COLUMBUS TOTAL KNEE SYSTEM AS COLUMBUS REVISION KNEE SYSTEM ENDURO KNEE SYSTEM
Applicant
Aesculap Implant Systems, LLC
Product Code
JWH · Orthopedic
Decision Date
Dec 17, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3560
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Columbus Total Knee System is indicated for use in reconstruction of the diseased knee joint caused by osteoarthritis, rheumatoid arthritis, post-traumatic arthritis, the need to revise failed arthroplasties or osteotomies where pain, deformity or dysfunction persist, and for patients suffering from correctable valgus or varus deformity and moderate flexion contracture. The Columbus Knee is designed for use with bone cement. The EnduRo Knee System is indicated for use in reconstruction of the diseased knee joint caused by osteoarthritis, rheumatoid arthritis, post-traumatic arthritis, the need to revise failed arthroplasties or osteotomies where pain, deformity or dysfunction persist, and for patients suffering from correctable valgus or varus deformity and moderate flexion contracture. Hinge knee systems are designed for use in patients in primary or revision surgery, where the posterior cruciate ligament and one or both of the collateral ligaments are absent or insufficient. The femoral and tibial augments are to be attached to their respective components with a fixation screw or screws. The EnduRo Knee System is intended for cemented use only.
Device Story
Columbus Total Knee System AS and EnduRo Knee System are cemented orthopedic prostheses for knee reconstruction. Columbus AS includes CR and PS femoral/tibial components, tibial hemi-spacers, and ZrN-coated surfaces. EnduRo is a rotating hinge design featuring CoCrMo femoral/tibial components, UHMWPE inserts/patellae, and PEEK Optima tibial masks/bushings. Devices are used by orthopedic surgeons in clinical settings for primary or revision arthroplasty. Components are implanted to restore joint function, alleviate pain, and correct deformities. Columbus AS provides dimensional modifications (narrow femurs) and size extensions. EnduRo utilizes a locking nut design for screw-connection stability. Both systems are single-use and sterile.
Clinical Evidence
Bench testing only. Design verification performed per FDA guidance for orthopedic devices. Mechanical testing for EnduRo evaluated screw connection loosening resistance over 5 million cycles, demonstrating substantial equivalence to the predicate.
Technological Characteristics
Columbus AS: CoCrMo with ZrN coating, UHMWPE inserts. EnduRo: CoCrMo femoral/tibial components, UHMWPE inserts/patellae, PEEK Optima (LT1/LT1CA30) tibial masks/bushings. Both are cemented, semi-constrained (Columbus) or rotating hinge (EnduRo) prostheses. Sterile, single-use.
Indications for Use
Indicated for patients requiring knee joint reconstruction due to osteoarthritis, rheumatoid arthritis, post-traumatic arthritis, or revision of failed arthroplasties/osteotomies with persistent pain, deformity, or dysfunction. Includes patients with correctable valgus/varus deformity and moderate flexion contracture. EnduRo specifically indicated for primary or revision surgery where posterior cruciate and/or collateral ligaments are absent/insufficient.
Regulatory Classification
Identification
A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).
{1}------------------------------------------------
K120955
510(k) Premarket Notification
Columbus Total Knee System AS
Page 1 of 2
DEC 1 7 2002 |
# 510(k) SUMMARY (as required by 21 CFR 807.92)
# Columbus Total Knee System AS December 17, 2012
COMPANY:
Aesculap® Implant Systems, Inc. 3773 Corporate Parkway Center Valley. PA 18034 Establishment Registration Number: 3005673311
CONTACT:
Julie Tom Wing 800-234-9179 (phone) 610-791-6882 (fax) julie tomwing@aesculap.com (email)
| TRADE NAME: | Columbus Total Knee System AS |
|----------------------|-------------------------------------------------------------------------------------------|
| COMMON NAME: | Total Knee System |
| CLASSIFICATION NAME: | Knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis |
| REGULATION NUMBER: | 888.3560 |
#### PRODUCT CODE: JWH
## SUBSTANTIAL EQUIVALENCE
The Columbus Total Knee System Alternate Surface (AS) is substantially equivalent to Aesculap's Columbus Total Knee System (K071220) and the previously cleared Columbus Total Knee System (CRA/PSA) (K053390).
#### DEVICE DESCRIPTION
The Columbus Total Knee System AS includes both "Cruciate Retaining" (CR) and "Posterior Stabilized" (PS) variants of the femoral and tibial components for cemented use with the previously cleared UHMWPE Columbus tibial inserts and patellae. The tibial hemi-spacers (wedges) are the same design as the previously cleared Columbus Total Knee System (CRA/PSA) (K053390). All components manufactured from CoCrMo are coated with a ZrN (Zirconium nitride) coating which has been cleared in Aesculap Columbus AS Knee submission (K071220).
{2}------------------------------------------------
Page 2 of 2
## INDICATIONS FOR USE
The Columbus Total Knee System is indicated for use in reconstruction of the diseased knee joint caused by osteoarthritis, rheumatoid arthritis, post-traumatic arthritis, the need to revise failed arthroplasties or osteotomies where pain, deformity or dysfunction persist, and for patients suffering from correctable valous or varus deformity and moderate flexion contracture.
The Columbus Knee is designed for use with bone cement.
#### TECHNOLIGICAL CHARACTERISTICS (compared to Predicate(s))
The base materials used for the new Aseculap implants are the same as that used to manufacture the predicate Aesculap devices. The only differences are a line extension introducing narrow CR and PS femurs and an introduction of AS coated tibial hemispacers including a smaller size (0/0+) hemi-spacer.
The Columbus narrow femurs are a dimensional modification that is available in right and left configurations in sizes that fall within the size range of femoral components previously cleared in Columbus AS Total Knee System (AS) (K071220).
The AS coated tibial hemi-spacers are a design change to the previously approved hemispacers cleared in K053390. The smaller size (0/0+) hemispacer is designed to mate with the same standard size or respective plus-size trays (i.e. TO wedge/T0+ tibial tray). The AS coating was cleared previously in K071220.
#### PERFORMANCE DATA
All required testing per "Draft Guidance for the Preparation of Premarket Notifications (510(k)s) Applications for Orthopedic Devices-The Basic Elements" were done where applicable.
Design verification of the Columbus Total Knee System AS was performed as a result of the risk assessment which were determined to be substantially equivalent to the predicates.
{3}------------------------------------------------
510(k) Premarket Notification
EnduRo Knee System
Page 1 of 2
# 510(k) SUMMARY (as required by 21 CFR 807.92)
EnduRo Knee System December 17, 2012
COMPANY:
Aesculap® Implant Systems, Inc. 3773 Corporate Parkway Center Valley, PA 18034 Establishment Registration Number: 3005673311
CONTACT:
Julie Tom Wing 800-234-9179 (phone) 610-791-6882 (fax) julie tomwing@aesculap.com (email)
TRADE NAME: EnduRo Knee System
COMMON NAME: Total Knee System
CLASSIFICATION NAME: Prosthesis. Knee, Femorotibial, Constrained, Cemented, Metal/Polymer
REGULATION NUMBER: 888.3510
PRODUCT CODE: KRO
#### SUBSTANTIAL EQUIVALENCE
The EnduRo Knee System is substantially equivalent to previously cleared, Aesculap Implant Systems EnduRo Knee System (K101815).
#### DEVICE DESCRIPTION
The EnduRo Knee System is a cemented prosthesis with a rotating hinge design. The femoral component, tibial plateau and extension stems are manufactured from CoCrMo. The tibial "gliding surfaces" (insert) and patella are manufactured from UHMWPE. The tibial mask is made from PEEK Optima® (LT1). The axial sleeve and femoral bushing components are produced from PEEK Optima® (LT1CA30). The system is made up of numerous components available in various sizes. The femoral, tibial plateau, and extension stems are also available with a ZrN (Zirconium nitride) coating. All components are sterile and for single use only.
{4}------------------------------------------------
Page 2 of 2
#### INDICATIONS FOR USE
The EnduRo Knee System is indicated for use in reconstruction of the diseased knee ioint caused by osteoarthritis, rheumatoid arthritis, post-traumatic arthritis, the need to revise failed arthroplasties or osteotomies where pain, deformity or dysfunction persist. and for patients suffering from correctable valgus or varus deformity and moderate flexion contracture.
Hinge knee systems are designed for use in patients in primary or revision surgery, where the posterior cruciate ligament and one or both of the collateral ligaments are absent or insufficient. The femoral and tibial augments are to be attached to their respective components with a fixation screw or screws.
The EnduRo Knee System is intended for cemented use only.
#### TECHNOLIGICAL CHARACTERISTICS (compared to Predicate(s))
The base materials used for the new Aseculap implants are the same as that used to manufacture the predicate Aesculap devices. The only difference is a design modification to the locking nut.
#### PERFORMANCE DATA
All required testing per "Draft Guidance for the Preparation of Premarket Notifications (510(k)s) Applications for Orthopedic Devices-The Basic Elements" were done where " applicable.
Mechanical testing to evaluate the resistance against loosening of the screw connection of the tibia component and the tibia locking ring of Aesculap EnduRo Knee System was performed for 5 million cycles. Results were found to be substantially equivalent to the predicate.
{5}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol that resembles three overlapping human figures or abstract shapes, possibly representing health and human services.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Letter dated: December 17, 2012
Aesculap Implant System, LLC % Ms. Julie Tom Wing Regulatory Affairs Specialist 3773 Corporate Parkway Center Valley, Pennsylvania 18034
Re: K120955
Trade/Device Name: COLUMBUS Total Knee System AS and EnduRo Knee System Regulation Number: 21 CFR 888.3560
Regulation Name: Knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis
Regulatory Class: II Product Codes: JWH, KRO Dated: December 13, 2012 Received: December 14, 2012
Dear Ms. Tom Wing:
We have reviewed your Section 510(k) premarket notification of intent to market the devices referenced above and have determined the devices are substantially equivalent (for the indications for use stated in the enclosures) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the devices, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your devices are classified (see above) into either class II (Special Controls) or class III (PMA), they may be subject to additional controls. Existing major regulations affecting your devices can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your devices in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your devices comply with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{6}------------------------------------------------
#### Page 2 -- Ms. Julie Tom Wing
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Mark N. Melkerson
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
Columbus Total Knee System AS
Page 1 of 1
#### INDICATIONS FOR USE STATEMENT A.
510(k) Number:_
Device Name: Columbus Total Knee System AS
#### Indications for Use:
The Columbus Total Knee System is indicated for use in reconstruction of the diseased knee joint caused by osteoarthritis, rheumatoid arthritis, post-traumatic arthritis, the need to revise failed arthroplasties or osteotomies where pain, deformity or dysfunction persist, and for patients suffering from correctable valgus or varus deformity and moderate flexion contracture.
The Columbus Knee is designed for use with bone cement.
× and/or Over-the-Counter Use Prescription Use (per 21 CFR 801.109)
(PLEASE DO NOT WRITE BELOW THIS LINEACONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Surgical, Orthopedic,
and Restorative Devices
K120955
510(k) Number .
{8}------------------------------------------------
Page 1 of 1
# A.
510(k) Number:_
Device Name: EnduRo Knee System
The EnduRo Knee System is indicated for use in reconstruction of the diseased knee joint caused by osteoarthritis, rheumatoid arthritis, post-traumatic arthritis, the need to revise failed arthroplasties or osteotomies where pain, deformity or dysfunction persist, and for patients suffering from correctable valgus or varus deformity and moderate flexion contracture.
Hinge knee systems are designed for use in patients in primary or revision surgery, where the posterior cruciate ligament and one or both of the collateral ligaments are absent or insufficient. The femoral and tibial augments are to be attached to their respective components with a fixation screw or screws.
The EnduRo Knee System is intended for cemented use only.
| Prescription Use | X |
|-----------------------------|---|
| and/or Over-the-Counter Use | |
| (per 21 CFR 801.109) | |
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH; Office of Device Evaluation (ODE)
and Restorative Devices
| 510(k) Number | K120955 |
|---------------|---------|
|---------------|---------|
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.