K091014 · Smith & Nephew, Inc. · JWH · Sep 29, 2009 · Orthopedic
Device Facts
Record ID
K091014
Device Name
JOURNEY BCS KNEE SYSTEM
Applicant
Smith & Nephew, Inc.
Product Code
JWH · Orthopedic
Decision Date
Sep 29, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3560
Device Class
Class 2
Attributes
Therapeutic
Intended Use
Total knee components are indicated for rheumatoid arthritis; post-traumatic arthritis, osteoarthritis, or degenerative arthritis in older patients whose age, weight, and activity level are compatible with an adequate long-term result; failed osteotomies, unicompartmental replacement, or total knee replacement. Posterior stabilized knee systems are designed for use in patients in primary and revision surgery, where the anterior and posterior cruciate ligaments are absent or incompetent and the collateral ligaments remain intact. Smith & Nephew, Inc. Journey BCS Knee components are indicated for use only with cement and are single use devices.
Device Story
The Journey BCS Knee System is a total knee prosthesis designed for primary and revision knee arthroplasty. This submission adds kinematics claims to the previously cleared system (K042515). The device is a semi-constrained, cemented, single-use implant. It is intended for use in patients with absent or incompetent cruciate ligaments but intact collateral ligaments. The system is implanted by orthopedic surgeons in a clinical/surgical setting. The device functions as a mechanical replacement for the knee joint to restore function and mobility. No new components were introduced; the submission focuses on characterizing the kinematic performance of the existing implant design.
Clinical Evidence
No clinical data provided. The submission is a 510(k) notification for additional kinematics claims for a previously cleared device.
Technological Characteristics
Semi-constrained, cemented, metal/polymer total knee prosthesis. Designed for primary and revision surgery. Single-use. No changes to materials or design from the predicate (K042515).
Indications for Use
Indicated for patients with rheumatoid, post-traumatic, or osteoarthritis, or degenerative arthritis requiring total knee replacement, including failed prior surgeries. Suitable for primary and revision surgery where anterior/posterior cruciate ligaments are absent/incompetent but collateral ligaments remain intact. For use in older patients with compatible age, weight, and activity levels.
Regulatory Classification
Identification
A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).
K060569 — BALANCED KNEE SYSTEM MODULAR REVISION KNEE SYSTEM · Ortho Development Corp. · Jul 6, 2006
K043440 — SMITH & NEPHEW, INC. REVISION KNEE SYSTEM · Smith & Nephew, Inc. · Feb 18, 2005
K110837 — REVISION PROTHESIS · Smith & Nephew, Inc. · Apr 22, 2011
K152726 — JOURNEY II XR Knee System · Smith & Nephew, Inc. · Oct 21, 2015
K023546 — MAXIM ACCEL KNEE SYSTEM · Biomet, Inc. · Jan 17, 2003
Submission Summary (Full Text)
{0}------------------------------------------------
510(k) Summary of Safety and Effectiveness Smith & Nephew JOURNEY BCS Knee System
# SEP 2 9 2009
K09 1014
Contact Person and Address
## Date of Summary: April 8, 2009
Jason Sells Manager, Regulatory Affairs Smith & Nephew, Inc. Orthopaedic Division 1450 E. Brooks Road Memphis, Tennessee 38116 (901) 399-5520
Name of Device: Smith & Nephew, Inc. Journey BCS Knee System Common Name: Total Knee Prosthesis
Device Classification Name and Reference: 21 CFR 888.3560 Knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis - Class II
Device Product Code: JWH
#### Device Description
This premarket notification seeks only to add kinematics claims for the existing Journey BCS Knee System cleared via K042515. No new total knee components being introduced as a result of this premarket notification.
#### Intended Use
Total knee components are indicated for rheumatoid arthritis; post-traumatic arthritis, osteoarthritis, or degenerative arthritis in older patients whose age, weight, and activity level are compatible with an adequate long-term result; failed osteotomies, unicompartmental replacement, or total knee replacement. Posterior stabilized knee systems are designed for use in patients in primary and revision surgery, where the anterior and posterior cruciate ligaments are absent or incompetent and the collateral ligaments remain intact. Smith & Nephew, Inc. Journey BCS Knee components are indicated for use only with cement and are single use devices.
# Substantial Equivalence Information
The total knee system subject of this 510(k) premarket notification has been previously cleared by FDA via K042515. This premarket notification seeks only to add kinematics claims for the existing devices. The kinematics claims being made about the device are similar to those being made for the Stryker Triathlon Total Knee System (K053514).
{1}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the circle is a stylized symbol that resembles a person embracing another person, or a stylized representation of a human figure.
SEP 2 9 2009
Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002
Smith & Nephew, Inc. % Mr. Jason Sells Manager, Regulatory Affairs 1450 East Brooks Road Memphis, Tennessee 38116
### Re: K091014
Trade Name: JOURNEY BCS Knee System Regulation Number: 21 CFR 888.3560
Regulation Name: Knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis
Regulatory Class: II Product Code: JWH Dated: August 12, 2009 Received: August 13, 2009
#### Dear Mr. Sells:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{2}------------------------------------------------
### Page 2- Mr. Jason Sells
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Mark A. Millican
Mark N. Melkerson Director Division of Surgical, Orthopedic, and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## 510(k) Number (if known):
Device Name: JOURNEY BCS Knee System
Indications for Use:
Total knee components are indicated for rheumatoid arthritis; post-traumatic arthritis, osteoarthritis, or degenerative arthritis in older patients whose age, weight, and activity level are compatible with an adequate long-term result; failed osteotomies, unicompartmental replacement, or total knee replacement. Posterior stabilized knee systems are designed for use in patients in primary and revision surgery, where the antenor and posterior cruciate ligaments are absent or incompetent and the collateral ligaments remain intact. Smith & Nephew, Inc. Journey BCS Knee components are indicated for use only with cement and are single use devices.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Mark Miller
(Division Sign-Oft) Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K09/014
Page 1 of
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.