VAPORIZED HYDROGEN PEROXIDE (VHP) STERILIZATION OF UHMWPE COMPONENTS IN THE GENESIS KNEE SYSTEM, THE GENESIS II KNEE SYS
K012778 · Smith & Nephew, Inc. · JWH · Mar 15, 2002 · Orthopedic
Device Facts
Record ID
K012778
Device Name
VAPORIZED HYDROGEN PEROXIDE (VHP) STERILIZATION OF UHMWPE COMPONENTS IN THE GENESIS KNEE SYSTEM, THE GENESIS II KNEE SYS
Applicant
Smith & Nephew, Inc.
Product Code
JWH · Orthopedic
Decision Date
Mar 15, 2002
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3560
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Components in the Genesis, Genesis II, and Profix Knee Systems are indicated for: 1) Rheumatoid arthritis; 2) Post-traumatic arthritis, osteoarthritis, or degenerative arthritis in older patients whose age, weight, and activity level are compatible with an adequate long-term result. 3) Failed osteotomies, unicompartmental replacement, or total knee replacement. 4) The posterior stabilized knee system is designed for use in patients in primary and revision surgery, where the anterior and posterior cruciate ligaments are incompetent and the collateral ligaments remain intact. 5) The constrained knee system is designed for use in patients in primary and revision surgeries, where the posterior cruciate ligament and one or both of the collateral ligaments (i.e. medial collateral and/or lateral collateral ligaments) are absent or incompetent. The devices that are the subject of this premarket notification are intended for cemented use only and are for single use only. The Orthopaedic Division of Smith & Nephew, Inc. will utilize the VHP® sterilization process to terminally sterilize all orthopaedic implant components manufactured from UHMWPE material in the Genesis Knee System, the Genesis II Knee System, and the Profix Knee System (tibial articular inserts, all polyethylene tibial bases, patellae, and flex-lok pegs).
Device Story
Subject devices are UHMWPE knee system components (tibial inserts, tibial bases, patellae, flex-lok pegs) for Genesis, Genesis II, and Profix systems. Device function remains unchanged; notification concerns implementation of VHP (vaporized hydrogen peroxide) sterilization process. Sterilization involves H2O2 gas injections at deep vacuum set points followed by chamber evacuation aeration. Components are packaged in Tyvek Mylar pouches. Process achieves Sterility Assurance Level (SAL) of 10^-6. Intended for use by surgeons in clinical/hospital settings for knee arthroplasty. Benefits include terminal sterilization of UHMWPE implants while maintaining material properties.
Clinical Evidence
Bench testing only. Validation studies confirmed microbicidal effectiveness of VHP sterilization, material compatibility with UHMWPE, and achievement of a Sterility Assurance Level (SAL) of 10^-6. No clinical data provided.
Indicated for patients with rheumatoid, post-traumatic, or osteoarthritis; failed osteotomies; or failed previous knee replacements. Posterior stabilized systems indicated for primary/revision surgery with incompetent cruciate ligaments but intact collateral ligaments. Constrained systems indicated for primary/revision surgery with absent/incompetent posterior cruciate and one or both collateral ligaments. Intended for cemented, single-use only.
Regulatory Classification
Identification
A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).
Predicate Devices
Genesis Knee System
Genesis II Knee System
Profix Knee System
Submission Summary (Full Text)
{0}------------------------------------------------
**MAR 1 5 2002** K0127
Orthopaedic Division
Smith & Nephew, Inc.
1450 Brooks Rd., Memphis, TN 38116 U.S.A.
901-396-2121, For information: 1-800-821-5700
For orders and order inquiries: 1-800-238-7538
**Smith+Nephew**
# 510(k) Summary of Safety and Effectiveness
| Submitter's name: | Smith & Nephew, Inc., Orthopaedic Division |
|-----------------------------------|------------------------------------------------------------------------------------------------------------------------------------|
| Submitter's address: | 1450 Brooks Road, Memphis, TN 38116 |
| Submitter's telephone number: | Direct phone: 901-399-6487 or FAX: 901-398-5146 |
| Contact person: | David Henley, Senior Clinical/Regulatory Affairs Specialist |
| Date summary prepared: | March 15, 2002 |
| Trade or proprietary device name: | UHMWPE Components of the Genesis, the Genesis II, and the<br>Profix Knee Systems Sterilized with the VHP® Sterilization<br>Process |
| Common or usual name: | Genesis, Genesis II, and Profix Knee System UHMWPE<br>Components |
| Classification name: | 21 CFR 888.3560, knee joint patellofemorotibial<br>polymer/metal/polymer, semi-constrained cemented prosthesis |
# Substantially Equivalent Legally Marketed Devices
- The Genesis Knee System Smith & Nephew, Inc. .
- The Genesis II Knee System Smith & Nephew, Inc. .
- The Profix Knee System Smith & Nephew, Inc. �
#### Device Description
There have been no changes in indications for use, design, or material property changes to any Geness, Genesis II, or Profix Knee System UHMWPE components that will be sterilized using the VHP® process.
The VHP sterilization process uses hydrogen peroxide (H2O2) vapor for sterilization Genesis Knee System, Genesis II Knee System, and Profix Knee System components (tibial articular inserts, all polyethylene tibial bases, patellar, and flex-lok pegs) manufactured from UHMWPE using the Century SL VHP® Sterilizer. Sterilization is achieved by a series of H2O2 gas injections at deep vacuum set points. Aeration of the medical devices after sterilization is conducted by a series of chamber evacuations.
VHPGenPro510kSumS&E
{1}------------------------------------------------
# Device Intended Use
Components in the Genesis, Genesis II, and Profix Knee Systems are indicated for:
1) Rheumatoid arthritis; 2) Post-traumatic arthritis, osteoarthritis, or degenerative arthritis in older patients whose age, weight, and activity level are compatible with an adequate long-term result. 3) Failed osteotomies, unicompartmental replacement, or total knee replacement. 4) The posterior stabilized knee system is designed for use in patients in primary and revision surgery, where the anterior and posterior cruciate ligaments are incompetent and the collateral ligaments remain intact. 5) The constrained knee system is designed for use in patients in primary and revision surgeries, where the posterior cruciate ligament and one or both of the collateral ligaments (i.e. medial collateral and/or lateral collateral ligaments) are absent or incompetent.
KOID + +
2012
The devices that are the subject of this premarket notification are intended for cemented use only and are for single use only.
The Orthopaedic Division of Smith & Nephew, Inc. will utilize the VHP® sterilization process to terminally sterilize all orthopaedic implant components manufactured from UHMWPE material in the Genesis Knee System, the Genesis II Knee System, and the Profix Knee System (tibial articular inserts, all polyethylene tibial bases, patellae, and flex-lok pegs).
# Technological Characteristics
The VHP sterilization process is similar to the predicate sterilization processes listed above. Both of these predicate sterilization processes are intended to terminally sterilize orthopaedic implants/medical devices to a Sterility Assurance Level (SAL) of 10 . When compared to the VHP process, the predicate processes also have similar technological characteristics.
## Performance Characteristics
The Orthopaedic Division of Smith & Nephew, Inc. has conducted numerous tests as supporting evidence that the VHP® sterilization process is qualified for sterilization of all UHMWPE orthopaedic implants in the Genesis, Genesis II, and Profix Knee Systems by demonstrating the following:
- Microbicidal effectiveness of the vaporized hydrogen peroxide .
- The effects of the VHP sterilization cycle on UHMWPE materials in which the . product is packaged
- Process validation efforts to demonstrate that the VHP® sterilization process is effective and . reproducible resulting in a SAL of at least 100
Test results demonstrated that the VHP® sterilization process is capable of terminally sterilizing UHMWPE orthopaedic implants and verifies achievement of a SAL of 10 . The VHP sterilization process was also demonstrated to be safe, reproducible, predictable and effective in sterilizing UHMWPE orthopaedic implants packaged and sealed in Tyvek Mylar pouches.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with its wings spread, symbolizing protection and care. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAR 1 5 2002
Mr. David Henley Senior Clinical/Regulatory Affairs Specialist Orthopedic Division Smith & Nephew, Inc. 1450 Brooks Road Memphis, Tennessee 38116
Re: K012778
Trade/Device Name: UHMWPE Components of the Genesis II, and the Profix Knee Systems Regulation Number: 21 CFR 888.3560
Regulation Name: Knee Joint Patellofemorotibial Polymer/Metal/Polymer Semi-constrained Cemented Prosthesis
Regulatory Class: Class II Product Code: JWH Dated: December 14, 2001 Received: December 17, 2001
Dear Mr. Henley:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for associated in the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean I that FDA has made a determination that your device complies with other requirements of the Act that I Dr. I has haatutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the 0017; adoning (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
#### Page 2 - Mr. David Henley
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
for Mark N Milkunas
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
### Premarket Notification Indications Enclosure
510(k) Number (if known): Kol2778
Device Name: UHMWPE Components of the Genesis, Genesis II, and Profix Knee Systems Sterilized with the VHP® Sterilization Process
#### Indications for Use:
Components in the Genesis, Genesis II, and Profix Knee Systems are indicated for:
1) Rheumatoid arthritis; 2) Post-traumatic arthritis, osteoarthritis, or degenerative arthritis in older patients whose age, weight, and activity level are compatible with an adequate long-term result. 3) Failed osteotomies, unicompartmental replacement, or total knee replacement. 4) The posterior stabilized knee system is designed for use in patients in primary and revision surgery, where the anterior and posterior cruciate ligaments are incompetent and the collateral ligaments remain intact. 5) The constrained knee system is designed for use in patients in primary and revision surgeries, where the posterior cruciate ligament and one or both of the collateral ligaments (i.e. medial collateral and/or lateral ligaments) are absent or incompetent.
The devices that are the subject of this premarket notification are intended for cemented use only and are for single use only.
The Orthopaedic Division of Smith & Nephew, Inc. will utilize the VHP sterilization process to terminally sterilize all orthopaedic implant components manufactured from UHMWPE material in the following product systems:
- . Genesis Knee System (i.e. tibial articular inserts, all polyethylene tibial bases, patellac, and flex-lok pegs)
- Genesis II Knee System (i.e. tibial articular inserts, all polyethylene tibial bases, patellae, and flex-lok ● pegs)
- . Profix Knee System (i.e. tibial articular inserts, all polyethylene tibial bases, patellae)
#### (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
I. Mach A Milken
Division of General, Restorative and No ological Devices
510(k) Number _
Prescription Use (Per 21 CFR 801.109)
OR
Over-the-Counter Use
(Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.