DISPOSABLE CEMENT SPACER MOLDS FOR TEMPORARY KNEE PROSTHESES
Applicant
Biomet, Inc.
Product Code
JWH · Orthopedic
Decision Date
Jul 19, 2005
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3560
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The intended use of the Biomet disposable cement spacer molds is to provide the surgeon with a means to mold a temporary knee prosthesis at the point of care that is substantially equivalent to the Exactech Spacer-k temporary knee prosthesis cleared in K032522. The temporary knee prosthesis made with the Biomet disposable cleared in R0325221. The same indication for use as the Exactech Spacer-k.
Device Story
Disposable silicone molds used at point-of-care by surgeons to create temporary knee prosthesis components. Process involves filling molds with PMMA/gentamicin bone cement via dispenser/gun or pouring; cement hardens within mold; temporary prosthesis removed and implanted into joint space during first stage of two-stage TKR procedure. Device remains in situ for up to 180 days until second-stage implantation of conventional prosthesis. Benefits include temporary joint stabilization and local antibiotic delivery for patients with septic TKR. Surgeon manages entire workflow; device is not implanted itself.
Clinical Evidence
No clinical testing was performed. Evidence is based on bench testing, including knee joint simulator studies comparing the Biomet-molded temporary prosthesis to the predicate device, demonstrating equivalent strength, wear, and gentamicin elution.
Technological Characteristics
Medical grade silicone molds; femoral and tibial configurations in various sizes. Principle of operation: manual casting of PMMA/gentamicin bone cement. Non-implanted, sterile, disposable. No electronic components or software.
Indications for Use
Indicated for skeletally mature patients undergoing a two-stage total knee replacement (TKR) process due to a septic process. The molded temporary knee prosthesis is indicated for an implantation period of 180 days or less. Contraindicated for patients with systemic clinical conditions precluding two-stage arthroplasty, inadequate bone stock/support, deficient peripheral ligamentous/extensor mechanism, muscular/nervous/vascular deficiencies, poor bone quality (osteoporosis), unconfirmed TKR infection, non-removable infected TKR, gentamicin-resistant pathogens, sensitivity to gentamicin/aminoglycosides/PMMA, systemic/remote infection, infection secondary to trauma/other surgery, neuromuscular disorders, or patient factors (age/weight/activity) predicting early system failure.
Regulatory Classification
Identification
A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).
{0}------------------------------------------------
JUL 1 9 2005
## 510(k) Summary
Applicant/Sponsor: Biomet Manufacturing Corp. P.O. Box 587 Warsaw, Indiana 46581-0587
- Contact Person: Lonnie Witham Telephone: (574) 267-6639 Fax: (574) 372-1683
Proprietary Name: Disposable Cement Spacer Molds for Temporary Knee Prosthesis
Common Name: Bone Cement Spacer Mold
Classification Name: Knee joint, patellofemorotibial, polymer/metal/polymer, semiconstrained, cemented prosthesis (21 CFR 888.3560)
Legally Marketed Devices To Which Substantial Equivalence Is Claimed: Eogany Marriousary Knee Prosthesis cleared by the FDA in K032522 and marketed by Exactech, Inc., Gainesville, FL.
Device Description: The femoral cement spacer molds are offered in four sizes (60mm, 65mm, 70mm & 75mm). The tibial cement spacer molds are offered in four sizes (65mm, 70mm, 75mm & 80mm). The disposable cement spacer molds are not implanted. (Osmith, Tonin, Tonin Conethylmethacrylate /gentamicin bone cement, or equivalent, either by injecting with a dispenser/gun, or by pouring the prepared cement into the mold. enther by mjocant hardens, the temporary knee prosthesis components are removed from Aner the ochlent and placed into the joint space. The temporary knee prosthesis remains in place the mores and played the second stage of the two-stage procedure is performed to implant a conventional knee joint prosthesis.
Intended Use: The intended use of the Biomet disposable cement spacer molds is to provide the surgeon with a means to mold a temporary knee prosthesis at the point of care that is substantially equivalent to the Exactech Spacer-k temporary knee prosthesis cleared in K032522. The temporary knee prosthesis made with the Biomet disposable cleared in R0325221. The same indication for use as the Exactech Spacer-k.
Indication for Use: Disposable cement spacer molds are indicated for use to mold a temporary total knee replacement (TKR) for skeletally mature patients undergoing a twotemporary total the replaceting process. The molded temporary knee prosthesis is indicated for an implantation period of 180 days or less. Because of inherent mechanical limitations of the device material (polymethylmethacrylate/gentamicin), the molded temporary prosthesis is only indicated for patients who will consistently use traditional mobility assist devices (e.g. crutches, walkers) throughout the implant period.
{1}------------------------------------------------
## Contraindications:
The temporary knee prosthesis made with the disposable cement spacer molds is contraindicated for the following situations:
- ndreated is to 's condition is such that a two-stage arthroplasty procedure is . I he patient's contrasted immune response or other relevant systemic clinical conditions.
- Bone loss precluding adequate support of the prosthesis. .
- Lack of adequate competence (anatomical and functional) of peripheral . ligamentous apparatus and extensor mechanism.
- The procedure is unjustified due to deficiencies in the patient's muscular, nervous . or vascular systems.
- or vasural by title . fracture host bone.
- Infection of the TKR cannot be confirmed. .
- The infected TKR devices cannot be removed. .
- The infecting pathogens are resistant to gentamicin. .
- The patient is sensitive (allergic) to gentamicin, aminoglycosides or PMMA bone . cement.
- A systemic or secondary remote infection is expected or confirmed. .
- The patient does not have a TKR and the infection is secondary to trauma, septic . arthritis or other surgical procedures.
- The patient does not have sufficient bone stock to allow insertion and fixation of . the prosthesis.
- The patient has neuromuscular disorders that do not allow control of the knee . ioint.
- Joins.
The patient's age, weight, or activity level would cause the surgeon to expect . early failure of the system.
Summary of Technologies: The bone cement molds (femoral and tibial) are sterile disposables made of medical grade silicone. The disposable cement spacer molds produce temporary knee prosthesis components that are composed of similar bone cement in similar sizes as the predicate.
Non-Clinical Testing: Comparative testing was performed utilizing a knee joint Nor. Onlined. 16 both the temporary knee prosthesis made with the Biomet disposable cement sumulator of oour the Exactech Spacer-K device. The temporary knee prostheses were spacer mbe equivalent in strength and wear characteristics. Elution testing demonstrated equivalent gentamicin release.
Clinical Testing: No clinical testing was performed.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the seal of the Department of Health and Human Services (HHS) of the United States. The seal features a stylized eagle with three stripes forming its wing, symbolizing health, services, and people. The text "DEPARTMENT OF HEALTH AND HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
JUL 1 9 2005
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. Lonnie Witham Biomet, Inc. 56 East Bell Drive Warsaw, Indiana 46581
Re: K050210
K030210
Trade/Device Name: Disposable Spacer Molds for Temporary Knee Prosthesis Regulation Number: 21 CFR 888.3560 Regulation Name: Knee joint patellofemorotibial polymer semi-constrained cemented prosthesis Regulatory Class: II Product Code: JWH Dated: June 30, 2005 Received: July 1, 2005
Dear Mr. Witham:
We have reviewed your Section 510(k) premarket notification of intent to market the device We have reviewed your Section 3 rots) premained is substantially equivalent (for the indications ferenced and nave decembined the are are are are are are and see marketed in interstate for use stated in the encrosule for tegally manaties provice Americal Device Ameral Read, Drug commerce prior to May 20, 1970, the encordance with the provisions of the Federal Food, Drug, devices that have been reciassince in accordance was a premarket approval application (PMA).
and Cosmetic Act (Act) that do not require approval of a premarket approval appli and Cosment Act (Act) that do not require special controls provisions of the Act. The You may, therefore, market the devices, boyer to the ments for annual registration, listing of general controls provisions of the rise, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it if your device is classified (see above) into exist in the subs are of the device can
may be subject to such additional controls. Existing major regulations FDA may may be subject to such additional controller in the 21, Parts 800 to 898. In addition, FDA may be found in the Code of Pouchal Ing.
publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean Please be advised that FDA 3 issualled of a backers of a backer requirements of the Act
that FDA has made a determination that your device complies with other requirements of that FDA has made a delorimiation administered by other Federal agencies. You must of any Federal statutes and regulations and admited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set CPR Part 807), labening (21 CFR Part 820); and if applicable, the electronic forth in the quality systems (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
# Page 2 - Mr. Lonnie Witham
This letter will allow you to begin marketing your device as described in your Section 510(k) I his letter will anow you to begin manieting your and equivalence of your device to a legally prematicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific advice for your aconce to the 0.000. Also, please mote the regulation entitled, a Colliact the Office of Compuner of the (21CFR Part 807.97). You may obtain Misolalling by reference to premail.com bilities under the Act from the Division of Small other general information on your response Assistance at its toll-free number (800) 638-2041 or Manufacturers, International and Ochess http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Muriam C. Provost
Miriam C. Provost, Ph.D Acting Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
# 510(k) Number (if known):
Device Name: Disposable Cement Spacer Molds for Temporary Knee Prosthesis
## Indications For Use:
Indications For Use:
Disposable cement spacer mold a temporary total kneereduced van a sentic process. The molded Disposable cement spacer molds are nutleated to as wonedure due to a septic process. The molded for skeletally mature patients undergolity a two-stage prodot of 180 days of inherent
temporary knee prosthesis is indicated for an implantation period of 180 days of inheren temporary knee prosthesis is mulce material (gentamicin/polyombility), the molded
mechanical limitations of the device material (gentamicin/polyments), is a si mechanical limitations of the device material (gentainensportions) and mobility assist devices (e.g. crutches, walkers) throughout the implant period.
### Contraindications:
Contraindications:
The temporary knee prosthesis made with the disposable cement spacer molds is contraindicated for the following situations:
- ng situations:
The patient's condition is such that a two-stage arthroplasty procedure is contraindicated due to . I he patient's condition is sack and systemic clinical conditions.
- Bone loss precluding adequate support of the prosthesis.
- . Bone loss preciuding adequate support of the problemal) of peripheral ligamentous apparatus �
- and extensor mechanism. The procedure is unjustified due to deficiencies in the patient's muscular, nervous or vascular . systems.
- systems.
Poor bone quality (as in osteoporosis) could cause the prosthesis to migrate or to fracture host . bone.
- Infection of the TKR cannot be confirmed. ●
- The infected TKR devices cannot be removed. .
- The infecting pathogens are resistant to gentamicin. .
- The intecting patiogens are resiblant to grain.
The patient is sensitive (allergic) to gentamicin, aminoglycosides or PMMA bone cement. .
- I he pation is sensitere (remote infection is expected or confirmed.
- . A systemic of secondary renote infection is oxposited or collary to trauma, septic arthritis or other . surgical procedures.
- surgical procedures.
The patient does not have sufficient bone stock to allow insertion and fixation of the prosthesis.
- . The patient has neuromuscular disorders that do not allow control of the knee joint.
- . I he patient's age, weight, or activity level would cause the surgeon to expect early failure of the . system.
Prescription Use YES (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-USE ANOTHER PAGE IF NEEDED)
2
Concurrence of CDRH, Office of Device Evaluation (ODE)
8050210
iriam C. Provost
(Division Sign-Off) Division of General, Restorative and Neurological Devices
Page 1 of 1
510(k) Number --
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.