SCORPIO CR SUPERFLEX TIBIAL INSERT

K011643 · Howmedica Osteonics Corp. · JWH · Jun 28, 2001 · Orthopedic

Device Facts

Record IDK011643
Device NameSCORPIO CR SUPERFLEX TIBIAL INSERT
ApplicantHowmedica Osteonics Corp.
Product CodeJWH · Orthopedic
Decision DateJun 28, 2001
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 888.3560
Device ClassClass 2
AttributesTherapeutic

Intended Use

The Scorpio® CR Superflex™ Tibial Bearing Inserts are single use devices. They are intended for mating with commercially available Howmedica Osteonics tibial trays and patellas and Scorpio® Total Knee CR femoral components, which are intended for cemented fixation.

Device Story

Scorpio® CR Superflex™ Tibial Insert is a knee prosthesis component; designed for use with Howmedica Osteonics tibial trays, patellas, and cruciate retaining femoral components. Device features barb and wire locking mechanism; modifications include increased anterior radius of sagittal curve and added patella recess to improve manufacturability and internal rotation constraint. Intended for cemented fixation in patients with degenerative, rheumatoid, or post-traumatic arthritis; or revision of prior knee procedures. Device is single-use; implanted by orthopedic surgeons during total knee arthroplasty to restore joint function and stability.

Clinical Evidence

Bench testing only. Static testing performed to demonstrate substantial equivalence to the predicate device.

Technological Characteristics

Polymer tibial insert; semi-constrained; cemented fixation. Features barb and wire locking mechanism, modified anterior sagittal curve radius, and patella recess. Materials and manufacturing methods identical to predicate.

Indications for Use

Indicated for patients with painful, disabling knee joint disease due to degenerative, rheumatoid, or post-traumatic arthritis; post-traumatic loss of knee joint function; moderate varus, valgus, or flexion deformity with stable ligamentous structures; or revision of previous unsuccessful knee replacement.

Regulatory Classification

Identification

A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ 510(k) Summary Scorpio® CR Superflex TM Tibial Inserts # JUN 2 8 2001 ## 510(k) Summary Scorpio® CR Superflex™ Tibial Insert ## Submission Information Name and Address of the Sponsor of the 510(k) Submission: Contact Person: Howmedica Osteonics Corp. 59 Route 17 Allendale, NJ 07401-1677 Mary-Catherine Dillon Regulatory Affairs Specialist Date of Summary Preparation: June 21, 2001 Device Identification Proprietary Name: Scorpio® CR Superflex™ Tibial Insert Common Name: Knee Prosthesis Classification Name and Reference: Knee Joint, Patellofemorotibial, Polymer/Metal/Polymer, Semi-Constrained, Cemented Prosthesis 21 CFR '888.3560 ### Predicate Device Identification The Scorpio® CR Superflex™ Tibial Bearing Inserts are substantially equivalent to the Scorpio® CR Tibial Bearing Inserts cleared via K974556. #### Device Description The Scorpio® CR Superflex™ tibial inserts share critical design features as the predicate Scorpio® CR tibial inserts. They have a barb and wire locking mechanism for mating with the appropriate Howmedica Osteonics tibial trays and are compatible with the current Scorpio® patella and CR femoral components. The subject inserts will be । ਕ {1}------------------------------------------------ 510(k) Summary Scorpio® CR SuperflexTM Tibial Inserts available in the same range of sizes and thicknesses as the current Scorpio® CR inserts. The modifications incorporated with the Scorpio® CR Superflex™ inserts address ease of manufacturability and constraint in internal rotation. These modifications include increasing the anterior radius of the sagittal curve and adding a patella recess feature. ### Intended Use The intended use of the modified tibial bearing inserts is identical to that of the unmodified tibial bearing inserts. As with the predicate inserts, the modified inserts are single use devices. They are intended for mating with commercially available Howmedica Osteonics tibial trays and patellas and corresponding Howmedica Osteonics cruciate retaining femoral components. Howmedica Osteonics Scorpio® CR Total Knee components are intended for cemented fixation. ### Indications The indications for the modified Scorpio® CR Superflex™ tibial inserts will remain the same as the indications for the current Scorpio® CR inserts. These are: - Painful, disabling joint disease of the knee resulting from: degenerative arthritis, . rheumatoid arthritis or post-traumatic arthritis - Post-traumatic loss of knee joint configuration and function . - Moderate varus, valgus, or flexion deformity in which the ligamentous structures can . be returned to adequate function and stability - Revision of previous unsuccessful knee replacement or other procedure . {2}------------------------------------------------ 510(k) Summary Scorpio® CR SuperflexTM Tibial Inserts ## Performance Data Static testing has been performed to demonstrate the substantial equivalence of the subject inserts to the predicate inserts. ## Statement of Technological Comparison The fundamental scientific technology of the current Scorpio® CR tibial inserts has not changed with regard to the modified inserts. The modified inserts employ the same basic design concepts, the same materials, and the same manufacturing methods. {3}------------------------------------------------ Image /page/3/Picture/2 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the department's name around the perimeter. In the center of the seal is a stylized image of an eagle with three wavy lines representing its wings. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 # JUN 2 8 2001 Ms. Mary-Catherine Dillon Regulatory Affairs Team Member Howmedica Osteonics Corporation 59 Route 17 Allendale, New Jersey 07401-1677 Re: K011643 Trade Name: Scorpio® Cruciate Retaining (CR) Superflex™ Tibial Inserts Regulation Number: 888.3560 Regulatory Class: II Product Code: JWH Dated: May 25, 2001 Received: May 29, 2001 #### Dear Ms. Dillon: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the sured in the encreater) to 'cal Device Amendments, or to devices that have been reclassified in encerdance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. {4}------------------------------------------------ ## Page 2 - Ms. Mary-Catherine Dillon If you desire specific advice for your device on our labeling regulation (21 CFF Part 801 and If you desire specific advice for your dones contact the Office of Compliance at variance at additionally 809.10 for m Viro diagnosus on the promotion and advertising of your device, (301) 594-4659. Addinonally, for questions of the presence the regulation prease contact the Onlive or Comphanee as (st a station" (21 CFR 807.97). Other generall entitled, "Misuranding by releited to prema Act may be obtained from the Division of Small information on your responsibilities under the Act 12012 142, 6507, and title information on your responsionnes and its not no not (300) 638-2041 or at (301) 443-6597, or at its Internet address "http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours, itchell MDfor Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ 510(k) Number (if known): Koll643 Device Name: Scorpio® CR Superflex™ Tibial Inserts The Scorpio® CR Superflex™ Tibial Bearing Inserts are single use devices. They are intended for mating with commercially available Howmedica Osteonics tibial trays and patellas and Scorpio® Total Knee CR femoral components, which are intended for cemented fixation. ## Indications The indications for the modified Scorpio® CR Superflex™ tibial inserts will remain the same as the indications for the current Scorpio® CR inserts. These are: - Painful, disabling joint disease of the knee resulting from: degenerative arthritis, . rheumatoid arthritis or post-traumatic arthritis - Post-traumatic loss of knee joint configuration and function . - Moderate varus, valgus, or flexion deformity in which the ligamentous structures can . be returned to adequate function and stability - Revision of previous unsuccessful knee replacement or other procedure . (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) | Prescription Use | X | |------------------|---| | (801.109) | | OR | Over-The-Counter Use | | |----------------------|--| | (Per 21 CFR | | (Optional Format 1-2-96) Dsmichell, MD onmul (Division Sign-Off) Division of General, Restorative and Neurological Devices | 510(k) Number | K011643 | |---------------|---------| |---------------|---------|
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...