including osteoarthritis, traumatic arthritis, or avascular necrosis; 2. Inflammatory degenerative joint disease, including rheumatoid arthritis; 3. Correction of functional deformity; 4. Revision procedures where other treatments or devices have failed; and treatment of fractures that are unmanageable using other techniques. The EVOLUTION" Medial-Pivot Total Knee System Nonporous implants are for cemented use only.

## Device Story

EVOLUTION® Revision CCK (Constrained Condylar Knee) system provides surgical options for revision or complex primary total knee replacements. System includes femoral components, augments, stem offset adapters, tibial inserts, and stem caps. Used in clinical orthopedic settings by surgeons to replace damaged knee joint surfaces. Device components are implanted to restore joint function and stability in patients with degenerative disease or failed prior arthroplasty. Output is a reconstructed knee joint; clinical decision-making relies on surgeon assessment of patient anatomy and joint pathology. Benefits include improved joint stability and function for patients with complex orthopedic needs.

## Clinical Evidence

No clinical data provided. Evidence consists of bench testing: bacterial endotoxin levels, patellofemoral/tibiofemoral contact area, tibial post shear fatigue, taper disassembly, and taper fatigue. Results showed performance comparable or superior to predicate K990030 and K162026.

## Technological Characteristics

Materials: Cobalt chrome alloy (ASTM F75), titanium alloy (ASTM F136), UHMWPE (ASTM F648). Semi-constrained cemented prosthesis. Available in 8 standard sizes and 2 plus sizes. Includes femoral component, augments, stem offset adapters, tibial insert, and stem cap. Cemented use only.

## Regulatory Identification

A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).

## Predicate Devices

- EVOLUTION® MP Total Knee System ([K093552](/device/K093552.md))
- EVOLUTION® MP Revision Femur ([K142550](/device/K142550.md))
- EVOLUTION® Revision Tibial System ([K162026](/device/K162026.md))

## Reference Devices

- ADVANCE® Revision (CCK System) ([K990030](/device/K990030.md))
- ADVANCE® STATURE Femoral Component ([K063731](/device/K063731.md))
- Whiteside Ortholoc Modular Post. Stab. Knee System ([K901992](/device/K901992.md))
- AXIOM Total Knee System ([K932677](/device/K932677.md))
- Omnifit Series I Total Knee System ([K863668](/device/K863668.md)/[K871769](/device/K871769.md)/[K862837](/device/K862837.md))
- MG II Total Knee System ([K892800](/device/K892800.md)/[K895639](/device/K895639.md)/[K911572](/device/K911572.md)/[K930831](/device/K930831.md))
- Continuum (CKS) Primary Total Knee System ([K882322](/device/K882322.md)/[K896542](/device/K896542.md)/[K901457](/device/K901457.md)/[K914072](/device/K914072.md))
- Ultrack (ADVANCE®) Total Knee ([K953439](/device/K953439.md))
- ADVANCE® Knee System ([K972626](/device/K972626.md))
- EVOLUTION® MP CS/CR and PS Tibial Inserts ([K093552](/device/K093552.md))
- ADVANCE® Revision CCK Tibial Insert ([K990030](/device/K990030.md))
- EVOLUTION® Revision Tibial Base and Screw ([K162026](/device/K162026.md))
- ADVANCE® Modular Tibial Component, Screw ([K973524](/device/K973524.md))
- ADVANCE® Ultra-Congruent Tibial Insert ([K972770](/device/K972770.md))
- EVOLUTION® MP Total Knee System ([K102380](/device/K102380.md))
- EVOLUTION® BIOFOAM Tibial Base ([K152298](/device/K152298.md))
- ADVANCE® Spiked Porous Tibial Base ([K063128](/device/K063128.md))
- EVOLUTION® Revision Stem Extensions ([K162026](/device/K162026.md))
- EVOLUTION® Revision Straight Stem Extension Adapters ([K162026](/device/K162026.md))
- EVOLUTION® Revision Distal and Posterior Femoral Augments ([K142550](/device/K142550.md))
- ADVANCE® Patella ([K122218](/device/K122218.md)/[K953439](/device/K953439.md))
- EVOLUTION® Revision Tibial Base ([K162026](/device/K162026.md))
- EVOLUTION® Revision Tibial Augments ([K162026](/device/K162026.md))
- EVOLUTION® Revision Modular Keel ([K162026](/device/K162026.md))
- EVOLUTION® Revision Tibial Stem Offset Adapters ([K162026](/device/K162026.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

#### DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, one behind the other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

MicroPort Orthopedics Inc. Matt Paul Regulatory Affairs Project Manager 5677 Airline Road Arlington, Tennessee 38002

August 8, 2017

Re: K171389

Trade/Device Name: EVOLUTION Revision CCK System Regulation Number: 21 CFR 888.3560 Regulation Name: Knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis Regulatory Class: Class II Product Code: JWH Dated: May 10, 2017 Received: May 11, 2017

Dear Matt Paul:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR

{1}------------------------------------------------

Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely.

# Mark N. Melkerson -S

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known)

#### K171389

Device Name

## EVOLUTION® Revision CCK System

Indications for Use (Describe)

The EVOLUTION® Revision CCK System is indicated for use in knee arthroplasty in skeletally mature patients with the following conditions:

1. Non-inflammatory degenerative joint disease: including osteoarthritis, traumatic arthritis, or avascular necrosis

- 2. Inflammatory degenerative joint disease, including rheumatoid arthritis;
- 3. Correction of functional deformity

4. Revision procedures where other treatments or devices have failed; and treatment of fractures that are unmanageable using other techniques.

The EVOLUTION® Medial-Pivot Total Knee System Nonporous implants are for cemented use only.

Type of Use (Select one or both, as applicable)

2 Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

# PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for MicroPort Orthopedics. The logo features a blue rounded square with a white circle inside, containing a red diamond shape. To the right of the icon, the word "MicroPort" is written in blue, with "Orthopedics" written in red below it. The tagline "Full Function. Faster." is visible below the company name.

EVOLUTION® Revision CCK System 510(k) Summary

# 510(k) Summary of Safety and Effectiveness

In accordance with the Food and Drug Administration Rule to implement provisions of the Safe Medical Devices Act of 1990 and in conformance with 21 CFR 807, this information serves as a Summary of Safety and Effectiveness for the use of EVOLUTION® Revision CCK System.

| Submitted by:                           | MicroPort Orthopedics Inc.<br>5677 Airline Rd, Arlington TN, 38002<br>Phone: 866-872-0211<br>Fax: 855-446-2247                                                          |
|-----------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Date:                                   | August 2, 2017                                                                                                                                                          |
| Contact Person:                         | Matt Paul                                                                                                                                                               |
| Proprietary Name:                       | EVOLUTION® Revision CCK  System                                                                                                                                         |
| Common Name:                            | Total knee implant                                                                                                                                                      |
| Classification Name<br>and Reference:   | 21 CFR 888.3560 - Knee joint patellofemorotibial metal/polymer/metal semi-<br>constrained cemented prosthesis.<br>Class II                                              |
| Subject Product Code<br>and Panel Code: | Orthopedics/87/JWH                                                                                                                                                      |
| Predicate Device:                       | K093552 EVOLUTION® MP Total Knee System<br>K142550 EVOLUTION® MP Revision Femur<br>K162026 EVOLUTION® Revision Tibial System<br>K093552 EVOLUTION® MP Total Knee System |

{4}------------------------------------------------

K171389 p.2/5

Image /page/4/Picture/1 description: The image is the logo for MicroPort Orthopedics. The logo features a blue square with a white circle inside, and a red diamond in the center of the circle. To the right of the square is the text "MicroPort" in blue, with "Orthopedics" in red below it. Below that is the text "Full Function. Faster" in blue.

EVOLUTION® Revision CCK System 510(k) Summary

| Reference Devices: | Femoral Component:                                                        |
|--------------------|---------------------------------------------------------------------------|
|                    | K990030 ADVANCE® Revision (CCK System)                                    |
|                    | K142550 EVOLUTION® MP Revision Femur                                      |
|                    | K063731 ADVANCE® STATURE Femoral Component                                |
|                    | K901992 Whiteside Ortholoc Modular Post. Stab. Knee System                |
|                    | K932677 AXIOM Total Knee System                                           |
|                    | K863668/K871769 Omnifit Series I Total Knee System                        |
|                    | K892800/K895639/K911572/K930831 MG II Total Knee System – Flat, AP Lipped |
|                    | K862837 Omnifit Series I Total Knee System                                |
|                    | K882322/K896542/K901457/K914072 Continuum (CKS) Primary Total Knee System |
|                    | K953439 Ultrack (ADVANCE®) Total Knee                                     |
|                    | K972626 ADVANCE® Knee System                                              |

Femoral Augment: K990030 ADVANCE® Revision CCK System

#### Tibial Insert:

K093552 EVOLUTION® MP CS/CR and PS Tibial Inserts K990030 ADVANCE® Revision CCK Tibial Insert K162026 EVOLUTION® Revision Tibial Base and Screw K973524 ADVANCE® Modular Tibial Component, Screw K063731 ADVANCE® STATURE Femoral Component K953439 Ultrack (ADVANCE®) Total Knee System K932677 AXIOM Total Knee System K972770 ADVANCE® Ultra-Congruent Tibial Insert K102380 EVOLUTION® MP Total Knee System K152298 EVOLUTION® BIOFOAM Tibial Base K063128 ADVANCE® Spiked Porous Tibial Base

MicroPort Orthopedics BV Hoogoorddreef 5, 1101 BA Amsterdam, The Netherlands +31 20 545 01 00

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image contains the logo for MicroPort Orthopedics. The logo consists of a blue square with rounded corners, inside of which is a white circle containing a red star-like shape. To the right of the square is the text "MicroPort" in blue, with "Orthopedics" underneath in red. Below that is the phrase "Full Function. Faster."

#### Device Information

#### A. Intended Use

The EVOLUTION® Revision CCK system is indicated for use in knee arthroplasty in skeletally mature patients with the following conditions:

- 1. Non-inflammatory degenerative joint disease: including osteoarthritis, traumatic arthritis, or avascular necrosis;
- 2. Inflammatory degenerative joint disease, including rheumatoid arthritis;
- 3. Correction of functional deformity;
- 4. Revision procedures where other treatments or devices have failed; and treatment of fractures that are unmanageable using other techniques.

The EVOLUTION" Medial-Pivot Total Knee System Nonporous implants are for cemented use only.

#### B. Device Description

EVOLUTION® Revision CCK (Constrained Condylar Knee) is being introduced to supplement MicroPort Orthopedics' knee product lines to provide options in the case of revision or complex primary total knee replacements. The design features are summarized below:

- Components manufactured from cobalt chrome alloy, titanium alloy, and UHMWPE conforming to ASTM F75, F136, and F648, respectively
- . Available in 8 standard sizes, left and right
- Available in 2 plus sizes, left and right
- . System includes femoral component, augments, stem offset adapters, tibial insert and stem cap

## C. Substantial Equivalence Information

The design features and materials of the subject devices are substantially equivalent to those of the predicate devices. The indications for use are identical to the predicate devices. The fundamental scientific technology of the modified devices has not changed relative to the predicate device. The safety and effectiveness of the subject devices are adequately supported by the substantial equivalence information, materials information, and analysis data provided within this Premarket Notification.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image contains the logo for MicroPort Orthopedics. The logo features a blue square with rounded corners, inside of which is a white circle containing a red star-like shape. To the right of the square is the text "MicroPort" in blue, with "Orthopedics" in red below it. Below that is the text "Full Function, Faster" in a smaller font.

K171389 p.4/5 EVOLUTION® Revision CCK System

510(k) Summary

# D. Nonclinical Testing

The subject EVOLUTION® Revision CCK system was evaluated for bacterial endotoxins and found to be less than the USP endotoxin limit of 20 EU/device.

The subject EVOLUTION® Revision CCK system was evaluated for patellofemoral contact area, tibiofemoral contact area, tibial post shear fatigue, taper disassembly, and taper fatigue.

Results for patellofemoral contact area testing concluded the subject EVOLUTION® Revision CCK system to perform as well as the predicate K990030 implants in respect to measured contact area. Results for tibiofemoral contact area testing concluded the subject EVOLUTION® Revision CCK System to possess greater maximum contact area than the predicate K990030 implants. Results of tibial post shear fatigue testing concluded the subject EVOLUTION® Revision CCK system to demonstrate greater fatigue resistance than the predicate K990030. Results of taper disassembly testing indicate the subject EVOLUTION® Revision CCK system will withstand typical in vivo disassociation forces as reported in peer reviewed literature. Results of taper fatigue testing demonstrated the subject EVOLUTION® Revision CCK Femoral constructs to achieve taper disassembly values greater than those of the predicate K162026.

# E. Clinical Testing

Clinical data was not provided for the subject devices.

{7}------------------------------------------------

K171389 p.5/5

Image /page/7/Picture/1 description: The image is the logo for MicroPort Orthopedics. The logo features a blue square with rounded corners on the left side, inside of which is a white circle with a red diamond shape in the center. To the right of the square is the text "MicroPort" in blue, with "Orthopedics" in red below it. Below that is the text "Full Function. Faster."

EVOLUTION® Revision CCK System 510(k) Summary

## F. Component and Accessory Compatibility

The subject devices are compatible with the following previously cleared MicroPort Orthopedics products.

#### Femoral

| K162026 | EVOLUTION® Revision Stem Extensions                       |
|---------|-----------------------------------------------------------|
| K162026 | EVOLUTION® Revision Straight Stem Extension Adapters      |
| K142550 | EVOLUTION® Revision Distal and Posterior Femoral Augments |

#### Patella

| K122218 | ADVANCE® Patella |
|---------|------------------|
| K953439 | ADVANCE® Patella |

#### Tibial

| K162026 | EVOLUTION® Revision Tibial Base                      |
|---------|------------------------------------------------------|
| K162026 | EVOLUTION® Revision Tibial Augments                  |
| K162026 | EVOLUTION® Revision Modular Keel                     |
| K162026 | EVOLUTION® Revision Tibial Stem Offset Adapters      |
| K162026 | EVOLUTION® Revision Straight Stem Extension Adapters |
| K162026 | EVOLUTION® Revision Stem Extensions                  |

The EVOLUTION® Revision CCK System instrumentation includes reamers, trials, guides, broaches, bushings, spacers, drivers, and clamps.

#### G. Conclusion

The design features, materials information, predicate testing and analysis data provided in this premarket notification adequately support the substantial equivalence of EVOLUTION® Revision CCK.

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/JWH/K171389](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/JWH/K171389)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
