← Product Code [JWH](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/JWH) · K153595

# Optetrak Logic Metaphyseal Cones (K153595)

_Exactech, Inc. · JWH · Mar 25, 2016 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/JWH/K153595

## Device Facts

- **Applicant:** Exactech, Inc.
- **Product Code:** [JWH](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/JWH.md)
- **Decision Date:** Mar 25, 2016
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3560
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

The Optetrak Logic Metaphyseal Cones are indicated for use in skeletally mature individuals undergoing primary surgery for total knee replacement due to osteoarthritis, osteonecrosis, rheumatoid arthritis and/or post-traumatic degenerative problems. They are also indicated for revision of failed previous reconstructions where sufficient bone stock and soft tissue integrity are present. The augments are intended for press-fit fixation with the proximal tibial or distal femur. The final implant construct is completed by cementing a tibial tray or femoral component in place.

## Device Story

Optetrak Logic Metaphyseal Cones are orthopedic implants designed to fill tibial and femoral bone voids during complex primary or revision total knee arthroplasty (TKA). The device addresses severe bone loss (AORI type II and III defects) by providing stability to total knee implants via the metaphysis. The cones are intended for press-fit fixation within the proximal tibia or distal femur, serving as a foundation for the final implant construct, which is completed by cementing a tibial tray or femoral component. The device is used in a surgical setting by orthopedic surgeons. It benefits patients by enhancing implant fixation in cases of compromised bone stock, potentially improving the longevity and stability of the knee reconstruction.

## Clinical Evidence

No clinical data. Evidence consists of non-clinical bench testing, including fatigue testing, mating analysis, cadaveric implantation and surgeon assessment, biocompatibility assessment, compressive plastic deformation testing, abrasion resistance, static tensile strength testing, shear resistance, and porous structure characterization.

## Technological Characteristics

Metaphyseal cones designed for press-fit fixation in TKA. Materials and porous structure are characterized for bone void filling and stability. The device is a passive orthopedic implant; no energy source, software, or connectivity. Sterilization method not specified.

## Regulatory Identification

A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).

## Predicate Devices

- Triathlon® Tritanium® Cone Augments ([K143393](/device/K143393.md))

## Reference Devices

- Novation® Crown Cup® With InteGrip™ ([K102975](/device/K102975.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized representation of three human profiles facing right, stacked on top of each other. The profiles are rendered in a dark color, creating a sense of depth and unity. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the profiles.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

March 25, 2016

Exactech Incorporated Mr. Patrick Hughes Senior Regulatory Affairs Specialist 2320 N.W. 66th Court Gainesville, Florida 32653

Re: K153595

Trade/Device Name: Optetrak Logic Metaphyseal Cones Regulation Number: 21 CFR 888.3560 Regulation Name: Knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis Regulatory Class: Class II Product Code: JWH Dated: February 24, 2016 Received: February 26, 2016

Dear Mr. Hughes:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{1}------------------------------------------------

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# Mark N. Melkerson -S

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

| article of the program and the services of the series of the series of the series of the series of the series of the series of the series of the series and                                                                                                                                                                                                                                                                                                                                                 |                                                                                                                                                                                                     |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 2301V932 ИАМИН ОНА НТТАЭН ТО ТИЭМТЯАРЭQ<br>Food and Drug Administration                                                                                                                                                                                                                                                                                                                                                                                                                                     | Form Approved: OM 8MO. 09/04/01/20<br>xpiration Date: January 31, 2017                                                                                                                              |
| ndications for Use                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | woled thememal below.                                                                                                                                                                               |
| 500(k) Number (if known                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |                                                                                                                                                                                                     |
| All Sales of Artistics of Article Station of Article Station of Annon Comparis of Annone September 1978 September 1998 Andress of Annon Comparis Comparis Comparis Comparis Co                                                                                                                                                                                                                                                                                                                              |                                                                                                                                                                                                     |
| Optetrak Logic Metaphyseal Conne<br>Device Name                                                                                                                                                                                                                                                                                                                                                                                                                                                             |                                                                                                                                                                                                     |
| for for and the stock and soft titles integrity are prossess.<br>ອາອຸໄພ ຂໍ້ເອີ້ນປີ 2003 ເມືອງ ບໍລິເຣີ ໃດ ແດ່ຮັບອາ ເດີນ ນອກສະວັງ ເທີນ ເຂົ້າຮອບອົງ ອະໄປສາຫາລາວລາວ ວມັກສາມາສາ) ຂອງ<br>ə bir siyahısı və bir və qalında çəkində olan məməsi qər əsaxlı istifadə və yemirdə yarında güngündən<br>ələrinin ənələri vahidində aid əsr növləri əra sənəl ilə sənələ isə və istemində isə 1 ilə 1970-cü ildə 1990-cü ildə 1990-cü ildə 1990-cü ildə 1990-cü ildə 1990-cü ildə 1990<br>Indications for Use (Describe) |                                                                                                                                                                                                     |
| າກໄຊຫນ່ ໄສເຖີ ອປີ  .ເນກາອີ ໄຣໄປນ້ 10 ໄຄ່ເປັນ ແລະມີຄະນີ ອອກເອຫລາ 10 ກິໂລະຣາທຸ 10 ກິໂລະອາທຸ 10 ປີ ອອກສອນນາ ອາໄປ                                                                                                                                                                                                                                                                                                                                                                                               |                                                                                                                                                                                                     |
| .50 pm ການປະເທດໄທ ປີ 70 ບຸຮງການສາຫະກຳ ປີ ປະເທດຈາກອນກາງການ ຄົນ ນາງງາມການ ກາ                                                                                                                                                                                                                                                                                                                                                                                                                                  |                                                                                                                                                                                                     |
|                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |                                                                                                                                                                                                     |
| Type of Use (Select one or both, as applicable<br>(0 heading too sail (2 the 9) est noulding seen (2                                                                                                                                                                                                                                                                                                                                                                                                        | (O heddus 108 972 12) ອະປາຈາກການປະຈຸບັນດາວລາວ (O                                                                                                                                                    |
| odation in a sola at ara ana and sumithoo                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |                                                                                                                                                                                                     |
| a would and and and and and and and of the on and and the comment of the only<br>See to to to to in the sees and to stremerings of the selliges notices single                                                                                                                                                                                                                                                                                                                                              |                                                                                                                                                                                                     |
| ເວລາຊອບ ການ ນອນອກ ຄ.ສ. ອາໄປ ຄ.ສ. 1972 ປີ ຄ.ສ. 1978 ອາໄສຣ໌ ຄ.ສ. 1978 ອົງສາຕາ ແລະ ອຳນາ ແລະ ອຳນາ ແລະ ອາຫານ<br>:o) , newsdad sint publishing suggestion , nottomoris side to<br>əd birinci ilə sənəqsən vəq sunda olan başların səsi nəlibəli ilə qalıb və sinə nə bir nə əmil rayonunun ən                                                                                                                                                                                                                     |                                                                                                                                                                                                     |
| D<br>Food and Dung Administration<br>Star<br>Office of Chief Information Officer<br>vob.syy epi@gggg.chgg                                                                                                                                                                                                                                                                                                                                                                                                   |                                                                                                                                                                                                     |
| to notisellos s, ot broqzer of beniupen ton zi noznoda 10 toubuco Jon Vem vonege nA<br>a production BMO bilev viluenno e sveldsib it sselun norfermostni                                                                                                                                                                                                                                                                                                                                                    |                                                                                                                                                                                                     |
| (ANB) I 888 AQ MAOT<br>1 to t appy                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | นุมพลุน วระ<br>وال الحصرينجية ( ( ( 1 ) 1 4 3 - 6 - 1 1 ) 1 4 3 - 6 - 1 1 - 1 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1<br>Fili |

{3}------------------------------------------------

#### K153595 - Page 1 of 2 Exactech® Optetrak Logic® Metaphyseal Cones Traditional 510(k) - 510(k) Summary of Safety and Effectiveness

| Sponsor: | Exactech, Inc.<br>2320 N.W. 66th Court<br>Gainesville, FL 32653 |
|----------|-----------------------------------------------------------------|
|          | Phone: (352) 327-4762                                           |
|          | Fax: (352) 378-2617                                             |
|          | FDA Establishment Number 1038671                                |
| Contact: | Patrick Hughes<br>Senior Regulatory Affairs Specialist          |
| Date:    | March 25, 2016                                                  |

Trade of Proprietary or Model Name(s):

Exactech® Optetrak Logic® Metaphyseal Cones

## Common Name:

Knee System Augments

## Classification Name:

Prosthesis, knee, patellofemorotibial, semi-constrained, cemented, polymer/metal/polymer (CFR 888. 888.3560; Knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis, Class II, Product Code JWH)

## Information on devices to which substantial equivalence is claimed:

Acetabular Shell, Cluster-Hole

| 510(k) Number      | Trade or Proprietary Model Name     | Manufacturer   |
|--------------------|-------------------------------------|----------------|
| K143393            | Triathlon® Tritanium® Cone Augments | Stryker        |
| Reference devices: |                                     |                |
| 510(k) Number      | Trade or Proprietary Model Name     | Manufacturer   |
| K102975            | Novation® Crown Cup® With InteGrip™ | Exactech, Inc. |

{4}------------------------------------------------

#### K153595 - Page 2 of 2 Exactech® Optetrak Logic® Metaphyseal Cones Traditional 510(k) - 510(k) Summary of Safety and Effectiveness

### Indications for Use:

The Optetrak Logic Metaphyseal Cones are indicated for use in skeletally mature individuals undergoing primary surgery for total knee replacement due to osteoarthritis, osteonecrosis, rheumatoid arthritis and/or post-traumatic degenerative problems. They are also indicated for revision of failed previous reconstructions where sufficient bone stock and soft tissue integrity are present.

The augments are intended for press-fit fixation with the proximal tibial or distal femur. The final implant construct is completed by cementing a tibial tray or femoral component in place.

## Device Description:

Optetrak Logic Metaphyseal Cones are designed to fill tibial and femoral bone voids during revision or complex primary surgery in the case of severe bone loss. Specifically, Optetrak Logic Metaphyseal Cones are designed to enhance fixation in primary or revision total knee arthroplasty (TKA) patients who have Anderson Orthopaedic Research Institute (AORI) type II and III tibial and femoral defects by providing stability to total knee implants via the metaphysis.

## Testing:

This submission includes or references the following non-clinical testing:

- Fatigue testing ●
- Mating analysis
- Cadaveric implantation and surgeon assessment ●
- Biocompatibility assessment
- Compressive plastic deformation testing
- Abrasion resistance ●
- Static tensile strength testing ●
- Shear resistance .
- Porous structure characterization

## Substantial Equivalence Conclusion:

A comparison of key features and attributes included in this submission demonstrates the proposed Optetrak Logic Metaphyseal Cones devices are substantially equivalent to predicate Triathlon Tritanium Cone Augments, legally marketed per 510(k) submission K143393.

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/JWH/K153595](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/JWH/K153595)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
