The Steinmann Pins are intended to be used in conjunction with a compatible Stryker external supporting frame for stabilization of open and/or unstable fractures and where soft tissue injury may preclude the use of other fracture treatments such as IM rodding, casting and other means of internal fixation. The Kirschner Wires are intended for use in skeletal traction for alignment and reduction of long bone fractures, and as guide wires in hip pinning, and as fracture fixation devices in certain other small bone fractures. • For use as guide wires in hip pinning procedures, • For use in aligning and reducing long bone fractures, • For use in securing temporary stabilization of bone fractures such as olecranon fractures; tibia plateau fractures; hand and foot bone fractures; humeral, radial and ulnar fractures. • For use with cerclage wire/cable in treating greater trochanter fractures.
Device Story
Steinmann Pins and Kirschner Wires are metallic bone fixation fasteners; used by orthopedic surgeons in clinical settings. Pins/wires inserted into bone to provide skeletal traction, fracture alignment, reduction, or temporary stabilization. Steinmann Pins used with external supporting frames for open/unstable fractures; Kirschner Wires used as guide wires for hip pinning or fixation for small bone fractures. Devices feature varying diameters, lengths, and tip geometries (trocar or diamond); some are threaded. Devices provide mechanical support to bone segments; aid in guiding cannulated instruments or evaluating screw trajectories. Single-use, sterile devices; facilitate fracture healing by maintaining bone position.
Clinical Evidence
No clinical data provided. Substantial equivalence demonstrated via bench testing and comparison to predicate devices. Packaging validated per ISO 11607-1 and ISO 11607-2; bench testing performed per ASTM standards.
Technological Characteristics
Material: Stainless steel (1.4441). Design: Smooth or threaded pins/wires with trocar or diamond tip geometries. Dimensions: Variable diameters and lengths. Sterilization: Radiation. Standards: ISO 5838-1 (implants for surgery), ISO 11607-1/2 (packaging).
Indications for Use
Indicated for patients requiring skeletal traction, fracture alignment, reduction, or temporary stabilization of long bone and small bone fractures (e.g., olecranon, tibia plateau, hand, foot, humeral, radial, ulnar, greater trochanter). Also indicated for use as guide wires in hip pinning procedures and with external fixation frames for open/unstable fractures where other internal fixation methods are precluded.
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 5, 2021
Stryker GmbH Keith Neligan Senior Staff Regulatory Affairs Specialist 325 Corporate Drive Mahwah, New Jersey 07430
Re: K211508
Trade/Device Name: Steinmann Pins and Kirschner Wires Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or Threaded Metallic Bone Fixation Fastener Regulatory Class: Class II Product Code: HTY, JDW Dated: August 18, 2021 Received: August 24, 2021
Dear Keith Neligan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{2}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Brent L. Showalter, Ph.D. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K211508
Device Name Steinmann Pins
Indications for Use (Describe)
The Steinmann Pins are intended to be used in conjunction with a compatible Stryker external supporting frame for stabilization of open and/or unstable fractures and where soft tissue injury may preclude the use of other fracture treatments such as IM rodding, casting and other means of internal fixation.
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------------------|---------------------------------------------------------------------------------|
| <div> <span> ☑ Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
510(k) Number (if known) K211508
Device Name Kirschner Wires
## Indications for Use (Describe)
The Kirschner Wires are intended for use in skeletal traction for alignment and reduction of long bone fractures, and as guide wires in hip pinning, and as fracture fixation devices in certain other small bone fractures.
· For use as guide wires in hip pinning procedures,
· For use in aligning and reducing long bone fractures,
· For use in securing temporary stabilization of bone fractures such as olecranon fractures; tibia plateau fractures; hand and foot bone fractures; humeral, radial and ulnar fractures.
· For use with cerclage wire/cable in treating greater trochanter fractures.
Type of Use (Select one or both, as applicable)
| <div> <input checked="true" type="checkbox"/> <span>Prescription Use (Part 21 CFR 201 Subpart D)</span> </div> |
|----------------------------------------------------------------------------------------------------------------|
| <div> <input type="checkbox"/> <span>Over-The-Counter Use (Part 21 CFR 201 Subpart C)</span> </div> |
> Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
## 510(k) Summary
| Proprietary Names: | Steinmann Pins<br>Kirschner Wires |
|---------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name: | Pin, Fixation, Threaded; Pin, Fixation, Smooth |
| Regulation Description: | Smooth or threaded metallic bone fixation fastener |
| Regulation Number: | 21 CFR 888.3040 |
| Classification Product Code: | JDW, HTY |
| Device Class: | II |
| Submitter: | Stryker GMBH<br>Bohnackerweg 1<br>2545 Selzach, Switzerland |
| Contact Person: | Keith Neligan<br>Senior Staff Regulatory Affairs Specialist<br>IDA Business & Technology Park, Carrigtwohill,<br>Co. Cork, Ireland T45 HX08<br>Phone: +00 353 87 7995645 |
| Date: | August 09, 2021 |
| Primary Predicate Device | Apex Fixation Pins (K001886) |
| Additional Predicate Device | Osteo Kirschner Wires (K971962) |
| Device Description: | The Steinmann Pins and Kirschner Wires are designed for multiple purposes<br>which allow the use in skeletal traction for alignment and reduction of bone<br>fractures, function as fracture fixation devices in certain fractures, preliminary<br>fixation for implants (for example plates) and for evaluation of screw<br>trajectories or guidance for cannulated instruments.<br><br>The Steinmann Pins and Kirschner Wires are available in different dimensions<br>and provide several design features. The dimensions vary in diameter and<br>length. The design features are the tip geometry which can be a three-sided<br>trocar tip or a two-sided diamond tip on either one end or both ends of the pin.<br>Additionally, some pins are threaded.<br><br>The pins and wires are made from stainless steel (1.4441). |
| Indications for Use: | The Steinmann Pins are intended to be used in conjunction with a compatible<br>Stryker external supporting frame for immobilization of open and/or unstable<br>fractures and where soft tissue injury may preclude the use of other fracture<br>treatments such as IM rodding, casting and other means of internal fixation. |
| | The Kirschner Wires are intended for use in skeletal traction for alignment and<br>reduction of long bone fractures, and as guide wires in hip pinning, and as<br>fracture fixation devices in certain other small bone fractures.<br>• For use as guide wires in hip pinning procedures,<br>• For use in aligning and reducing long bone fractures,<br>• For use in securing temporary stabilization of bone fractures such as<br>olecranon fractures; patella fractures; tibia plateau fractures; small hand and<br>foot bone fractures; humeral, radial and ulnar fractures.<br>• For use with cerclage wire/cable in treating greater trochanter fractures. |
| Summary of Technologies: | Steinmann Pins:<br>The intended use of the subject devices (Steinmann Pins) are similar to those<br>detailed in the predicate device (Apex Fixation Pins). There is no change in the<br>fundamental scientific technology shared by both the subject device and<br>predicate device. |
| | Kirschner Wires<br>The intended use of the subject devices (Kirschner Wires) are similar to those<br>detailed in the predicate device (Osteo Kirschner Wires). There is no change in the<br>fundamental scientific technology shared by both the subject device and<br>predicate device. |
| Performance Data (Nonclinical): | Non-Clinical Performance and Conclusions:<br>The Steinmann Pins and Kirschner wires are single use instruments, sterilized<br>by means of radiation; they are sterile stainless steel wires with diameter and<br>lengths comparable to those of the predicate devices. The design is<br>substantially equivalent to that of the predicate devices; the indications for use<br>are included in those of the predicate devices; the material of the Steinmann<br>Pins and Kirschner wire are the same as that of the predicate device. Moreover,<br>Steinmann Pins and Kirschner wires conform to the international standard ISO<br>5838-1. Testing, therefore, is not needed to demonstrate that the subject devices<br>are substantially equivalent to other legally marketed Steinmann Pins and<br>Kirschner wires. |
| | Packaging tests were performed according ISO 11607-1 and ISO 11607-2. All<br>bench tests performed in accordance with ASTM standards. |
| | Clinical Performance and Conclusions:<br>Clinical data and conclusions were not needed for these devices to demonstrate<br>substantial equivalence to the predicate devices. |
| Conclusion: | The subject devices have the same intended use and similar indications for use<br>as the predicate devices. The subject devices use the same operating principle,<br>incorporate the same basic design and labeling and are manufactured and<br>sterilized using the same materials and processes as the predicate devices.<br>The performance data and analyses demonstrate that:<br>• any differences do not raise new questions of safety and effectiveness as<br>established with performance testing; and<br>• the subject devices are at least as safe and effective as the legally marketed<br>predicate devices |
{6}------------------------------------------------
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.