The EVOLUTION™ MP Total Knee System is indicated for use in knee arthroplasty in skeletally mature patients with the following conditions: 1) noninflammatory degenerative joint disease including osteoarthritis, traumatic arthritis, or avascular necrosis; inflammatory degenerative joint disease including rheumatoid arthritis; 2) correction of functional deformity; 3) 4) revision procedures where other treatments or devices have failed; and treatment of fractures that are unmanageable using other techniques. The EVOLUTION™ Total Knee System is for cemented use only.
Device Story
EVOLUTION™ MP Total Knee System is a semi-constrained cemented knee prosthesis; intended for knee arthroplasty. Modification increases functional range of motion; no new components introduced. Device consists of existing metal/polymer implants. Evaluated via mechanical testing and engineering analyses, including static stability and contact area testing. Used by orthopedic surgeons in clinical settings to replace damaged knee joints; restores joint function; corrects deformities; manages fractures. Benefits patients through improved mobility and pain relief.
Clinical Evidence
Bench testing only. No clinical data provided. Performance evaluated via mechanical testing and engineering analyses, including static stability and contact area testing, to confirm equivalence to predicate devices.
Technological Characteristics
Semi-constrained cemented knee prosthesis. Materials: Metal/Polymer. Components: Patellofemorotibial. Regulatory Class II. Device is a mechanical implant system; no software or electronic components.
Indications for Use
Indicated for skeletally mature patients undergoing knee arthroplasty for noninflammatory degenerative joint disease (osteoarthritis, traumatic arthritis, avascular necrosis), inflammatory degenerative joint disease (rheumatoid arthritis), functional deformity correction, revision procedures, or unmanageable fractures. Cemented use only.
Regulatory Classification
Identification
A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).
K020114 — 3D KNEE · Encore Orthopedics, Inc. · Jul 12, 2002
K182251 — EVOLUTION NitrX Medial-Pivot Knee · Mircoport Orthopedics, Inc. · May 16, 2019
K023546 — MAXIM ACCEL KNEE SYSTEM · Biomet, Inc. · Jan 17, 2003
Submission Summary (Full Text)
{0}------------------------------------------------
K102380
JAN 7 2011
### 510(k) Summary of Safety and Effectiveness
In accordance with the Food and Drug Administration Rule to implement provisions of the Safe Medical Devices Act of 1990 and in conformance with·21 CFR 807, this information serves as a Summary of Safety and Effectiveness for the use of the EVOLUTION™ Total Knee System.
| Submitted By: | Wright Medical Technology, Inc.<br>5677 Airline Rd, Arlington TN, 38002<br>(800) 238-7188 |
|--------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Date: | August 16, 2010 |
| Contact Person: | Ryan Ross<br>Senior Regulatory Affairs Specialist |
| Proprietary Name: | EVOLUTION™ MP Total Knee System |
| Common Name: | Total Knee System |
| Classification Name and Reference: | 21 CFR 888.3560 Knee joint Patellofemorotibial<br>Polymer/Metal/Polymer Semi-Constrained<br>Cemented Prosthesis Class II |
| Subject Product Code and Panel Code: | 21 CFR 888.3530 Knee joint Femorotibial<br>Metal/Polymer Semi-Constrained Cemented<br>Prosthesis Class II<br>Orthopedics/87/ JWH, HRY |
| Predicate Devices: | EVOLUTION™ MP Total Knee System, K093552<br>ADVANCE® Total Knee System, K063731,<br>K033890, K973524, K972626, K960617<br>AXIOM® Total Knee System, K894334 |
#### DEVICE INFORMATION
A. Intended Use
The EVOLUTION™ MP Total Knee System is indicated for use in knee arthroplasty in skeletally mature patients with the following conditions:
- 1) noninflammatory degenerative joint disease including osteoarthritis, traumatic arthritis, or avascular necrosis;
- inflammatory degenerative joint disease including rheumatoid arthritis; 2)
- correction of functional deformity; 3)
- 4) revision procedures where other treatments or devices have failed; and treatment of fractures that are unmanageable using other techniques.
The EVOLUTION™ Total Knee System is for cemented use only.
Page 1 of 2
{1}------------------------------------------------
#### B. Device Description
The subject of this Special 510(k)-Premarket Notification is to request a performance specification modification to increase the functional range of motion for the Wright Medical Technology, Inc. EVOLUTION™ MP Total Knee System. The EVOLUTION™ MP Total Knee System consists of existing total knee implant devices. No new total knee components are being introduced as a result of this Special 510(k) premarket notification.
The EVOLUTION™ MP Total Knee System was evaluated via mechanical testing and engineering analyses; including static stability and contact area testing. A review of these results indicates that the EVOLUTION™ MP Total Knee System is equivalent to predicate devices and are capable of withstanding expected in vivo loading without failure.
## C. Substantial Equivalence Information
The indications for use of the EVOLUTION™ MP Total Knee System are identical to the previously cleared predicate devices. The design features and materials of the subject devices are substantially equivalent to those of the predicate devices. The fundamental scientific technology of the modified devices has not changed relative to the predicate devices. The safety and effectiveness of the EVOLUTION™ MP Total Knee System are adequately supported by the substantial equivalence information, materials information, and analysis data provided within this Premarket Notification.
Page 2 of 2
{2}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services USA. The seal is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. In the center of the seal is an abstract image of an eagle. The eagle is facing to the right and has three lines representing its wings.
7 2011
JAN
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Wright Medical Technology, Inc. % Mr. Ryan Ross Senior Regulatory Affairs Specialist 5677 Airline Road Arlington Tennessee, 38002
Re: K102380
Trade/Device Name: EVOLUTION™ MP Total Knee System Regulation Number: 21 CFR 888.3560 Regulation Name: Knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis Regulatory Class: Class II Product Code: JWH, HRY Dated: December 7, 2010 Received: December 8, 2010
Dear Mr. Ross:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{3}------------------------------------------------
Page 2 - Mr. Ryan Ross
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
Sincerely yours,
Mark A. Matheson
Mark N. Melkerson Director Division of Surgical, Orthopedic, and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known): 12 02380
Device Name: EVOLUTION™ MP Total Knee System
Indications For Use:
The EVOLUTION™ MP Total Knee System is indicated for use in knee arthroplasty in skeletally mature patients with the following conditions:
- 1) noninflammatory degenerative joint disease including osteoarthritis, traumatic arthritis, or avascular necrosis;
- 2) inflammatory degenerative joint disease including rheumatoid arthritis;
- 3) correction of functional deformity;
- 4) revision procedures where other treatments or devices have failed; and treatment of fractures that are unmanageable using other techniques.
The EVOLUTION™ Total Knee System is for cemented use only.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use _ (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
for n. Mellekon
Page 1 of 1
Orthopedic, Tot Sure
Storative Devices
510(k) Number
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.