COLUMBUS TOTAL KNEE SYSTEM AS

K071220 · Aesculap Implant Systems, Inc. · JWH · Jun 1, 2007 · Orthopedic

Device Facts

Record IDK071220
Device NameCOLUMBUS TOTAL KNEE SYSTEM AS
ApplicantAesculap Implant Systems, Inc.
Product CodeJWH · Orthopedic
Decision DateJun 1, 2007
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 888.3560
Device ClassClass 2
AttributesTherapeutic

Intended Use

The Columbus Total Knee System is indicated for use in reconstruction of the diseased knee joint caused by osteoarthritis, rheumatoid arthritis, post-traumatic arthritis, the need to revise failed arthroplasties or osteotomies where pain, deformity or dysfunction persist, and for patients suffering from correctable valgus or varus deformity and moderate flexion contracture. The Columbus Knee is designed for use with bone cement.

Device Story

Columbus Total Knee System AS is a semi-constrained cemented knee prosthesis. It consists of femoral and tibial components made of Cobalt Chrome alloy (CoCrMo) with an added Zirconium Nitrate (ZrN) PVD coating. The system is used in orthopedic surgery for knee joint reconstruction. The device is intended for use with bone cement and previously cleared UHMWPE tibial inserts. The addition of the ZrN coating is the only modification to the predicate Columbus Total Knee System. The device is implanted by a surgeon in a clinical/OR setting to restore joint function and alleviate pain in patients with degenerative or post-traumatic knee conditions.

Clinical Evidence

Bench testing only. Testing was conducted in accordance with the "Draft Guidance for the Preparation of Premarket Notifications for Orthopedic Devices-The Basic Elements."

Technological Characteristics

Semi-constrained cemented knee prosthesis. Materials: Cobalt Chrome alloy (CoCrMo) with Zirconium Nitrate (ZrN) PVD coating. Components: Femoral and tibial. Connectivity: None. Software: None. Sterilization: Not specified.

Indications for Use

Indicated for patients requiring knee joint reconstruction due to osteoarthritis, rheumatoid arthritis, post-traumatic arthritis, or revision of failed arthroplasties/osteotomies with persistent pain, deformity, or dysfunction. Includes patients with correctable valgus or varus deformity and moderate flexion contracture.

Regulatory Classification

Identification

A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Columbus Total Knee System AS # 4071220 (1/2) | Page 1 of 2<br>B. | 510(k) SUMMARY (as required by 21 CFR 807.92) | JUN - 1 2007 | |----------------------|---------------------------------------------------------------------------------------------------------------------------------------|--------------| | | Columbus Total Knee System AS<br>2 May 2007 | | | COMPANY: | Aesculap® Implant Systems, Inc.<br>3773 Corporate Parkway<br>Center Valley, PA 18034<br>Establishment Registration Number: 3005673311 | | | CONTACT: | Matthew M. Hull<br>800-258-1946 (phone)<br>610-791-6882 (fax)<br>matt.hull@aesculap.com (email) | | | TRADE NAME: | Columbus Total Knee System AS | | | COMMON NAME: | Total Knee System | | | CLASSIFICATION NAME: | Knee joint patellofemorotibial polymer/metal/polymer semi-<br>constrained cemented prosthesis | | | REGULATION NUMBER: | 888.3560 | | | PRODUCT CODE: | JWH | | ## SUBSTANTIAL EQUIVALENCE Aesculap®, Inc. believes that the Columbus Total Knee System Alternate Surface (AS) is substantially equivalent to Aesculap's Columbus Total Knee System PS (K030367), CR (K022672), Smith & Nephew's Genesis II Zirconium "Oxinium" knee (K962557), and Endotec's (B-P) New Jersey Total knee (K012702). ## DEVICE DESCRIPTION The Columbus Total Knee System AS includes both CR and PS variants of the femoral and tibial components for cemented use with the previously cleared UHMWPE Columbus tibial inserts. The Zirconium Nitrate ZrN coating is the only change to the previously cleared Columbus Total Knee System. The coated components are still made from Cobalt Chrome alloy (CoCrMo). ## INDICATIONS FOR USE The Columbus Total Knee System is indicated for use in reconstruction of the diseased knee joint caused by osteoarthritis, rheumatoid arthritis, post-traumatic arthritis, the need to revise failed arthroplasties or osteotomies where pain, deformity or dysfunction persist, and for patients suffering from correctable valgus or varus deformity and moderate flexion contracture. The Columbus Knee is designed for use with bone cement. {1}------------------------------------------------ # (22) ס22170) Page 2 of 2 ## TECHNOLIGICAL CHARACTERISTICS(compared to Predicate(s)) The base material used for the new Aseculap implants is the same as that used to r re base material adsculap devices. The only difference is the addition of a The nium Nitrate (ZrN) coating to the CoCrMo implants. The PVD coating method is Elroomian Nitrate (EPT) council devices as well including the other predicate implants. ## PERFORMANCE DATA All required testing per "Draft Guidance for the Preparation of Premarket Notifications All roquired (50.mg per or Orthopedic Devices-The Basic Elements" were done where applicable. {2}------------------------------------------------ Public Health Service Image /page/2/Picture/2 description: The image is a seal for the Department of Health & Human Services - USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract image of an eagle. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Aesculap Implant Systems, Inc. % Mr. Matthew M. Hull Regulatory Affairs Manager 3773 Corporate Parkway Center Valley, Pennsylvania 18034 JUN - 1 2007 Re: K071220 Trade/Device Name: Columbus Total Knee Systems AS Regulation Number: 21 CFR 888.3560 Regulation Name: Knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prothesis Regulatory Class: II Product Code: JWH Dated: May 2, 2007 Received: May 10, 2007 Dear Mr. Hull: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set {3}------------------------------------------------ Page 2 – Mr. Matthew Hull forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at toll-free number (800) 638-2041 or (240) 276-3150 or on the Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Mark A. Millman Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ Page 1 of 1 #### INDICATIONS FOR USE STATEMENT A. 510(k) Number: 14071220 ## Device Name: Columbus Total Knee System AS ### Indications for Use: The Columbus Total Knee System is indicated for use in reconstruction of the diseased knee joint caused by osteoarthritis, theumatoid arthritis, post-traumatic arthritis, the need to revise failed arthroplasties or osteotomies where pain, deformity or dysfunction persist, and for patients suffering from correctable valgus or varus deformity and moderate flexion contracture. The Columbus Knee is designed for use with bone cement. Mark N. Millheim (Division Sign-Off) Division of General, Restorative, and Neurological Devices 510(k) Number Prescription Use***_***_ Prescription Use (per 21 CFR 801.109) (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...