This device is indicated for patients with severe knee pain and disability due to: - Rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis. - Collagen disorders and/or avascular necrosis of the femoral condyle. - Post-traumatic loss of joint configuration, particularly when there is patellofemoral erosion, dysfunction or prior patellectomy. - Moderate valgus, varus or flexion deformities. - The salvage of previously failed surgical attempts or for a knee in which satisfactory stability in flexion cannot be obtained at the time of surgery. This device is indicated for use when the posterior cruciate ligament is absent or deficient. This device is intended for use as part of a cemented or uncemented knee prosthesis.
Device Story
NexGen Complete Knee Solution UC-Flex is a fixed-bearing articular surface component for total knee arthroplasty; part of a semiconstrained, nonlinked condylar knee system. Used in clinical settings by orthopedic surgeons; implanted as part of a cemented or uncemented knee prosthesis. Designed for patients with severe knee pain, disability, or ligamentous deficiency (posterior cruciate ligament absent/deficient). Device provides structural support and joint articulation to restore knee function and stability. Substantial equivalence established via comparison to predicate; no clinical data required.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Fixed-bearing articular surface component for total knee prosthesis. Semiconstrained, nonlinked design. Manufactured using same materials and processes as predicate device. Intended for cemented or uncemented use.
Indications for Use
Indicated for patients with severe knee pain/disability due to RA, OA, traumatic arthritis, polyarthritis, collagen disorders, avascular necrosis of femoral condyle, or post-traumatic joint loss (with patellofemoral erosion/dysfunction/prior patellectomy). Also for moderate valgus, varus, or flexion deformities, salvage of failed surgeries, or knees with unstable flexion. Indicated when posterior cruciate ligament is absent/deficient.
Regulatory Classification
Identification
A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).
K031061 — NEXGEN POROUS, UNCEMENTED FEMORAL AND TIBIAL BASEPLATE COMPONENTS · Zimmer, Inc. · Oct 9, 2003
K042271 — NEXGEN COMPLETE KNEE SOLUTION LPS-FLEX PROLONG HIGHLY CROSSLINKED POLYETHYLENE ARTICULAR SURFACES · Zimmer, Inc. · Oct 13, 2004
K023211 — NEXGEN COMPLETE KNEE SOLUTION CRUCIATE RETAINING (CR)-FLEX FEMORAL COMPONENTS · Zimmer, Inc. · Oct 17, 2002
K052879 — NEXGEN COMPLETE KNEE SOLUTION MIS MODULAR TIBIAL PLATES AND KEELS · Zimmer, Inc. · Nov 22, 2005
K062768 — NEXGEN COMPLETE KNEE SOLUTION LEGACY POSTERIOR STABILIZED FLEX FIXED BEARING AND TIVANIUM TI-6AL-4V ALLOY FEMORAL · Zimmer, Inc. · Oct 13, 2006
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the Zimmer logo. The logo consists of a stylized letter "Z" inside a circle, followed by the word "zimmer" in lowercase letters. The logo is in black and white.
# Summary of Safety and Effectiveness
| Submitter: | MAY 26 2006<br>Zimmer, Inc.<br>P.O. Box 708 |
|---------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Warsaw, IN 46581-0708 |
| Contact Person: | Brandon Hipsher<br>Associate, Corporate Regulatory Affairs<br>Telephone: (574) 371-8083<br>Fax: (574) 372-4605 |
| Date: | March 16, 2006 |
| Trade Name: | <i>NexGen</i> ® Complete Knee Solution Ultracongruent<br>(UC-Flex) Fixed Bearing Articular Surface<br>Component |
| Common Name: | Total Knee Prosthesis |
| Classification Name<br>and Reference: | Knee joint patellofemorotibial polymer/metal/<br>polymer semi-constrained cemented prosthesis<br>21 CFR § 888.3560 |
| | Knee joint patellofemorotibial metal/polymer<br>porous-coated uncemented prosthesis<br>21 CFR § 888.3565 |
| Predicate Device: | <i>Prolong</i> ™ Highly Crosslinked Polyethylene<br>Cruciate Retaining (CR) Articular Surface<br>Component, manufactured by Zimmer, Inc.,<br>K013991, cleared December 27, 2001 |
| Device Description: | This device is part of the <i>NexGen</i> system of<br>semiconstrained, nonlinked condylar knee<br>prostheses. |
| Intended Use: | This device is indicated for patients with severe<br>knee pain and disability due to:<br>- Rheumatoid arthritis, osteoarthritis, traumatic<br>arthritis, polyarthritis.<br>- Collagen disorders and/or avascular necrosis of<br>the femoral condyle.<br>- Post-traumatic loss of joint configuration |
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the title "Traditional 510(k) Premarket Notification". The text is in bold font. The title is centered on the page.
particularly when there is patellofemoral erosion, dysfunction or prior patellectomy.
- Moderate valgus, varus or flexion deformities. ﮯ
- The salvage of previously failed surgical attempts or for a knee in which satisfactory stability in flexion cannot be obtained at the time of surgery.
This device is indicated for use when the posterior cruciate ligament is absent or deficient. This device is intended for use as part of a cemented or uncemented knee prosthesis.
Comparison to Predicate Device:
Performance Data (Nonclinical and/or Clinical):
The proposed device is manufactured, packaged and sterilized using the same materials and processes as the predicate device.
Non-Clinical Performance and Conclusions:
Non-clinical testing demonstrated that this device met performance requirements and is as safe and effective as the predicate device.
Clinical Performance and Conclusions:
Clinical data and conclusions were not needed for this device.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH AND HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles a stylized eagle or bird in flight, composed of three curved lines.
#### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
## MAY 2 6 2006
Zimmer, Inc. % Mr. Brandon Hipsher, RAC Associate, Corporate Regulatory Affairs P.O. Box 708 Warsaw, Indiana 46581-0708
Re: K060722
Trade/Device Name: NexGen® Complete Knee Solution Ultracongruent (UC-Flex) Fixed Bearing Articular Surface Component Regulation Number: 21 CFR 888.3560 Regulation Name: Knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis Regulatory Class: Class II Product Codes: JWH, MBH
Dated: March 16, 2006 Received: March 17, 2006
Dear Mr. Hipsher:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set
{3}------------------------------------------------
Page 2 - Mr. Brandon Hipsher
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Herbert Lerner us
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
# 510(k) Number (if known): KO607222
### Device Name:
NexGen® Complete Knee Solution Ultracongruent (UC-Flex) Fixed Bearing Articular Surface Component
### Indications for Use:
This device is indicated for patients with severe knee pain and disability due to:
- Rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis. l
- Collagen disorders and/or avascular necrosis of the femoral condyle. ।
- । Post-traumatic loss of joint configuration, particularly when there is patellofemoral erosion, dysfunction or prior patellectomy
- Moderate valgus, varus or flexion deformities. -
- -The salvage of previously failed surgical attempts or for a knee in which satisfactory stability in flexion cannot be obtained at the time of surgery.
This device is indicated for use when the posterior cruciate ligament is absent or deficient.
This device is intended for use as part of a cemented or uncemented knee prosthesis.
Prescription Use X (Part 21 CFR 801 Subpart D)
### AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(Please do not write below this line - Continue on another page if needed)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Helene Lehner
Division of General, Restorative, and Neurological Devices
Page 1 of 1
**510(k) Number** K060722
002
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.