K030301 · Smith & Nephew, Inc. · JWH · Feb 25, 2003 · Orthopedic
Device Facts
Record ID
K030301
Device Name
UNICONDYLAR FEMORAL COMPONENT
Applicant
Smith & Nephew, Inc.
Product Code
JWH · Orthopedic
Decision Date
Feb 25, 2003
Decision
SESE
Submission Type
Special
Regulation
21 CFR 888.3560
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The Unicondylar Femoral Components are indicated for restoring either compartment of a knee that has been affected by the following: 1) noninflammatory degenerative joint disease including osteoarthritis, traumatic arthritis, or avascular necrosis; 2) correction of functional deformity; 3) revision procedures where other treatments or devices have failed; and 4) treatment of fractures that are unmanageable using other techniques. The Unicondylar Femoral Components are single use only and are intended for implantation with bone cement.
Device Story
Unicondylar Femoral Components are metal alloy orthopedic implants designed for use with Genesis I Knee System tibial components. Devices undergo proprietary oxidation process. Intended for surgical implantation in knee joint to restore compartment function; requires bone cement. Used by orthopedic surgeons in clinical/hospital settings. Provides structural replacement for damaged femoral condyles; enables restoration of joint mechanics and patient mobility.
Clinical Evidence
Bench testing only. Mechanical test data indicates the components are equivalent to devices currently used clinically and capable of withstanding expected in vivo loading.
Technological Characteristics
Metal alloy femoral components processed via proprietary oxidation. Designed for use with Genesis I Knee System tibial components. Single-use, cemented fixation.
Indications for Use
Indicated for patients requiring restoration of a knee compartment due to noninflammatory degenerative joint disease (osteoarthritis, traumatic arthritis, avascular necrosis), functional deformity, revision of failed treatments, or unmanageable fractures. For use with bone cement.
Regulatory Classification
Identification
A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).
Related Devices
K063515 — TITANIUM REPICCI II UNICONDYLAR FEMORAL KNEE COMPONENTS COCR REPICCI II UNICONDYLAR KNEE · Biomet Manufacturing, Inc. · Feb 2, 2007
K061011 — COMPETITOR UNICONDYLAR KNEE TIBIAL BASEPLATE · Smith & Nephew, Inc. · Jul 11, 2006
K100197 — GENESIS UNICONDYLAR UHMWPE ARTICULAR INSERTS · Smith & Nephew, Inc. · Feb 19, 2010
K081351 — MODIFICATION TO JOURNEY UNICONDYLAR FEMORAL IMPLANT · Smith & Nephew, Inc. · Jul 24, 2008
K120698 — GENESIS II PS NON-MODULAR FEMORAL COMPONENT · Smith & Nephew, Inc. · May 18, 2012
Submission Summary (Full Text)
{0}------------------------------------------------
Kososol pge 11
Summary of Safety and Effectiveness Smith & Nephew, Inc. Unicondylar Femoral Components
FEB 2 5 2003
# Contact Person and Address
Kim Kelly Project Manager, Clinical and Regulatory Affairs Smith & Nephew, Inc. Orthopaedic Division 1450 East Brooks Road Memphis, TN 38116 (991) 399-6566
### Device Description
The Unicondylar Femoral Components are designed for use with tibial components of the Genesis I Knee System. The Unicondylar Femoral Components are metal alloy devices processed via a proprietary oxidation process.
### Device Classification Name
21 CFR 888.3520 Knee joint femorotibial metal/polymer/non-constrained cemented prosthesis - Class II
# Indications for Use
The Unicondylar Femoral Components are indicated for restoring either compartment of a knee that has been affected by the following:
- 1) noninflammatory degenerative joint disease including osteoarthritis, traumatic arthritis, or avascular necrosis;
- 2) correction of functional deformity;
- 3) revision procedures where other treatments or devices have failed; and
- 4) treatment of fractures that are unmanageable using other techniques.
The Unicondylar Femoral Components are single use only and are intended for implantation with bone cement.
# Mechanical and Clinical Data
A review of the mechanical test data indicated that the Unicondylar Femoral Components are equivalent to devices currently used clinically and are capable of withstanding expected in vivo loading without failure.
## Substantial Equivalence Information
The substantial equivalence of the Unicondylar Femoral Components is substantiated by its similarities in design features, overall indications, and material composition as existing components distributed by Smith & Nephew, Inc.
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circle around the caduceus symbol. The logo is black and white.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
FEB 2 5 2003
Ms. Kim P. Kelly Project Manager, Regulatory and Clinical Affairs Smith & Nephew, Inc Orthopaedic Division 1450 Brooks Road Memphis. TN 38116
Re: K030301 Trade/Device Name: Unicondylar Femoral Component Regulation Number: 21 CFR 888.3520 Regulation Name: Knee joint femorotibial metal/polymer non-constrained cemented prosthesis Regulatory Class: II Product Code: JWH Dated: January 28, 2003 Received: January 29, 2003
Dear Ms. Kelly:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{2}------------------------------------------------
Page 2 - Ms. Kim P. Kelly
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely vours,
Mark N Milkman
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
K0}030/
# Unicondylar Femoral Components Indications Statement
The Unicondylar Femoral Components are indicated for restoring either compartment of a knee that has been affected by the following:
- 5) noninflammatory degenerative joint disease including osteoarthritis, traumatic arthritis, or avascular necrosis;
- 6) correction of functional deformity;
- 7) revision procedures where other treatments or devices have failed; and
- 8) treatment of fractures that are unmanageable using other techniques.
The Unicondylar Femoral Components are single use only and are intended for implantation with bone cement.
O Mark N. Milherson
Division
Division
and
K03030
(k) Number
Concurrence of CDRH, Office of Device Evaluation
Prescription Use
OR (Per 21 CFR 801.109)
Over-The Counter Use No
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.