← Product Code [JWH](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/JWH) · K012172

# DURACON PS LIPPED TIBIAL INSERT (K012172)

_Howmedica Osteonics · JWH · Oct 10, 2001 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/JWH/K012172

## Device Facts

- **Applicant:** Howmedica Osteonics
- **Product Code:** [JWH](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/JWH.md)
- **Decision Date:** Oct 10, 2001
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3560
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

The Duracon® Posteriorly Stabilizer (PS) Lipped Tibial Inserts are intended to be used with legally marketed Duracon® tibial baseplates and wedges, femoral components, and patellar components as part of a total knee system in primary or revision cemented total knee arthroplasty. These inserts are intended to be used in cases where there is destruction of the joint surfaces with or without bone deformity, where the cruciate ligaments are inadequate, not present, or cannot be preserved during the operative procedure, especially when anterior-posterior stability is impaired due to absence of the patella. The collateral ligaments may or may not be used for revision of a failed prosthesis.

## Device Story

Duracon® PS Lipped Tibial Insert is a component of a total knee prosthesis system. It functions as a bearing surface for cemented total knee arthroplasty. Device features include an anterior/posterior (A/P) lipped design for enhanced stability in full extension and a patellar cut-out for improved tracking. It utilizes a snap-fit locking mechanism with an auxiliary locking screw for attachment to Duracon® tibial baseplates. Used by orthopedic surgeons in clinical/hospital settings during primary or revision knee replacement procedures. The device replaces damaged joint surfaces, restores joint anatomy, and provides stability in cases of cruciate ligament deficiency or patellar absence. It is intended to improve patient mobility and reduce pain associated with degenerative or traumatic joint conditions.

## Clinical Evidence

Bench testing only. No clinical data provided.

## Technological Characteristics

Polymer tibial insert component for a semi-constrained, cemented knee prosthesis. Features snap-fit locking mechanism with locking screw. Includes A/P lipped geometry and patellar cut-out. Materials and manufacturing methods are identical to predicate Duracon® Stabilized inserts. Designed for use with existing Duracon® femoral and patellar components.

## Regulatory Identification

A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).

## Predicate Devices

- Duracon® Stabilized Tibial Inserts ([K932070](/device/K932070.md))
- Duracon® Stabilized Tibial Inserts ([K936292](/device/K936292.md))

## Reference Devices

- Duracon® tibial baseplates ([K915512](/device/K915512.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

OCT 1 0 2001

510(k) Summarv Duracon® PS Lipped Tibial Inserts

Image /page/0/Picture/2 description: The image shows handwritten text that appears to be a combination of letters and numbers. The first line reads 'K012172', and the second line reads '1 OF 4'. The handwriting is somewhat stylized, with distinct shapes for each character, and the text is presented in a clear, legible manner against a white background.

#### 510(k) Summary Duracon® PS Lipped Tibial Insert

#### Submission Information

| Name and Address of the Sponsor<br>of the 510(k) Submission: | Howmedica Osteonics Corp.<br>59 Route 17<br>Allendale, NJ 07401-1677                                                     |
|--------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------|
| Contact Person:                                              | Mary-Catherine Dillon<br>Regulatory Affairs Specialist<br>and<br>Susan Krasny, PhD<br>Director, Regulatory Affairs       |
| Date of Summary Preparation:                                 | July 10, 2001                                                                                                            |
| Device Identification                                        |                                                                                                                          |
| Proprietary Name:                                            | Duracon® PS Lipped Tibial Insert                                                                                         |
| Common Name:                                                 | Knee Prosthesis                                                                                                          |
| Classification Name and Reference:                           | Knee Joint, Patellofemorotibial,<br>Polymer/Metal/Polymer, Semi-<br>Constrained, Cemented Prosthesis<br>21 CFR '888.3560 |

#### Predicate Device Identification

The Duracon® PS Lipped Tibial Bearing Inserts are substantially equivalent to the Duracon® Stabilized Tibial Inserts cleared via K932070 and K936292.

#### Device Description

The Duracon® PS Lipped tibial inserts share the same critical design features as the predicate Duracon® Stabilized tibial inserts. They both have the same snap-fit locking

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows handwritten text on a white background. The text on the top line reads 'K012172'. The text on the bottom line reads '20F4'. The handwriting appears to be somewhat stylized and the numbers are clearly written.

mechanism, augmented by a locking screw, for mating with the current Duracon® tibial baseplates (K915512). The subject Duracon inserts are intended to be used with current Duracon® femoral and patellar components. The subject inserts will be available in the same range of sizes and thicknesses and are manufactured from the same materials as the current Duracon® Stabilized inserts.

The subject inserts have an anterior/posterior (A/P) lipped feature to enhance femoral/tibial implant anterior stability in full extension. In addition, the subject inserts incorporate a patellar cut-out feature to improve patellar tracking.

#### Intended Use

The intended use of the Duracon® PS Lipped tibial inserts is identical to that of the predicate Duracon® Stabilized tibial bearing inserts. As with the predicate inserts, the modified inserts are single use devices for use as part of a total knee system in primary or revision cemented total knee arthroplasty. These inserts are intended to be used in cases where there is destruction of the joint surfaces with or without bone deformity, where the cruciate ligaments are inadequate, not present, or cannot be preserved during the operative procedure, especially when anterior-posterior stability is impaired due to absence of the patella. The collateral ligaments may or may not be used for revision of a failed prosthesis.

#### Indications

- Noninflammatory degenerative joint disease including osteoarthritis, traumatic . arthritis or avascular necrosis,
- . Rheumatoid arthritis,
- . Correction of functional deformity,
- Revision procedures where other treatments or devices have failed, u

{2}------------------------------------------------

K012172
3 OF 4

510(k) Summary Duracon® PS Lipped Tibial Inserts

- adon't To Elped notal more
 Post-traumatic loss of joint anatomy, particularly when there is patello-femoral
erosion, dysfunction or prior patellectomy; and,

- Irreparable fracture of the knee .
# Contraindications

#### Absolute contraindications include:

- 6 Overt infection,
- Distant foci of infections (which may cause hematogenous spread to the implant I site),
- Rapid disease progression as manifested by joint destruction or bone absorption . apparent on roentgenogram,
- l Skeletally immature patients
- Cases where there is poor bone stock which would make the procedure # unjustifiable,

### Conditions presenting an increased risk of failure include:

- Uncooperative patient or patient with neurological disorders who is incapable of l following instructions,
- I Osteoporosis,
- Metabolic disorders which may impair bone formation, l
- . Osteomalacia, and
- 트 Previous arthrodesis

# Performance Data

Mechanical testing has been performed to demonstrate the substantial equivalence of the subject inserts to the predicate inserts.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows handwritten text on a white background. The text "K012172" is written in the top half of the image. Below that, the text "4 OF 4" is written in a smaller font size.

# Statement of Technological Comparison

The fundamental scientific technology of the current Duracon® PS Lipped tibial inserts has not changed with regard to the modified inserts. The modified inserts employ the same basic design concepts, the same materials, and the same manufacturing methods.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image is a black and white circular seal. The seal contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the edge of the circle. Inside the circle is a stylized image of an eagle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

OCT 1 0 2001

Ms. Mary-Catherine Dillon Regulatory Affairs Howmedica Osteonics Corp. 59 Route 17 Allendale, New Jersey 07401-1677

Re: K012172 Trade Name: Duracon® Posteriorly Stabilized Lipped Tibial Insert Regulation Number: 21 CFR 888.3560 Regulation Name: Knee joint patellofemorotibial, polymer/metal/polymer semi-constrained cemented prosthesis Regulatory Class: Class II Product Code: JWH Dated: June 22, 2001 Received: July 12, 2001

Dear Ms. Dillon:

We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{5}------------------------------------------------

#### Page 2 - Ms. Mary-Catherine Dillon

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours,

Sincerely yours,

Mark N Millerson

Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

510(k) Number (if known): K O la | 7 à

Device Name: Duracon® Posteriorly Stabilized Lipped Tibial Insert

The Duracon® Posteriorly Stabilizer (PS) Lipped Tibial Inserts are intended to be used with legally marketed Duracon® tibial baseplates and wedges, femoral components, and patellar components as part of a total knee system in primary or revision cemented total knee arthroplasty. These inserts are intended to be used in cases where there is destruction of the joint surfaces with or without bone deformity, where the cruciate ligaments are inadequate, not present, or cannot be preserved during the operative procedure, especially when anterior-posterior stability is impaired due to absence of the patella. The collateral ligaments may or may not be used for revision of a failed prosthesis.

#### Indications

- Noninflammatory degenerative joint disease including osteoarthritis, traumatic 미 arthritis or avascular necrosis,
- Rheumatoid arthritis, 트
- Correction of functional deformity, l
- Revision procedures where other treatments or devices have failed, 1
- Post-traumatic loss of joint anatomy, particularly when there is patello-femoral . erosion, dysfunction or prior patellectomy; and,
- . Irreparable fracture of the knee

(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

| Prescription Use<br>(CFR 801.109) | X | OR |
|-----------------------------------|---|----|
|-----------------------------------|---|----|

Over-the Counter Use (Per 21

Musth N millersen
Division Sign-Off)

eral. Restorative

**510(k) Number** K012172

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/JWH/K012172](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/JWH/K012172)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
