K051640 · United Orthopedic Corp. · JWH · Mar 1, 2006 · Orthopedic
Device Facts
Record ID
K051640
Device Name
U2 TOTAL KNEE SYSTEM
Applicant
United Orthopedic Corp.
Product Code
JWH · Orthopedic
Decision Date
Mar 1, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3560
Device Class
Class 2
Attributes
Therapeutic
Intended Use
This device is indicated in knee arthroplasty for reduction or relief of pain and/or improved knee function in skeletally mature patients with severe knee pain and disability due to rheumatoid arthritis, osteoarthritis, primary and secondary traumatic arthritis, polyarthritis, collagen disorders, avascular necrosis of the femoral condyle or pseudogout, posttraumatic loss of joint configuration, particularly when there is patellofemoral erosion, dysfunction or prior patellectomy, moderate valgus, varus, or flexion deformities. This device may also be indicated in the salvage of previously failed surgical attempts if the knee can be satisfactorily balanced and stabilized at the time of surgery. This device system is designed for cemented use only.
Device Story
U2 Total Knee System is a semi-constrained, fixed-bearing, cemented knee prosthesis. Components include Co-Cr-Mo alloy femoral component, Ti-6Al-4V alloy tibial baseplate, and UHMWPE tibial insert and patellar component. Modular snap-fit locking mechanism secures insert to baseplate. Femoral component design mimics natural anatomy to restore rotation, extension, and flexion. Patellar component features dome shape for even stress distribution. Device implanted by orthopedic surgeons in clinical settings to replace damaged knee joints. Benefits include pain reduction, improved joint function, and restoration of joint configuration. System supports cruciate retaining and posterior stabilized configurations.
Clinical Evidence
Bench testing only. Testing included range of motion (135°), constraint testing, contact area analysis (femorotibial and patellofemoral), locking mechanism strength (anteroposterior motion at 100N), and cyclic fatigue testing of the tibial baseplate (5 million cycles at 200N-2.0KN). Porous coating testing included static tensile strength (>5,000 psi) and shear strength (>20 MPa). No clinical data provided.
Indicated for skeletally mature patients requiring knee arthroplasty for pain relief or improved function due to rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis, collagen disorders, avascular necrosis, pseudogout, or posttraumatic joint loss (including patellofemoral erosion/dysfunction or prior patellectomy) and moderate valgus, varus, or flexion deformities. Also indicated for salvage of failed surgical attempts if knee can be balanced/stabilized. Cemented use only.
Regulatory Classification
Identification
A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).
K082424 — UNITED U2 TOTAL KNEE SYSTEM · United Orthopedic Corp. · Dec 16, 2008
K021657 — UKNEE TOTAL KNEE SYSTEM · United Orthopedic Corp. · Aug 12, 2002
K180446 — b-ONE Total Knee System · B-One Ortho, Corp. · Oct 12, 2018
K221705 — U2 Total Knee System-PF+ · United Orthopedic Corporation · Feb 28, 2023
K183344 — The Progressive Orthopaedic Company Total Knee System II · The Progressive Orthopaedic Company, LLC · Mar 18, 2019
Submission Summary (Full Text)
{0}------------------------------------------------
2006 MAR I
U2 Total Knee System
Summary
# 510(k) Summary of Safety and Effectiveness
United UKNEE® Total Knee System ( K021657 )
| Company: | United Orthopedic Corporation |
|----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Address: | No 57, Park Ave. 2, Science Park, Hsinchu 300, Taiwan |
| Phone Number: | +886-3-5773351 |
| Fax Number: | +886-3-5777156 |
| Date Prepared: | May 12, 2005 |
| Device Name: | U2 Total Knee System |
| Common Name: | Semi-constrained total knee prostheses |
| Classification Name: | Knee joint patellofemorotibial polymer/metal/polymer<br>semi-constrained cemented prosthesis per 21CFR 888.3560<br>This falls under the Orthopedics panel/87 |
Predicate Device:
The data that United Orthopedic Corporation included in this submission show the U2 Total Knee System (Cruciate retaining type and posterior stabilized type) is safe and effective. The U2 Total Knee System comprises a femoral component, which articulates with a polyethylene insert The underside of the insert component is flat and is snapped into the metal tibial component. The modular (snap-fit) locking mechanism of the insert components has proven to be component. safe and effective in the clinical area. The design and sizing of the femoral component correspond to the natural femoral anatomy, enhancing stress distribution and contributing to restore original femoral dimensions and normal rotation, extension and flexion. Each size of femur has the same intercondylar distance and radius of curvature; this feature is replicated on the insert component, thus allowing any size of femur to be matched with any size of tibial component. The dome shape all UHMWPE patellar design provides excellent contact and even distribution of stresses, simplifies implantation by eliminating need for rotational orientation. Test data indicate the U2 Total Knee System performs as well or better than the predicate device in all areas tested. Materials used in the manufacture of the U2 Total Knee System meet the property requirements of the ASTM standards associated with each material.
United Orthopedic Corporation, Taiwan.
Page: 1 / III
{1}------------------------------------------------
#### (S) U2 Total Knee System
Summary
### Device Description:
U2 Total Knee system is the fixed bearing type with cruciate retaining and posterior stabilized It is a Patellofemorotibia, polymer/metal/polymer, semi-constrained, cemented knee design. prosthesis, which has a cobalt-chromium-molybdenum (Co-Cr-Mo) alloy femoral component and Ti alloy tibial component composed of a insert, machined from compressed molded The tibial baseplate UHMWPE flat, mechanical locked with metallic tibial baseplate. component is made of Ti-6A1-4V alloy which has groove for cement fixation. The cruciate retaining type of femoral component has w/ and w/o beaded porous coating design, and the posterior stabilized type has only w/o. The porous coating with two layers of beads has an about 55% of porosity, a 440 um of average pore size, and its thickness is 1.0 +0/+0.3 mm. The patella is machined from extruded UHMWPE bar. This device have 3 pegs and cement groove designed for cemented use only.
## Intended Use:
This device is indicated in knee arthroplasty for reduction or relief of pain and/or improved knee function in skeletally mature patients with severe knee pain and disability due to rheumatoid arthritis, osteoarthritis, primary and secondary traumatic arthritis, polyarthritis, collagen disorders, avascular necrosis of the femoral condyle or pseudogout, posttraumatic loss of joint configuration, particularly when there is patellofemoral erosion, dysfunction or prior patellectomy, moderate valgus, varus, or flexion deformities. This device may also be indicated in the salvage of previously failed surgical attempts if the knee can be satisfactorily balanced and stabilized at the time of surgery. This device system is designed for cemented use only.
### Basis for Substantial Equivalence:
Features comparable to predicate devices, United UKNEE® Total Knee System ( K021657 ) , include femoral components, patellar components, tibial trays and tibial inserts. The femoral components are available in cruciate retaining and posterior stabilized designs. Tibial inserts are available in a range of thickness and in cruciate retained and posterior stabilized designs. The patellar components are available in all plastic on-set designs with dome shape configurations.
United Orthopedic Corporation, Taiwan.
{2}------------------------------------------------
#### Summary
# Test Results:
The U2 Total Knee System was analyzed following draft guidance for ccmented. semi-constrained total knee prostheses. The range of motion for the U2 Total Kncc System is 135°, satisfying necessary needs of daily life and general activities which exclude placing the patient in condition of severe loading caused higher risk for failure of the knee replacement. such as obesity, heavy labor, active sports participation, high levels of patient activity. likelihood of falls, alcohol or drug addiction and other disabilities, as appropriate.
Constraint testing compares with the UKNEE® Total Knee System and showed the U2 Total Knee System to be slightly more constraint.
Contact area between the femorotibial joint and patellofemoral joint is conducted respectively, and compared with other two products, AMK® and UKNEE®. From the data, we can find U2 Total Knee System could provide similar or greater contact area than others.
Comparing the tibial insert and tibial tray locking with UKNEE®, the average anteroposterior motion at 100N load of U2 Total Knee System is less than the predicated device, which means the locking mechanism of U2 Total Knee System is stronger than UKNEE®.
Cyclic test is performed on U2 system to evaluate the fatigue properties of tibial baseplate. Compressive cyclic load between 200N and 2.0KN, 5 million cycles is applied to the test system. There is no fracture on the tibial baseplate.
The static tensile strength of the beaded porous coating was conformed to endure 5,000 psi min. The shear strength of the beaded porous coating was more than 20 MPa. The porous coating has volume porosity between 30 and 70 percent, an average pore size between 100 and 1000 microns and a porous coating thickness between 1,000 and 1,300 microns.
Based on previous test results, the U2 Total Knee System is substantial equivalent :o UKNEE® Total Knee System.
nited Orthopedic Corporation, Taiwan.
Page: III / III
යි. පිහිටි පිහිටි පිහිටි පිහිටි විසින් පිහිටි පිහිටි පිහිටි පිහිටි පිහිටි පිහිටි පිහිටි පිහිටි පිහිටි පිහිටි පිහිටි පිහිටි පිහිටි පිහිටි පිහිටි පිහිටි පිහිටි පිහිටි පිහිටි ප
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH AND HUMAN SERVICES, USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with three curved lines above them.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAR I 2006
Mr. Gene Huang Regulatory Affairs Coordinator United Orthopedic Corporation No. 57, Park Ave. 2, Science Park Hsinchu 300, Taiwan
Re: K051640
Trade/Device Name: U2 Total Knee System Regulation Number: 21 CFR 888.3560 Regulation Name: Knee joint patellofemorotibial polymer/metal/polymer semiconstrained cemented prosthesis Regulatory Class: II Product Codes: JWH Dated: February 24, 2006 Received: February 27, 2006
Dear Mr. Huang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{4}------------------------------------------------
Page 2 - Mr. Gene Huang
CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
Heilut Lemmer us
Image /page/4/Picture/5 description: The image shows a handwritten word that appears to be "for". The word is written in a cursive style, with the letters connected. The "f" has a loop extending above the main body of the letter, and the "o" and "r" are joined together in a flowing manner. The writing is simple and clear, making the word easily readable.
Mark N. Melkerson, M.S. Acting Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
| 510 (k) Number (if known): | K051640 |
|----------------------------|---------|
|----------------------------|---------|
Device Name: __U2 Total Knee System
# Indications for Use:
This device is indicated in knee arthroplasty for reduction or relief of pain and/or improved knee function in skeletally mature patients with severe knee pain and disability due to rheumatoid arthritis, osteoarthritis, primary and secondary traumatic arthritis, polyarthritis, collagen disorders, avascular necrosis of the femoral condyle or pseudogout, posttraumatic loss of joint configuration, particularly when there is patellofemoral erosion, dysfunction or prior patellectomy, moderate valgus, varus, or flexion deformities. This device may also be indicated in the salvage of previously failed surgical attempts if the knee can be satisfactorily balanced and stabilized at the time of surgery.
This device system is designed for cemented use only.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
1. 1. 1. 1. 1. 2. 2. 2. 2. 2. 2. 2. 2. 2. 2. 2. 2. 2. 2. 2. 2.
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF
NEEDED)
510(k) Number.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.